Abstract
In chronic myeloid leukemia (CML), the duration of deep molecular response (MR) before treatment cessation (MR4 or deeper, corresponding to BCR-ABL1 ≤ 0.01% on the International Scale (IS)) is considered as a prognostic factor for treatment free remission in stopping trials. MR level determination is dependent on the sensitivity of the monitoring technique. Here, we compared a newly established TaqMan (TM) and our so far routinely used LightCycler (LC) quantitative reverse transcription (qRT)-PCR systems for their ability to achieve the best possible sensitivity in BCR-ABL1 monitoring. We have comparatively analyzed RNA samples from peripheral blood mononuclear cells of 92 randomly chosen patients with CML resembling major molecular remission (MMR) or better and of 128 CML patients after treatment cessation (EURO-SKI stopping trial). While our LC system utilized ABL1, the TM system is based on GUSB as reference gene. We observed 99% concordance with respect to achievement of MMR. However, we found that 34 of the 92 patients monitored by TM/GUSB were re-classified to the next inferior MR log level, especially when LC/ABL1-based results were borderline to thresholds. Thirteen patients BCR-ABL1 negative in LC/ABL1 became positive after TM/GUSB analysis. In the 128 patients included in the EURO-SKI trial identical molecular findings were achieved for 114 patients. However, 14 patients were re-classified to the next inferior log-level by the TM/GUSB combination. Eight of these patients relapsed after treatment cessation; two of them were re-classified from MR4 to MMR and therefore did not meet inclusion criteria anymore. In conclusion, we consider both methods as comparable and interchangeable in terms of achievement of MMR and of longitudinal evaluation of clinical courses. However, in LC/ABL1 negative samples, slightly enhanced TM/GUSB sensitivity may lead to inferior classification of clinical samples in the context of TFR.
Citation: Spiess B, Rinaldetti S, Naumann N, Galuschek N, Kossak-Roth U, Wuchter P, et al. (2019) Diagnostic performance of the molecular BCR-ABL1 monitoring system may impact on inclusion of CML patients in stopping trials. PLoS ONE 14(3): e0214305. https://doi.org/10.1371/journal.pone.0214305
Editor: Susan Branford, SA Adelaide, AUSTRALIA
Received: October 25, 2018; Accepted: March 11, 2019; Published: March 21, 2019
Copyright: © 2019 Spiess et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript.
Funding: SS received honoraria from Novartis Pharma GmbH, Bristol-Myers Squibb (BMS), Pfizer, ARIAD and research funding from Novartis Pharma GmbH and BMS. This does not alter our adherence to PLOS ONE policies on sharing data and materials. The funders had no role in study design, collection, analysis and interpretation of data; writing of the paper and/or decision to submit for publication. We further acknowledge financial support by Deutsche Forschungsgemeinschaft within the funding programme Open Access Publishing, by the Baden-Württemberg Ministry of Science, Research and the Arts and by Ruprecht-Karls-Universität Heidelberg.
Competing interests: SS received honoraria from Novartis Pharma GmbH, Bristol-Myers Squibb (BMS), Pfizer, ARIAD and research funding from Novartis Pharma GmbH and BMS. This does not alter our adherence to PLOS ONE policies on sharing data and materials. The funders had no role in study design, collection, analysis and interpretation of data; writing of the paper and/or decision to submit for publication. We further acknowledge financial support by Deutsche Forschungsgemeinschaft within the funding programme Open Access Publishing, by the Baden-Württemberg Ministry of Science, Research and the Arts and by Ruprecht-Karls-Universität Heidelberg.
Introduction
The occurrence of the reciprocal translocation between chromosomes 9 and 22 (t(9;22)(q34;q11)) resulting in a translocation of the genes BCR and ABL1 is causal for development of chronic myeloid leukemia (CML).
The majority of patients express different BCR-ABL1 mRNA fusion variants (most commonly e13-a2, e14-a2, e1-a2) resulting in the expression of abnormal BCR-ABL1 fusion tyrosine kinase. Molecular diagnostics of CML and patient monitoring is based on quantification of BCR-ABL1 transcript levels in peripheral blood (PB) or bone marrow (BM) aspirates of patients by various qRT-PCR technologies as the number of BCR-ABL1 transcripts serves as surrogate indicator for the amount of residual BCR-ABL1 positive tumor cells and/or their proliferative potential during therapy with tyrosine kinase inhibitors (TKI). Therapy regimen employing TKI directed at the abnormal BCR-ABL1 fusion tyrosine kinase can achieve durable cytogenetic and molecular remissions (MRs) and improve survival in the majority of patients which is approaching that of the general population [1–3]. Since a high percentage of the patients achieve deep molecular response (DMR) under TKI treatment the conception of treatment free remission (TFR) was supported [4–6]. TKI treatment has been stopped successfully in approximately half of the patients with DMR in cessation trials [7, 8]. The criteria for discontinuation differ in the various studies [9]. Most have defined MR4 or MR4.5 for at least one year and TKI treatment duration of at least three years [7]. Currently, the European EURO-SKI trial evaluating molecular data of 755 patients after TKI discontinuation is the first study to define the best criteria for cessation of TKI treatment in patients with CML in deep molecular response [10]. The interim multivariable analysis revealed that the probability of MMR maintenance was significantly associated with overall TKI treatment duration, interferon pre-treatment, and DMR duration, the latter showing the largest effect on the success of treatment cessation [10]. 373/755 (49%) patients analyzed so far in EURO-SKI trial have lost MMR after 24 months of TKI cessation.
