A new inertial navigation system for guiding implant placement. An in-vitro proof-of-concept study

  • G. Esteve-Pardo ,

    Contributed equally to this work with: G. Esteve-Pardo, L. Esteve-Colomina, E. Fernández

    Roles Conceptualization, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing

    guillemjoes@gmail.com

    Current address: Clínica Dental Esteve, Alicante, Spain

    Affiliation Bioengineering Institute, Miguel Hernández University of Elche, Group Aula Dental Avanzada, Alicante, Spain

  • L. Esteve-Colomina ,

    Contributed equally to this work with: G. Esteve-Pardo, L. Esteve-Colomina, E. Fernández

    Roles Conceptualization, Data curation, Formal analysis, Supervision, Validation, Writing – original draft, Writing – review & editing

    Current address: Clínica Dental Esteve, Alicante, Spain

    Affiliation Bioengineering Institute, Miguel Hernández University of Elche, Group Aula Dental Avanzada, Alicante, Spain

  • E. Fernández

    Contributed equally to this work with: G. Esteve-Pardo, L. Esteve-Colomina, E. Fernández

    Roles Conceptualization, Data curation, Resources, Software, Supervision, Validation, Visualization, Writing – review & editing

    Current address: Bioengineering Institute, Miguel Hernández University of Elche, Elche, Alicante, Spain

    Affiliation Bioengineering Institute, Miguel Hernández University of Elche and CIBER-BBN, Elche, Spain

Abstract

The aim of this study was to assess the potential use of a new advanced inertial navigation system for guiding dental implant placement and to compare this approach with standard stereolithographic template guiding. A movement processing unit with a 9-axis absolute orientation sensor was adapted to a surgical handpiece and wired to a computer navigation interface. Sixty implants were placed by 10 operators in 20 jaw models. The 30 implants of the test group were placed in 10 models guided by the new inertial navigation prototype. The 30 implants of the control group were placed in another 10 models using a CAD-CAM template. Both groups were subdivided into experienced and non-experienced operators. Pre- and postoperative computer tomography images were obtained and matched to compare the planned and final implant positions. Four deviation parameters (global, angular, depth, and lateral deviation) were defined and calculated. The primary outcome was the angular deviation between the standard stereolithographic approach and the new inertial navigation system. Results showed no significant differences between both groups, suggesting that surgical navigation based on inertial measurement units (IMUs) could potentially be useful for guiding dental implant placement. However, more studies are still needed to translate this new approach into clinical practice.

Introduction

It is generally believed that prosthetically driven implant placement is a prerequisite to achieve the precise positioning required for the final restoration [1]. The proper placement of implants is also necessary to prevent future technical and biological complications [2]. However, the accuracy of implant placement is inexorably subjected to the so-called “human factors” [3]. In this sense, virtual planning software and computer-assisted surgery have significantly improved the accurate transfer of implants from the pre-planned position to the final surgical position.

Computer-guided implant placement can be performed either statically via splints or by the dynamic navigation devices. Guiding by both computer-aided design and computer-aided manufacturing (CAD-CAM) surgical guides [48] or by dynamic navigation [913] was found to achieve a higher precision of the target implant position compared with conventional free-hand implant placement.

In the last few decades, several dynamic navigation systems based on different tracking technologies have been developed. Optical tracking systems are currently the most used in implant dentistry [1214]. Navigation systems based on electromagnetic sensors are well established mostly in minimally invasive surgery [15], but their application to implant dentistry has the drawback of distortions from interference of neighbouring medical devices or ferromagnetic objects [16]. Hybrid devices combining both trackers in a magneto-optic hybrid system have been validated for maxillofacial surgery [17,18].

As in the case of template guiding, implant placement by surgical navigation has been shown to be more accurate, faster, and easier than conventional non-guided surgery [19]. Despite these benefits, the added cost, the complex ergonomics required, and the unavoidable learning curve [19] result in these technologies being dismissed in the clinical routine. According to market analysis (iData Research Inc.), in the US, computer-guide surgery will reach only 3.7% of the total projected 2021 implant market size [20].

