Index extraction for electromagnetic field evaluation of high power wireless charging system

Index extraction for electromagnetic field evaluation of high power wireless charging system

  • SangWook Park
PLOS
x

Abstract

This paper presents the precise dosimetry for highly resonant wireless power transfer (HR-WPT) system using an anatomically realistic human voxel model. The dosimetry for the HR-WPT system designed to operate at 13.56 MHz frequency, which one of the ISM band frequency band, is conducted in the various distances between the human model and the system, and in the condition of alignment and misalignment between transmitting and receiving circuits. The specific absorption rates in the human body are computed by the two-step approach; in the first step, the field generated by the HR-WPT system is calculated and in the second step the specific absorption rates are computed with the scattered field finite-difference time-domain method regarding the fields obtained in the first step as the incident fields. The safety compliance for non-uniform field exposure from the HR-WPT system is discussed with the international safety guidelines. Furthermore, the coupling factor concept is employed to relax the maximum allowable transmitting power. Coupling factors derived from the dosimetry results are presented. In this calculation, the external magnetic field from the HR-WPT system can be relaxed by approximately four times using coupling factor in the worst exposure scenario.

Introduction

Recently, wireless power transfer (WPT) technique using electromagnetic resonance phenomena has been proposed by MIT research team [1]. This technique can achieve efficient WPT over relatively long distances between transmitter and receiver compared to conventional inductive power transfer techniques [28]. The highly resonant (HR) WPT systems typically operate in the low MHz range of approximately 1 to 20 MHz, because this range is proper to reduce wire loss caused by the skin depth and to realize high resonant coils. Since this HR-WPT technique has been reported, there has been increasing interest in the commercialization of HR-WPT system.

The HR-WPT systems using the strong reactive near fields may induce high internal electric fields in a human body in the vicinity to the systems. In particular, the HR-WPT systems can produce excessively higher fields than those used in wireless communications systems. Therefore, it is important to evaluate the internal electric fields induced in the human body in the vicinity of the HR-WPT system and to confirm compliance with international safety guidelines. The related studies have been conducted in [911].

For commercialization, one of the most interesting factors is the maximum allowable transmitting power under the compliance with the safety guidelines. Safety guidelines/standards for non-ionizing electromagnetic field exposure of occupational and general public have been provided by the International Commission on Non-Ionizing Radiation Protection (ICNIRP 1998 [12] and 2010 [13]) and the Institute of Electrical and Electronics Engineers (IEEE C95.1–2005 [14]). Test methods and procedures for electromagnetic exposure safety of products have been provided by IEEE ICES/Technical Committee (TC) 34 and International Electrotechnical Commission (IEC) TC 106. There is no experimental compliance procedure for electromagnetic fields exposure between 400 kHz and 30 MHz except for IEC 62311 [15], which describes only general aspects. Thus, when evaluating the compliance of non-uniform electromagnetic field exposure with the safety guidelines/standards, the measured maximum external field strength is first compared to a limit of reference level or maximum permissible exposure, provided by ICNIRP guidelines and IEEE standards. Generally, if the measured external fields are not satisfied for the reference levels, the internal fields induced in the human body should be obtained and compared to the limit of the basic restrictions provided by the safety guidelines/standards. Above 100 kHz, these guidelines/standards provide basic restrictions for electromagnetic fields in metric of the specific absorption rate (SAR) in the body to prevent whole-body heat stress and excessive localized tissue heating. For the general public, a whole-body averaged SAR limit of 0.08 W/kg and a localized SAR limit of 2 W/kg averaged over 10 g of tissue are recommended. The frequency range of interest for HR-WPT system is in a transition region where the other basic restriction is provided to protect electrostimulation of nerve tissues below 10 MHz. It is difficult to comply with the reference level in the case of applications requiring high transmitting power such as electric vehicle charging. Therefore, the use of coupling factor concept, which is defined by IEC 62311, will be the reasonable procedure to evaluate the compliance by measuring external field and comparing to the reference levels.

In this paper, the HR-WPT system is designed to operate at 13.56 MHz frequency, which is one of Industry-Science-Medical band. The dosimetry for the system is conducted by the two-step approach [10] in the condition of misalignment as well as alignment between transmitter and receiver, and along the various location of human body with respect to the system. Coupling factors are derived from the dosimetry results for the HR-WPT system.

Materials and methods

HR-WPT system

A highly resonant wireless power transfer (HR-WPT) system is compactly designed to operate at 13.56 MHz frequency (Fig 1). The HR-WPT system consists of two resonant coils and two feeding loops. The inner loops play the role of a matching circuit. The outer coil radius of the HR-WPT system and the power transfer distance are set by 150 mm. Copper wire with a radius of 2 mm is used for the system. In the conditions of misalignment (Fig 1b) as well as alignment (Fig 1a) between transmitter and receiver, the power transfer efficiency is investigated because of often the situation when charging. The power transfer efficiency in the misalignment case generally decreases without matching. However, the practical HR-WPT system should be matched for the power transfer efficiency. In this work, the power transfer efficiency is investigated with various misaligned distances, t, in the matching and mismatching conditions. The designed HR-WPT system is simulated by commercial electromagnetic solver (FEKO) based on the hybrid technique of method of moment (MOM) and the finite element method (FEM), to obtain the power transfer efficiency and the electric and magnetic field nearby the system.

thumbnail
Fig 1. Configuration of highly resonant wireless power transfer system.

(a) alignment, (b) misalignment.

https://doi.org/10.1371/journal.pone.0180019.g001

Computational method

To investigate the compliance with the safety guidelines for the HR-WPT system used in the vicinity of the human body, dosimetry for the system is conducted with the anatomically realistic Japanese adult male model, TARO [16]. This human model possesses 2-mm spatial resolution and 51 tissues and organs, based on the accumulated magnetic resonant imaging (MRI) images of adult Japanese volunteer. The electrical properties of the TARO model are taken from Gabriel’s Cole-Cole models [17]. The SAR in the human body is calculated by using a two-step computation method [10]. In the first step, the electric fields, which are produced by the HR-WPT system in the absence of the human body, are calculated by using the HFSS. In the second step, by considering the fields obtained in the previous step as the incident fields, the SARs in the human body are computed by using the scattered-field FDTD method [18]. It should be noted that this two-step computation method does not include the interaction effects between the HR-WPT system and the human body. However, the interaction effects are negligible because of the weak backscattering from the human body [19].

Coupling factor

The applications of WPT technique will gradually require higher transmitting powers. This means that a human will be exposed to strong fields in proximity to the HR-WPT system. This condition makes the compliance of the safety guidelines to be difficult. The reasonable procedure is necessary for both companies and consumers. Employing coupling factor concept is able to increase the maximum allowable transmitting power under the compliance of the safety guidelines. That is, the factor compensates the spatial-peak field strength measured at one point for equivalent comparison to the basic restrictions. The general definition of the coupling factor in [15] is employed for the compliance with a HR-WPT system. In this work, the external field strength to compare with the reference levels is considered only the magnetic field because the magnetic field strength is dominant rather than magnetic field strength in a WPT system, which the electromagnetic field energies is transferred by magnetic coupling. Coupling factors for the SAR and 2mm-spatial peak SAR have been derived by the following equations, respectively. (1) (2) where SARcal and SARcal_peak are the peak values of the SAR averaged over 10 g of tissue and the SAR averaging volume of 2mm× 2mm × 2mm cube in the human model, while SARlim is their basic restrictions provided by the safety guidelines. Hcal and Hlim are the spatial maximum magnetic field calculated by HFSS and the reference level for the magnetic field strength in the safety guidelines, respectively. κpeak is always larger than κ because SARcal_peak averaging 2-mm spatial volume is larger than SARcal averaging 22-mm spatial volume at the peak point. Employing κpeak expects the conservative estimation rather than κ. When measuring the magnetic field for a practical WPT system, the calculated value of Hcal can be replaced by measured value. The measure or calculated magnetic field strength multiplied by coupling factors is relaxed and can be directly compared to the reference level.

Results and discussion

Characteristics and analysis of the HR-WPT system

The power transfer efficiency (|S21|2) of the HR-WPT system designed here is 98% in the alignment case. The power transfer efficiencies along with the misaligned distance (t) are shown in Fig 2, for matching and mismatching condition. The matching condition can be achieved by adjusting the radius of inner loop (r).

thumbnail
Fig 2. Power transfer efficiency comparison of HR-WPT system along with matching and mismatching condition.

https://doi.org/10.1371/journal.pone.0180019.g002

The input impedance is changed with the coupling coefficient between the inner loop and the outer coil. This coupling coefficient depends on the mutual magnetic flux between the outer coil and the inner loop. Thus, we can easily find the matching condition as the input impedance is swept by adjusting the loop size. Without the matching, the power transfer efficiency rapidly drops as the misaligned distance increases. However, the power transfer efficiencies improve with the matching compared to those without the matching. Specially, the efficiency improvement of 83.6% is the best at misaligned distance of 250 mm. At this misaligned distance, the difference of radiation from the HR-WPT system between the alignment and misalignment situation is the largest among the other misaligned distance.

The electric field strength and magnetic field strength for reference level of the 1998 ICNIRP guideline are 28 V/m and 0.073 A/m, respectively, at 13.56 MHz frequency [12]. The electric and magnetic field strength nearby the HR-WPT system are calculated to investigate the compliance with the reference level of the system. The magnetic and electric field strength distributions nearby the system of 1 W input power for the alignment (case1) and misalignment (case2) are shown in Figs 3 and 4, respectively, with the reference level, which is indicated by black solid line. The field distributions are the results in the absence of human body. However, the field distribution results are shown with the human body model to confirm the compliance with the ICNIRP guidelines. As shown in the results, the area over the reference level of magnetic field strength is wider than that of electric field strength. We can also find that the area over the reference level in the misalignment condition is wider than in the alignment condition. This result suggests that we should consider dosimetry for the misalignment situation.

thumbnail
Fig 3. Magnetic field strength distribution nearby the high resonant wireless power transfer system.

The black solid line indicates the reference level defined as ICNIRP guideline.

https://doi.org/10.1371/journal.pone.0180019.g003

thumbnail
Fig 4. Electric field strength distribution nearby the high resonant wireless power transfer system.

The black solid line indicates the reference level defined as ICNIRP guideline.

https://doi.org/10.1371/journal.pone.0180019.g004

Dosimetry and coupling factors

Alignment and misalignment.

WPT technology is being applied in various fields. In the case of an electronic device requiring a low output power, the level of electromagnetic field radiation is not so large. However, in high-power applications such as electric vehicles, the emission of electromagnetic fields is very serious. In this paper, the exposure in the proximity of the WPT systems on the ground is investigated, assuming the application of the most serious electric vehicle. Anatomically realistic human voxel model, TARO, stands in front of a HR-WPT system as shown in Fig 5(a). Dosimetry scenarios are one case for alignment and four cases for misalignment as shown in Fig 5(b). In order to investigate dosimetry for misalignment with regard to that for alignment, the human body is located nearby transmitter (case2 and case3) and nearby receiver (case4 and case5) because strong magnetic and electric fields are constructed around the transmitting and receiving resonant coils. At these cases, distance between a HR-WPT system and a human body is 50 mm. SARs results are shown in Table 1. These results show that the SAR in a misalignment case is generally larger than that in alignment case. It is found that the SAR in case 2 is the largest, that is to say the worst case among the exposure scenarios.

thumbnail
Fig 5.

(a) configuration of human voxel model with respect to a highly resonant wireless power transfer system, (b) exposure scenarios for alignment and misalignment conditions.

https://doi.org/10.1371/journal.pone.0180019.g005

thumbnail
Table 1. Specific absorption rate in a human body for alignment and misalignment.

https://doi.org/10.1371/journal.pone.0180019.t001

The log scale SAR distributions for case 1 and case 2 are shown in Fig 6, respectively. As you can see, the SARs are strongly distributed at lower half of body because a HR-WPT system is close to feet of human model. It is also found that the distributed SAR for case 2 is entirely stronger than that for case 1. We can confirm these SAR results corresponding to the field distributions in the previous section. The maximum allowable transmitting powers (MAP) for those cases are calculated with basic restrictions of the ICNIRP guidelines and are shown in Fig 7. This calculation suggests that the MAP of the HR-WPT system is 6.36 kW from the SAR10g result in the worst exposure scenario (case 2).

thumbnail
Fig 6. SAR distributions of cross section in xy plane.

(a) case 1, (b) case 2.

https://doi.org/10.1371/journal.pone.0180019.g006

Various distances.

Dosimetry along with transition of distances (d) between the human body and the HR-WPT system for alignment condition is also conducted and compared to each other. The distances are 10 mm, 30 mm, 50 mm, 90 mm, and 110 mm. The SAR results are listed in Table 2. As we expect, the SARs decreases along with the distance. Fig 8 shows MAP along with distances between a human body and a HR-WPT system. At 10 mm distance of the worst exposure scenario in these cases, MAP is 27.8 kW, which is larger than that of misalignment case 2. The result represents that the factor of misalignment between the two transmitting and receiving coils is significant for electromagnetic field exposure.

thumbnail
Table 2. Specific absorption rate in a human body along with distances from a HR-WPT system.

https://doi.org/10.1371/journal.pone.0180019.t002

thumbnail
Fig 8. Maximum allowable power (MAP) along with distances between a human body and a highly resonant wireless power transfer system.

https://doi.org/10.1371/journal.pone.0180019.g008

Coupling factors.

Coupling factors κpeak and κ are computed with calculated magnetic fields (Hcal) in the center of a human body. From the coupling factors, the compensated magnetic fields, which mean maximum allowable external magnetic fields to satisfy reference levels, are derived. In this paper, the respect relaxation factors, ∝mp and ∝m, are defined as inverse coupling factors 1/κpeak and 1/κ. The results are listed in Table 3.

thumbnail
Table 3. Coupling factors and maximum allowable external magnetic fields.

https://doi.org/10.1371/journal.pone.0180019.t003

As shown in Fig 3, near field from a HR-WPT system is non-uniform. Gradient of the field decreases along with distance from a HR-WPT system. Therefore, we can expect that coupling factor decreases along with distance as shown in Table 3. In case 2, which has the largest gradient of the field, coupling factor is the largest among cases at the same distance of 50 mm. The external magnetic field can be relaxed four to nigh times along with distance range of 10 mm to 110 mm, and four to twenty two times in the alignment and misalignment cases at 50 mm distance. Finally, in this calculation, the relaxation factor of the worst exposure scenario, four times, can be applied to the HR-WPT system for all cases.

Conclusions

Dosimetry for a 13.56 MHz highly resonant wireless power transfer system (HR-WPT) has been conducted in the situations of misalignment and transition of distance between a human and the system. Anatomically realistic voxel human model was used to investigate specific absorption rates (SAR) in the model computed by the scattered field finite difference time domain method. From the results, maximum allowable transmitting powers and coupling factors in the various exposure scenarios are derived. It is found that the SAR for the misalignment case is larger than that for the alignment case. The results suggest that MAP is 6.36 kW and relaxation factor is 4 in this calculation.

Acknowledgments

This work was supported by a grant “Development of Induction/magnetic resonance type 6.6kW, 90% EV Wireless Charger (No. 10052912)” from the Ministry of Trade, Industry and Energy of Korea.

Author Contributions

  1. Conceptualization: SP.
  2. Funding acquisition: SP.
  3. Methodology: SP.
  4. Software: SP.
  5. Supervision: SP.
  6. Validation: SP.
  7. Writing – original draft: SP.
  8. Writing – review & editing: SP.

References

  1. 1. Kurs A, Karalis A, Moffatt R, Joannopoulos JD, Fisher P, Soljacic M. Wireless Power Transfer via Strongly Coupled Magnetic Resonances. Science (80-) 2007;317:83–6. pmid:17556549
  2. 2. Schuder J, Stephenson H, Townsend J. High-level electromagnetic energy transfer through a closed chest wall. Inst Radio Engrs Int Conv Rec 1961.
  3. 3. Hurley W, Kassakian J. