Evaluation of quantitative accuracy among different scatter corrections for quantitative bone SPECT/CT imaging

  • Kenta Miwa ,

    Roles Conceptualization, Writing – original draft

    kenta5710@gmail.com

    Affiliations Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan, Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Reo Nemoto,

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition

    Affiliation Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Hirotsugu Masuko,

    Roles Data curation, Formal analysis

    Affiliation Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Tensho Yamao,

    Roles Formal analysis

    Affiliations Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan, Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Rinya Kobayashi,

    Roles Data curation

    Affiliation Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Noriaki Miyaji,

    Roles Formal analysis

    Affiliation Department of Nuclear Medicine, Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan

  • Kosuke Inoue,

    Roles Data curation

    Affiliation Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

  • Hiroya Onodera

    Roles Data curation

    Affiliation Department of Radiological Sciences, School of Health Sciences, International University of Health and Welfare, Tochigi, Japan

Abstract

Although scatter correction improves SPECT image contrast and thus image quality, the effects of quantitation accuracy under various conditions remain unclear. The present study aimed to empirically define the conditions for the optimal scatter correction of quantitative bone SPECT/CT images. Scatter correction was performed by applying dual and triple energy windows (DEW and TEW) with different sub-energy window widths, and effective scatter source estimation (ESSE) to CT-based scatter correction. Scattered radiation was corrected on images acquired using a triple line source (TLSP) phantom and an uniform cylinder phantom. The TLSP consisted of a line source containing 74.0 MBq of 99mTc in the middle, and a background component containing air, water or a K2HPO4 solution with a density equivalent to that of bone. The sum of all pixels in air, water and the K2HPO4 solution was measured on SPECT images. Scatter fraction (SF) and normalized mean square error (NMSE) based on counts from the air background as a reference were then calculated to assess quantitative errors due to scatter correction. The uniform cylinder phantom contained the same K2HPO4 solution and 222.0 MBq of 99mTc. The coefficient of variation (CV) was calculated from the count profile of this phantom to assess the uniformity of SPECT images across scatter correction under various conditions. Both SF and NMSE in SPECT images of phantoms containing water in the background were lower at a TEW sub-window of 3% (TEW3%), than in other scatter corrections, whereas those in K2HPO4 were lower at a DEW sub-window of 20% (DEW20%). Larger DEW and smaller TEW sub-energy windows allowed more effective correction. The CV of the uniform cylinder phantom, DEW20%, was inferior to all other tested scatter corrections. The quantitative accuracy of bone SPECT images substantially differed according to the method of scatter correction. The optimal scatter correction for quantitative bone SPECT was DEW20% (k = 1), but at the cost of slightly decreased image uniformity.

Introduction

Bone scintigraphy using 99mTc-labeled phosphate compounds is the most prevalent means of detecting bone metastases of prostate and breast cancer [1]. Planar whole-body bone scintigraphy has high sensitivity, although the specificity is limited to characterizing bone metastases. Adding single photon emission computed tomography (SPECT) to planar acquisition has improved diagnostic confidence [2], and when combined with computed tomography (CT), bone SPECT/CT provides better specificity with more precise localization, and better contrast between hot and cold lesions [3].

Skeletal 99mTc uptake can now be quantified in absolute units (amount of radioactivity per unit volume; kBq/mL), thanks to recent advances in SPECT/CT technology [4]. The use of 3D iterative reconstruction has increased accuracy to within ± 5% of the true radioactivity concentration [57]. However, corrections for photon attenuation and scattering, resolution recovery, instrumental dead time, radioactive decay and cross-calibration are usually required to generate precise quantitative SPECT images [8]. Quantitative bone SPECT data provide more objective information that facilitates the discrimination of bone metastases [9], degenerative joint disease [10], medication-related osteonecrosis of the jaw [11] and rheumatic disease [12].

Correction for attenuation and scattered photons has the greatest impact on SPECT image quantitation [7, 13, 14]. Integrated SPECT/CT scanner can correct non-uniform attenuation using CT images. CT acquisition parameters have negligible effect in the accurate attenuation correction of the SPECT images [15]. The CT-based attenuation correction improves the quantitative accuracy of SPECT images [16]. In contrast, scatter correction for quantitative SPECT/CT has not been established. Identifying and subtracting the true number of photons that undergo scatter and lose energy before entering the detector remains a major obstacle to achieving accurate SPECT image quantitation [7, 17].

Scattered counts acquired within photopeak window varies between ~ 25% and 40% in 99mTc imaging, which leads to deteriorated image contrast and poor SPECT image quantitation [18]. The most popular methods of correcting scatter in SPECT images are usually based on energy window-based scatter correction (EWSC), the simplest of which use dual (DEW) or triple (TEW) energy windows [19, 20]. Empirical and simulated studies using 99mTc have found that DEW and TEW improve contrast rather than quantitative accuracy in SPECT images [2123]. Previous studies reported that contrast and image resolution were considerably improved after DEW and TEW so physicians could look at defects better in 99mTc cardiac perfusion SPECT images [24, 25]. However, differences in methods of scatter correction and among conditions such as the width of sub-energy windows can lead to over- or under-correction that increases image noise and results in poor quantitation [26]. The appropriate determination of sub-energy window width on energy spectra should be optimal to improve bone SPECT image quality and quantitative accuracy [27]. Moreover, effective scatter source estimation (ESSE) is combined with measured transmission data for commercial scatter correction [28]. This is because an effective scatter source is estimated for each projection view based on object-independent scatter kernels obtained using Monte Carlo simulations [29].

Although all scatter correction methods should improve contrast, and thus the quality of SPECT images, the effect of quantitation accuracy under various conditions remains unclear. Phantoms are useful and essential to define the characteristics of the quantitative accuracy of scatter correction that cannot be determined in clinical studies because they can be constructed with known materials and desirable activity concentrations. The present study aimed to empirically define the optimal conditions for correcting scatter on quantitative bone SPECT/CT images by measuring the quantitative accuracy of correction methods under various conditions using phantoms.

Materials and methods

Data acquisition and image reconstruction

All imaging data were acquired using a Brightview XCT SPECT/CT system (Philips Healthcare, Cleveland, OH, USA) with a high-resolution flat panel x-ray detector (40 × 30 cm2) with low-dose cone-beam CT imaging for localization, and attenuation correction of images. The detector is mounted on the same gantry as the SPECT camera, which allows the acquisition of SPECT and cone-beam CT images. The SPECT images were acquired under a ± 10% main energy window at 140 keV with ⅜″ crystal thickness, a low-energy high-resolution collimator (LEHR), 128 × 128 matrix with 4.8-mm pixels, and 60 projections of 20 s/view over 360° in circular orbit continuous acquisition mode. The Brightview XCT SPECT/CT system allows the simultaneous setting of 16 energy windows. With a 360° rotation of the gantry, a 47-cm diameter transverse field of view (FOV) and a 14.4-cm axial length can be visualized along a patient during slow rotation (60 s per rotation) as a cone-beam CT image [30]. We reconstructed the SPECT images using the Philips Astonish (Philips Healthcare) 3D iterative method, with combinations of 15 subsets and 2 iterations, attenuation, scatter correction, and resolution recovery.

Phantoms

The inner diameter and inner length of a NEMA SPECT Triple Line Source Phantom (TLSP; Data Spectrum Corp., Durham, NC, USA) were both 20 cm and those of an uniform cylinder phantom were 16 and 15 cm, respectively. The TLSP consisted of triple line sources with a diameter of ~ 1 mm and a height of 18.4 cm. We removed two of the three line sources, leaving the remaining source in the center (Fig 1). Thus, the TLSP consisted of the line source containing 74.0 MBq of 99mTc at the center of the phantom where the attenuating medium was air (reference), water or a K2HPO4 solution with a density equivalent to that of bone. The uniform cylinder phantom contained the same K2HPO4 and 222.0 MBq of 99mTc.

thumbnail
Fig 1. Setup of NEMA SPECT triple line source phantom.

Overview (a), and CT axial (b) and coronal (c) views of triple line source phantom.

https://doi.org/10.1371/journal.pone.0269542.g001

Scatter correction

We applied the following scatter correction methods and the energy window settings for EWSC shown in Table 1 and Fig 2.

thumbnail
Table 1. Overview of energy window widths and scatter correction factors for energy window scatter correction.

https://doi.org/10.1371/journal.pone.0269542.t001

Dual-Energy Window (DEW).

Data were simultaneously recorded within main energy window and within a secondary lower Compton sub-window in DEW. The spatial distribution of the scattered radiation acquired in the Compton window is considered a good estimate of the distribution of scattered radiation in the main energy window [31]. We applied 5%, 10%, 15%, 20% Compton sub-windows. The correction consisted of subtracting a fraction of the Compton data from the photopeak data as follows [19] (Eq 1): (1) where Cprim is the estimated unscattered radiation counts in the photopeak window, Ctotal is the total counts in the photopeak window, and Cs is the total counts in the Compton window. The scatter multiplier (k) has been determined heuristically for our SPECT acquisition system and reconstruction algorithm. Koral et al. and Luo et al. reported that the best agreement for experimental, compared with predicted counts in 99mTc images is k = 1.0 rather than 0.5. Therefore, we applied k = 1.0 in the DEW [3234].

Triple Energy Window (TEW).

The TEW includes two small sub-windows around the main energy window that are used to estimate the amount of scatter in the main energy window. Because various widths of energy sub-windows have been applied [20, 35], we investigated the effects of sub-windows with widths of 3%, 5%, 7%, 10%. Scatter correction was calculated as follows (Eqs 2 and 3): (2) (3) where Cprim is the estimated count of unscattered radiation in the photopeak window, Ctotal is the total count in the photopeak window, klower and khigher are the respective multiplication factors, and Clower and Chigher are counts in the lower and higher energy sub-windows, respectively, and Wprim and Wsub are widths (in keV) of the photopeak and energy sub-windows, respectively.

Effective Scatter Source Estimation (ESSE).

The ESSE is the standard Philips method of correcting scatter [28, 32]. Briefly, this method produces an effective scatter source obtained by convolving the radioactivity distribution with several kernels [36]. The effective scatter source kernel and the relative scatter attenuation coefficient kernel for convolution are derived from Monte Carlo simulations and depend on photon energy [37]. Scatter projection data are then estimated by forward-projecting the effective scatter source.

Data analysis

A circular region of interest (ROI) was placed at the center of the TLSP. The sum of all pixel counts per slice in air, water and the K2HPO4 solution were measured on SPECT images generated by the TLSP. The counts per slice were acquired from the average of the counts in 10 slices. The scatter fraction (SF) and normalized mean square error (NMSE) based on counts from the air background as the reference were calculated to assess quantitative errors due to scatter correction as follows (Eqs 4 and 5): (4) (5) where Tair and Twater or bone were the average counts of the ROI in the phantoms containing air and in that containing water or bone equivalent solution, respectively.

The coefficient of variation (CV) was evaluated at an 80% circular ROI placed at the center of an uniform cylinder phantom. We calculated the CV from the count profile of this phantom to assess SPECT image uniformity across all scatter corrections as follows (Eq 6): (6) where SD and mean are the standard deviation and means of counts, respectively, in different scatter corrections. The CV represents the amount of statistical noise in SPECT images, which reflects uniformity after scatter correction.

Results

Fig 3A and 3B shows the SF and NMSE, respectively, for each scatter correction. Both SF and NMSE in SPECT images of phantoms containing water in the background were lower at a TEW sub-window of 3% (TEW3%) than any other scatter correction, whereas those in the K2HPO4 solution were lower at a DEW sub-window of 20% (DEW20%). The SF and NMSE became smaller with larger DEW and smaller TEW sub-energy windows, and ESSE overcorrected the scatter radiation in the K2HPO4 solution. Table 2 shows the CV of the uniform cylinder phantom. The CV of ESSE was superior, whereas the CV of DEW20% was inferior to that of any other assessed method of scatter correction. We visually confirmed the ability of different scatter correction technologies in scatter radiation images after scatter correction (Fig 4). Scattered radiation images were obtained by subtracting the reference image with DEW20% (considered as the most accurate in bone equivalent solutions) from the SPECT images with scatter corrected under different conditions. The remaining scattered radiation was visually almost identical between DEW20% and ESSE, although the scattered radiation of ESSE might be overcorrected.

thumbnail
Fig 3.

Scatter fraction (a) and normalized mean-square error (b) of SPECT images after scatter correction in phantoms containing water or bone equivalent solution. DEW 5, 10, 15, 20, are dual energy window sub-windows of 5%, 10%, 15%, 20%; ESSE, effective scatter source estimation; non SC, no scatter correction; TEW 3, 5, 7, 10 are triple energy window sub-windows of 3%, 5%, 7%, 10%.

https://doi.org/10.1371/journal.pone.0269542.g003

thumbnail
Fig 4. Images of scattered radiation after scatter correction.