The inertial measurement unit (IMU) is a combination of three sensors simultaneously detecting the velocity, orientation, and gravitational force of an object. The linear acceleration is measured by accelerometers, the rotational rate is measured by gyroscopes, and additionally the heading reference can be provided by a magnetometer that can be included in the same microchip. Micro Electro Mechanical Systems (MEMS) technology has successfully reduced the size of these sensors to a sub-millimetre scale, and their cost to the same extent.

IMUs have been widely used in navigation, to operate air vehicles, and in areas such as manufacturing and robotics. They still show a great potential for further developments and applicability due to their simplicity, miniature size, low cost, and easy of use [21]. The raw measurements of the three kinds of sensors are filtered and merged by the software to calculate attitude, angular rates, linear velocity, and relative position within a magnetic field. In short, the computer can follow the movement of the object in real time for each of the three axes known as pitch, roll, and yaw in a vehicle [22].

Until now, inertial sensing has been used in various hybrid surgical navigation systems. In these systems, IMUs functioned to compensate for electromagnetic sensor errors, or to support for partial occlusions in optical tracking devices. Preliminary clinical studies have been conducted in various surgical fields like endoscopic and laparoscopic surgery [2325] and orthopedic surgery [2629].

Due to their characteristics, IMUs can also be applicable to implant surgery. But, to our knowledge, its presumed possibilities have not yet been explored in this field, beyond an in vitro study [30].

The present study aims to validate the potential usefulness of IMU technology to assist in the placement of dental implants. The accuracy in transferring implants to the planned position with the help of an inertial device was evaluated and compared with a “gold-standard” control, the placement of implants using stereolithographic templates. Our results suggest that an inertial guide system could be as accurate as standard surgical templates.

Materials and methods

Twenty identical jaw models were designed using Autodesk Inventor 2016 (Autodesk Inc.), manufactured with Fused Filament Fabrication (FFF) 3D printer (BCN3D Sigma R17) and polylactic acid (PLA) resin (BCN PLA 2,85 mm). Prints were made with 0,1 mm layer height and positioning resolution (X/Y/Z) of 12,5 μm/12,5 μm/1 μm.

Since maxillary models are more difficult to align we only used mandibular models. No attempt was made to emulate the anatomy in detail, since our main goal was to assess the precision of implant placement with respect to the planned positions. The models had an extra posterior space to place a digital inclinometer BDJK model SH-5339-90 (Shenzhen Jieshun Science and Technology Industry Co., Ltd). An inclinometer was used to assure the reproducibility during the scanning radiology and later during the drilling process at 0°. Three shallow holes were prepared in the models to facilitate the location of the planned osteotomies’ starting points. Furthermore, three triangular protrusions enabled a firm hold on the splint used for drilling and included two vertical pillars with concavities to position and calibrate the IMU adapter to the surgical handpiece.

Computer scans of the models were obtained using cone beam computed tomography (CBCT) scans with the help of Planmeca ProMax 3D Plus (Planmeca Oy), in a 0° position and the following parameters: 90 kV, 10 mA, FOV 200 x 100 mm and 600 μm voxel size. Five hundred one slices were obtained per CBCT image and model. The digital imaging and communication on medicine (DICOM) files (Fig 1) were then exported to the software planner Simplant Pro v. 17.01 (Dentsply Sirona Inc.), where three 4.2 x 13 mm Dentsply Astra Tech EV implants were planned with different angulations. According to these planned parameters, a Simplant Safe Guide was fabricated (Materialise NV). The drilling protocol of the Astra Tech Implant System EV was followed to place the implants into the models using the fully guiding template.

thumbnail
Fig 1. Different views of the jaw model.

Initial digital design of the model (A) with the internal honeycomb design (B) to allow implant bed preparation and insertion. DICOM views (Sagittal (C), coronal (D), axial (E) and tridimensional rendering (F) from the jaw model prior to planning implant positions with the Simplant Pro v.17.01 software planner.

https://doi.org/10.1371/journal.pone.0255481.g001

A high-performance orientation sensor, able to estimate 3D orientation in space, was used for real-time surgical navigation. The selected device was the Adafruit BNO 055 (Adafruit Industries), which contains a microprocessor and 3 specific microsensors: a triaxial 16-bit gyroscope, a triaxial 14-bit accelerometer, and a triaxial geomagnetic sensor, totalling 9 degrees of freedom (DoF). The microprocessor is a 32-bit a high-speed ARM Cortex-M0 microcontroller that collects raw data from the 3 sensors. The Adafruit BNO 055 was adapted to the surgical handpiece (contra-angle) with which the drilling of the models was performed.

A programmable electronic board based on the microcontroller ATmega328P (Arduino Uno R3) was connect to the Adafruit BNO055 absolute orientation sensor by a serial data bus using the I2C protocol. Specific algorithms for controlling the position of the contra-angle were developed using the Integrated Development Environment Arduino v. 1.8.1 (S4 File). Using this approach, the angulation of the handpiece, and therefore of the drill, can be obtained and easily transferred to the navigator in real time (Fig 2).

thumbnail
Fig 2. Surgical template and navigator´s graphical interface.

(A) Surgical template adapted on the model. (B) Screen of the real time navigation software with IMU adapted to handpiece for drilling the models. The graphical interface includes the location of the handpiece and the drill together with the planned implant axis, enabling the operator to align both indications in real time. (C and D) Pictures of implant placements using the navigator’s graphical interface.

https://doi.org/10.1371/journal.pone.0255481.g002

The study was performed in accordance with the guidelines and regulations within the training framework of the Master in Implantology programme and approved by the Research Ethics Board of the University Miguel Hernández (Elche, Alicante, Spain). Ten subjects were randomly recruited among teachers and trainees enrolled in this programme. We obtained informed consent of all the subjects before any study procedure was conducted. According to their prior experience in placing implants, the subjects were distributed into two subgroups of 5 persons each: the “experienced operators” (ExO) who had previously placed more than 100 implants in their private practice and the “non-experienced operators” (NExO), who had previously placed less than 20 implants. This second subgroup was composed of recent graduate students enrolled in the programme.

Each subject performed the simulated osteotomy in 2 jaw models, placing 3 implants in each model. The first surgeries were performed using the standard CAD-CAM guiding templates. After finishing the first 3 implants, the subjects had a 30-minute break. Then, they were instructed to use the real-time navigation device to place the remaining 3 implants in the other model.

All the implants were placed the same day, under the same circumstances, and following the same drilling protocol. The drilling sequence used was that recommended by the manufacturer for the Astra Tech EV implant system. For the 4.2 mm wide implant used in this study, drills 1, 3, and 4 were successively used. When drilling through the template, the respective metal sleeves were used for each drill width.

Before each drill, we checked the right alignment of the model as well as all the parameters of the surgical navigation guide and/or the absence of movement of the surgical guide. The reproducibility of the template position was facilitated by the design of the model with 3 triangular protrusions and the total rigidity and absence of mobility of the support.

Finally, a questionnaire was distributed to all participants to assess their opinions regarding (1) feeling of support during drilling, (2) ease of use of the navigation prototype, and (3) general satisfaction with the systems.

All 20 models with the implants placed were scanned with the Planmeca ProMax 3D Plus CBCT imaging unit. Then, DICOM files were imported one by one into Simplant Pro, v 17.01, to obtain the Sprite (SPR) files. Furthermore, an additional SPR file of the initial planning was obtained. The SPR files of the drilled models were superimposed one by one onto the initial planned file. The software Simplant Internal Master (Materialise NV) was used to merge the pre- and postoperative CBCT files using an iterative closest point algorithm (ICP) method. Then, the angular differences between the pre- and postoperative implant positions were compared and automatically analysed by the software. Global, angular, depth and lateral deviation parameters were defined and calculated between the planned and the placed implants, for the apical and coronal positions of the implants (Fig 3).

thumbnail
Fig 3. Measures to compare planned with placed implant positions.

α: angular deviation (degrees); a: Global deviation at coronal point (mm), b: Lateral deviation at coronal point (mm), c: Depth deviation at coronal point (mm); a1: Global deviation at apical point (mm), b1: Lateral deviation at apical point (mm) c1: Depth deviation at apical point (mm).

https://doi.org/10.1371/journal.pone.0255481.g003

Statistical analysis was performed using IBM SPSS Software (Version 25.0, IBM Corp.). Data were described with regard to mean values, standard deviations, and 95% confidence intervals. A one-way ANOVA test was used to determine whether there were statistically significant differences between the means of different values. In all cases, p < 0.05 was considered statistically significant.