Achieving a TFR is desirable, as it reduces treatment costs and avoids therapy-associated side effects and late sequelae of therapy. So far, it is unclear how deep the MR should be before a stopping attempt. In this context, the log reduction of detectable BCR-ABL1 fusion transcripts, is a crucial factor in the decision to TKI discontinuation and essentially depends on the definition of molecular response as well as on technical aspects of BCR-ABL1 transcript measurements. Therefore, a highly standardized assay system with the highest possible sensitivity has to be considered the most appropriate test system [11].
New revised definitions were introduced by the CML Working Group of the ELN that take into account the sensitivity of the molecular qRT-PCR test, that is, MR4 indicates ≥ 4-log reduction (BCR-ABL IS ≤ 0.01%), MR4.5 indicates ≥ 4.5-log reduction (BCR-ABL IS ≤ 0.0032%), and MR5 indicates ≥ 5-log reduction (BCR-ABL IS ≤ 0.001%), especially in negative qRT-PCR results. Mandatory for valid calculation of major responses (log reduction) is a sufficient RNA quality as given by the absolute copy numbers of housekeeping genes ABL1 and/or GUSB [11, 12]. qRT-PCR using LightCycler (LC, Roche Applied Science, Mannheim, Germany) and TaqMan (TM; ThermoFisher Scientific/Applied Biosystems, Darmstadt, Germany) technologies meet all the requirements for sensitive and reliable diagnostic tools to perform molecular monitoring in CML patients [13–15]. The well-established LC methodology and newly implemented TM system are routinously used in our laboratory, standardized according to IS [16, 17], in strict consent with the international guidelines [18, 19]. The high fidelity and longitudinal reproducibility of our measurements are granted by a stringent intramural quality and validation management [20].
Here, we report on the development of a novel protocol for BCR-ABL1 monitoring using TM/GUSB technology. Moreover we present data on the BCR-ABL1 monitoring of 220 CML patients (92 and 128 patients) with MMR or better comparatively tested with TM and LC. We evaluated accomplishment of the best possible sensitivity in BCR-ABL1 transcript level assaying and we discuss the impact of our results in terms of involvement of patients in TKI stopping trials.
Materials and methods
Clinical sample preparation and controls
For RNA extraction, the clinical samples were processed using the automated Maxwell MDx technology (Promega, Mannheim, Germany). A PB sample of a healthy person processed in parallel to patients’ samples served as negative control for each processing cycle. In case of false positivity the qRT-PCR data of all corresponding (processed in parallel) clinical samples was rejected and RNA was prepared again employing frozen backup material. cDNA synthesis was performed using random hexamers and Superscript II reverse transcriptase (ThermoFisher Scientific, Darmstadt, Germany) as described previously [13].
The serial diluted standard plasmid provided an external positive control for each LC run. A PCR mix control served as negative control for each PCR experiment. For quality control of the sample material the internal reference transcript of ABL1 or GUSB of each sample was used. All controls and samples were performed in duplicate. Assay stability and longitudinal performance of the qRT-PCR assays were assessed by the use of quality control charts (QCC) as described previously [20].
qRT- PCR on LC platform
The LC PCR and detection system (version 2, Roche Applied Science) was used for amplification and quantification of ABL1 control and BCR-ABL1 fusion genes. The PCR reactions were performed in glass capillaries employing a LC “Fast Start DNA Master Hybridization Probes” kit (Roche Applied Science), BCR and ABL1 specific primer and fluorescent probes as described by Emig et al., 1999 [13]. qRT-PCR for BCR-ABL1 and ABL1 transcripts was performed in duplicates using cDNA of diagnostic specimens prepared according to a standardized operating procedure (SOP), (Emig et al., 1999), and of seven serial dilutions of 4, 10, 40, 400, 4000, 40,000 and 400,000 copies of standard plasmid pME-2 per reaction [21]. The limit of detection (95%) (LOD95) is 4 copies using serial dilutions of pME-2.
For a LC experiment to be acceptable the generated data has to meet the passing criteria described in detail by Cross et al. [11]. The tolerable difference of the two single measurements of the standard and sample duplicates were evaluated according to Foroni et al., 2011 [19].
