Comparable respiratory activity in attached and suspended human fibroblasts

Abstract

Measurement of oxygen consumption of cultured cells is widely used for diagnosis of mitochondrial diseases, drug testing, biotechnology, and toxicology. Fibroblasts are cultured in monolayers, but physiological measurements are carried out in suspended or attached cells. We address the question whether respiration differs in attached versus suspended cells using multiwell respirometry (Agilent Seahorse XF24) and high-resolution respirometry (Oroboros O2k), respectively. Respiration of human dermal fibroblasts measured in culture medium was baseline-corrected for residual oxygen consumption and expressed as oxygen flow per cell. No differences were observed between attached and suspended cells in ROUTINE respiration of living cells and LEAK respiration obtained after inhibition of ATP synthase by oligomycin. The electron transfer capacity was higher in the O2k than in the XF24. This could be explained by a limitation to two uncoupler titrations in the XF24 which led to an underestimation compared to multiple titration steps in the O2k. A quantitative evaluation of respiration measured via different platforms revealed that short-term suspension of fibroblasts did not affect respiratory activity and coupling control. Evaluation of results obtained by different platforms provides a test for reproducibility beyond repeatability. Repeatability and reproducibility are required for building a validated respirometric database.

Introduction

Studies of cells attached in a monolayer or suspended in the medium have wide-ranging applications and implications. These include metabolic profiling [1], substrate diffusion [2], cell morphology and rheology [3], macrophage adherence [2], suspension culture as mimetic of circulating tumor cells [4], metastatic potential [4], therapeutic cell reimplantation, and cell culture biotechnology and pharmacology [5].

Fibroblast cell lines are established models routinely applied in studies of mitochondrial diseases [69]. These cells can be investigated in culture either attached to the surface of an experimental chamber or in suspension after detachment.

The structure of cells growing in culture changes from the attached to the suspended state. After trypsinization, fibroblasts undergo membrane reorganization and attain a spherical shape with a so-called blebbed surface morphology to prevent membrane loss by providing transient membrane storage [3]. In rabbit lung macrophages, transport of lysine and adenosine across the plasma membrane is faster in suspended cells compared to adherent ones [2]. Moreover, mouse macrophages oxidize glucose six times faster when in suspension than in monolayers [1]. Suspending anchorage‐dependent fibroblasts results in an abrupt drop of mRNA production, while protein synthesis declines slowly but extensively and its recovery requires surface contact [10, 11]. Taken together, these observations raise the physiological question whether cell respiration differs in suspended versus attached fibroblasts. Stimulatory or suppressive effects may be exerted on aerobic ATP demand and consequently respiration may be regulated differently in suspended and attached states.

The first aim of the present study was the evaluation of respiration in attached compared to suspended fibroblasts. The Seahorse XF Analyzer (Agilent, US) is designed for studying respiration of attached cells (ace), whereas the Oroboros O2k (Oroboros Instruments, Austria) is optimized for high-resolution respirometry with suspended cells (sce). Therefore, a platform comparison was required as the second aim of our study. We applied comparable and standardized protocols for monitoring respiratory activity in living cells under near-physiological conditions. Specific titration steps in the protocol disrupt the physiological control of respiration, which then allows for instrumental comparison of performance of the two platforms. Respiratory activity was normalized per cell and expressed in identical SI units for a quantitative comparison of the data. The experimental period of suspended cells was limited to less than one hour. The respiratory activity of suspended and attached cells was not distinguishable during such short time intervals. Comparative studies provide the basis for extending databases of cellular respiration by including critically evaluated results obtained with different platforms.

Materials and methods

Reagents

DMEM5030, Carbonyl cyanide 4-(trifluoromethoxy)phenylhydrazone (FCCP, C2920-10MG), oligomycin(O4876-5MG), antimycin A (A8674-25MG), rotenone (R8875-1G), and glutamine (G7513-100mL) were purchased from Sigma Aldrich. Trypsin- ethylenediaminetetraacetic acid EDTA 10x (XC-T1717/100) and antibiotics-antimycotics 100X (XC-A4110/100) were obtained from Biosera, Dulbecco’s modified Eagle’s medium (DMEM, P04-04510) from Pan Biotech, fetal bovine serum (SV30160.03) from HyClone, and XF Calibrant Solution 100840–000 from Agilent.

The instruments

The Seahorse XF Analyzer (Agilent, Santa Clara, US) provides multi-well plate analysis of two processes in real time: oxygen consumption rate (OCR) as an indicator of cell respiration, and extracellular acidification rate largely dependent on glycolytic processes. Cellular O2 consumption causes changes in the concentration of dissolved dioxygen O2 in so-called ‘transient microchambers’. O2 pressure is measured by solid-state fluorescent probes and converted to O2 concentration. Every measurement step takes 5–8 min during which O2 concentration is measured every few seconds, providing data for OCR calculation. Then the probes are lifted, and the larger volume of medium is mixed to restore O2 levels to baseline conditions. Chemical compounds are injected pneumatically, limited to four sequential injections per well [12].

The Oroboros O2k (Oroboros Instruments, Innsbruck, Austria) is a two-chamber high-resolution respirometer used in cell and mitochondrial research to measure respiration in mitochondrial preparations and living cells. The O2k supports multi-sensor modules and measurement is performed in the experimental chamber, where suspended cells or mitochondrial preparations are continuously mixed by a stirrer at 750 rotations per minute. O2 consumption in nearly diffusion-tight closed chambers is calculated in real time from O2 partial pressure measured by polarographic oxygen sensors. The O2k provides the option of practically unlimited titrations and, therefore, the possibility to apply various Substrate-Uncoupler-Inhibitor-Titration (SUIT) protocols designed to address specific research questions [13, 14]. Table 1 summarizes the comparison between the two instruments.

thumbnail
Table 1. Comparative specifications of Seahorse XF24 and Oroboros O2k applied in the present coupling control protocol.

https://doi.org/10.1371/journal.pone.0264496.t001

Cell culture

Two human dermal fibroblast cell lines were purchased, HDF 1 (Primary Dermal Fibroblast Normal; Human, Neonatal HDFn, PCS_201_010, ATCC, NHDF-Neo), and HDF 2 (Human Dermal Fibroblasts, Neonatal, CC-2509, Lonza). A human dermal fibroblast cell line (HDF 3) was derived from a disease-free control at age of 5 months upon verbal informed parentals’ consent obtained in General Hospital in Prague with ethics committee approval No 92/18 (18.10.2018) for the project GAUK (Grant Agency of Charles University) 110119. All cell lines were cultured in Dulbecco’s modified Eagle medium (DMEM, Pan Biotech) with 25 mM glucose, 10% fetal bovine serum and 1% antibiotics-antimycotics 100X at 37°C under 5% CO2 atmosphere. Fibroblast cultures at passage number 13 to 15 were grown to approximately 80% confluence. Suspended cells were counted by a Handheld Automated Cell Counter (Millipore). Cells were harvested by incubation in trypsin 0.05% w/V with EDTA 0.02%, w/V for 5 min at 37°C, washed, and centrifuged at 300 g (5 min, 24°C).

Sample preparation and respirometry

DMEM5030 was the basis of DMEM respiration medium with addition of 3.9 mM glutamate, 5 mM glucose and 2 mM pyruvate, adjusted to pH 7.4 at 37°C. DMEM respiration medium was freshly prepared on the day of use.

Mitochondrial respiration was measured in the Department of Pediatrics and Inherited Metabolic Disorders, Prague. The Agilent Seahorse XF Analyzer (XF24) according to [15] with slight modifications. The day before measurement, cells were harvested after trypsinization, resuspended in DMEM culture medium, counted, and 35 000 cells were seeded on 20 wells of a 24-well plate for overnight incubation. 4 wells were used as blanks without cells. The plates were incubated overnight in a 5% CO2 atmosphere at 37°C. In parallel, the Sensor Cartridge was hydrated in wells filled with Seahorse XF Calibrant Solution (Agilent) by incubation without CO2 over-night prior to use. The XF24 was switched on during the day before experiments to equilibrate at 37°C. On the following day, wells were washed twice with 1 mL DMEM respiration medium. 450 μL DMEM respiration medium was added to each well and incubated without CO2 for one hour at 37°C. In the meantime, 50 μL oligomycin (stock 20 μM; experimental concentration 2 μM) was added to cartridge port A, 55 μL FCCP (stock 5 μM) to port B, 61 μL FCCP (stock 2 μM) to port C (experimental concentrations 0.5 and 0.7 μM, respectively), and 67 μL rotenone (stock 20 μM) with antimycin A (stock 10 μM) to port D (experimental concentrations 2 and 1 μM). Sensor cartridges were incubated without CO2 for 30–40 min at 37°C, transferred to the XF24 for equilibration and calibration in Seahorse XF Calibrant Solution. Then the calibration well plate was exchanged for the cell plate. Before starting the respiratory protocol with living cells, wells were mixed for 3 min and left idle for 2 min. Respiratory flux was measured in each state three times for 3 min. The transient microchamber for measurement had a volume of approximately 7 μL, with sensors positioned 200 μm above the well bottoms.

Air calibration in the Oroboros O2k was performed daily before measurement [16]. The O2 partial pressure for air calibration is calculated (DatLab software) for air saturated 100% with water vapor at experimental temperature and local barometric pressure recorded real-time by the electronic pressure transducer of the O2k. The O2 solubility of DMEM respiration medium was assumed to be 92% relative to the O2 solubility of pure water at 37°C [17]. Calibrations at zero O2 concentration were performed before the experimental series. Instrumental O2 background flux [18] was measured each day before the experiment with cells in the range of O2 concentration from air saturation at 190 μM to 100 μM.

Experimental oxygen concentrations near air saturation (190 μM to 140 μM; Fig 1B and 1D) are much higher than extracellular oxygen concentrations in various tissues in vivo. Respiration of living cells, however, is independent of oxygen pressure from air saturation to tissue-level oxygen concentration, the latter ranging between 50 and 10 μM [19]. Therefore, even if high experimental oxygen concentrations are not physiological, respiratory measurements are not affected. This is in direct contrast to cellular hydrogen peroxide production, which increases with oxygen concentration over the entire experimental oxygen concentration range [20].

thumbnail
Fig 1. Representative traces of respiration and O2 concentration or O2 pressure in the protocol with living human dermal fibroblasts (HDF 1).

(a) O2 flux per volume in two chambers measured simultaneously in the Oroboros O2k. (b) O2 concentration and partial O2 pressure corresponding to the traces in panel a. (c) O2 consumption rate (OCR) measured simultaneously in six wells of one plate of the XF24. Three time intervals of respiration per respiratory state (R1, R2, R3; L1, L2, L3; etc.). (d) O2 partial pressure corresponding to the traces in panel c.

https://doi.org/10.1371/journal.pone.0264496.g001

Respiratory protocol

Comparable respiratory protocols were used with the Agilent XF24 and Oroboros O2k using the same DMEM respiration medium (Fig 1). Every cell line was measured on the same day in parallel, splitting the cells from one culture flask into two O2k -chambers and 20 wells of the XF24. In the coupling control protocol for living cells (SUIT-003), four respiratory states are distinguished [21]. We adhere to the platform-independent terminology of the MitoEAGLE consensus paper [22] and add the terms widely used in conjunction with the XF24 or XF96 [15] in parentheses (Table 2).

thumbnail
Table 2. Harmonization of terminology on respiratory states.

https://doi.org/10.1371/journal.pone.0264496.t002

First, ROUTINE respiration R (‘basal respiration’) was measured in attached or suspended cells. The ROUTINE state is a physiological state, in which respiration is controlled by cellular energy demand. Next, the ATP synthase inhibitor oligomycin was added to induce LEAK respiration L (‘proton leak’). In the non-phosphorylating LEAK state, a low rate of respiration is maintained mainly to compensate for the proton leak at a high protonmotive force. Afterwards, the uncoupler FCCP was titrated in at least two steps to obtain a maximum rate of O2 consumption reflecting the electron transfer capacity E (‘maximal respiration’) in the noncoupled state. Finally, antimycin A and rotenone were added. These inhibitors of Complexes CI and CIII, respectively, inhibit mitochondrial electron transfer and thus induce residual oxygen consumption Rox (‘nonmitochondrial respiration’) due to oxidative side reactions. Importantly, the plasma membrane is permeable for these inhibitors and the uncoupler which, therefore, can be applied in living cells. The concentrations applied in the XF24 and O2k adhered closely to the respective manuals (Table 3; Fig 1).