Results

Sixty implants placed in 20 identical jaw models were included in this study. Each subject placed 3 implants using standard CAD-CAM stereolithographic templates and 3 implants using the novel real-time navigation system. Table 1 shows the mean values for the global, angular, depth, and lateral deviation parameters between the planned and the placed implants for the apical and coronal positions.

thumbnail
Table 1. Mean deviations between implant planned and placed positions in the control (SG) and experimental (IMU) groups.

https://doi.org/10.1371/journal.pone.0255481.t001

The primary goal of the study was to comparatively assess the mean angular deviation of the implants from the planned positions related to its final placement for both techniques, stereolithographic guide (SG control group) and novel real-time navigation device (IMU experimental group). The average coronal angle deviation was 5.63 ± 1.41° (SD 3.94; 95% CI, 4.22–7.04) with the surgical template (SG control group) and 7.13 ± 1.47° (SD 4.1; 95% CI, 5.66–8.6) using the navigation prototype (IMU experimental group).

The linear deviations were very similar for implants placed using CAD-CAM templates and the new IMU devices, independently of users’ previous experience (p = 0.319). For example, the mean global coronal lateral deviations, between the 2 centre points of the implant platforms were 1.42±0.2 mm and 1.48±0.2 mm for the SG control group and IMU experimental group, respectively. There were also no significant differences in the linear deviations in the apex (p = 0.777). Thus, linear deviations in the apex were 2.07±0.35 mm for the SG and 2.00±0.33 mm, for the IMU group.

Table 2 shows the averaged mean angular deviations according to the type of implant for experienced and non-experienced users. No statistically significant differences could be detected between each type of implant for implants placed using standard CAD-CAM stereolithographic templates and implants placed using the novel real-time navigation system based on IMUs (p>0.13). This was valid for both experienced and non-experienced subjects.

thumbnail
Table 2. Mean angular deviations in the SG (stereolithographic guide) control group and in the IMU (inertial navigation) test group.

https://doi.org/10.1371/journal.pone.0255481.t002

On the other hand, and regarding the use of the new IMU device, 70% of the subjects found it highly satisfactory. However, 30% of the subjects considered that previous training was necessary to achieve better results.

Discussion

The aim of the present work was to evaluate the possible usefulness of IMU sensors to assist surgical navigation to place implants. We compared the precision achieved by the testing inertial navigation device (IMU) with implant placement guided by CAD-CAM templates as a control group (SG).

Because the highest clinical accuracy was reported using the tooth-supported guide [31,32], 3 projections were added to the models to stabilize the support of the guide during the osteotomy, thus simulating a tooth-supported guide. Although significantly greater deviations have been shown in partially guided, compared with fully guided, surgery [33], in the present study, partial guiding was used to provide a better comparison setting between static and dynamic guiding.

The average deviation of the implant-achieved positions using the CAD-CAM guides was found to be in the range of 1 and 2 mm and about 5° of angulation [5,10,12]. In the present experiment, the mean intergroup measures were 6.38° of angular deviation, 1.45 mm of global coronal deviation, and 2.00 mm in the apex, figures consistent with mean data reported in the literature.

However, the average data are not totally relevant; it would be more clinically significant to consider the maximum deviations occasionally reached using the procedure, given the potential complications that could involve. In the present study, the maximum angular deviations were 17.34° (IMU) and 16.68° (SG), and the maximum lateral deviations in the apex were 4.3 mm (IMU) and 4.5 mm (SG), all by a non-experienced operator. Although high, these figures are in line with the reported maximal inaccuracies in the literature. Indeed, 24.9° of deviation in implant angulation [6] and more than 7 mm of deviation in the implant apex [5] have been reported.