qRT- PCR on TM platform
The TM PCR detection system (TaqMan 7500 Fast Real-Time PCR System, ThermoFisherScientific/Applied Biosystems) was used for amplification and quantification of GUSB control and BCR-ABL1 fusion genes (e13-a2 and e14-a2). qRT-PCR for BCR-ABL1 and ABL1 transcripts was performed in 96-well plates in duplicates using cDNA of diagnostic specimens and of eight serial dilutions of 3, 15, 30, 300, 3000, 30,000 and 300,000 and 3 x 106 copies of commercially available standard plasmid pIRMM0099 (ERM-AD623/”Institute for Reference Materials and Measurements”) per reaction. Available TM primers were for BCR-ABL1: ENF 501 (sense) 5’-TCCGCTGACCATCAAYAAGGA-3’ and ENR 561 (antisense) 5’-CACTCAGACCCTGAGGCTCAA-3’ [15, 22]. The expected sizes of BCR-ABL1 amplicons for b2a2 and b3a2 fusion points transcripts are 228 bp and 153 bp in length, repectively. As BCR-ABL1 specific probe ENP541-MGB sense (6FAM-CCCTTCAGCGGCCAGT-MGB) was used. GUSB specific primers were ENF 1102 (sense) 5’-GAAAATATGTGGTTGGAGAGCTCATT-3’ and ENR1162 (antisense) 5’-CCGAGTGAAGATCCCCTTTTTA-3’. The GUSB fragment is 101 bp in length. For detection of GUSB the specific probe ENPr1142 (antisense)-MGB was used (NED-CCAGCACTCTCGTCGGTGACTGTTCA-MGB). The specific primer pair for ABL1 was: ENF 1003 (sense) 5’-TGGAGATAACACTCTAAGCATAACTAAACCT-3’ and ENR 1062 (antisense) 5’-GATGTAGTTGCTTGGGACCCA-3’. The PCR fragment is 124 bp long; the ABL1 specific probe was ABL-1043V-MGB antisense (VIC-CATTTTTGGTTTGGGCTTC-MGB). For BCR-ABL1 detection FAM was used as reporter dye and NFQ-MGB as quencher, for GUSB the reporter dye was NED, NFQ-MGB was used as quencher, ABL1 was detected with the reporter dye VIC, as quencher NFQ-MGB was used, too. Primers and probes were used in a final concentration of 10 μM per PCR reaction. The amplification was performed in a total volume of 20 μl per reaction, whereby BCR-ABL1 and GUSB amplification was performed as duplex PCR in one well. For each reaction 10 μl of a TaqMan Fast Advances Master Mix (ThermoFisherScientific/Applied Biosystems) were added. The PCR program implied at first a holding stage at 50°C for 2 min, than an additional holding stage at 95°C for 20 s, the cycling stage comprised 95°C for 3 s and 60°C for 45 s, the cycling stage was repeated for 45 times.
Criteria for passing for each TM run were: determination of gene specific different threshold values which were for GUSB = 0.06, for ABL1 = 0.08 and for BCR-ABL1 = 0.1. Achievements of further technical values were prerequisites: R2 > 0.98, Slope = -3.2 to -3.6 and intercept ≤ 42. At least four standards of the standard series have to be evaluable, whereby at least for BCR-ABL1 measurements one duplicate of standard 3 or standard 15 has to be positive. For GUSB and ABL1 at least one of the standard 300 duplicates has to be positive. All duplicates were evaluated according to the guidelines of Foroni., 2011 [19]. The cut off for the evaluation of positive signals from negative control blood samples and patients’ samples was Ct = 42. Housekeeping genes and internal references ABL1 and GUSB are commonly used. For MMR, MR4, MR4.5 and MR5, at least 24,000; 24,000; 77,000 and 240,000 GUSB copies, or 10,000; 10,000; 32,000 and 100,000 ABL1 copies are necessary, respectively [11].
Quality management for qRT-PCR data
Ct values (Cycle threshold) of the standard dilution series of the respective qRT-PCR experiments (LC and TM) were exported into a Microsoft Excel sheet. These data was compared to the mean values and the respective warning and control (intervention) limits given by the respective QCC currently valid. Standard dilution data for each experiment must meet QCC requirements for passing. The analysis and documentation of the Ct values and the comparison to warning and control (intervention) limits is done automated by an algorithm implemented in our proprietary lab information management software (LIMS). Exceeding the limits is documented by our LIMS database (LeukoDB2) [20].
Statistics
Statistical calculations were performed using Microsoft Excel and GraphPad Prism (V 6.0) according to standard procedures. We have applied simple column statistics (mean, median, range). In addition unpaired (T-test) and paired analysis (Wilcoxon signed rank test) were employed for calculation of variation ranges between LC/ABL1 and TM/GUSB analysis.
Ethical considerations
Control blood samples of healthy donors were obtained with written informed consent in accordance with the declaration of Helsinki from the local blood bank in fully anonymized manner.
Analyses were done according to Good Clinical Practice (GCP) guidelines as well as in concordance with the Declaration of Helsinki. The investigations were approved by the local Ethics Committee (Medizinische Ethikkommision II der Medizinischen Fakultät Mannheim der Ruprecht Karls-Universität Heidelberg, 2013-509N-MA and 212-247-AWB-MA). This analysis included 128 CML patients from the European Stop Tyrosine Kinase Inhibitor Study (EURO-SKI) trial, a prospective multicenter TKI discontinuation trial (NCT01596114). In accordance with the Declaration of Helsinki, written informed consent was obtained from all patients.
Results
Establishment and validation of TM PCR methodology
For establishing and implementing the TM technology, various parameters of the technology were tested within our quality management as part of method validation [23].
These were:
- Inter and intra experimental accuracy
- Robustness
- Limit of detection (LOD95)
(i) Inter and intra experimental accuracy
For determination of the accuracy, six experiments were performed in parallel using the standard series (pIRM 3–3,000,000), generating a total of 60 data points and 10 wells to quantify ABL1. Results are shown graphically in Fig 1A–1C.
Assessment of inter- and intraexperimental accuracy by TM PCR using pIRM standard dilutions representing 3–3,000,000 plasmid molecules per assay. (A) Six independent experiments with 10 replicates each revealed 60 consecutive data points for each of the 8 standard dilutions as depicted by the symbols in horizontal order. (B) Calculated variance of standard dilution measurements with regard to the defined/expected plasmid molecule numbers as theoretically deployed in the assay. (C) Standard deviation (in % of mean) of measurements shown in (A).