thumbnail
Table 3. Recommended and applied experimental concentrations of inhibitors and uncoupler [14; 23].

https://doi.org/10.1371/journal.pone.0264496.t003

Data analysis

Measurements were normalized for the cell count. Traces were analyzed in DatLab 7 software (Oroboros Instruments) where each O2k -chamber contained 375 000 cells. O2 flow is expressed per cell [amol∙s-1∙x-1] equivalent to [pmol∙s-1∙(106 x)-1]. The Wave 2 software (Agilent) presents results in units [pmol∙min-1] in a well which contained 35 000 cells. Data were converted to the same units [amol∙s-1∙x-1], where x represents the unit cell [14, 24]. In regressions between respiration for different states, variables in X and Y have comparable errors of measurement. To minimize the residuals of both variables, Y and X, slopes bY and βX and intercepts aY and αX are calculated for the Y/X and X/Y inverted linear regressions, respectively. The mean slope and mean intercept are used, where bX = 1/βX and aX = −αX/βX [25]. Further statistical evaluation was performed using Prism (GraphPad Software, California, USA). Rox-corrected rates were symmetrically distributed, and logarithmic transformation was not required. One outlier was removed in the XF24 with negative LEAK respiration irrespective or Rox-correction. There were no outliers in the O2k.

Results

Respiration normalized for cell count

Respiratory rates measured in the XF24 (ace) and O2k (sce) are expressed per cell in units for O2 flow [amol∙s-1∙x-1] and shown on identical scales in Fig 2A and 2B. The ranking of respiratory capacities of the three cell lines followed different patterns in the XF24 and O2k, which was taken as an argument for pooling all results in the scatter plots. The variability within cell lines was greater in the XF24 (n = 20) than the O2k (n = 4). Averaging five wells of the XF24 to obtain n = 4 per cell line did not reduce the coefficient of variation (SD/average), which was 0.29 and 0.26 for R and E (n = 20) to 0.27 and 0.27 (n = 4, each for 5 pooled wells), compared to the coefficient of variation of 0.10 and 0.10 for R and E in the O2k (Table 4A).

thumbnail
Fig 2.

Sequence of respiratory states in the coupling control protocol applied in the XF24 for ace (a) and O2k for sce (b). Circles: individual wells or chambers; diamonds and bars: median ± SD. ROUTINE respiration R; LEAK respiration L; electron transfer capacity E; residual oxygen consumption Rox.

https://doi.org/10.1371/journal.pone.0264496.g002

thumbnail
Table 4. Respiration of three human fibroblast cell lines attached (ace; XF24) or suspended (sce; O2k).

https://doi.org/10.1371/journal.pone.0264496.t004

Rates of O2 consumption are expected to be identical in adherent and suspended cells in the LEAK, ET, and ROX states when physiological control of cellular energy demand is eliminated. This is achieved by inhibition of mitochondrial electron transfer using rotenone and antimycin A. These block the Complexes I and III, respectively, and thus induce the state of residual oxygen consumption ROX. Similarly, inhibition of ATP synthase by oligomycin induces LEAK respiration L in a state that is not controlled by cellular ATP turnover. ROX and LEAK are, therefore, states of minimum O2 consumption, providing a comparison of instrumental resolution of the two instrument types, independent of using attached cells (ace) in the XF24 and suspended cells (sce) in the O2k. Residual oxygen consumption Rox and total LEAK respiration L’tot not corrected for Rox were higher with a larger scatter in the XF24 (ace) compared to the O2k (sce) (Fig 3A and 3B). Rox-correction of LEAK respiration L, however, eliminated the differences observed in the two platforms (Fig 3C). Bioenergetic cluster analysis BCA [25] shows the pairwise correlation between Rox and total LEAK respiration (Fig 3D and 3E).

thumbnail
Fig 3. Residual oxygen consumption Rox, LEAK respiration L, ROUTINE respiration R, and electron transfer capacity E in attached and suspended fibroblast cells (ace and sce).