The subject’s experience and the learning curve have been considered factors influencing the precision of the guided surgery. Dynamically guided surgery has been shown to require a longer learning curve prior to use than splint surgery [34]. Two in vitro studies using surgical templates showed significantly greater deviations by non-experienced users [35,36]. However, in a similar setting to that of the present study, the subject’s experience was not demonstrated to be a critical factor in achieving the accuracy of implant position [37]. This lack of consistency could be explained by the design of the guide. In the former studies, guides were not fully supported in their distal ends, whereas the latter study used a tooth-supported guide. A similar in-vitro study using fully guiding tooth-supported templates, found only 0.5° more angular deviation for non-experienced users. The authors concluded that these slight differences could be clinically irrelevant [38].

However, there are many other confounding factors potentially influencing these differences, such as the hardness of the resin model, the congruence between sleeves and drills, the position of the model on the bench, or the way the operator holds the guide in place while drilling, to name just a few. With regard to the shape of the jaw models, the present study only focused on evaluating the levels of precision for IMU navigation in vitro and not the implication of anatomical limitations. This is why a simple anatomy has been used without taking into account the detailed characteristics of the jaws after the various situations of edentulism.

Likewise, the placement of the model on a table differs from the real situation in the clinical setting. Despite this, this position allows us a strict correlation with the radiographic acquisition and the planned angulation of the implants.

In future developments of the system, the use of manikin-anchored jaws could be relevant in the approximation of a more realistic environment.

In the present experiment, using a tooth-supported partial guide, non-significant differences between experienced and non-experienced users were found. However, the most deviated values, or outliers, were obtained in the non-experienced users’ group. In this context, it is possible that experience and training may have some influence, at least in preventing the extreme deviations of implant positions.

With the inertial navigation prototype, the perception regarding its ease of use was “satisfactory” for 80% of non-experienced subjects. This qualitative evaluation was consistent with the study by Goodacre et al, where a similar navigation device was valued as “a positive aid to place implants by non-experienced operators [23]. However, these authors only performed a comparison between freehand and inertial dynamic guided drilling. Neither splits were used, nor were implants placed, and the models had a middle line pin as a visual orientation for operators.

In the present study, the inertial navigation device only informed the subjects about the proper angulation of the drilling. The initial point of drilling was determined by a mark in the model, and operators had to match the drill tip with the mark. Although there was no exact transfer of these points, the present study focused mainly on angular deviations. In the clinical setting, the previously planned initial point to drill and the depth of the implant are 2 easily reproducible factors. Given that dynamic navigation is an open process, not blinded like a surgical template, the implantation area is directly visualized, and these 2 planned aspects of the implant position can be easily decided at the time of surgery [39]. By contrast, the osteotomy to place an implant must be adapted to the different angulations of the anatomy. Therefore, the tridimensional angulation of drilling could be considered highly relevant and the factor that will determine the clinical outcome to a greater extent [11], provided that the initial point of insertion and the final depth were properly achieved.

With regard to costs, in all cases, guided surgery presents an added cost. Static guiding involves the cost of manufacturing the CAD-CAM stent. Some authors described different analogical methods trying to reduce the manufacturing cost of the guide, but all should still be scientifically validated. Currently, the cost of the guide can be definitely lowered by the generalization of the 3D printing technology, paving the way to design and manufacture the templates directly in the dental office [4042]. On the other hand, in dynamic navigation, cost derives from the acquisition of the equipment and its software that, in the case of optical navigation, can be as high as several tens of thousands of euros [43]. Furthermore, an additional cost of about 50 euros per patient should be added for manufacturing the individual plates with the fiduciary markers [44]. Such higher costs are perhaps the main limitation for an extensive use of dynamic navigation guided-surgery. Generally speaking, the additional costs of guided-surgery procedures may exceed the patient’s financial resources [45].

It might be thought that simplifying the guidance system, while maintaining its precision, would facilitate its wide acceptance by clinicians. Thus, the better implant positioning obtained with guided-surgery would result in fewer complications for patients on a larger scale. Undoubtedly, future research in the scope of computer-assisted implant surgery will aim to reduce the costs and increase the simplicity of the systems. This is the final objective of testing the potential clinical application of inertial navigation in the present study. The prototype device has been shown to be feasible to place implants with precision similar to the stereolithographic guides, the current standard. The new system can be simpler, less expensive, and easier to use, provided that the interface can be improved and the operator respects the necessary learning curve of use.