The acceptance criteria were met after evaluation of the data. The correctness was given because the device software was able to generate a calibration straight line with R2 > 0.98 [19] for each of the six evaluations and the standard deviations (%) were not greater than the permitted fluctuations (within warning and control limits) [20]. Naturally, the standard deviation increases with decreasing number of molecules in the sample (standards 3, 15, 30) as the influence of pipetting inaccuracy increases. For standard 3, the mean for the measured real value was 3.9 BCR-ABL1 copies (+/- 1.95 (49%)), for standard 15, 14.2 copies (+/- 3.77 (26%)), for standard 30, 28.6 copies (+/- 6.56 (23%)), for standard 300, 315 copies (+/- 54 (17%)), for standard 3,000, 3,015 copies (+/- 423 (14%)), for standard 30,000, 29,563 copies (+/- 3,698 (12%)), for standard 300,000, 292,894 copies (+/- 31,350 (10%)) and mean for standard 3,000,000 was 2,943,016 copies (+/- 436,236 (14%)) (Fig 1B and 1C).
(ii) Robustness
A diagnostic method can be considered robust if the same results are obtained even though slight variations (operator- or hardware-related) in the experimental procedure may have occurred. To test the TM PCR for robustness, five independent experiments were performed using the plasmid standard dilution series (pIRM 3–3,000,000) by varying the following parameters:
- Altered annealing temperature: the assay was carried out by one single operator but at varying annealing temperatures (1°C below and 1°C above the standard annealing temperature of 60°C).
- Variations by different operator: two identical standard experiments were performed in parallel by two different lab operators.
- Altered volume of master mix added to PCR reaction: from the master mix only 16 μl were used instead of 17 μl, so the PCR reaction was carried out in 1 μl less volume than according to standard protocol (= 17 μl master mix + 3 μl cDNA sample).
The Ct values output by the TM software were compared to the Ct values of the intra-experimental variation (IEV = upper and lower warning limits according to QCC) [20]. For robustness, the values should be within the range specified as warning limits (= average values +/- 2 x standard deviation). As shown in Fig 2 this was achieved in the majority of experiments. Only at the standard dilutions 300 and 30.000 the upper warning limit was hit when the annealing temperature was decreased by 1° C (59° C instead of 60° C). The values were still below the control limit (+/- 3 x standard deviation) and therefore within the inter-experimentally acceptable variation.
Annealing temperatures (59°C, 60°C, 61°C), operators (n = 2) and volumes of master mix added to PCR template were varied. All experiments were performed in duplicates. Blue and red lines above data columns depict warning (+/- 2fold standard deviation) and control limits (+/- 3fold standard deviation), respectively.
(iii) Limit of detection (LOD95)
For reliable detection of low target molecule numbers the assay validation should include the determination of the assay’s detection limit (LOD95). LOD95 represents the number of target molecules that will be reliably detected in 95% of all tested samples. For calculation of LOD95 for the TM-based qRT-PCR the data of 40 consecutive measurements of standard dilution series (pIRM 3–3,000,000) were evaluated. For standard dilution 3 and 15 the expected signals were positively detected in 38 and 40 cases corresponding to 95% and 100% of detection, respectively. The detection limit of LOD95 was thus 3 molecules in the sample. The detection limit in the diagnostic system can be considered < 4 target molecules in the reaction mixture, which can be rated as good. It is in line with the internationally achievable sensitivity standards for TM and LC technology [11].
Comparison of LC results to TM results
For the comparative assessment of the BCR-ABL1 quantification by means of the TM and LC method with regard to the reliable achievement of the diagnostic statement "achievement of MMR—yes / no", 92 patient consecutive samples of daily routine diagnostic operation were analyzed in parallel using both qRT-PCR technologies. The calculated values are based on the means of duplex experiments. A comparison of the findings produced in parallel with LC and TM revealed the following similarities and changes of log levels (Table 1) including 16 of 19 (TM/GUSB) and 11 of 19 (TM/ABL1) patients who lost their MR5 status. Thirteen of 16 patients with formerly undetected BCR-ABL1 transcripts were analyzed positive using TM/GUSB. Four cases changed from MR5 to MR4.5, three, four and two cases changed from MR5 to MR4.0, MR4.5 to MR4 and from MR4.5 to MMR, respectively.
The differences in diagnostic findings for BCR-ABL1 monitoring between LC using ABL1 and TM using ABL1/GUSB as housekeeping genes is graphically shown in Fig 3.
A. Assay-related shifts (n = 34) in the diagnostic outcome of 92 CML patient samples were analyzed with the TM/GUSB combination (pink squares) instead of LC/ABL1 combination (blue diamonds). B. Assay-related shifts (n = 30) in the diagnostic outcome of 92 CML patient samples were analyzed with the TM/ABL1 combination (pink squares) instead of LC/ABL1 combination (blue diamonds). Abbreviations: MMR, major molecular response; MR4 indicates ≤ 4-log reduction (BCR-ABL1 IS ≤ 0.01%), MR4.5 indicates ≤ 4.5-log reduction (BCR-ABL1 IS ≤ 0.0032%), and MR5 indicates ≤ 5-log reduction (BCR-ABL1 IS ≤ 0.001%). The prerequisite for valid calculation of major responses (log reduction) is a sufficient RNA quality as given by the absolute copy numbers of housekeeping gene GUSB.
