Abstract
Quantitative diagnosis of glucose-6-phosphate dehydrogenase (G6PD) deficiency is essential for the safe administration of 8-aminoquinoline based radical cure for the treatment of Plasmodium vivax infections. Here, we present the PreQuine Platform (IVDS, USA), a quantitative biosensor that uses a dual-analyte assay for the simultaneous measurement of Hemoglobin (Hgb) levels and G6PD enzyme activity within the same sample. The platform relies on a downloadable mobile application. The device requires 10μl of whole blood and works with a reflectance-based meter. Comparing the G6PD measurement normalized by Hgb of 12 samples from the PreQuine Platform with reference measurements methods (spectrophotometry, Pointe Scientific, USA and hemoglobin meter, HemoCue, Sweden) showed a positive and significant agreement with a slope of 1.0091 and an intercept of -0.0379 under laboratory conditions. Next steps will be to conduct field trials in Bangladesh, Cambodia, and the USA to assess diagnostic performance, user friendliness and acceptance.
Citation: Harper R, Ley B, Kabir MA, Matulis G, von Seidlein L, Alam MS, et al. (2024) The PreQuine Platform: A novel diagnostic tool for measuring glucose-6-phosphate dehydrogenase (G6PD) activity and hemoglobin concentration. PLoS ONE 19(5): e0297918. https://doi.org/10.1371/journal.pone.0297918
Editor: Germana Bancone, Shoklo Malaria Research Unit, THAILAND
Received: October 4, 2023; Accepted: January 15, 2024; Published: May 10, 2024
This is an open access article, free of all copyright, and may be freely reproduced, distributed, transmitted, modified, built upon, or otherwise used by anyone for any lawful purpose. The work is made available under the Creative Commons CC0 public domain dedication.
Data Availability: All relevant data are within the manuscript.
Funding: This project was supported by NIH SBIR Phase I award # 1R43AI129057-01A1, Phase II 2R44AI129057-02A1 & 5R44AI129057-03, and 2SB1AI129057-04. The funder had no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: I have read the journal’s policy and the authors of this manuscript have the following competing interests: RH is the Founder and CEO of In Vitro Diagnostic Solutions, MAK is a scientist currently employed by IVDS, BL, LvS, MSA, BA, BAO, and MEV are receiving project funding for the field trials in the United States, Bangladesh and Cambodia. Both MEV and BL serve as academic editors for PLoS One. This does not alter our adherence to PLoS ONE policies on sharing data and materials.
Introduction
Plasmodium vivax causes between 5–14 million malaria cases each year, with over two billion people at risk of infection [1,2], and is the predominant cause of malaria in areas outside of the African continent. In contrast to other malaria parasites, P. vivax (and P. ovale) can form dormant liver stages (hypnozoites) which reactivate weeks to months after the initial infection accounting for up to 80% of all cases of vivax malaria in some areas [3].
Only one class of drugs with hypnozoitocidal properties is currently licensed. 8-aminoquinoline compounds (8AQ), namely primaquine (PQ) and tafenoquine (TQ), can clear hypnozoites, and in combination with a blood schizontocidal are referred to as radical cure [4–6]. While well tolerated in most patients, PQ and TQ can cause hemolysis in individuals with glucose-6-phosphate dehydrogenase (G6PD) deficiency, an enzymopathy that affects between 400 to 500 million people worldwide [7–9]. Since routine testing for G6PD deficiency is rarely available in many endemic settings, clinicians often do not prescribe hypnozoitocidal treatment due to concerns of drug induced acute hemolytic anemia in patients with unknown G6PD deficiency. Hence, there is a critical need to Identify G6PD deficiency in patients and monitor hemoglobin (Hgb) concentrations at point-of-care, to guide radical cure and support overall clinical management of patients infected with P. vivax [1]. Wider use of G6PD diagnostics for the safe administration of radical cure will be critical to reduce relapse and the transmission of vivax malaria in high-risk populations. Scaled availability of point-of-care G6PD diagnostic kits for rural tropical regions, which suffer the largest burden of vivax malaria, will protect individuals vulnerable to 8-aminoquinolone drug-induced adverse events and enable individuals with normal G6PD levels to receive curative treatment [10].
The PreQuine Platform is a novel, point-of-care diagnostic kit developed by In Vitro Diagnostics Solutions (IVDS, USA) for the simultaneous, quantitative determination of G6PD enzyme activity and Hgb concentration. We report the first findings from the device when tested under laboratory conditions.