In conclusion, this proof-to-concept study did not find significant differences between test and control groups. Therefore, the precision of the novel surgical navigation tested in the study, based on IMUs, can be said to be comparable to the precision data from the current gold-standard guiding system. Within its limitations, the present study opened the way to consider surgical navigation based on IMUs as a potential tool to guide implant placement. However, more studies are still needed to translate this new approach from an in vitro setting into clinical practice.

Supporting information

S1 File. Descriptive data and distribution box plot charts.

https://doi.org/10.1371/journal.pone.0255481.s001

(PDF)

Acknowledgments

We would like to thank Alejandro García and Juan Sebastian Peña for all his help with the development of specific algorithms for controlling the position of the contra-angle.

References

  1. 1. Garber DA, Belser UC. Restoration-driven implant placement with restoration-generated site development. Compend Contin Educ Dent. 1995 Aug;16(8):796–804. pmid:8620398
  2. 2. Berglundh T, Persson L, Klinge B. A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years. J Clin Periodontol. 2002;29 Suppl 3:197–212; discussion 232–3. pmid:12787220
  3. 3. Renouard F, Amalberti R, Renouard E. Are "Human Factors" the Primary Cause of Complications in the Field of Implant Dentistry? Int J Oral Maxillofac Implants. 2017 Mar/Apr;32(2):e55–e61. pmid:28291846
  4. 4. Arisan V, Karabuda CZ, Mumcu E, Özdemir T. Implant positioning errors in freehand and computer-aided placement methods: a single-blind clinical comparative study. Int J Oral Maxillofac Implants. 2013 Jan-Feb;28(1):190–204. pmid:23377066
  5. 5. Tahmaseb A, Wismeijer D, Coucke W, Derksen W. Computer technology applications in surgical implant dentistry: a systematic review. Int J Oral Maxillofac Implants. 2014;29 Suppl:25–42. pmid:24660188
  6. 6. Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implants Res. 2012 Oct;23 Suppl 6:112–23. pmid:23062136
  7. 7. Vercruyssen M, Hultin M, Van Assche N, Svensson K, Naert I, Quirynen M. Guided surgery: accuracy and efficacy. Periodontol 2000. 2014 Oct;66(1):228–46. pmid:25123771
  8. 8. Smitkarn P, Subbalekha K, Mattheos N, Pimkhaokham A. The accuracy of single-tooth implants placed using fully digital-guided surgery and freehand implant surgery. J Clin Periodontol. 2019 Sep;46(9):949–957. pmid:31241782
  9. 9. Jung RE, Schneider D, Ganeles J, Wismeijer D, Zwahlen M, Hämmerle CH, et al. Computer technology applications in surgical implant dentistry: a systematic review. Int J Oral Maxillofac Implants. 2009;24 Suppl:92–109. pmid:19885437
  10. 10. Somogyi-Ganss E, Holmes HI, Jokstad A. Accuracy of a novel prototype dynamic computer-assisted surgery system. Clin Oral Implants Res. 2015 Aug;26(8):882–890. pmid:24837492
  11. 11. Emery RW, Merritt SA, Lank K, Gibbs JD. Accuracy of Dynamic Navigation for Dental Implant Placement-Model-Based Evaluation. J Oral Implantol. 2016 Oct;42(5):399–405. pmid:27267658
  12. 12. Block MS, Emery RW, Lank K, Ryan J. Implant Placement Accuracy Using Dynamic Navigation. Int J Oral Maxillofac Implants. 2017 Jan/Feb;32(1):92–99. pmid:27643585
  13. 13. Stefanelli LV, Mandelaris GA, DeGroot BS, Gambarini G, De Angelis F, Di Carlo S. Accuracy of a Novel Trace-Registration Method for Dynamic Navigation Surgery. Int J Periodontics Restorative Dent. 2020 May/Jun;40(3):427–435. pmid:32233198