We have calculated the variation ranges of LC/ABL1 (mean: 0.003; median: 0.00; SEM: 0.0085; range: 0 to 0.047) versus TM/GUSB (mean: 0.011; median: 0.004; SEM: 0.027; range: 0 to 0.157) using the Wilcoxon signed rank test (p<0.0001) based on those samples where changes in log levels have been found (n = 34).
All clinical course data before January 2018 was LC/ABL1 based. Since then, TM/GUSB was used for BCR-ABL1 patient monitoring. After evaluation of the overall clinical courses of all 92 patients under consideration of LC/ABL1 and the available TM/GUSB data, it turned out that single TM/GUSB data points have no impact on the overall clinical course. However, we found slightly enhanced sensitivity for the TM/GUSB combination. This is statistically valid as pairwise testing (Wilcoxon signed rank test) revealed p<0.0001 when comparing the 92 data points of LC/ABL1 (mean 1.828 +/- 0.9983, range 0.00 to 58.00) to the corresponding 92 data points of TM/GUSB (mean 1.901 +/- 1.078, range 0.00 to 82.75). In about 30% of patients, this led to inferior staging as exemplarily shown for eight patients (#1 to #8) in S1 Fig. The differences in sensitivity between LC/ABL1 and TM/GUSB measurements become clearer the more TM/GUSB data points were available as depicted in the clinical courses of patients (#4 to #8). Here, a clear sensitivity shift can be observed nevertheless concurring with the respective trends of the clinical courses established by the preceding LC/ABL1 data points. The enhanced sensitivity of TM/GUSB is reflected by the fact that many sample points that were negative in LC/ABL1 analysis became positive when TM/GUSB method was applied (patients #1, 2, 5, 6, 8). This phenomenon is independent of the conversion factor (CF) applied for calculation of the quotient BCR-ABL1 IS [%].
Re-evaluation of EURO-SKI samples using TM technology
In order to assess the potential impact of the higher sensitive TM/GUSB combination on inclusion of patients in stopping trials we performed a “what if” analysis on a panel of 128 patients of the EURO-SKI stopping trial that have been previously monitored by employment of the LC/ABL1 combination. One sample of each patient representative of the screening period before inclusion in the stopping trial (i.e. MR4 mandatory for at least one year) was included in the analysis. Identical results were obtained for 114 patients irrespective whether LC/ABL1 or TM/GUSB was used for monitoring concerning MR4 in four and MR4.5/MR5 in 110 patients. Fourteen patients were re-classified to the next inferior log-level when using the TM/GUSB system. Altogether 12 patients were re-classified from MR4.5 to MR4. Of these 12, six patients showed relapse and the other six are still in TFR but are potential candidates for relapse and are under special attention. Upon hypothetic application of our “what if” scenario, the six relapsing patients were “saved” when inclusion into the stopping trial was restricted to MR5 and MR4.5 patients. Two patients of the relapse group were re-classified from MR4 to MMR and would not have led to cessation of treatment (Fig 4).
Comparison of EURO-SKI stopping trial interim TFR analysis data (left pie) with a “what if” scenario (right pie) where patient inclusion was based on more sensitive TM/GUSB instead of LC/ABL1 assays. Of 128 patients who were evaluated by LC/ABL1 (left pie), 58 (45%) relapsed within 24 months of TKI discontinuation. According to data based on TM/GUSB monitoring of 128 EURO-SKI patients (right pie), two of the relapsing patients (2/58 = 3.4%) were re-classified from MR4 to MMR and, therefore, both specimen analyzed did not meet the inclusion criteria anymore (yellow sector). The same holds true for another 6 patients (6/58 = 10%) when MR4 was excluded as inclusion criterion (orange sector). Six samples of patients in remission were re-classified from MR4.5 to MR4 but are currently still in TFR (light green sector).
We have added the variation ranges of actual BCR-ABL1 quotient IS [%] ratio for the 14 patients who were diagnosed differently when LC/ABL1 and TM/GUSB techniques were applied. As LC/ABL1 showed a mean variation of 0.0027 (+/- 0.0021) (range: 0.00097 to 0.0097) whereas the TM/GUSB displayed a mean variation of 0.006 (+/- 0.0035) (range 0 to 0.0144). Wilcoxon signed rank test revealed significance (p = 0.0023) between both groups.
Discussion
We demonstrated that our newly established TM/GUSB system is a robust and reliable qPCR method for monitoring CML patients. The TM technology for BCR-ABL1 quantitative measurements in our laboratory using the diagnostic combination TM/GUSB improved the sensitivity when compared to LC/ABL1 technology [13]. We found that 34 (37%) and 30 (33%) of the 92 patients monitored by TM/GUSB and TM/ABL1, respectively, were re-classified to the next inferior log level when compared to LC/ABL1 results. This includes 16 of 19 patients who lost their MR5 status in TM/GUSB analysis, indicating that some (13 of 92) of the “BCR-ABL1-negative” samples tested may not be really negative if higher sensitive methods are applied. Albeit the slightly observed enhanced sensitivity of TM/GUSB that in some cases may lead to inferior classification of single data points, we consider both methods as absolutely comparable and interchangeable in terms of evaluating longitudinal clinical courses (compare S1 Fig).
