Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

  • Loading metrics

Detection of BRAF mutations in malignant melanoma and colorectal cancer by SensiScreen® FFPE BRAF qPCR assay

  • Anna Lahn Sørensen ,

    Roles Validation, Visualization, Writing – original draft, Writing – review & editing

    als@pentabase.com

    Affiliation PentaBase A/S, Odense, Denmark

  • Mariann Guldmann-Christensen,

    Roles Data curation, Investigation, Resources, Writing – original draft, Writing – review & editing

    Affiliation Department of Pathology, Laboratory of Research and Development, Aarhus University Hospital, Aarhus, Denmark

  • Michael Børgesen,

    Roles Formal analysis, Methodology, Writing – original draft, Writing – review & editing

    Affiliation PentaBase A/S, Odense, Denmark

  • Rasmus Koefoed Petersen,

    Roles Data curation, Formal analysis, Validation, Visualization, Writing – review & editing

    Affiliation PentaBase A/S, Odense, Denmark

  • Katharina Flugt,

    Roles Visualization

    Affiliation PentaBase A/S, Odense, Denmark

  • Julie Mejer Holmgaard Duelund,

    Roles Data curation, Investigation, Resources

    Affiliation Department of Pathology, Laboratory of Research and Development, Aarhus University Hospital, Aarhus, Denmark

  • Majbritt Hauge Kyneb,

    Roles Investigation, Methodology, Resources

    Affiliation Danish Technological Institute, Aarhus, Denmark

  • Jan Lorenzen,

    Roles Conceptualization, Funding acquisition, Project administration

    Affiliation Danish Technological Institute, Aarhus, Denmark

  • Emma Pipó-Ollé,

    Roles Data curation, Investigation

    Affiliation PentaBase A/S, Odense, Denmark

  • Samantha Epistolio,

    Roles Data curation, Investigation, Resources, Writing – review & editing

    Affiliation Institute of Pathology EOC, Ente Ospedaliero Cantonale, Locarno, Switzerland

  • Alice Riva,

    Roles Data curation, Investigation, Resources

    Current address: Clinical Chemistry and Immunometry Department, Ospedale Generale di Zona Moriggia Pelascini, Gravedona, Italy

    Affiliation Institute of Pathology EOC, Ente Ospedaliero Cantonale, Locarno, Switzerland

  • Giulia Dazio,

    Roles Data curation, Investigation, Resources, Writing – review & editing

    Affiliation Institute of Pathology EOC, Ente Ospedaliero Cantonale, Locarno, Switzerland

  • Elisabetta Merlo,

    Roles Resources

    Affiliation Institute of Pathology EOC, Ente Ospedaliero Cantonale, Locarno, Switzerland

  • Tine Meyer,

    Roles Data curation, Investigation, Resources

    Affiliation Department of Pathology, Laboratory of Research and Development, Aarhus University Hospital, Aarhus, Denmark

  • Ulf Bech Christensen,

    Roles Conceptualization, Project administration

    Affiliation PentaBase A/S, Odense, Denmark

  •  [ ... ],
  • Milo Frattini

    Roles Conceptualization, Funding acquisition, Project administration, Supervision, Writing – review & editing

    Affiliation Institute of Pathology EOC, Ente Ospedaliero Cantonale, Locarno, Switzerland

  • [ view all ]
  • [ view less ]

Abstract

Mutations in BRAF exon 15 lead to conformational changes in its activation loops, resulting in constitutively active BRAF proteins which are implicated in the development of several human cancer types. Different BRAF inhibitors have been developed and introduced in clinical practice. Identification of BRAF mutations influences the clinical evaluation, treatment, progression and for that reason a sensitive and specific identification of BRAF mutations is on request from the clinic. Here we present the SensiScreen® FFPE BRAF qPCR Assay that uses a novel real-time PCR-based method for BRAF mutation detection based on PentaBases proprietary DNA analogue technology designed to work on standard real-time PCR instruments. The SensiScreen® FFPE BRAF qPCR Assay displays high sensitivity, specificity, fast and easy-to-use. The SensiScreen® FFPE BRAF qPCR Assay was validated on two different FFPE tumour biopsy cohorts, one cohort included malignant melanoma patients previously analyzed by the Cobas® 4800 BRAF V600 Mutation Test, and one cohort from colorectal cancer patients previously analyzed by mutant-enriched PCR and direct sequencing. All BRAF mutant malignant melanoma patients were confirmed with the SensiScreen® FFPE BRAF qPCR Assay and additional four new mutations in the malignant melanoma cohort were identified. All the previously identified BRAF mutations in the colorectal cancer patients were confirmed, and additional three new mutations not identified with direct sequencing were detected. Also, one new BRAF mutation not previously identified with ME-PCR was found. Furthermore, the SensiScreen® FFPE BRAF qPCR Assay identified the specific change in the amino acid. The SensiScreen® FFPE BRAF qPCR Assay will contribute to a more specific, time and cost saving approach to better identify and characterize mutations in patients affected by cancer, and consequently permits a better BRAF characterization that is fundamental for therapy decision.