  14. 14. Chen X, Lin Y, Wu Y, Wang C. Real-time motion tracking in image-guided oral implantology. Int J Med Robot. 2008 Dec;4(4):339–47. pmid:18803338
  15. 15. Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L. Electromagnetic tracking in medicine-a review of technology, validation, and applications. IEEE Trans Med Imaging. 2014 Aug;33(8):1702–25. pmid:24816547
  16. 16. Sorriento A, Porfido MB, Mazzoleni S, Calvosa G, Tenucci M, Ciuti G, et al. Optical and Electromagnetic Tracking Systems for Biomedical Applications: A Critical Review on Potentialities and Limitations. IEEE Rev Biomed Eng. 2020;13:212–232. pmid:31484133
  17. 17. Nakada K, Nakamoto M, Sato Y, Konishi K, Hashizume M, Tamura S. A rapid method for magnetic tracker calibration using a magneto-optic hybrid tracker. In: Ellis RE, Peters ™, editors. Medical Image Computing and Computer-Assisted Intervention—MICCAI 2003. MICCAI 2003: Lecture Notes in Computer Science vol 2879. Berlin: Springer; 2003. p. 285–93.
  18. 18. Berger M, Nova I, Kallus S, Ristow O, Eisenmann U, Freudlsperger C, et al. Electromagnetic navigated positioning of the maxilla after Le Fort I osteotomy in preclinical orthognathic surgery cases. Oral Surg Oral Med Oral Pathol Oral Radiol. 2017 Mar;123(3):298–304. pmid:28049608
  19. 19. Block MS, Emery RW. Static or Dynamic Navigation for Implant Placement-Choosing the Method of Guidance. J Oral Maxillofac Surg. 2016 Feb;74(2):269–77. pmid:26452429
  20. 20. iData Research. Computer-Guided Surgery Market Size, Share & COVID19 Impact Analysis | United States. 2017| 2021–2027. https://idataresearch.com/product/computer-guided-surgery-market-united-states.
  21. 21. Caldas R, Mundt M, Potthast W, Buarque de Lima Neto F, Markert B. A systematic review of gait analysis methods based on inertial sensors and adaptive algorithms. Gait Posture. 2017 Sep;57:204–210. pmid:28666178
  22. 22. Szczęsna A, Skurowski P, Lach E, Pruszowski P, Pęszor D, Paszkuta M, et al. Inertial Motion Capture Costume Design Study. Sensors (Basel). 2017 Mar 17;17(3):612. pmid:28304337
  23. 23. Goodacre B, Mashni J, Yankee J, Goodacre C, Lozada J, Won J. The effect of angulation sensors on implant placement. J Calif Dent Assoc. 2013 Jan;41(1):47–50. pmid:23437606
  24. 24. Ando T, Kiyomatsu H, Kobayashi E, Sakuma I. In vivo evaluation of a new tool tracking system for laparoscopic surgery based on inertial sensor. Int J Comput Assist Radiol Surg. 2015;10(Suppl 1):S54–5
  25. 25. Höller, K, Penne J, Schneider A, et al. Endoscopic orientation correction. In: International Conference on Medical Image Computing and Computer-Assisted Intervention 2009. (pp. 459–466). Springer, Berlin, Heidelberg).
  26. 26. Lee HC, Jung CW, Kim HC. Real-time endoscopic image orientation correction system using an accelerometer and gyrosensor. Plos one, 2017;12(11), e0186691. pmid:29099845
  27. 27. Laoruengthana A, Rattanaprichavej P, Tantimethanon T et al. Usefulness of an accelerometer-based navigation system in bilateral one-stage total knee arthroplasty. BMC Musculoskelet Disord 2021;22:164. pmid:33568132
  28. 28. Suda Y, Takayama K, Ishida K et al. Improved implant alignment accuracy with an accelerometer-based portable navigation system in medial unicompartmental knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 2020;28:2917–23. pmid:31451844
  29. 29. Ueyama H, Minoda Y, Sugama R, Ohta Y, Yamamura K, Nakamura S, et al. An accelerometer-based portable navigation system improved prosthetic alignment after total knee arthroplasty in 3D measurements. Knee Surg Sports Traumatol Arthrosc. 2019;27(5):1580–6. pmid:30066018