This The observed enhanced sensitivity of TM/GUSB is important, as one could consider that employment of more sensitive diagnostic methods could push the limits for inclusion of patients in stopping trials and finally may reduce the number of patients incurring loss of MMR after TKI discontinuation. Therefore, it seems mandatory to use the most efficient logistics and most sensitive techniques for sample collection and transportation, RNA preparation and qRT-PCR.
In fact, when comparing both PCR systems and generating a “what if” scenario for retrospectively selected cDNA samples of 128 EURO-SKI patients identical results were obtained only for 114 patients concerning MR4 in four and MR4.5/MR5 in 110 patients. TM/GUSB PCR findings were different from LC/ABL1 findings in 14 patients. Of these, eight patients were from the relapse group; in two of these patients the TM/GUSB PCR results would not have led to cessation of therapy, suggesting that an employment of more sensitive PCR systems may spare patients disease progression. It is conceivable that patients that are borderline to threshold ranges may shift to the next inferior log level (MR4 -> MMR) and therefore lose their inclusion criterion.
Six of the 14 patients re-classified from MR4.5 to MR4 had no relapse during TFR in a follow up period of at least 24 month [10]. Further monitoring will show future loss or maintenance of TFR in these patients. There could be manifold reasons for higher TM/GUSB qRT-PCR sensitivity. In the TM/GUSB PCR assay, there are differences in primer combinations and generated PCR amplicons. In determining the sensitivity of a PCR assay, primer selection and the length of the resulting PCR amplicons may play a crucial role. Shorter PCR fragments (length: 100–150 bp) are amplified in the TM system with higher efficiency. Furthermore TM and LC are different PCR systems with different PCR performance and kinetics. The TM technology uses only one labelled probe as a 5’ nuclease assay to generate a specific fluorescence signal instead of two probes, resulting in a simplification of the PCR kinetics which contributes to an increase in sensitivity. In addition, another type of Taq Polymerase used in TM technology could also influence PCR sensitivity and efficiency. Also the choice of the housekeeping gene may influence the diagnostic outcome. After careful testing and evaluating so far published data, we prefer GUSB as housekeeping gene for TM-based analysis. While Beillard et al. [24] recommended ABL1 as control gene for RQ-PCR-based diagnosis and MRD detection in leukemic patients because of more stable, uniform expression and lack of pseudogenes, GUSB transcript levels were later confirmed as a suitable alternative [25]. Other authors described the suitability of ABL1, BCR and GUSB for BCR-ABL1 quantification [15] or even considered beta-glucoronidase (GUSB) as the most suitable genes tested [26] because the expression pattern of beta-glucoronidase (GUSB) was found more homogenous than that of ABL1 or β2 microglobulin (B2M) [27]. Not to be neglected the main practical advantage of the use GUSB over ABL1 as housekeeping genes is the performance of BCR-ABL1 and GUSB quantification in one single well (duplex PCR) avoiding the bias introduced by setting-up two spatially distinct TM qPCR reactions.
One could argue that the observed shift in log levels may purely reflect the variability of inter or intra-experimental variation or just pipetting inaccuracy, naturally observed for every assay system in particular when testing was performed for very low numbers of target molecules (compare Fig 1C). This may be the case for samples in log levels MR4.5 and MR5 where very few numbers (<10) of BCR-ABL1 cDNA molecules decide on sample positivity. This becomes evident in clinical courses of some representative patients shown in S1 Fig, where positivity and negativity fluctuates in consecutive measurements. These fluctuations seem to be normal and do not impact the overall clinical courses as depicted in S1 Fig but may be important in TFR-related settings. However this variation is expected to follow a random Poisson distribution as described previously [20]. In contrast, in our methodological comparison the majority of samples increased in log levels pointing to a non-random change in assay sensitivity (= assay variation).
One could further argue that the increased sensitivity of TM/GUSB versus LC/ABL1 is merely due to an incorrect CF that may lead to erroneously high BCR-ABL1 IS quotients. Our CFs were established using Bland Altman method by the Australian laboratory of Susan Branford and/or the EUTOS consortium. However, the detection of BCR-ABL1 copies in 13 cases where LC/ABL1 did not succeed in detecting residual BCR-ABL1 transcripts (0.000% IS) is independent of the CF and therefore, our data point to higher sensitivity of TM/GUSB when compared to LC/ABL1. Therefore it seems not advantageous to us to change the CF by own calculation. The clinical relevance of the detection of minimal amounts of BCR-ABL1 copies using a more sensitive PCR system remains to be unclear and could be discussed controversially. Our preliminary data are a hint that increased sensitivity of the PCR system could play an important role in the scenario of stopping trials as being the basis for determination of MR4 and MR4.5 and therefore for the decision of treatment discontinuation. The duration and the log range of molecular remission is the factor with the highest predictive value for deciding on discontinuation of therapy. Therefore, the methods used to detect molecular remission play a crucial role. The deepening of the sensitivity of the diagnostic technologies depends on many different methodological factors. The quality of the RNA and the efficiency of the cDNA synthesis have an influence on the detected copy numbers of housekeeping gene and BCR-ABL1. A more efficient cDNA synthesis, together with a more sensitive PCR system, provides higher absolute copy numbers of BCR-ABL1 and housekeeping gene, which have an influence on the generation of the respective log level for the individual patient.
The increased sensitivity of the TM/GUSB diagnostic system resulted in detection of BCR-ABL1 copies in 13 cases where LC/ABL1 did not succeed in detecting residual BCR-ABL1 transcripts (0.000% IS). On the other hand, more sensitive systems generate the possibility to detect very high amounts of housekeeping gene copy numbers creating log level MR5.5 or even MR6, when a patient is in fact BCR-ABL1 negative by qRT-PCR. It should be emphasized that inclusion criteria should grant the lowest number of relapsing patients at long sight. Aiming at this, it is conceivable that omitting the MR4 log level as inclusion criterion may be advantageous for the outcome of future stopping trials.
In conclusion, although enhanced sensitivity of TM/GUSB combination may be advantages the BCR-ABL1 monitoring in some cases we consider both methods as comparable and interchangeable in terms of MMR achievement and evaluation of clinical courses. However, in LC/ABL1 negative samples, slightly enhanced TM/GUSB sensitivity may lead to positivity and inferior classification of clinical samples.
Our data suggests that in clinical terms the diagnostic discrimination of MR5 and MR4.5 BCR-ABL1 log levels may be clinically less important for the outcome of a stopping trial than the discrimination between MR4.5 and MR4, since the latter is nearby MMR. Our study is not meant to give statements about “correct” and “incorrect” methods, but is intended to serve optimization of molecular methods in terms of TFR-related settings.
The clinical impact of resulting changes needs to be further supported in large scale studies within stopping trials. We believe that the development of higher sensitive detection methods will positively impact on diagnostics and treatment of CML patients.
Supporting information
S1 Fig. Evaluation of longitudinal clinical courses of eight patients with CML.
https://doi.org/10.1371/journal.pone.0214305.s001
(TIF)
Acknowledgments
For excellent experimental assistance and data collection we are thankful to Carolin Hölting, Elena Felde, Vanessa Leins, Melanie Bauer, Susanne Keppler-Gerner and Heike Braun.
We are thankful to Susan Branford and team (Institute of Medical and Veterinary Science, Adelaide, Australia) for generating and validating the conversion factors used in this study.
References
- 1. Gambacorti-Passerini C, Antolini L, Mahon FX, Guilhot F, Deininger M, Fava C, et al. Multicenter independent assessment of outcomes in chronic myeloid leukemia patients treated with imatinib. J Natl Cancer Inst. 2011;103(7):553–61. Epub 2011/03/23. pmid:21422402.
- 2. Bjorkholm M, Ohm L, Eloranta S, Derolf A, Hultcrantz M, Sjoberg J, et al. Success Story of Targeted Therapy in Chronic Myeloid Leukemia: A Population-Based Study of Patients Diagnosed in Sweden From 1973 to 2008. J Clin Oncol. 2011;29(18):2514–20. WOS:000291684600027. pmid:21576640
- 3. Hehlmann R, Lauseker M, Jung-Munkwitz S, Leitner A, Mueller MC, Pletsch N, et al. Tolerability-Adapted Imatinib 800 mg/d Versus 400 mg/d Versus 400 mg/d Plus Interferon-alpha in Newly Diagnosed Chronic Myeloid Leukemia. J Clin Oncol. 2011;29(12):1634–42. WOS:000289714600037. pmid:21422420
- 4. Hehlmann R, Müller MC, Lauseker M, Hanfstein B, Fabarius A, Schreiber A, et al. Deep molecular response is reached by the majority of patients treated with imatinib, predicts survival, and is achieved more quickly by optimized high-dose imatinib: results from the randomized CML-study IV. J Clin Oncol. 2014;32(5):415–23. Epub 2013/12/04. pmid:24297946.
- 5. Castagnetti F, Gugliotta G, Breccia M, Stagno F, Iurlo A, Albano F, et al. Long-term outcome of chronic myeloid leukemia patients treated frontline with imatinib. Leukemia. 2015;29(9):1823–31. Epub 2015/06/20. pmid:26088952.
- 6. Hehlmann R, Lauseker M, Saußele S, Pfirrmann M, Krause SW, Kolb H-J, et al. Assessment of imatinib as first-line treatment of chronic myeloid leukemia: 10–year survival results of the randomized CML study IV and impact of non-CML determinants. Leukemia. 2017;31(11):2398–406. Epub accepted for publication in July 4, 2017. pmid:28804124
- 7. Saussele S, Richter J, Hochhaus A, Mahon FX. The concept of treatment-free remission in chronic myeloid leukemia. Leukemia. 2016;30(8):1638–47. Epub 2016/05/03. pmid:27133824.
- 8. Hughes TP, Ross DM. Moving treatment-free remission into mainstream clinical practice in CML. Blood. 2016;128(1):17–23. pmid:27013442
- 9. Dulucq S, Mahon FX. Deep molecular responses for treatment-free remission in chronic myeloid leukemia. Cancer medicine. 2016;5(9):2398–411. Epub 2016/07/02. pmid:27367039; PubMed Central PMCID: PMCPMC5055167.
- 10. Saussele S, Richter J, Guilhot J, Gruber FX, Hjorth-Hansen H, Almeida A, et al. Discontinuation of tyrosine kinase inhibitor therapy in chronic myeloid leukaemia (EURO-SKI): a prespecified interim analysis of a prospective, multicentre, non-randomised, trial. Lancet Oncol. 2018:Epub ahead of print May 4, 2018. Epub 2018/05/08. pmid:29735299.
- 11. Cross NC, Hochhaus A, Müller MC. Molecular monitoring of chronic myeloid leukemia: principles and interlaboratory standardization. Ann Hematol. 2015;94 Suppl 2:S219–25. Epub 2015/03/31. pmid:25814088.