Increased red intensity indicates more residual scattered radiation. DEW, dual energy window; ESSE, effective scatter source estimation; SC, scatter correction; TEW, triple energy window.

https://doi.org/10.1371/journal.pone.0269542.g004

thumbnail
Table 2. Coefficients of variation at center of uniform cylinder phantom.

https://doi.org/10.1371/journal.pone.0269542.t002

Discussion

Scatter correction for quantitative bone SPECT images has not yet been standardized. We evaluated various scatter correction methods based on EWSC and ESSE using phantoms. Our findings indicated that the accuracy of scatter correction for SPECT image quality and quantitation depends on the method applied, sub-energy window width and the background material that produces scattered radiation. We also found that the scatter correction for quantitative bone SPECT was optimal with a DEW20% sub-window (k = 1).

We applied a K2HPO4 solution with a density equivalent to that of bone [38, 39]. The trends of SF and NMSE between water and the K2HPO4 solution in the background component of the phantom considerably differed (Fig 3). The effects of photon scatter by bone and the accuracy of scatter correction for bone SPECT/CT images in clinical practice can be simulated and investigated using a bone equivalent solution. Miyaji et al. [40] validated the accuracy of a new reconstruction method for bone SPECT/CT (xSPECT Bone; Siemens Healthineers), using a K2HPO4 solution, and Yoshii et al. [41] developed a new phantom containing a K2HPO4 solution that allows consideration of scatter and photon attenuation due to bone in 18F-NaF PET/CT images.

Both SF and NMSE in SPECT images of phantoms containing background K2HPO4 solution were lower at DEW15% or DEW20% than with TEW scatter correction (Fig 3). An experimental study of the effects of backscattered radiation from various materials and 99mTc radionuclide showed that photons were backscattered in energy regions below ~ 120 keV and did not affect the photopeak [22]. We considered that DEW method, which contained enough multi-forward and backscattered radiation due to bone in the sub-energy window, would be an appropriate choice for scatter correction. Several studies using Monte Carlo simulation [27, 42, 43] and clinical data [7, 44] have found that DEW method with an optimized sub-energy window width and k value offered the most appropriate scatter correction of 99mTc SPECT images in terms of improved contrast, higher SNR and good quantitation.

The SF and NMSE decreased with wider DEW sub-energy windows. However, CV was higher in DEW sub-window 20% (Table 2). A wide sub-energy window can measure largely radiations with large deflection angle and higher order scatter. The correction can improve image contrast and quantitation by removing too many photons far from the actual source location, but not enough at the source location. Thus, image uniformity in SPECT images were considered to be slightly degraded [18]. Moreover, DEW and TEW assume that the sub-energy window of scattered radiation are minimally contaminated by unscattered radiations, and the presence of unscattered radiations in the sub-energy window leads to an overestimate of the scatter. Subtraction of this estimate thus reduces scatter radiation in the photopeak, but also reduces the primary signal and thereby increases the noise (CV) of the scatter corrected data [25].

Although the SPECT image corrected using ESSE appeared essentially identical to that of DEW20% sub-window, the SF of ESSE was overestimated in SPECT images acquired from phantoms containing water and background K2HPO4 solution. This might be due to errors specific to our phantom. The ESSE generates good results in a wide range of application [31, 32, 45, 46] and improves image contrast as well as the quantitative accuracy of radioactivity estimation. However, our findings were inconsistent with these results. The large attenuation between the source and the point of Compton interaction in the large phantom used herein might have resulted in slight errors in the calculation of an effective scatter source for ESSE. Moreover, the CT value of the background might be higher than actual bone due to the high concentration of K2HPO4 in the solution.

The present study has several limitations. We analyzed data generated by a phantom simulating patient-specific bone. Further studies should investigate bone 99mTc SPECT images acquired from actual patients. We validated only scatter correction in SPECT images for quantitative accuracy. Most corrections such as photon attenuation and scattering, as well as resolution recovery influence quantitative SPECT images [7]. The combination effects of various correction methods in quantitative SPECT images requires further investigation.

Conclusions

The quantitative accuracy of bone SPECT images considerably differed according to the method of scatter correction. The optimal scatter correction for quantitative bone SPECT was DEW20% (k = 1) according to our phantom study using water and K2HPO4 solution as scatter radiation components representing soft tissue or bone, respectively, but at the cost of slightly decreased image uniformity. The present findings provide useful information about how to confirm optimal scatter correction for quantitative evaluations of bone SPECT images. Compton scatter is object-dependent and spatially varying. Further studies should validate bone 99mTc SPECT images acquired from more complex geometry and realistic phantom, e.g. NEMA IEC body phantom, and actual patients using optimal scatter correction obtained from our phantom study.

Acknowledgments

For valuable contributions in the data collection, we would like to thank Mr. Yukinori Fukuhara and Mr. Shinichiro Terada from Tokyo Women’s Medical Hospital.

References

  1. 1. Beck M, Sanders JC, Ritt P, Reinfelder J, Kuwert T. Longitudinal analysis of bone metabolism using SPECT/CT and 99mTc-diphosphono-propanedicarboxylic acid: comparison of visual and quantitative analysis. EJNMMI Res. 2016;6(1):60. pmid:27464623; PubMed Central PMCID: PMC4963336.
  2. 2. Ichikawa H, Miwa K, Okuda K, Shibutani T, Kato T, Nagaki A, et al. Current state of bone scintigraphy protocols and practice in Japan. Asia Ocean J Nucl Med Biol. 2020;8(2):116–22. Epub 2020/07/28. pmid:32714999; PubMed Central PMCID: PMC7354239.
  3. 3. Israel O, Pellet O, Biassoni L, De Palma D, Estrada-Lobato E, Gnanasegaran G, et al. Two decades of SPECT/CT—the coming of age of a technology: An updated review of literature evidence. Eur J Nucl Med Mol Imaging. 2019;46(10):1990–2012. Epub 2019/07/06. pmid:31273437; PubMed Central PMCID: PMC6667427.
  4. 4. Miyaji N, Miwa K, Motegi K, Yamashita K, Terauchi T, Onoguchi M. Patient arm position during quantitative bone single-photon emission computed tomography/computed tomography acquisition can affect image quality and quantitative accuracy: a phantom study. Nucl Med Commun. 2021;42(3):267–75. pmid:33323866.
  5. 5. Ritt P, Vija H, Hornegger J, Kuwert T. Absolute quantification in SPECT. Eur J Nucl Med Mol Imaging. 2011;38 Suppl 1:S69–77. pmid:21484383.
  6. 6. Bailey DL, Willowson KP. An evidence-based review of quantitative SPECT imaging and potential clinical applications. J Nucl Med. 2013;54(1):83–9. Epub 2013/01/04. pmid:23283563.
  7. 7. Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51(6):921–8. pmid:20484423.
  8. 8. Bailey DL, Willowson KP. Quantitative SPECT/CT: SPECT joins PET as a quantitative imaging modality. Eur J Nucl Med Mol Imaging. 2014;41 Suppl 1:S17–25. pmid:24037503.
  9. 9. Yamane T, Fukushima K, Shirotake S, Nishimoto K, Okabe T, Oyama M, et al. Test-retest repeatability of quantitative bone SPECT/CT. Ann Nucl Med. 2021;35(3):338–46. Epub 20210105. pmid:33400148.
  10. 10. Mohd Rohani MF, Mat Nawi N, Shamim SE, Wan Sohaimi WF, Wan Zainon WMN, Musarudin M, et al. Maximum standardized uptake value from quantitative bone single-photon emission computed tomography/computed tomography in differentiating metastatic and degenerative joint disease of the spine in prostate cancer patients. Ann Nucl Med. 2020;34(1):39–48. Epub 20191014. pmid:31612417.
  11. 11. Okui T, Kobayashi Y, Tsujimoto M, Satoh K, Toyama H, Matsuo K. Quantitative evaluation of anti-resorptive agent-related osteonecrosis of the jaw using bone single photon emission computed tomography in clinical settings: relationship between clinical stage and imaging. Ann Nucl Med. 2020;34(9):620–8. Epub 20200615. pmid:32557015.
  12. 12. Lim CH, Kim HS, Lee KA, Kim J, Park SB. Clinical significance of quantitative bone SPECT/CT in the evaluation of hand and wrist pain in patients with rheumatic disease. Sci Rep. 2022;12(1):327. Epub 20220110. pmid:35013365; PubMed Central PMCID: PMC8748937.
  13. 13. El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med. 2000;41(8):1400–8. Epub 2000/08/17. pmid:10945534.
  14. 14. Ichikawa H, Onoguchi M, Shibutani T, Kato T, Ito T, Shimada H. Optimization of cross-calibration factor for quantitative bone SPECT without attenuation and scatter correction in the lumbar spine: head-to-head comparison with attenuation and scatter correction. Nucl Med Commun. 2021;42(12):1404–16. pmid:34456319.