Methods
The PreQuine Platform
The PreQuine Platform consists of a portable reflectometer, dual-analyte test strips, blood collection kits, fixed-volume, single-use pipettes, lysing buffer, and a mobile application (Fig 1). Each lot of test strips is validated by scanning a QR code which contains information concerning the lot number, expiration date, and calibration constants needed for calculating Hgb and G6PD levels. A total of 10 μl of capillary or venous blood is added to the lysis buffer using a single-use pipette, and the resulting suspension is mixed thoroughly. Using a second pipette, 20 μl of the blood-buffer suspension are then added to a test strip inserted into the meter cassette guide. A precise formulation of enzymes, bioactive components, surfactants, and excipients immobilized within the test strip membranes produce a colorimetric signal proportional to analyte levels in the sample. The reflectometer detects and transmits the colorimetric data to the mobile app for display, storage, and dissemination. The mobile app also displays step-by-step directions of use. A normalized G6PD measurement is displayed within 6 minutes, the measurement of Hgb concentration is done within 1 minute.
Dual analyte test strip design
The design schema of the dual analyte test strip for simultaneous detection of Hgb concentration and G6PD enzyme activity is shown in Fig 2. A hydrophilic spreading layer (transport membrane), which contains a chelating agent, helps to filter cell debris while evenly distributing the sample laterally across the underlying membrane layers. The applied sample then migrates vertically through the Hgb and G6PD reaction layers, eventually reaching the read zones positioned over the 570nm (for Hgb detection) and 670nm (for G6PD detection) LEDs, which are embedded in the reflectometer.
a) Exploded view of the G6PD dual analyte test strip. (1) a spreading layer (transport membrane), (2) a G6PD reagent membrane (hollow fiber membrane with electron mediator, inhibitor, and substrate), (3) a G6PD detection membrane (mixed mesh < 1 μm) facing the LED, (4) a reagent membrane (polyether sulfone membrane containing sodium azide and sodium nitrite) and (5) an optical membrane (polyether sulfone membrane < 1 μm) facing the LED. b) Top and bottom view of the fully assembled test strip with representative color change after test strip chemical reaction (far right). c) PreQuine Platform’s reflectance-based meter shows both 570 and 670 nm LEDs that are used for the simultaneous detection of Hgb and G6PD, respectively.
G6PD detection mechanism
The reaction mechanism for the G6PD measurement is shown in Fig 3A. The chemistry utilizes a substrate β-D-glucose-6-phosphate disodium salt (G6PD substrate), NADP+, diaphorase (electron mediator), and a tetrazolium salt, which eventually produce formazan, a colored end-reaction product. The chosen formazan reaction product is outside of the spectral wavelength maxima of hemoglobin and bilirubin, reducing the potential for assay interference from these two compounds (Fig 3B). Maleimide is also added to the membrane as a sequestering compound to reduce the creation of secondary NADPH resulting from reactions with 6-phosphogluconate dehydrogenase. The PreQuine Platform incorporates a Temperature Correction Factor (TCF) for G6PD activity measurements, similar to other G6PD point-of-care quantitative devices [11], which is critical given that the G6PD enzyme activity is temperature dependent.
a) G6PD assay reaction mechanism. b) PreQuine Platform’s unique indicator showing free of interference with Bilirubin and Hemoglobin.
Hgb detection mechanism
The reaction mechanism for Hgb detection is shown in Fig 4. Free Hgb is liberated from the blood sample by mixing with the lysing buffer. After traversing the transport membrane, the sample flows through a polyether sulfone (PES) membrane coated with an oxidizer and sodium azide to convert the free Hgb to azide methemoglobin (Fig 2A, layer 5). The optical membrane is an anisotropic PES membrane facing the LED, which collects the colored reaction product and serves as the Hgb read zone.
Preparation of blood samples
A total of 10 ml of deidentified healthy donor venous blood (heparinized) were purchased from BioIVT (USA) and aliquoted. The native G6PD enzyme activity was assayed by spectrophotometry on a temperature-controlled Genesys 10S (Thermo Scientific, Waltham, MA) at 37°C using commercially available reagents (Cat# G7583180, Pointe Scientific, Canton, MI) and the Hgb value of the blood was determined using the HemoCue HB 201 System (HemoCue, Angelholm, Sweden). G6PD enzyme activity was adjusted via freeze/thaw cycles and/or spiking with recombinant G6PD enzyme (Cat# 078A0020, Calzyme, San Luis Obispo, CA) while Hgb levels were adjusted through the addition and subtraction of plasma. This generated blood samples with G6PD enzyme activity ranging from 0–300 U/dL and with Hgb values ranging from 0–25 g/dL. Prepared samples were assayed in triplicate using the PreQuine Platform and the average values were used to generate a best-fit polynomial equation relating reflectance (%R) to Hgb (g/dL) or G6PD (U/dL). These polynomial equations were used to calculate G6PD and Hgb levels from the %R data. The calculated values were then compared to measurements obtained using reference assays (spectrophotometry with Pointe Scientific reagents and HB 201 HemoCue).