- 12. Cross NCP, White HE, Müller MC, Saglio G, Hochhaus A. Standardized definitions of molecular response in chronic myeloid leukemia. Leukemia. 2012;26(10):2172–5. pmid:22504141
- 13. Emig M, Saussele S, Wittor H, Weisser A, Reiter A, Willer A, et al. Accurate and rapid analysis of residual disease in patients with CML using specific fluorescent hybridization probes for real time quantitative RT-PCR. Leukemia. 1999;13(11):1825–32. WOS:000083664000022. pmid:10557058
- 14. Hochhaus A, Weisser A, La Rosee P, Emig M, Muller MC, Saussele S, et al. Detection and quantification of residual disease in chronic myelogenous leukemia. Leukemia. 2000;14(6):998–1005. Epub 2000/06/24. pmid:10865964.
- 15. White H, Deprez L, Corbisier P, Hall V, Lin F, Mazoua S, et al. A certified plasmid reference material for the standardisation of BCR-ABL1 mRNA quantification by real-time quantitative PCR. Leukemia. 2015;29(2):369–76. Epub 2014/07/19. pmid:25036192; PubMed Central PMCID: PMCPMC4320294.
- 16. Branford S, Fletcher L, Cross NCP, Müller MC, Hochhaus A, Kim D-W, et al. Desirable performance characteristics for BCR-ABL measurement on an international reporting scale to allow consistent interpretation of individual patient response and comparison of response rates between clinical trials. Blood. 2008;112(18684859):3330–8.
- 17. Müller MC, Cross NC, Erben P, Schenk T, Hanfstein B, Ernst T, et al. Harmonization of molecular monitoring of CML therapy in Europe. Leukemia. 2009;23(11):1957–63. Epub 2009/08/28. pmid:19710700.
- 18. Hughes T, Deininger M, Hochhaus A, Branford S, Radich J, Kaeda J, et al. Monitoring CML patients responding to treatment with tyrosine kinase inhibitors: review and recommendations for harmonizing current methodology for detecting BCR-ABL transcripts and kinase domain mutations and for expressing results. Blood. 2006;108(1):28–37. Epub 2006/03/09. pmid:16522812; PubMed Central PMCID: PMC1895821.
- 19. Foroni L, Wilson G, Gerrard G, Mason J, Grimwade D, White HE, et al. Guidelines for the measurement of BCR-ABL1 transcripts in chronic myeloid leukaemia. Br J Haematol. 2011;153(2):179–90. Epub 2011/03/09. pmid:21382019.
- 20. Spiess B, Naumann N, Galuschek N, Rinaldetti S, Kossak-Roth U, Tarnopolscaia I, et al. The benefit of quality control charts (QCC) for routine quantitative BCR-ABL1 monitoring in chronic myeloid leukemia. PLoS One. 2018;13(4):e0196326. Epub 2018/04/25. pmid:29689094; PubMed Central PMCID: PMCPMC5916859.
- 21. Müller MC, Erben P, Saglio G, Gottardi E, Nyvold CG, Schenk T, et al. Harmonization of BCR-ABL mRNA quantification using a uniform multifunctional control plasmid in 37 international laboratories. Leukemia. 2008;22(17943168):96–9102.
- 22. Gabert J, Beillard E, van der Velden VH, Bi W, Grimwade D, Pallisgaard N, et al. Standardization and quality control studies of 'real-time' quantitative reverse transcriptase polymerase chain reaction of fusion gene transcripts for residual disease detection in leukemia—a Europe Against Cancer program. Leukemia. 2003;17(12):2318–57. Epub 2003/10/17. pmid:14562125.
- 23.
Tague NR. Seven basic tools. The quality toolbox. 15. Milwaukee, Wisconsin: American Society for Quality; 2004.
- 24. Beillard E, Pallisgaard N, van der Velden VH, Bi W, Dee R, van der Schoot E, et al. Evaluation of candidate control genes for diagnosis and residual disease detection in leukemic patients using 'real-time' quantitative reverse-transcriptase polymerase chain reaction (RQ-PCR)—a Europe against cancer program. Leukemia. 2003;17(12):2474–86. Epub 2003/10/17. pmid:14562124.
- 25. Alikian M, Gale RP, Apperley JF, Foroni L. Molecular techniques for the personalised management of patients with chronic myeloid leukaemia. Biomolecular detection and quantification. 2017;11:4–20. Epub 2017/03/24. pmid:28331814; PubMed Central PMCID: PMCPMC5348117.
- 26. Wang YL, Lee JW, Cesarman E, Jin DK, Csernus B. Molecular monitoring of chronic myelogenous leukemia: identification of the most suitable internal control gene for real-time quantification of BCR-ABL transcripts. J Mol Diagn. 2006;8(2):231–9. Epub 2006/04/29. pmid:16645210; PubMed Central PMCID: PMCPMC1867593.
- 27. Fantasia F, Di Capua EN, Cenfra N, Pessina G, Mecarocci S, Rago A, et al. A highly specific q-RT-PCR assay to address the relevance of the JAK2WT and JAK2V617F expression levels and control genes in Ph-negative myeloproliferative neoplasms. Ann Hematol. 2014;93(4):609–16. Epub 2013/11/01. pmid:24173087; PubMed Central PMCID: PMCPMC3945640.