  15. 15. Hulme KW, Kappadath SC. Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging. Med Phys. 2014;41(4):042502. Epub 2014/04/04. pmid:24694155.
  16. 16. Keidar Z, Israel O, Krausz Y. SPECT/CT in tumor imaging: technical aspects and clinical applications. Semin Nucl Med. 2003;33(3):205–18. Epub 2003/08/22. pmid:12931322.
  17. 17. Robinson AP, Tipping J, Cullen DM, Hamilton D. The influence of triple energy window scatter correction on activity quantification for (1 7 7)Lu molecular radiotherapy. Phys Med Biol. 2016;61(14):5107–27. pmid:27351914.
  18. 18. Hutton BF, Buvat I, Beekman FJ. Review and current status of SPECT scatter correction. Phys Med Biol. 2011;56(14):R85–112. pmid:21701055.
  19. 19. Jaszczak RJ, Greer KL, Floyd CE Jr., Harris CC, Coleman RE. Improved SPECT quantification using compensation for scattered photons. J Nucl Med. 1984;25(8):893–900. Epub 1984/08/01. pmid:6611390.
  20. 20. Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34(12):2216–21. Epub 1993/12/01. pmid:8254414.
  21. 21. Gustafsson A, Arlig A, Jacobsson L, Ljungberg M, Wikkelso C. Dual-window scatter correction and energy window setting in cerebral blood flow SPECT: a Monte Carlo study. Phys Med Biol. 2000;45(11):3431–40. Epub 2000/12/01. pmid:11098915.
  22. 22. Kojima A, Matsumoto M, Takahashi M. Experimental analysis of scattered photons in Tc-99m imaging with a gamma camera. Ann Nucl Med. 1991;5(4):139–44. Epub 1991/11/01. pmid:1797068.
  23. 23. Noori-Asl M, Sadremomtaz A, Bitarafan-Rajabi A. Evaluation of three scatter correction methods based on estimation of photopeak scatter spectrum in SPECT imaging: a simulation study. Phys Med. 2014;30(8):947–53. Epub 2014/07/06. pmid:24985135.
  24. 24. Changizi V, Takavar A, Babakhani A, Sohrabi M. Scatter correction for heart SPECT images using TEW method. J Appl Clin Med Phys. 2008;9(3):2767. Epub 2008/08/22. pmid:18716588.
  25. 25. Pourmoghaddas A, Vanderwerf K, Ruddy TD, Glenn Wells R. Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state camera. J Nucl Cardiol. 2015;22(2):334–43. pmid:25342215.
  26. 26. Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shoji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39(1):181–9. pmid:9443759.
  27. 27. Asgari A, Ashoor M, Sohrabpour M, Shokrani P, Rezaei A. Evaluation of various energy windows at different radionuclides for scatter and attenuation correction in nuclear medicine. Ann Nucl Med. 2015;29(4):375–83. Epub 20150123. pmid:25613356.
  28. 28. Frey EC, Tsui BMW, editors. A new method for modeling the spatially-variant, object-dependent scatter response function in SPECT. 1996 IEEE Nuclear Science Symposium Conference Record; 1996 2–9 Nov 1996.
  29. 29. Du Y, Tsui BM, Frey EC. Model-based compensation for quantitative 123I brain SPECT imaging. Phys Med Biol. 2006;51(5):1269–82. Epub 2006/02/17. pmid:16481693.
  30. 30. Ljungberg M, Pretorius PH. SPECT/CT: an update on technological developments and clinical applications. Br J Radiol. 2018;91(1081):20160402. Epub 2016/11/16. pmid:27845567; PubMed Central PMCID: PMC5966195.
  31. 31. de Nijs R, Lagerburg V, Klausen TL, Holm S. Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections. Nucl Med Commun. 2014;35(5):522–33. Epub 2014/02/15. pmid:24525900; PubMed Central PMCID: PMC3969156.
  32. 32. Lagerburg V, de Nijs R, Holm S, Svarer C. A comparison of different energy window subtraction methods to correct for scatter and downscatter in I-123 SPECT imaging. Nucl Med Commun. 2012;33(7):708–18. Epub 2012/04/20. pmid:22513883.