  30. 30. Minoda Y, Hayakawa K, Hagio K, Konishi N, Tamaki T, Iwakiri K. Multicenter RCT Group Usefulness of an Accelerometer-Based Portable Navigation System for Total Knee Arthroplasty. J Bone Joint Surg. 2020;102(22):1993–2000. pmid:33208642
  31. 31. Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implants Res. 2017 May;28(5):602–612. pmid:27062555
  32. 32. El Kholy K, Lazarin R, Janner SFM, Faerber K, Buser R, Buser D. Influence of surgical guide support and implant site location on accuracy of static Computer-Assisted Implant Surgery. Clin Oral Implants Res. 2019 Nov;30(11):1067–1075. pmid:31381178
  33. 33. Zhou W, Liu Z, Song L, Kuo CL, Shafer DM. Clinical Factors Affecting the Accuracy of Guided Implant Surgery-A Systematic Review and Meta-analysis. J Evid Based Dent Pract. 2018 Mar;18(1):28–40. pmid:29478680
  34. 34. Wu D, Zhou L, Yang J, Zhang B, Lin Y, Chen J, et al. Accuracy of dynamic navigation compared to static surgical guide for dental implant placement. Int J Implant Dent. 2020 Nov 24;6(1):78. pmid:33230662
  35. 35. Cushen SE, Turkyilmaz I. Impact of operator experience on the accuracy of implant placement with stereolithographic surgical templates: an in vitro study. J Prosthet Dent. 2013 Apr;109(4):248–54. pmid:23566606
  36. 36. Toyoshima T, Tanaka H, Sasaki M, Ichimaru E, Naito Y, Matsushita Y, et al. Accuracy of implant surgery with surgical guide by inexperienced clinicians: an in vitro study. Clin Exp Dent Res. 2015 Jul 14;1(1):10–17. pmid:29744135
  37. 37. Lin CC, Ishikawa M, Huang BH, Huang MS, Cheng HC, Maida T, et al. In Vitro Accuracy of Static Guided Implant Surgery Measured by Optical Scan: Examining the Impact of Operator Experience. Appl Sci. 2020; 10(8):2718.
  38. 38. Fernández-Gil Á, Gil HS, Velasco MG, Moreno Vázquez JC. An In Vitro Model to Evaluate the Accuracy of Guided Implant Placement Based on the Surgeon’s Experience. Int J Oral Maxillofac Implants. 2017 May/Jun;32(3):151–154. pmid:28494035
  39. 39. Jorba-García A, Figueiredo R, González-Barnadas A, Camps-Font O, Valmaseda-Castellón E. Accuracy and the role of experience in dynamic computer guided dental implant surgery: An in-vitro study. Med Oral Patol Oral Cir Bucal. 2019 Jan 1;24(1):e76–e83. pmid:30573712
  40. 40. Koch GK, James B, Gallucci GO, Hamilton A. Surgical Template Fabrication Using Cost-Effective 3D Printers. Int J Prosthodont. 2019 Jan/Feb;32(1):97–100. pmid:30677120
  41. 41. Chen L, Lin WS, Polido WD, Eckert GJ, Morton D. Accuracy, reproducibility, and dimensional stability of additively manufactured surgical templates. J Prosthet Dent. 2019 Sep;122(3):309–314. pmid:30948293
  42. 42. Herschdorfer L, Negreiros WM, Gallucci GO, Hamilton A. Comparison of the accuracy of implants placed with CAD-CAM surgical templates manufactured with various 3D printers: An in vitro study. J Prosthet Dent. 2020 Jun 1:S0022-3913(20)30235-3. Epub 2020 Jun 2.
  43. 43. Gulati M, Anand V, Salaria SK, Jain N, Gupta S. Computerized implant-dentistry: Advances toward automation. J Indian Soc Periodontol. 2015 Jan-Feb;19(1):5–10. pmid:25810585
  44. 44. Collyer J. Stereotactic navigation in oral and maxillofacial surgery. Br J Oral Maxillofac Surg. 2010 Mar;48(2):79–83. pmid:20061072
  45. 45. Arisan V, Karabuda CZ, Ozdemir T. Implant surgery using bone- and mucosa-supported stereolithographic guides in totally edentulous jaws: surgical and post-operative outcomes of computer-aided vs. standard techniques. Clin Oral Implants Res. 2010 Sep;21(9):980–8. pmid:20497439