Interdevice variability of central corneal thickness measurement

  • Peter M. Maloca ,

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing

    peter.maloca@moorfields.nhs.uk

    Affiliations OCTlab, Department of Ophthalmology, University Hospital Basel, Basel, Switzerland, Department of Ophthalmology, University of Basel, Basel, Switzerland, Institute of Molecular and Clinical Ophthalmology Basel (IOB), Basel, Switzerland, Moorfields Eye Hospital, London, United Kingdom

  • Harald P. Studer,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliations OCTlab, Department of Ophthalmology, University Hospital Basel, Basel, Switzerland, Swiss Eye Research Foundation, Reinach, Switzerland

  • Renato Ambrósio Jr.,

    Roles Methodology, Supervision, Writing – original draft, Writing – review & editing

    Affiliations Department for Ophthalmology, Federal University of Sao Paulo, Sao Paulo, Brazil, Rio de Janeiro Corneal Topography and Biomechanics Study Group, Rio de Janeiro, Brazil, Instituto de Olhos Renato Ambrósio, Rio de Janeiro, Brazil

  • David Goldblum,

    Roles Conceptualization, Data curation, Methodology, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Department of Ophthalmology, University of Basel, Basel, Switzerland

  • Simon Rothenbuehler,

    Roles Conceptualization, Methodology, Visualization, Writing – original draft, Writing – review & editing

    Affiliations OCTlab, Department of Ophthalmology, University Hospital Basel, Basel, Switzerland, Department of Ophthalmology, University of Basel, Basel, Switzerland

  • Daniel Barthelmes,

    Roles Visualization, Writing – original draft, Writing – review & editing

    Affiliation Department of Ophthalmology, University Hospital Zurich, Switzerland

  • Sandrine Zweifel,

    Roles Data curation, Writing – original draft, Writing – review & editing

    Affiliation Department of Ophthalmology, University Hospital Zurich, Switzerland

  • Hendrik P. N. Scholl,

    Roles Data curation, Methodology, Writing – original draft, Writing – review & editing

    Affiliations Department of Ophthalmology, University of Basel, Basel, Switzerland, Institute of Molecular and Clinical Ophthalmology Basel (IOB), Basel, Switzerland, Wilmer Eye Institute, Johns Hopkins University, Baltimore, United States of America

  • Konstantinos Balaskas,

    Roles Data curation, Writing – original draft, Writing – review & editing

    Affiliations Moorfields Eye Hospital, London, United Kingdom, Moorfields Ophthalmic Reading Centre, London, United Kingdom

  • Adnan Tufail,

    Roles Data curation, Methodology, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Moorfields Eye Hospital, London, United Kingdom

  • Pascal W. Hasler

    Roles Conceptualization, Data curation, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliations OCTlab, Department of Ophthalmology, University Hospital Basel, Basel, Switzerland, Department of Ophthalmology, University of Basel, Basel, Switzerland

Abstract

Purpose

To evaluate variability of central corneal thickness measurement (CCT) devices using a hitherto unprecedented number of CCT devices.

Methods

CCT was measured consecutively in 122 normal corneas of 61 subjects with seven different devices using three distinct measurement technologies: Scheimpflug, Ultrasound, and Optical Coherence Tomography (OCT). Per device deviation from the mean CCT value per eye was used to determine which of the devices performed best, compared to the mean value.

Results

Cirrus OCT yielded the lowest deviation. Deviations of the individual devices from the mean CCT of each eye were (OS/OD) 12.8±5.0/14.9±9.4 μm for Topcon noncontact specular microscopy (NCSM), 11.3±5.9/10.6±7.3 μm for Pentacam, 10.7±5.2/10.4±4.8 μm for Spectralis OCT, 6.0±3.9/6.2±4.9 μm for Topcon DRI OCT, 5.1±3.4/5.9±10.3 μm for AngioVue OCT, 4.8±4.1/5.7±4.6 μm for US pachymetry, and 4.2±3.2/5.7±4.6 μm for Cirrus OCT. The maximum differences between US pachymetry and the other devices were very high (up to 120 μm).