  33. 33. Koral KF, Swailem FM, Buchbinder S, Clinthorne NH, Rogers WL, Tsui BM. SPECT dual-energy-window Compton correction: scatter multiplier required for quantification. J Nucl Med. 1990;31(1):90–8. Epub 1990/01/01. pmid:2404095.
  34. 34. Luo JQ, Koral KF, Ljungberg M, Floyd CE Jr., Jaszczak RJ. A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT. Phys Med Biol. 1995;40(1):181–99. Epub 1995/01/01. pmid:7708840.
  35. 35. Nakajima K, Matsubara K, Ishikawa T, Motomura N, Maeda R, Akhter N, et al. Correction of iodine-123-labeled meta-iodobenzylguanidine uptake with multi-window methods for standardization of the heart-to-mediastinum ratio. J Nucl Cardiol. 2007;14(6):843–51. Epub 2007/11/21. pmid:18022111.
  36. 36. Bai C, Zeng GL, Gullberg GT. A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D scatter compensation in parallel and converging beam SPECT. Phys Med Biol. 2000;45(5):1275–307. Epub 2000/06/08. pmid:10843105.
  37. 37. Kadrmas DJ, Frey EC, Karimi SS, Tsui BM. Fast implementations of reconstruction-based scatter compensation in fully 3D SPECT image reconstruction. Phys Med Biol. 1998;43(4):857–73. Epub 1998/05/08. pmid:9572510; PubMed Central PMCID: PMC2808130.
  38. 38. Iida H, Hori Y, Ishida K, Imabayashi E, Matsuda H, Takahashi M, et al. Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour. Ann Nucl Med. 2013;27(1):25–36. Epub 2012/09/27. pmid:23011903; PubMed Central PMCID: PMC3549246.
  39. 39. Dreuille OD, Strijckmans V, Almeida P, Loc’h C, Bendriem B, editors. Bone equivalent liquid solution to assess accuracy of transmission measurements in SPECT and PET. 1996 IEEE Nuclear Science Symposium Conference Record; 1996 2–9 Nov 1996.
  40. 40. Miyaji N, Miwa K, Tokiwa A, Ichikawa H, Terauchi T, Koizumi M, et al. Phantom and clinical evaluation of bone SPECT/CT image reconstruction with xSPECT algorithm. EJNMMI Research. 2020;10(1):71. pmid:32601770
  41. 41. Yoshii T, Miwa K, Yamaguchi M, Shimada K, Wagatsuma K, Yamao T, et al. Optimization of a Bayesian penalized likelihood algorithm (Q.Clear) for (18)F-NaF bone PET/CT images acquired over shorter durations using a custom-designed phantom. EJNMMI Phys. 2020;7(1):56. Epub 2020/09/12. pmid:32915344.
  42. 42. Asl MN, Sadremomtaz A, Bitarafan-Rajabi A. Evaluation of six scatter correction methods based on spectral analysis in (99m)Tc SPECT imaging using SIMIND Monte Carlo simulation. J Med Phys. 2013;38(4):189–97. pmid:24672154; PubMed Central PMCID: PMC3958999.
  43. 43. Knoll P, Rahmim A, Gultekin S, Samal M, Ljungberg M, Mirzaei S, et al. Improved scatter correction with factor analysis for planar and SPECT imaging. Rev Sci Instrum. 2017;88(9):094303. Epub 2017/10/02. pmid:28964205; PubMed Central PMCID: PMC5848721.
  44. 44. Huang K, Feng Y, Liu D, Liang W, Li L. Quantification evaluation of (99m)Tc-MDP concentration in the lumbar spine with SPECT/CT: compare with bone mineral density. Ann Nucl Med. 2020;34(2):136–43. Epub 2019/11/27. pmid:31768820.
  45. 45. Du Y, Frey EC. Quantitative evaluation of simultaneous reconstruction with model-based crosstalk compensation for 99mTc/123I dual-isotope simultaneous acquisition brain SPECT. Med Phys. 2009;36(6):2021–33. pmid:19610291; PubMed Central PMCID: PMC2736700.
  46. 46. Farncombe TH, Gifford HC, Narayanan MV, Pretorius PH, Frey EC, King MA. Assessment of scatter compensation strategies for (67)Ga SPECT using numerical observers and human LROC studies. J Nucl Med. 2004;45(5):802–12. Epub 2004/05/12. pmid:15136630.