1. Introduction

In 2002 Davies and colleagues demonstrated that the occurrence of a specific mutation in exon 15, codon 600 (formerly identified as 599) can constitutively activate the BRAF (B-rapidly accelerated fibrosarcoma protein) [1]. Codon 600 encodes a part of the activation loop in the kinase domain of the BRAF protein, and mutations in this region result in a substitution of valine by different amino acids, such as glutamate (V600E), arginine (V600R), lysine (V600K) or aspartate (V600D). These substitutions, especially the V600E, can lead to an abnormal increase in the catalytic activity of the BRAF protein by inducing conformational changes or mimicking a change in the phosphorylation status in its activation loop and leading to a BRAF protein with constitutively active kinase activity. As a consequence of these mutations, the BRAF -MEK (Mitogen-activated protein kinase kinase)–ERK (Extracellular signal-regulated kinase)—pathway is constitutively active, leading to uncontrolled cell growth, cell division and subsequently cancer development [13]. Almost all mutations in the BRAF gene occur at the hotspot codon 600 and roughly 90% are represented by the V600E change. However, although rarer, non-V600E mutations like V600K, V600D and V600R substitutions are observed as well [1].

BRAF mutations have been described in about 50% of malignant melanoma (MM) [4], 50% of papillary thyroid carcinoma, 5–10% of colorectal cancer (CRC) [5], and, in a lesser extent, in lung adenocarcinomas about 2–4% of cases and low-grade glioma especially the pilocytic astrocytoma subtype [1, 612]. In MM and CRC, BRAF mutations play the most relevant clinical role. Even in lung adenocarcinoma BRAF mutations play a clinical role, although limited by the low frequency [13].

At the clinical level, the treatment of MM has greatly been improved by the development of selective inhibitors against BRAF mutations in codon 600 (e.g., vemurafenib) [14]. To initiate a treatment with targeted therapy in patients with stage IV metastatic melanoma the identification of a BRAF mutation in codon 600 is mandatory [15]. The Food and Drug Administration (FDA) approved BRAF inhibitors for patients harboring the BRAF V600E mutation only, while the European Medicine Agency (EMA) for all patients with a mutation in the same codon. The identification of the BRAF mutation is clinically relevant since these patients are more unlikely to undergo metastasectomy as the metastases are present at sites not assessable for surgery, usually peritoneal disease, or distant lymph node metastasis [16, 17]. Moreover, identification of the BRAF V600E mutation in patients with metastatic CRC is associated with a poor prognosis for obtaining modest benefit from treatments with chemotherapy, particularly in the second- and third-line settings compared with BRAF WT cases [3, 5, 16]. An approved treatment of CRC is administration of monoclonal antibodies against the epidermal growth factor receptor (EGFR) (e.g., cetuximab and panitumumab). These antibodies bind the extracellular domain on EGFR and inhibit ligand binding and dimerization, rendering the downstream signaling pathway unsusceptible for activation. It is believed that BRAF mutations are responsible for 12–15% of patients who fail anti-EGFR treatment, similarly to what demonstrated for KRAS/NRAS mutations. Therefore, it is recommended only to use anti-EGFR drugs in patients whose tumours display BRAF/KRAS/NRAS WT sequences [16, 18].

At present, there exist various methodologies to analyze the BRAF mutations. The cheapest is immunohistochemistry. However, this method is only able to identify the V600E change [19, 20]. Further methodologies include direct sequencing (DS), next-generation sequencing (NGS), mass spectrometry and real-time PCR based methods such as Cobas® BRAF mutation test (Roche), Therascreen BRAF PCR kit (Qiagen), and Idylla™ BRAF Mutation Test (Biocartis), overall characterized by marked differences in limit-of-detection (LOD): 1–10%, down to 5%, around 5%, 0.5–1% and 1%, respectively. Before choosing the methodology for BRAF evaluation, handling time, time to answer, skills, the expertise of the technical personnel required for performing the analysis and finally, costs should be considered. All these aspects may have a significant impact in the clinical management of patients in terms of proper diagnosis, classification and, in the worst case, in terms of missing the opportunity of a targeted treatment.