Conclusion

Central corneal thickness may be under- or overestimated due to high interdevice variations. Measuring CCT with one device only may lead to inappropriate clinical and surgical recommendations. OCT showed superior results.

Introduction

Clinical decision making, whether for the Laser-in-situ-Keratomileusis (LASIK), or for monitoring and screening of various diseases, often relies on the results of CCT measurement. It has been found, that a thinner central corneal thickness (CCT) may represent a higher risk in primary open-angle glaucoma, and patients with a thinner CCT tend to exhibit a more severe glaucomatous defect. [1, 2]

CCT measurements helped diagnose subclinical keratoconus [3], identify contact-lens-induced corneal thinning [4], and monitor various corneal pathologies. [5, 6] Preoperative pachymetry assessment has been shown to be effective to avoid postsurgical corneal ectasia [7].Due to the high reproducibility for measuring CCT Ultrasound (US) pachymetry is still widely used and considered the gold-standard for CCT assessment. [8]

However, Doughty & Zaman reported that ultrasonic measurements tend to overestimate the cornea thickness than optical measurements. [9] Rotating Scheimpflug cameras have proven to be valuable tools, although the CCT values obtained from them have been shown to be slightly lower than those measured by US pachymetry. [10] Both Scheimpflug imaging and spectral-domain OCT (SDOCT) have been demonstrated with high precision, while the SDOCT possesses a higher repeatability. [11] Noncontact specular microscopy (NCSM) can image the corneal endothelial morphology, but underestimates CCT compared to US pachymetry. [12, 13]

Swept-source OCT (SSOCT) has shown a higher reliability than Scheimpflug imaging and allows for monitoring corneal thickness after photorefractive keratectomy. [14, 15]

To improve surgical procedures or treatment, this study compares hitherto unprecedented number of devices for central corneal thickness measurements. From seven (7) devices we calculated the lowest deviation from the mean CCT value.

Materials and methods

Subjects

Subject age, race, sex and diurnal fluctuations were not considered in this study as our goal was to collect the most possible CCT data from many devices in short time.

Sixty-one subjects who were visiting the clinic were randomly selected for CCT measurements using seven different devices. Inclusion criteria were as follows: Subjects with normal visual acuity (20/20), good and steady fixation, and clear cornea. Exclusion criteria included a history of corneal refractive surgery, corneal scarring, degenerations, and dystrophies. Written informed consent was obtained from all subjects, and the study was approved by the Ethics Committee Zentral und Ostschweiz, Switzerland (EKNZ UBE-15/881).

Data acquisition

The order of measurements using different devices was random, apart from the US pachymetry that was always done last in order not to disturb the corneal surface and impair the non-contact measurements. All 14 measurements were conducted at the University of Basel, Switzerland, within a timeframe of 60 minutes. It was proposed to use central thickness measurements to increase comparability between devices. CCT was measured using the following seven devices: Oculus Pentacam version 1.19r11 (Oculus, Wetzlar, Germany), noncontact specular microscopy SP-1P version 1.21 (NCSM, Topcon, Tokyo, Japan), spectral-domain OCT with Cirrus HD-OCT 5000–2997 version 7.0 (SDOCT, Carl Zeiss Meditec, Dublin, CA, USA), Spectralis HRA 2 version 6.0.10.0 (Heidelberg Engineering, Heidelberg Germany), AngioVue OCT software version 2014.2.0.93 (Optovue, Inc., Fremont, CA, USA), swept-source OCT with prototype OCT DRI OCT-1 Atlantis software version 9.12.003.04 (SSOCT, Topcon, Tokyo, Japan), and US pachymetry ALCON RxP version 1.15 (Alcon, Fort Worth, TX, USA).