(a) Rox measured after inhibition by rotenone and antimycin A reflects instrumental resolution independent of cell physiological conditions for comparison of the two respirometers (ace: XF24, sce: O2k). (b) Total LEAK respiration Ltot uncorrected for Rox, higher in ace than sce. (c) Rox-correction of LEAK respiration L eliminates the difference between ace and sce. (d) Relation between Rox and Ltot (full range). is the mean slope of the ordinate and inverted slopes which are shown for ace (thin dashed lines). The low coefficient of determination r2 for sce is related to the very low scatter of the data. (e) Zoom into the positive range of values in panel d. (f) Rox-corrected ROUTINE respiration was similar in ace and sce. (g) Rox-corrected electron transfer capacity was lower in ace than sce.

https://doi.org/10.1371/journal.pone.0264496.g003

ROUTINE respiration R is under physiological control of living cells. Therefore, R reflects the possible changes in mitochondrial ATP demand induced by suspending cells that were grown attached in a monolayer. Rox-corrected R was similar in attached cells measured in the XF24 and suspended cells measured in the O2k (Fig 3F).

Electron transfer capacity E is supported by physiological substrates in the living cell but is entirely independent of respiratory control by cellular ATP demand. Therefore, E corrected for Rox was expected to be independent of cell physiology and comparable in suspended and attached cells. However, E was lower in ace than sce (Fig 3G). Results for the three cell lines are summarized in Table 4.

ROUTINE versus LEAK respiration and electron transfer capacity

Internal normalization eliminates any possible bias caused by methodological differences in determining the concentration of cells per experimental chamber volume in wells (attached; ace) and in suspension (sce).

ET capacity is frequently used as a reference and functional mt-marker (Table 4B). The higher R/E flux control ratio in ace, however, does not indicate a lower ROUTINE respiration in sce, which would contradict the results on O2 flow normalized for cell count (Fig 3). BCA illustrates a left shift of the R/E regression line due to lower ET capacity in ace compared to sce, at comparable ROUTINE respiration of suspended and attached cells (Fig 4A). The compensatory increase of R as a function of intrinsic uncoupling detected by increasing L in sce (Fig 4B) conforms to a pattern generally observed in cell respiration [25]. Noise in the ace data prevents resolution, yet the overlap of clusters supports the conclusion based on O2 flow (Fig 3F) that ROUTINE respiration was not different in ace and sce.

thumbnail
Fig 4. Bioenergetic cluster analysis of respiration in different respiratory coupling states in attached and suspended cells (ace and sce).

(a) R varies in direct proportion to ET capacity. The linear R/E regressions are different for ace and sce, thus the data points are arranged as separate heterolinear clusters with different and . E is lower in ace than sce at similar R (blue arrow pointing to the left from the median of sce to the median of ace). (b) ace and sce are in the same cluster in the plot of R versus L, with high noise in ace LEAK respiration. For sce, bY and bX indicate the slopes for the ordinary and inverted least squares regressions. The trend of R increasing linearly with L and a positive intercept agree with results on different human fibroblasts [25]. (c) A constant coupling control efficiency j at declining E (dotted line for sce) is predicted at constant mitochondrial quality with proportional decline of E and L. The relationship between coupling efficiency and E at constant L is hyperbolic (full line fitted for ace). (E-L)50 is the net ET capacity when coupling efficiency is 50%. The maximum coupling efficiency, jmax ≝ 1, is approached with increasing E.

https://doi.org/10.1371/journal.pone.0264496.g004

Low ET capacities in ace are explained by the high oligomycin concentration applied in the XF24, which may inhibit the ET capacity compared with the minimum oligomycin concentration optimized by titrations in each experiment conducted in the O2k. In addition, the trend of increasing respiration in response to the second and low uncoupler titration suggests that O2 flow would further increase in the actual ET state (Fig 2A).

Declining E at constant L lowers the E-L coupling efficiency, j = (E-L)/E, by an ET-linked mechanism. The distinction between ET-linked and uncoupling-linked effects, however, is not apparent from the ratios presented in Table 4B. When E and L are expressed as O2 flow per cell (Table 4A) and vary proportionally due to differences in cell size, mt-density, and noise in the cell count, then coupling efficiency remains constant at variable ET capacity, as seen in sce (Fig 4C). Even when E is underestimated progressively, the drop in (E-L)/E is small initially, since the decline in (E-L) is partially compensated for by the decline of E. The relationship between coupling efficiency and declining E at constant L is hyperbolic. The XF24 data follow this nonlinear model as a separate cluster overlapping with the O2k data at high coupling efficiency (Fig 4C).

Discussion

Cells cultured attached to a physical substrate undergo many changes when they are detached and maintained in suspension, but possible bioenergetic respiratory changes have not yet been addressed. The production of mRNA in fibroblasts is reduced to 20% within a few hours of suspension [10]. The mRNA is not degraded but inactivated and its amount remains constant [10]. The rate of protein synthesis declines more slowly, even slower than expected from the mRNA lifetime of about nine hours [26]. Nevertheless, the suspended cells contain ― after almost 3 days of reduced messenger RNA production ― unchanged levels of cytoplasmic polyadenylated RNA due to a stabilization of mRNA against normal turnover. A continuous decline in protein synthesis starts after 12 h in suspension [10]. The recovery of protein synthesis is rapid after reattachment of cells to a tissue culture dish and reactivation of the sequestered mRNA [27]. These responses of cells to suspension or attachment suggest the possibility that altered physical configuration and cell morphology may induce metabolic responses [10]. To our knowledge, the bioenergetic consequences of suspending fibroblasts grown in a monolayer have not yet been quantitatively assessed, despite numerous studies reporting fibroblast respiration of either suspended or attached cells. Our results showed that ROUTINE respiration was not different in attached cells studied after overnight seeding versus freshly suspended cells measured one hour after harvesting. These findings are relevant for interpreting respirometric short-term studies, but do not exclude severe bioenergetic alterations and even cell death as a consequence of prolonged suspension of fibroblasts. Importantly, cells do not immediately change their morphology when re-plated after trypsinization [28].

For a comparison of respiration of attached versus suspended fibroblasts, it was necessary to use two different platforms. Multiwell microchambers contain attached cells without stirring. In contrast, the 0.5-mL twin chambers that are stirred continuously to maintain cells in homogenous suspension and to avoid O2 diffusion gradients which would compromise high-resolution respirometry. Our approach covered a time range that is relevant for current techniques of measuring respiration in suspended fibroblasts. The duration of maintaining cells in suspension extends from trypsinization to the actual respirometric measurement, which limits the duration to less than two hours from detachment to monitoring of ROUTINE respiration. Importantly, even respiratory measurement of adherent cells in the XF24 did not represent a direct and undisturbed monitoring of oxygen consumption of attached cultured cells but required harvesting of the cells by detachment on the day before measurement. Our study, however, provides the rationale for extended studies of the stress-response of cells suspended for prolonged periods of time.

Primarily, therefore, we had to evaluate the quantitative agreement between measurements with the two instruments. Higher rates of residual oxygen consumption Rox and total LEAK respiration not corrected for Rox were obtained in the XF24. This may be attributed to variable instrumental background rather than the attached state of the cells, which is supported by the fact that Rox-corrected LEAK respiration was not different between attached and suspended cells (ace and sce). Our results can be compared with a large database on respiration of normal human dermal fibroblasts NHDF measured with the Seahorse XF96 Analyzer and used as controls for the diagnosis of inherited mitochondrial diseases [6]. NHDF data from 2630 wells are summarized in a meta-analysis based on BCA [25]. Respiration was normalized for the count of seeded cells (20 000 cells per well in the XF96, compared to 35 000 cells per well in the XF24 and 375 000 cells in the O2k chamber in the present study). After conversion of the NHDF raw data to SI units, deletion of runs with missing data points, further elimination of 7% of outliers ([6] report 17% outliers), and log transformation to account for positive skewness, the NHDF data are expressed as means ± SD after linear back-transformation. R, L and E are 37 ± 12; 6 ± 3; and 78 ± 26 amol∙s-1∙x-1, respectively [25]. This agrees with our results on suspended cells in the O2k and attached cells in the XF24 for R and L (Table 4). Our data on E in the XF24 after 2 μM oligomycin and uncoupler titration up to 0.7 μM were lower (52 ± 8) compared to 78 ± 25 in the XF96 at a lower oligomycin concentration (1 μM) and higher FCCP concentration (1 μM; [6]). This supports the interpretation that ET capacity was underestimated in the present study in the XF24 due to an insufficiently high uncoupler concentration (Fig 2A) combined with a high concentration of oligomycin inhibiting E.