Simultaneous G6PD and Hgb detection
G6PD enzyme activity must be normalized by Hgb concentration to provide a meaningful reading. To evaluate the performance of the PreQuine assay, we encoded the polynomial equations relating %R to G6PD (U/dL) and %R to Hgb (g/dL) into the QR code that is scanned using the PreQuine mobile app (the QR code also contains lot information and expiration date of the test strips) and programmed the results output to be expressed in normalized values of G6PD activity (U/g Hgb). We then obtained twelve de-identified blood samples with varying levels of G6PD enzyme activity (provided by PATH, USA, as G6PD references). The samples were assayed on the PreQuine Platform, HB 201, and using spectrophotometry to compare the G6PD enzyme activity values as determined by the PreQuine Platform with the values obtained from the reference assays.
Results
Performance of PreQuine system for simultaneous G6PD and Hgb measurement
Twelve de-identified blood samples with varying G6PD enzyme activities were assayed simultaneously using the HemoCue HB 201, Pointe Scientific reagents, and the PreQuine system. The linear relationship between the PreQuine Platform and combined HB 201/Pointe Scientific assays is shown in Fig 5. The PreQuine Platform showed a positive and significant linear agreement with the reference assays with a slope of 1.0091, an intercept of -0.0379, and R2 of 0.9936.
Dose response of G6PD enzyme activity using PreQuine platform
Fig 6 shows the dose response in %R of samples across the analytical range of 0–300 U/dL G6PD activity. Regression analysis revealed that a cubic fit polynomial described the dose response with R2 = 0.9638 and an average coefficient of variation of 2% (range 1–5%). The linear regression comparing the PreQuine Platform with the reference assay is shown in Fig 7. The PreQuine Platform showed a positive and significant linear agreement with spectrophotometry, with a slope of 1.0959, an intercept of -3.7386, and R2 of 0.9614.
Dose response of Hgb using PreQuine platform
Fig 8 shows the dose response in %R of samples across the analytical range of 0–25 g/dL Hgb. Regression analysis revealed that a cubic fit polynomial described the dose response with R2 = 0.9933 and an average coefficient of variation of 9% (range 3–18%). The linear regression comparing the PreQuine Platform with HB 201 is shown in Fig 9. The PreQuine Platform showed a positive and significant linear agreement with the HB 201, with a slope of 1.0262, an intercept of -0.1074, and R2 of 0.993.
Discussion
We found the PreQuine system to provide accurate measurements of G6PD activity levels up to 300 U/dL and Hgb values up to 25 g/dL, and a good correlation based on R2 with the corresponding reference methods, but further analysis of a larger sample size is required to determine statistically significant association (spectrophotometry, Pointe Scientific, USA; hemoglobin meter, HemoCue, Sweden). The PreQuine system is quantitative, does not require laboratory equipment, does not require cold chain for distribution or storage, and should remain functional at ambient tropical temperatures, but requires field validation to determine actual performance in vivo which will help us understand the impact of other environmental factors like humidity.
The dual-analyte design omits the need to run two assays in parallel and removes the need for health care providers to calculate G6PD activity. In contrast to other comparable devices the Prequine measures product formation from G6PD activity at 670nm and thereby avoids inference from the Hgb present in the blood sample, where Fe2+ from Hgb can yield the same color signal as the end reaction of the G6PD assay, resulting in an unwanted background signal. Furthermore, the chosen chemical indicator for G6PD detection has an absorbance maximum outside of the absorbance range of bilirubin.
The PreQuine Platform relies on a mobile application that connects the meter device via Bluetooth to a mobile phone. This allows users to store patient ID, test results, and geolocation information, while also allowing for the calculation of normalized G6PD activity in U/g Hgb. This approach reduces the hardware costs for the end user, simplifies software updates worldwide, and provides a real time, step-by-step guide to the user while performing a measurement. The app also provides a platform for additional features that can be added on later such as telehealth consultation, and if requested by the user uploading of test results to centralized databases. While various assays have been developed to determine G6PD deficiency, many of have limited utility in malaria-endemic regions, including: the need for expensive laboratory equipment, cold chain requirements, the need of a separate assay for Hgb quantification, cost, and inadequate assay sensitivities and specificities [12].
The reported results were performed by trained scientists in a laboratory setting, thus far the PreQuine Platform has not been tested under field conditions on samples from malaria patients. It is expected that the observed correlation may differ under more realistic conditions. To assess field feasibility, the PreQuine Platform will be trialed at field sites in Bangladesh, Cambodia, and the United States following standard procedures [13].