Automatic acquisition mode was used for the rotating Scheimpflug camera and for NSCM to capture CCT. For SDOCT, additional anterior lens systems were used in the Spectralis and AngioVue devices. The OCT scan patterns were: Cirrus anterior segment 5-line raster scan protocol without anterior segment lens (scan length 3 mm, interslice distance 0.25 mm, 4,096 A-scans per line), Spectralis anterior segment protocol with anterior segment lens (21 B-scans, scan area 8.3 mm x 5.6 mm, interslice distance 278 μm), AngioVue cornea pachymetry protocol with anterior segment lens (scan area 6 mm x 6 mm, measurement within central 3 mm zone), and Topcon SSOCT volume scan patterns over a 6.0 mm x 6.0 mm area centered on the corneal apex (256 cross scans with a scan density of 512 x 128 pixel). The manufacturer caliper tools were used to measure the CCT. The US pachymetry measurement of the cornea was performed as the last test a few minutes after the local application of preservative-free oxybuprocaine (0.4%). Three consecutive measurements were taken from the central cornea. For all measurements the thinnest corneal point was selected.

The axial resolution is 1μm for both the ultrasound device Alcon RxP [16] and for the Pentacam [17], 2 to 3μm for the AngioVue [18], 10μm for the NCSM [19], 3.5μm for the Heidelberg Spectralis [20], 5μm for the Cirrus HD-OCT [21], and 8μm OCT DRI OCT-1 Atlantis [22].

Statistical analysis

Statistical analysis was performed using Microsoft Office Excel 2013 (Microsoft Corp., Redmond, WA). Analysis of variance (ANOVA) and Chi-Square(Х2)-test were performed but not included in this manuscript as this study represents a method comparison study ranking pairs of methods on the basis of their extent of agreement. There were seven groups (Gi) of data, one group for each measurement device. Each group Gi thereby consists of 61 data points, one point per eye. Our statistical analysis assumes that the value of each of these 61 data points Gi,n corresponds to the following sum: the device-independent true CCT value tn of the eye, the device-dependent bias bi, and the device-dependent measurement error ei: (1) where the error ei is assumed to have a normal distribution N(0,vi). In order to determine a possible between-group difference, paired differences of Eq (1) were formed as follows (2) where i,j ∈ (Pentacam, NSCM, US, Cirrus, AngioVue, Spectralis, and DRI) and ij. For un-replicated data, the mean and standard deviation of the between-group differences Di usually convey the most useful information. Paired device-by-device differences between US pachymetry and the other six devices were computed and analyzed. As a summary statistic of inter-device variation, the mean absolute deviation from a central point, the mean CCT over all seven devices per eye, was calculated as (3) where xi,k are the 7 measurements (i ∈ Pentacam, NSCM, US, Cirrus, AngioVue, Spectralis, and DRI) for the k-th eye (one measurement value per device), and m(X) is the mean of the 7 values, to investigate deviation of the measurements. Mean and standard deviation, calculated as (4) (5) express the deviation of each individual device i.

Results

This study measured the central corneal thickness in 122 normal eyes from 61 subjects with an average age of 63.7 years (range 20–86 years). No study subject was excluded. Men represented 34% of the study subjects (66% females).

The overall mean CCT values were as follows (OS/OD): Pentacam 552±35/550±34 μm (SD), Topcon 529±34/530±30 μm (SD), Cirrus 544±34/546±33 μm (SD), AngioVue 540±32/540±32 μm (SD), Spectralis 552±34/553±35 μm (SD), DRI 537±34/537±33 μm (SD), and US pachymetry 539±35/539±36 μm (SD), (Fig 1).

thumbnail
Fig 1. Comparison of central corneal thickness (CCT) measurements.

Significant differences between devices are illustrated (Pentacam/US p<0.0001, Topcon/US p<0.0001, Cirrus/US p<0.0001, AngioVue/US p = 0.092, Spectralis/US p<0.0001, and DRI/US p = 0.004). However, no significant difference was found between left and right eyes (Pentacam p = 0.687, Topcon p = 0.864, Cirrus p = 0.745, AngioVue p = 0.709, Spectralis p = 0.973, DRI p = 0.955, and US pachymetry p = 0.866).

https://doi.org/10.1371/journal.pone.0203884.g001

CCT ranges measured by different devices in the left eyes are Pentacam 481–619 μm, Topcon 463–595 μm, Cirrus 470–608 μm, AngioVue 467–606 μm, Spectralis 474–622 μm, DRI 462–595 μm, and US pachymetry 465–610 μm.

CCT ranges measured by different devices in the right eyes are Pentacam 470–620 μm, Topcon 460–595 μm, Cirrus 470–618 μm, AngioVue 465–654 μm, Spectralis 473–626 μm, DRI 461–600 μm, and US pachymetry 461–604 μm.

Between-device differences

Between-device differences Di were analyzed by looking at the mean and standard deviation of paired CCT difference values between each device and the US pachymetry device (OS/OD). A “-”indicates that a device measured thicker, and a “+” indicates that a device measured thinner CCT values than the ultrasound device: US-Pentacam -14.5±7.8/-10.9±12.5 micron (SD), US-NSCM +9.4±8.0/+9.5±15.6 micron, US-Cirrus -5.3±7.5/-6.2±8.9 micron, US-AngioVue -1.3±8.9/-2.6±15.3 micron, US-Spectralis -14.1±7.6/12.8±8.9 micron, and US-DRI +1.9±8.6/+2.6±8.5 micron. These values and the histogram representations in Fig 2 show that Pentacam overestimates CCT with respect to US pachymetry.

thumbnail
Fig 2.

Histogram representations of paired between-device differences of the left eyes (2. A) and right eyes (2. B). Compared to the ultrasound device, mean and standard deviation difference clearly exist between devices. While some devices over-estimate CCT in average (e.g. Pentacam), other under-estimate it (e.g. NCSM). Furthermore, while some comparisons show rather large standard deviation (e.g. Pentacam), some match better with the US device (e.g. DRI).

https://doi.org/10.1371/journal.pone.0203884.g002

NCSM is similar to US pachymetry for small values, but underestimates higher values. Cirrus slightly overestimates lower values with respect to US pachymetry. For higher values, the two devices are in good agreement. AngioVue overestimates low values and underestimates higher values with respect to US pachymetry. Spectralis overestimates low values and is in good agreement with US pachymetry for higher values. DRI SSOCT overestimates low and underestimates higher values with respect to US pachymetry. Even more surprisingly, there are differences of up to 120 μm between US pachymetry and the NCSM device.

Correlation and agreement between devices

All devices were compared to the US pachymetry device, because US pachymetry has formerly been considered the gold standard in CCT measurement, and most ophthalmologists are familiar with this device. The 95% limit of agreement (LoA) of the differences between devices was (OS) -31.6μm to +2.7μm, -6.2μm to +25.0μm, -20.0μm to +9.5μm, -18.7μm to +16.2μm, -28.9μm to +0.7μm, and -14.8μm to 18.7μm; and (OD) -35.3μm to +13.6μm, -21.1μm to +40.1μm, -23.6μm to +11.2μm, -32.6μm to +27.4μm, -30.2μm to +4.6μm, and -14.0μm to 19.3μm, for US minus Pentacam, US minus NSCM, US minus Cirrus, US minus AngioVue, US minus Spectralis, and US minus DRI. Further, Bland-Altman difference plots were used to represent these comparisons (Fig 3), because they argue that any two methods that are designed to measure the same parameter should have good correlation.

thumbnail
Fig 3.

Bland-Altman analysis of all measured data of the left eye (A) and right eye (B) compared to US pachymetry. The results suggest that the Pentacam and the Spectralis SDOCT overestimate CCT with respect to the ultra-sound device. Further, the DRI SSOCT also slight overestimates the values, with respect to US measurements. The results further suggest a good agreement of both the Cirrus SDOCT and the AngioVue with US pachymetry measurements. On one hand, the NCSM device corresponds well with US pachymetry for lower values, but on the other hand shows underestimation for higher CCT values.

https://doi.org/10.1371/journal.pone.0203884.g003

Mean absolute deviation

Deviation of the individual devices from the mean CCT value of each eye was (OS/OD) for Pentacam 11.3±5.9/10.6±7.3 μm μm(SD), Topcon NCSM 12.8±5.0/14.9±9.4 μm μm(SD), Cirrus OCT 4.2±3.2/5.7±4.6 μm μm(SD), AngioVue OCT 5.1±3.4/5.9±10.3 μm μm(SD), Spectralis OCT 10.7±5.2/10.4±4.8 μm μm(SD), Topcon DRI OCT 6.0±3.9/6.2±4.9 μm μm(SD), and US pachymetry 4.8±4.1/5.7±4.6 μm μm(SD).

Discussion

This study compared an unprecedented number of devices for central corneal thickness measurements. The aim was to define the lowest deviation from the overall mean CCT value.

CCT measurement results in this study were in good agreement with previous reports. [23, 24] With respect to the inter-eye correlation, a considerably large inter-device variation was found. While, the results of this study indicate that Pentacam and spectral OCT are quite consistent [25], the results of all other OCT-based devices in this study seem to rather agree with the ultrasound device.

In contrast, previous studies have found that OCT devices underestimated CCT compared to US pachymetry.[26, 27] But since these studies were performed with a time-domain OCT and not with spectral- or swept-source technologies, their comparability to our study is limited. Swept-source OCT measured slightly higher values than US pachymetry, which, however, is in agreement with prior findings.[28] Chaudhry et al. presented CCT data from NCSM, which compared well with US pachymetry. However, they used a different generation of device than this study did and performed their study on a smaller population. [29] Additionally, to compare each device’s measurement results with the US value, measurement deviation was analyzed by calculating mean absolute deviation from mean CCT values per eye. Ultrasound, DRI, AngioVue and the Cirrus were all close to the mean, the other three devices (Pentacam, NCSM, and Spectralis) showed more than doubled deviation. Thus, a new insight of this study was that Cirrus OCT showed the lowest deviation from the mean CCT value.

This study could be limited because no repeated measurements were performed to assess device precision. However, repeatability of CCT measurements has already been reported earlier. [8, 10, 24, 30, 31] Furthermore, it has been discussed in literature that better axial resolutions might positively influence the accuracy of cornea thickness measurement, mainly because the individual layers (such as for example Bowman’s membrane) inside the cornea can be distinguished more clearly [32]. Hence, comparisons of CCT measurement devices may inherently be limited because of the differences in their respective resolutions.

Another limitation may be that this study did not compare corneal pachymetry values, because the US device does not provide thinnest pachymetry, as the decentered point cannot easily be detected by punctual measurements. Still it was crucial to include the US device, as clinical practice has traditionally been relying on that method for decision-making. Furthermore, the results may not be applicable to patients with pathological corneas or eyes that have had prior corneal surgery because of potentially change in biomechanical variables and corneal shape. Finally, this study demonstrates that the CCT measurement by OCT showed slightly superior results compared to the other measurement methods.

Supporting information

S1 Table. Central corneal thickness measurement data are included.

https://doi.org/10.1371/journal.pone.0203884.s001

(PDF)

Acknowledgments

The authors would like to thank Nadia Maloca and Laura Maloca, Jolanda Burri and Mona Haltmayer for their research assistance.

References

  1. 1. Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, et al. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Archives of ophthalmology. 2002;120(6):714–20; discussion 829–30. Epub 2002/06/07. pmid:12049575.
  2. 2. Herndon LW, Weizer JS, Stinnett SS. Central corneal thickness as a risk factor for advanced glaucoma damage. Archives of ophthalmology. 2004;122(1):17–21. Epub 2004/01/14. pmid:14718289.
  3. 3. Cui J, Zhang X, Hu Q, Zhou WY, Yang F. Evaluation of Corneal Thickness and Volume Parameters of Subclinical Keratoconus Using a Pentacam Scheimflug System. Current eye research. 2016;41(7):923–6. Epub 2015/11/26. pmid:26605668.
  4. 4. Pflugfelder SC, Liu Z, Feuer W, Verm A. Corneal thickness indices discriminate between keratoconus and contact lens-induced corneal thinning. Ophthalmology. 2002;109(12):2336–41. Epub 2002/12/06. pmid:12466180.
  5. 5. Shousha MA, Perez VL, Wang J, Ide T, Jiao S, Chen Q, et al. Use of ultra-high-resolution optical coherence tomography to detect in vivo characteristics of Descemet's membrane in Fuchs' dystrophy. Ophthalmology. 2010;117(6):1220–7. Epub 2010/02/19. pmid:20163865; PubMed Central PMCID: PMCPMC2881170.
  6. 6. Hillenaar T, van Cleynenbreugel H, Verjans GM, Wubbels RJ, Remeijer L. Monitoring the inflammatory process in herpetic stromal keratitis: the role of in vivo confocal microscopy. Ophthalmology. 2012;119(6):1102–10. Epub 2012/03/01. pmid:22361312.
  7. 7. Alio JL, Soria F, Abbouda A, Pena-Garcia P. Laser in situ keratomileusis for -6.00 to -18.00 diopters of myopia and up to -5.00 diopters of astigmatism: 15-year follow-up. Journal of cataract and refractive surgery. 2015;41(1):33–40. Epub 2014/12/04. pmid:25465210.
  8. 8. Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. The British journal of ophthalmology. 2004;88(2):174–7. Epub 2004/01/23. pmid:14736765; PubMed Central PMCID: PMCPMC1772014.
  9. 9. Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44(5):367–408. Epub 2000/03/29. pmid:10734239.
  10. 10. O'Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea. 2005;24(8):920–4. Epub 2005/10/18. pmid:16227833.
  11. 11. Huang J, Ding X, Savini G, Pan C, Feng Y, Cheng D, et al. A Comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness. Ophthalmology. 2013;120(10):1951–8. Epub 2013/05/16. pmid:23672973.
  12. 12. Almubrad TM, Osuagwu UL, Alabbadi I, Ogbuehi KC. Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clinical ophthalmology (Auckland, NZ). 2011;5:871–6. Epub 2011/07/16. pmid:21760714; PubMed Central PMCID: PMCPMC3133003.
  13. 13. Bao F, Wang Q, Cheng S, Savini G, Lu W, Feng Y, et al. Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea. 2014;33(6):576–81. Epub 2014/04/05. pmid:24699563.
  14. 14. Chan TC, Biswas S, Yu M, Jhanji V. Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik. Medicine. 2015;94(30):e1219. Epub 2015/07/30. pmid:26222852; PubMed Central PMCID: PMCPMC4554136.
  15. 15. Chan TC, Liu D, Yu M, Jhanji V. Longitudinal evaluation of posterior corneal elevation after laser refractive surgery using swept-source optical coherence tomography. Ophthalmology. 2015;122(4):687–92. Epub 2014/12/10. pmid:25487425.
  16. 16. Alcon L, Incorporation. Specifications OcuScan® RxP Measuring System Operator's Manual 2005 [2018, 12 June]. Available from: http://www.frankshospitalworkshop.com/equipment/documents/ophthalmology/user_manuals/Alcon%20OcuScan%20RxP%20Ophthamic%20Ultrasound%20System%20-%20User%20manual.pdf.
  17. 17. Schroder S, Eppig T, Langenbucher A. A Concept for the analysis of repeatability and precision of corneal shape measurements. Zeitschrift fur medizinische Physik. 2016;26(2):150–8. Epub 2016/02/15. pmid:26874793.
  18. 18. Optovue I. Specifications AngioVue axial scan resolution 2018, June, 12 [2018, 12 June]. Available from: https://www.optovue.com/products/angiovue.
  19. 19. Topcon C. Specifications TOPCON Specular Microscope SP-1P [PDF]. 2013 [2018, 12 June]. Available from: www.imemcogroup.com/SP-1P-um.pdf.
  20. 20. Helb HM, Charbel Issa P, Fleckenstein M, Schmitz-Valckenberg S, Scholl HP, Meyer CH, et al. Clinical evaluation of simultaneous confocal scanning laser ophthalmoscopy imaging combined with high-resolution, spectral-domain optical coherence tomography. Acta ophthalmologica. 2010;88(8):842–9. Epub 2009/08/27. pmid:19706019.
  21. 21. Legarreta JE, Gregori G, Punjabi OS, Knighton RW, Lalwani GA, Puliafito CA. Macular thickness measurements in normal eyes using spectral domain optical coherence tomography. Ophthalmic surgery, lasers & imaging: the official journal of the International Society for Imaging in the Eye. 2008;39(4 Suppl):S43–9. Epub 2008/09/10. pmid:18777876.
  22. 22. TopconCorporation. Specification Topcon Atlantis DRI 1 OCT 2018, June 12. Available from: http://www.topconmedical.com/_assets/cf/downloadFile.cfm?section=productLit&file=DRI%20OCT-1%20Brochure_Revised.pdf&parentUUID=E5DF4B3A-A099-BF4C-8D7BB7E4864C06CA.
  23. 23. Nam SM, Im CY, Lee HK, Kim EK, Kim TI, Seo KY. Accuracy of RTVue optical coherence tomography, Pentacam, and ultrasonic pachymetry for the measurement of central corneal thickness. Ophthalmology. 2010;117(11):2096–103. Epub 2010/06/19. pmid:20557940.
  24. 24. Ishibazawa A, Igarashi S, Hanada K, Nagaoka T, Ishiko S, Ito H, et al. Central corneal thickness measurements with Fourier-domain optical coherence tomography versus ultrasonic pachymetry and rotating Scheimpflug camera. Cornea. 2011;30(6):615–9. Epub 2011/02/02. pmid:21282999.
  25. 25. Randleman JB, Lynn MJ, Perez-Straziota CE, Weissman HM, Kim SW. Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. The British journal of ophthalmology. 2015;99(9):1176–81. Epub 2015/04/01. pmid:25824260.
  26. 26. de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol. 2007;144(5):712–8. Epub 2007/09/18. pmid:17869208.
  27. 27. Fukuda S, Kawana K, Yasuno Y, Oshika T. Anterior ocular biometry using 3-dimensional optical coherence tomography. Ophthalmology. 2009;116(5):882–9. Epub 2009/05/05. pmid:19410946.
  28. 28. Li H, Leung CK, Wong L, Cheung CY, Pang CP, Weinreb RN, et al. Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography. Ophthalmology. 2008;115(5):796–801.e2. Epub 2007/10/06. pmid:17916376.
  29. 29. Chaudhry IA. Measurement of central corneal thickness in health and disease. Saudi journal of ophthalmology: official journal of the Saudi Ophthalmological Society. 2009;23(3–4):179–80. Epub 2009/10/01. pmid:23960857; PubMed Central PMCID: PMCPMC3729847.
  30. 30. Ferrer-Blasco T, Dominguez-Vicent A, Esteve-Taboada JJ, Aloy MA, Adsuara JE, Montes-Mico R. Evaluation of the repeatability of a swept-source ocular biometer for measuring ocular biometric parameters. Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. 2017;255(2):343–9. Epub 2016/12/03. pmid:27900479.
  31. 31. Ramesh PV, Jha KN, Srikanth K. Comparison of Central Corneal Thickness using Anterior Segment Optical Coherence Tomography Versus Ultrasound Pachymetry. Journal of clinical and diagnostic research: JCDR. 2017;11(8):Nc08–nc11. Epub 2017/10/04. pmid:28969170; PubMed Central PMCID: PMCPMC5620811.
  32. 32. Ge L, Yuan Y, Shen M, Tao A, Wang J, Lu F. The role of axial resolution of optical coherence tomography on the measurement of corneal and epithelial thicknesses. Investigative ophthalmology & visual science. 2013;54(1):746–55. Epub 2012/11/10. pmid:23139281.