The variability of respiration normalized for the cell count is of interest from two points of view. (1) A methodological perspective: Is high variability mainly the result of respirometric noise, noise in the cell count, or variability introduced by the addition of cells to the respirometric chambers or wells? Is the reproducibility linked to specific techniques and to the magnitude of the cell count used in an assay? (2) A physiological perspective: Is the variability factual rather than artefactual [25]? A possible reason for higher variability of residual oxygen consumption obtained in the XF24 is the instrumental background O2 rate, which is not determined in the experimental wells but is measured in parallel only in four separate control wells without cells (Table 1).

Taken together, neither LEAK respiration nor ROUTINE respiration were different when comparing attached and suspended cells. Bioenergetic cluster analysis (BCA) identified the lower coupling control efficiency obtained in the XF24 to be caused by an underestimation of ET capacity as opposed to uncoupling (Fig 4C). Further studies of respiration in different attached and suspended cell types are of great interest in cell physiology, particularly in neuronal and blood cells, and importantly, in cancer cell lines and other cell models of disease. The present approach includes BCA and provides a guideline for extending databases on cell respiration across instrumental platforms, emphasizing the importance of harmonization of protocols.

References

  1. 1. Lazdins JK, Koech DK, Karnovsky ML. Oxidation of glucose by mouse peritoneal macrophages: a comparison of suspensions and monolayers. J Cell Physiol. 1980 Nov;105(2):191–6. pmid:6780573
  2. 2. Pofit JF, Strauss PR. Membrane transport by macrophages in suspension and adherent to glass. J Cell Physiol. 1977 Aug;92(2):249–55. pmid:881435
  3. 3. Harrison CJ, Allen TD. Cell surface morphology after trypsinisation depends on initial cell shape. Differentiation. 1979;15(1):61–6. pmid:527772
  4. 4. Park JY, Jeong AL, Joo HJ, Han S, Kim SH, Kim HY, et al. Development of suspension cell culture model to mimic circulating tumor cells. Oncotarget. 2017 Dec 7;9(1):622–640. pmid:29416640
  5. 5. Shen CF, Guilbault C, Li X, Elahi SM, Ansorge S, Kamen A, et al. Development of suspension adapted Vero cell culture process technology for production of viral vaccines. Vaccine. 2019 Nov 8;37(47):6996–7002. pmid:31288997
  6. 6. Yépez VA, Kremer LS, Iuso A, Gusic M, Kopajtich R, Koňaříková E, et al. OCR-Stats: Robust estimation and statistical testing of mitochondrial respiration activities using Seahorse XF Analyzer. PLoS One. 2018 Jul 11;13(7): e0199938. pmid:29995917
  7. 7. Danhelovska T, Kolarova H, Zeman J, Hansikova H, Vaneckova M, Lambert L, et al. Multisystem mitochondrial diseases due to mutations in mtDNA-encoded subunits of complex I. BMC Pediatr. 2020 Jan 29; 20(1):41. pmid:31996177
  8. 8. Burska D, Stiburek L, Krizova J, Vanisova M, Martinek V, Sladkova J, et al. Homozygous missense mutation in UQCRC2 associated with severe encephalomyopathy, mitochondrial complex III assembly defect and activation of mitochondrial protein quality control. Biochim Biophys Acta Mol Basis Dis. 2021 Aug 1;1867(8):166147. pmid:33865955
  9. 9. Ye F, Hoppel CL. Measuring oxidative phosphorylation in human skin fibroblasts. Anal Biochem. 2013; 437:52–8. pmid:23462540
  10. 10. Benecke BJ, Ben-Ze’ev A, Penman S. The control of mRNA production, translation and turnover in suspended and reattached anchorage-dependent fibroblasts. Cell. 1978 Aug; 14(4):931–9. pmid:688399
  11. 11. Farmer SR, Ben-Ze’ev A, Benecke BJ, Penman S. (1978). Altered translatability of messenger RNA from suspended anchorage-dependent fibroblasts: reversal upon cell attachment to a surface. Cell. 1978 Oct 1; 15(2):627–37. pmid:719755