In the present study, we report the development and testing of a real-time PCR assay based on a new technology, for the identification of mutations in the BRAF gene on formalin-fixed paraffin-embedded (FFPE) tissue, named SensiScreen® FFPE BRAF qPCR Assay. The SensiScreen® FFPE BRAF qPCR Assays are based on the PentaBase DNA analogue technology, where oligoes are modified with intercalator, also known as pentabase. Oligoes with pentabases are defined as intercalating nucleic acid (INA®). Electrons from the pentabase are participating in the stacking of DNA duplexes [21, 22]. The samples were tested by the SensiScreen® FFPE BRAF qPCR simplex and multiplex assays and the results compared to DS, mutant-enriched PCR (ME-PCR) or to the Cobas® 4800 BRAF V600 Mutation Test (Roche). The overall purpose of this study was to validate a highly sensitive, specific, short turnaround time, low cost and easy-to-use assay for BRAF characterization able to work on the most common real-time PCR instruments. Furthermore, the SensiScreen® FFPE BRAF qPCR Assay was developed to use a minimal input of DNA.

2. Material and methods

2.1. Cell lines

The cell lines used for the sensitivity studies contained four different mutations in BRAF codon 600 (V600E, V600K, V600D and V600R). The sensitivity tests were conducted using cell line DNA extracted from HT-29 (ATCC®; catalogue number: HTB-38TM), IGR-1 (DSMZ®; catalogue number: ACC 236TM), WM-115 (Rockland Immunochemicals®; catalogue number: WM115-01-0001TM) and HCT116 BRAF (V600R/+) (Horizon Discovery Group®; catalogue number: HD 104-037TM) (Table 1). Subcultures were made in appropriate media according to the manufacturer’s instructions and genomic DNA was isolated using the QIAmp Mini kit (Qiagen, Chatsworth, CA, USA).

thumbnail
Table 1. Features of the cell lines used for the sensitivity studies.

https://doi.org/10.1371/journal.pone.0281558.t001

2.2. Sensitivity studies

LOD of the SensiScreen® FFPE BRAF qPCR Assay was evaluated by serial dilutions of cell line DNA, harboring BRAF mutations, in a background of 50 ng WT human genomic DNA (Promega, Madison, Wisconsin, USA). Five different concentrations of mutated DNA were tested (10%, 5%, 2%, 1% and 0.5%) on MyGo Pro real-time PCR instruments (IT-IS Life Science Ltd, Mahon, Ireland) and Rotor-Gene 6000 (Corbett Research), using the SensiScreen® protocol previously described [22]. The SensiScreen® FFPE BRAF qPCR Assay limit of blank (LOB) was evaluated by application of 20 replica of 50 ng WT human genomic DNA (Promega, Madison, Wisconsin, USA).

2.3. SensiScreen® FFPE BRAF qPCR assay

To develop a real-time PCR assay able to specifically amplify few copies of mutated DNA in a large WT background we took advantage of PentaBases intercalating nucleic acid technology. The development of the SensiScreen® FFPE BRAF qPCR Assay was performed as previously described [22].

2.4. Patient samples

Patients included in this study were diagnoses with MM (cohort recruited at the Department of Pathology, Aarhus University Hospital, Denmark) or with CRC (cohort recruited at the Institute of Pathology EOC, Locarno, Switzerland). Only patients with sufficient tumour tissue material have been included. The tumour materials used in the study were surplus to requirements for routine testing and were obtained after surgery and clinicopathological assessment of the tumour as part of the standard clinical procedure. All data and materials were fully anonymized before assessment. This study was approved by the Institutional Ethical Committee of the Institute of Pathology EOC of Locarno, Switzerland, by the Central Denmark Region Committee on Health Research Ethics, and by the Danish Data Protection Agency. All procedures were performed in accordance with the ethical standards of the Helsinki Declaration.