Conclusions
Initial laboratory-based data suggest that the PreQuine device shows good correlation across the physiological range of 0–300 U/dL G6PD activity and 0–25 g/dL Hgb, when compared to the respective reference assays. Field studies will be critical to validate these findings, assess the overall device performance, monitor user friendliness and acceptance, and eventually examine PreQuine performance in comparison to other commercially available point-of-care G6PD devices.
References
- 1.
World Health Organization. WHO guidelines for malaria, 14 March 2023 [Internet]. World Health Organization; 2023 [cited 2023 Mar 30]. Report No.: WHO/UCN/GMP/2023.01. Available from: https://apps.who.int/iris/handle/10665/366432.
- 2. Battle KE, Lucas TCD, Nguyen M, Howes RE, Nandi AK, Twohig KA, et al. Mapping the global endemicity and clinical burden of Plasmodium vivax, 2000–17: a spatial and temporal modelling study. The Lancet. 2019 Jul;394(10195):332–43. pmid:31229233
- 3. Commons RJ, Simpson JA, Watson J, White NJ, Price RN. Estimating the Proportion of Plasmodium vivax Recurrences Caused by Relapse: A Systematic Review and Meta-Analysis. Am J Trop Med Hyg. 2020 Sep;103(3):1094–9. pmid:32524950
- 4. Edgcomb JH, Arnold J, Yount EH, Alving AS, Eichelberger L, Jeffery GM, et al. Primaquine, SN 13272, a new curative agent in vivax malaria; a preliminary report. J Natl Malar Soc. 1950 Dec;9(4):285–92. pmid:14804087
- 5. Taylor WRJ, Thriemer K, von Seidlein L, Yuentrakul P, Assawariyathipat T, Assefa A, et al. Short-course primaquine for the radical cure of Plasmodium vivax malaria: a multicentre, randomised, placebo-controlled non-inferiority trial. The Lancet. 2019 Sep;394(10202):929–38. pmid:31327563
- 6. Shanks GD, Oloo AJ, Aleman GM, Ohrt C, Klotz FW, Braitman D, et al. A New Primaquine Analogue, Tafenoquine (WR 238605), for Prophylaxis against Plasmodium falciparum Malaria. CLIN INFECT DIS. 2001 Dec 15;33(12):1968–74. pmid:11700577
- 7. Phru CS, Kibria MG, Thriemer K, Chowdhury MU, Jahan N, Aktaruzzaman MM, et al. Case Report: A Case of Primaquine-Induced Hemoglobinuria in Glucose-6-Phosphate Dehydrogenase Deficient Malaria Patient in Southeastern Bangladesh. The American Journal of Tropical Medicine and Hygiene. 2020 Jan 8;102(1):156–8. pmid:31701865
- 8. Chu CS, Bancone G, Nosten F, White NJ, Luzzatto L. Primaquine-induced haemolysis in females heterozygous for G6PD deficiency. Malar J. 2018 Dec;17(1):101. pmid:29499733
- 9.
World Health Organization. Control and elimination of Plasmodium vivax malaria: a technical brief. 27 August 2015. Geneva: World Health Organization; 2023 [cited 2023 Mar 30]. Available from: https://www.who.int/publications/i/item/9789241509244.
- 10. Gething PW, Elyazar IRF, Moyes CL, Smith DL, Battle KE, Guerra CA, et al. A Long Neglected World Malaria Map: Plasmodium vivax Endemicity in 2010. Carlton JM, editor. PLoS Negl Trop Dis. 2012 Sep 6;6(9):e1814. pmid:22970336
- 11. Ley B, Winasti Satyagraha A, Kibria MG, Armstrong J, Bancone G, Bei AK, et al. Repeatability and reproducibility of a handheld quantitative G6PD diagnostic. PLoS Negl Trop Dis. 2022 Feb 17;16(2):e0010174. pmid:35176015
- 12. Weppelmann TA, von Fricken ME, Wilfong TD, Aguenza E, Philippe TT, Okech BA. Field Trial of the CareStart Biosensor Analyzer for the Determination of Glucose-6-Phosphate Dehydrogenase Activity in Haiti. Am J Trop Med Hyg. 2017 Oct;97(4):1262–1270. pmid:28820691
- 13. Ley B, Bancone G, von Seidlein L, Thriemer K, Richards JS, Domingo GJ, et al. Methods for the field evaluation of quantitative G6PD diagnostics: a review. Malar J. 2017 Sep 11;16(1):361. pmid:28893237