  12. 12. Wu M, Neilson A, Swift AL, Moran R, Tamagnine J, Parslow D, et al. Multiparameter metabolic analysis reveals a close link between attenuated mitochondrial bioenergetic function and enhanced glycolysis dependency in human tumor cells. Am J Physiol Cell Physiol. 2007 Jan;292(1):C125–36. pmid:16971499
  13. 13. Doerrier C, Garcia-Souza LF, Krumschnabel G, Wohlfarter Y, Mészáros AT, Gnaiger E. High-Resolution FluoRespirometry and OXPHOS Protocols for Human Cells, Permeabilized Fibers from Small Biopsies of Muscle, and Isolated Mitochondria. Methods Mol Biol. 2018;1782:31–70. pmid:29850993
  14. 14. Gnaiger E. Mitochondrial pathways and respiratory control. An introduction to OXPHOS analysis (5th ed.). Bioenerg Commun. 2020; 2, 112 pp.
  15. 15. Zhang J, Nuebel E, Wisidagama DR, Setoguchi K, Hong JS, Van Horn CM, et al. Measuring energy metabolism in cultured cells, including human pluripotent stem cells and differentiated cells. Nat Protoc. 2012 May 10; 7(6):1068–85. pmid:22576106
  16. 16. Gnaiger E. Polarographic oxygen sensors, the oxygraph and high-resolution respirometry to assess mitochondrial function. In: Mitochondrial dysfunction in drug-induced toxicity (Dykens JA, Will Y, eds) John Wiley & Sons, Inc, Hoboken, NJ,2008; 327–52. https://doi.org/10.1002/9780470372531.ch12
  17. 17. Gnaiger E. O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondrial Physiology Network. 2019; 1–19. [posted 2019, March 6, revised 2021 January 19, cited 2021 September 10] Available from: https://wiki.oroboros.at/index.php/MiPNet06.03_POS-calibration-SOP
  18. 18. Gnaiger E. Bioenergetics at low oxygen: dependence of respiration and phosphorylation on oxygen and adenosine diphosphate supply. Respir Physiol. 2001 Nov 15;128(3):277–97. pmid:11718759
  19. 19. Scandurra FM, Gnaiger E. Cell respiration under hypoxia: facts and artefacts in mitochondrial oxygen kinetics. Adv. Exp. Med. 2010; 662: 7–25. pmid:20204766
  20. 20. Komlódi T, Sobotka O, Gnaiger E. Facts and artefacts on the oxygen dependence of hydrogen peroxide production using Amplex UltraRed. Bioenerg Commun. 2021; 4.
  21. 21. Hutter E, Renner K, Pfister G, Stöckl P, Jansen-Dürr P, Gnaiger E. Senescence-associated changes in respiration and oxidative phosphorylation in primary human fibroblasts. Biochem J. 2004 Jun 15;380(Pt 3):919–28. pmid:15018610
  22. 22. Gnaiger E.–MitoEAGLE task group. Mitochondrial physiology. Bioenergetics Communication, 2020; 1, 1–44.
  23. 23. Agilent Technologies . Mito Stress Test Kit, User Guide. 2019. [cited 2021 September 10]. Available from: https://www.agilent.com/cs/library/usermanuals/public/XF_Cell_Mito_Stress_Test_Kit_User_Guide.pdf
  24. 24. Gnaiger, E. (2021) The elementary unit—canonical reviewer’s comments on: Bureau International des Poids et Mesures (2019) The International System of Units (SI) 9th ed. MitoFit Preprints 2020.4.v2 [preprint]. Available from: https://doi.org/10.26124/mitofit:200004.v2
  25. 25. Gnaiger E. Bioenergetic cluster analysis–mitochondrial respiratory control in human fibroblasts. MitoFit Preprints [preprint]. 2021; 8.
  26. 26. Abelson HT, Johnson LF, Penman S, Green H. Changes in RNA in relation to growth of the fibroblast: II. The lifetime of mRNA, rRNA, and tRNA in resting and growing cells. Cell. 1974 Apr 01; 1(40), 161–5.
  27. 27. Ben-Ze’ev A., Reiss R., Bendori R., Gorodecki B. (1990). Transient induction of vinculin gene expression in 3T3 fibroblasts stimulated by serum-growth factors. Cell Regulation. 1990 Aug 1; 1(9), 621–6. pmid:2127699
  28. 28. Benecke BJ, Ben-Ze’ev A, Penman S. The regulation of RNA metabolism in suspended and reattached anchorage-dependent 3T6 fibroblasts. J Cell Physiol. 1980 May;103(2):247–54. pmid:6160160