2.4.1. Tissue analyses.

Quality and tumour cell content of the FFPE tissue block sections were evaluated by an experienced pathologist. Genomic DNA was extracted from representative FFPE tissue blocks, containing at least 50–70% neoplastic cells. In cases of low tumour tissue content, a macrodissection was performed to enrich the amount of tumour cells. Genomic DNA was extracted from a 5 μm thick section of MM tissue, using the Cobas® DNA Sample Preparation Kit (Roche) according to the manufacturer’s instructions. In CRC samples, genomic DNA was extracted from six 7 μm thick serial FFPE sections. DNA extraction was performed using the QIAmp Mini kit (Qiagen) according to the manufacturer’s instructions.

2.4.2. Cohort 1.

The first cohort included 127 patients diagnosed with histologically confirmed MM collected from 2012 to 2013 at the Department of Pathology, Aarhus University Hospital, Denmark. BRAF exon 15 was retrospectively analyzed using the Cobas® 4800 BRAF V600 Mutation Test (Roche) and the SensiScreen® FFPE BRAF qPCR Simplex and Multiplex Assays (PentaBase). Four patients were excluded from both tests, due to insufficient DNA. Clinical-pathological features are included in S1 Table.

2.4.3. Cohort 2.

The second cohort included 100 patients affected by histologically confirmed CRC and collected from 1996 to 2009 at the Institute of Pathology EOC of Locarno, Switzerland. The samples were retrospectively analyzed for BRAF exon 15 mutations by DS, ME-PCR and the SensiScreen® FFPE BRAF qPCR Simplex and Multiplex assays. Three patients were excluded from the SensiScreen® FFPE BRAF qPCR assay due to long term storage, resulting in insufficient and invalid DNA. Clinical-pathological features are included in S1 Table.

2.5. Mutational analysis

2.5.1. Direct sequencing

The DS analyses were performed at the Institute of Pathology EOC (Locarno, Switzerland) using the primers listed in Table 2. The samples were subjected to automated sequencing on an ABI PRISM 3130 Genetic Analyzer (Applied Biosystems, Foster City, CA, USA), and evaluated using the Sequencing Navigator Software (Applied Biosystems). The identified mutations were confirmed in two independent PCR reactions run on ProFlex PCR System (ThermoFisher Scientific, Waltham, MA, USA).

thumbnail
Table 2. Sequences of oligonucleotides used for construction of model templates for DS, ME-PCR and SensiScreen assays.

https://doi.org/10.1371/journal.pone.0281558.t002

2.5.2. Mutant-enriched PCR.

Analysis of BRAF exon 15 mutations by ME-PCR was performed at the Institute of Pathology EOC (Locarno, Switzerland), essentially as described previously [15, 23, 24] ME-PCR allows the enrichment of mutant alleles and the elimination of WT alleles by application of a restriction enzyme digestion on the PCR amplification product [2527]. The ME-PCR products were subsequently subjected to automated sequencing and analyzed as described above. All mutated cases were confirmed twice in independent PCR reactions.

2.5.3. Cobas® BRAF mutation test.

This BRAF exon 15 mutations analysis was performed at the Department of Pathology at Aarhus University Hospital (Denmark) using the Cobas® 4800 BRAF V600 Mutation Test on a Cobas® 480Z analyzer. DNA extraction was conducted as previously described [22]. The DNA concentration was measured using an Implen Nanophotometer (Germany). Subsequently, the DNA was diluted to a concentration of 5 ng/μl in water. Twenty-five μl mixture was added in each reaction tube.

2.5.4. SensiScreen® FFPE BRAF qPCR assay.

The SensiScreen® FFPE BRAF qPCR Simplex and Multiplex mutation analysis on MM and CRC were performed at the Department of Pathology, Aarhus University Hospital, Denmark and at the Institute of Pathology, Locarno, EOC, Switzerland, respectively.

Development of the SensiScreen® FFPE BRAF qPCR Assay for MM and CRC was performed on a Mx3005P qPCR system (Stratagene, CA, USA) and a CFX96 (Bio-Rad, Hercules, CA, USA), respectively. The reaction was conducted in 25 μL with Ampliqueen 2xMaster Mix (PentaBase, Odense, Denmark), 300–900 nM of each SuPrimers, 200 nM of each HydrolEasy probes, and 1000–5000 nM WT BaseBlocker. All SensiScreen® FFPE BRAF qPCR Assays contained an internal control targeting the CYP17A1 gene. The multiplex assay included SuPrimers for detection of V600E, V600D, V600K and V600R mutations in a single tube, and was used for the initial screening of BRAF mutations (Table 2). The individual V600 mutation was subsequently specified when using the simplex assay–each tube containing a single BRAF mutation. In vitro specificities were analyzed using 50 ng of cell line DNA or 50 ng human genomic WT DNA (Promega G304A) in the absence or presence of 1000 copies of mutated plasmid DNA carrying BRAF mutations. FFPE purified DNA was diluted to a concentration of 1–10 ng/ μL in water. Five μL was added to each reaction tube. The PCR program was: 2 min at 95°C, followed by 45 cycles of 15 sec at 94°C and 60 sec at 60°C.

2.5.5. Data processing.

The threshold cycle (Ct) was defined as 10% of the signal strength (RFU) of the reference assay at cycle 45. The ΔCt value was calculated for each mutation specific analysis and defined as the difference between the Ct value from the given mutation analysis subtracted the Ct value from the corresponding reference analysis. ΔCt was calculated for all mutation assays having a Ct value beneath 39.

The ΔCt cut-off was defined as the value that will avoid false positive results from amplification of WT samples and at the same time identify as many true positive samples as possible. Since a ΔCt at 12.1 was the lowest value identified when using WT DNA as template for mutation assay, we considered a ΔCt of 9 as the optimal value for subsequent analysis of assay sensitivities. A sample was defined as positive for a given mutation if Ct of the mutation assay was beneath 39 and the ΔCt value beneath 9.

To determine the LOD, a serial dilution experiments were made in duplicate for each plasmid to achieve concentrations from 16.000 to 0 copies/reaction. The serial dilutions were used to establish a calibration curve for each mutated DNA. The LODs values were determined and calculated from the calibration curve, where y is Ctreference +9. This was determined for both simplex and multiplex. PCR amplification efficiency was determined from the slope of the log-linear portion of the calibration curve (Efficiency = 10−1/slope-1).

3. Results

3.1. SensiScreen® FFPE BRAF qPCR assay sensitivity

To evaluate the sensitivity of the SensiScreen® FFPE BRAF qPCR Simplex and Multiplex assays, a serial dilution of cell line DNA harboring BRAF mutations in a background of human genomic WT DNA was performed in duplicate. The LOD for SensiScreen® FFPE BRAF qPCR Simplex ranges from 0.55% to 1.04%, with V600R showing the lowest value (Table 3). Looking at the SensiScreen® FFPE BRAF qPCR Multiplex assay the range is from 0.66% to 1.81% with V600D and V600E showing the lowest value. V600E shows identical LODs in SensiScreen® FFPE BRAF qPCR Simplex and Multiplex. The equation for the dilution series curve for alle PCR assays showed that the PCR amplification efficiencies and correlation coefficient (R2) were within the acceptable range. The efficiency ranged between 77% and 109%, and the R2 values were all greater than 0.96. Calibration curves for each target are included in the S1 Fig. SensiScreen® FFPE BRAF qPCR Assays LOB was evaluated for each variant by 20 replicas of human genomic WT DNA. Mean ΔCt and standard deviation for each SensiScreen® FFPE BRAF qPCR is stated in Table 3.

thumbnail
Table 3. Sensitivity and efficiency of SensiScreen® FFPE BRAF qPCR assay on mutated DNA in WT background.

https://doi.org/10.1371/journal.pone.0281558.t003

3.2. Clinical performance of the SensiScreen® FFPE BRAF qPCR assay

To validate the SensiScreen® FFPE BRAF qPCR Assay in a clinical setting, the assay was tested on DNA from FFPE tumour tissues from two different cohorts, S1 Table. In cohort 1, FFPE tumour biopsies from 127 Danish MM patients, previously analyzed by the Cobas® 4800 BRAF V600 Mutation Test were retrospectively analyzed by the SensiScreen® FFPE BRAF qPCR Simplex and Multiplex assays. In cohort 2, 100 FFPE tumour specimens from CRC patients previously analyzed by ME-PCR and DS were retrospectively tested by the SensiScreen® FFPE BRAF qPCR simplex Assay. In general, there was a good separation between the WT and the mutant group in both cohorts when using the SensiScreen® FFPE BRAF qPCR Assay analysis settings defined as positive for a given mutation if Ct of the mutation assay was beneath 39 and the ΔCt value beneath 9, Fig 1.

thumbnail
Fig 1. Dot plot of SensiScreen® FFPE BRAF qPCR assay clinical data.

(A) SensiScreen® FFPE BRAF qPCR multiplex assay applied to the MM cohort. BRAF Multiplex positive samples were BRAF V600 mutation subtyped by SensiScreen BRAF V600 simplex. (B) SensiScreen® FFPE BRAF qPCR Multiplex assay applied to the CRC cohort. Samples not resulting in a Ct value is not represented in the figures. The ΔCt (≤ 9) and Ct values (≤ 39) are indicated by the dotted line. Abbreviations; CRC: Colorectal cancer; MM: malignant melanoma; Ct: Threshold cycle.

https://doi.org/10.1371/journal.pone.0281558.g001

The SensiScreen® FFPE BRAF qPCR Assay confirmed the same mutations already identified by Cobas® 4800, ME-PCR and DS (Table 4). Furthermore, in cohort 1, the SensiScreen® FFPE BRAF qPCR Assays (both simplex and multiplex) identified four mutations not already identified with Cobas®, corresponding to an additional identification of 8% mutated cases at the MM cohort. Beside identification of not previously detected mutations, the SensiScreen® FFPE BRAF qPCR Assay also identified specific changes in amino acids, these variants were the V600K in two cases, and the V600E and the V600R in one case each (no V600D variants were identified). In cohort 2 the SensiScreen® FFPE BRAF qPCR Assay identified three new mutations not previously found with DS, all represented by a V600E change. Identification of the three new mutations corresponds to an additional 50% mutated cases in the CRC cohort not already identified with DS. The SensiScreen® FFPE BRAF qPCR Assay also identified a new mutation not previously identified with ME-PCR, corresponding to an additional 12.5% mutated cases in the CRC cohort not already identified with ME-PCR.

thumbnail
Table 4. Clinical performance of SensiScreen® FFPE BRAF qPCR assays in comparison to Cobas® 4800, ME-PCR and DS.

https://doi.org/10.1371/journal.pone.0281558.t004

4. Discussion

A high sensitivity and specificity for detection and identification of V600 mutations in the BRAF gene are of great importance, since it has essential consequences for the choice of proper treatment of the patients. Indeed, the characterization of this genetic loci helps to predict the response to the treatment and consequently avoid the unnecessary administration of expensive therapies related to huge side effects and decreasing the quality of life of the patients. An increase in the technology capability for identification of genetic variants occurring during tumourigenesis will increase our knowledge in the field of molecular biology, allowing in the end a better personalized treatment strategy. Many different BRAF test options are available in the clinical, such as immunohistochemistry, real-time PCR, sequencing, or next-generation sequencing (NGS). Direct sequencing is becoming a less commonly used method today, due to its high LOD. The National Comprehensive Cancer Network (NCCN) guidelines recommend BRAF testing in patients where targeted therapy is a possibility. NCCN guidelines does not recommend any specific methodology for BRAF detection [28]. The commercially available qPCR mutations tests, such as Cobas 4800 BRAF V600 mutation test, THxID BRAF kit (bioMérieux), and Therascreen BRAF V600E RGQ PCR kit cannot detect mutations levels below 1%. Further, several methods for BRAF mutations test do not discriminate between the V600E/D/K/R variants. A poor LOD may results in false negatives that deprive patients from optimal treatment.

In this work we have demonstrated how our technology makes it possible to discriminate between the V600 variants obtaining results with decreased LOD compared to other assays on the market. The comparison between other methods for BRAF mutation testing and the SensiScreen® FFPE BRAF qPCR Assay showed a high concordance and, in addition, demonstrated that the SensiScreen® FFPE BRAF qPCR Assay can identify more BRAF mutant cases. Indeed, we found additional mutations on MM patients and on CRC patients when compared to Cobas® 4800 BRAF V600 Mutation Test (Roche), ME-PCR and DS, respectively. Cobas 4800 is, according to the specification, able to detect V600E, V600D and V600K. The SensiScreen® FFPE BRAF qPCR Assay was able to identify one V600R, two V600K and one V600E not detected by Cobas® 4800 BRAF V600. Furthermore, The SensiScreen® FFPE BRAF qPCR Assay was able to identify one V600E not detected with ME-PCR, and three not detected with DS. We have designed and developed a novel BRAF assay for the analysis of DNA extracted from FFPE tissues, possessing high sensitivity and high specificity, which is suitable for use in daily laboratory routine. In addition, the short turnaround time for the SensiScreen® FFPE BRAF qPCR Assay is saving time and makes it attractive for clinical decision making. There is no need of expensive and time-consuming equipment with the requirement of special trained personnel since SensiScreen® FFPE BRAF qPCR Assay is low costs (similar to other PCR based methods)

The developed SensiScreen® FFPE BRAF qPCR Assay is of relevance especially in the case of genes with a hotspot site, such as codon 600 of the BRAF gene. More interestingly, the SensiScreen® FFPE BRAF qPCR Assay is able to identify not only the V600E variant, but also other changes that are relevant for MM patients.

In conclusion, the SensiScreen® FFPE BRAF qPCR Assay will permit to have a more specific, time and cost saving approach to better characterize, and consequently to better treat, patients affected by MM, CRC and the other cancer types, for which BRAF characterization is fundamental for therapy decision.

Supporting information

S1 Fig. Calibration curves of SensiScreen® FFPE BRAF qPCR assay for each target Ct value as function of log (copy number).

https://doi.org/10.1371/journal.pone.0281558.s001

(DOCX)

S1 Table. Clinical-pathological characteristics of cohort 1 and 2 used for analysis by Cobas® 4800, ME-PCR and DS.

Abbreviations; F: female; M: male.

https://doi.org/10.1371/journal.pone.0281558.s002

(DOCX)

Acknowledgments

The authors declare that there is none to list in the acknowledgments.

References

  1. 1. Davies H, Bignell GR, Cox C, Stephens P, Edkins S, Clegg S, et al. Mutations of the BRAF gene in human cancer. 2002. Available: www.nature.com/nature
  2. 2. Yaeger R, Corcoran RB. Targeting alterations in the RAF–MEK pathway. Cancer Discovery. American Association for Cancer Research Inc.; 2019. pp. 329–341. https://doi.org/10.1158/2159-8290.CD-18-1321 pmid:30770389
  3. 3. Kanaty O, Ertas H, Caner B. Contemporary treatment approaches for metastatic colorectal cancer driven by BRAF V600 mutations. World J Gastrointest Oncol. 2020;12.
  4. 4. Abraham J, Stenger M. Dabrafenib in advanced melanoma with BRAF V600E mutation. Journal of Community and Supportive Oncology. Frontline Medical Communications; 2014. pp. 48–49. pmid:24971404
  5. 5. Tol J, Nagtegaal I, Punt C. BRAF mutation in metastatic colorectal cancer. N Engl J Med. 2009. pmid:19571295
  6. 6. BRAF V600E Mutation Identifies a Subset of Low-Grade Diffusely Infiltrating Gliomas in Adults. 2013. Available: www.jco.org
  7. 7. Menzies AM, Long G v. Dabrafenib and Trametinib, alone and in combination for BRAF-Mutant metastatic melanoma. Clinical Cancer Research. 2014;20: 2035–2043. pmid:24583796
  8. 8. Olszanski AJ. Current and Future Roles of Targeted Therapy and Immunotherapy in Advanced Melanoma. Journal of Managed Care Pharmacy JMCP April. 2014. Available: www.amcp.org pmid:24684639
  9. 9. Cohen R, Cervera P, Svrcek M, Pellat A, Dreyer C, de Gramont A, et al. BRAF-Mutated Colorectal Cancer: What Is the Optimal Strategy for Treatment? Curr Treat Options Oncol. 2017;18: 9. pmid:28214977
  10. 10. Rotow J, Bivona TG. Understanding and targeting resistance mechanisms in NSCLC. Nat Rev Cancer. 2017;17: 637–658. pmid:29068003
  11. 11. Caputo F, Santini C, Bardasi C, Cerma K, Casadei-Gardini A, Spallanzani A, et al. BRAF-mutated colorectal cancer: Clinical and molecular insights. International Journal of Molecular Sciences. MDPI AG; 2019. pmid:31661924
  12. 12. O’Leary CG, Andelkovic V, Ladwa R, Pavlakis N, Zhou C, Hirsch F, et al. Targeting BRAF mutations in non-small cell lung cancer. Translational Lung Cancer Research. AME Publishing Company; 2019. pp. 1119–1124. pmid:32010589
  13. 13. Roviello G, D’angelo A, Sirico M, Pittacolo M, Felipe &, Conter U, et al. Advances in anti-BRAF therapies for lung cancer.
  14. 14. Chapman PB, Hauschild A, Robert C, Haanen JB, Ascierto P, Larkin J, et al. Improved Survival with Vemurafenib in Melanoma with BRAF V600E Mutation. New England Journal of Medicine. 2011;364: 2507–2516. pmid:21639808
  15. 15. Cheng LY, Haydu LE, Song P, Nie J, Tetzlaff MT, Kwong LN, et al. High sensitivity sanger sequencing detection of BRAF mutations in metastatic melanoma FFPE tissue specimens. Sci Rep. 2021;11: 9043. pmid:33907234
  16. 16. Clarke CN, Kopetz ES. BRAF mutant colorectal cancer as a distinct subset of colorectal cancer: Clinical characteristics, clinical behavior, and response to targeted therapies. Journal of Gastrointestinal Oncology. Pioneer Bioscience Publishing; 2015. pp. 660–667. https://doi.org/10.3978/j.issn.2078-6891.2015.077 pmid:26697199
  17. 17. Wang X, Wei Q, Gao J, Li J, Li J, Gong J, et al. Clinicopathologic features and treatment efficacy of Chinese patients with BRAF-mutated metastatic colorectal cancer: a retrospective observational study. Chin J Cancer. 2017;36. pmid:29037218
  18. 18. di Nicolantonio F, Martini M, Molinari F, Sartore-Bianchi A, Arena S, Saletti P, et al. Wild-type BRAF is required for response to panitumumab or cetuximab in metastatic colorectal cancer. Journal of Clinical Oncology. 2008;26: 5705–5712. pmid:19001320
  19. 19. Orchard G, Wojcik K, Rickaby W, Martin B, Semkova K, Shams F, et al. Immunohistochemical detection of V600E BRAF mutation is a useful primary screening tool for malignant melanoma. Br J Biomed Sci. 2019;76. pmid:30870099
  20. 20. Spagnolo F, Ghiorzo P, Orgiano L, Pastorino L, Picasso V, Tornari E, et al. BRAF-mutant melanoma: Treatment approaches, resistance mechanisms, and diagnostic strategies. OncoTargets and Therapy. Dove Medical Press Ltd.; 2015. pp. 157–168. pmid:25653539
  21. 21. Christensen UB, Pedersen EB. Intercalating nucleic acids containing insertions of 1-O-(1-pyrenylmethyl)glycerol: stabilisation of dsDNA and discrimination of DNA over RNA. pmid:12433995
  22. 22. Riva A, BØrgesen M, Guldmann-Christensen M, Kyneb MH, Voogd K, Andersen C, et al. SensiScreen® KRAS exon 2-sensitive simplex and multiplex real-time PCR-based assays for detection of KRAS exon 2 mutations. PLoS One. 2017;12. pmid:28636636
  23. 23. Bearzatto A, Conte D, Frattini M, Zaffaroni N, Andriani F, Balestra D, et al. p16 INK4A Hypermethylation Detected by Fluorescent Methylation-specific PCR in Plasmas from Non-Small Cell Lung Cancer 1. 2002.
  24. 24. Nollau P, Moser C, Weinland G, Wagener C. DETECTION OF K-ras MUTATIONS IN STOOLS OF PATIENTS WITH COLORECTAL CANCER BY MUTANT-ENRICHED PCR. Int J Cancer. Wiley-Liss, Inc; 1996.
  25. 25. Mitsudomi T, Viallet J, Mulshine J, Linnoila R, Minna J, Gazdar A. Mutations of ras genes distinguish a subset of non-small-cell lung cancer cell lines from small-cell lung cancer cell lines. Oncogene. 1991; 1353–1362. pmid:1679529
  26. 26. Kahn S, Jiang W, Culbertson T, Weinstein I, Williams G, Tomita N, et al. Rapid and sensitive nonradioactive detection of mutant K-ras genes via “enriched” PCR amplification. Oncogene. pmid:1676837
  27. 27. Chen J, Viola M. A method to detect ras point mutations in small subpopulations of cells. Anal Biochem. 1991;195: 51–56. pmid:1679614
  28. 28. Benson AB, Al-Hawary MM, Azad N, Chen Y-J, Ciombor KK, Cohen S, et al. NCCN Guidelines Version 2.2022 Colon Cancer Continue NCCN Guidelines Panel Disclosures. 2022. Available: https://www.nccn.org/home/member-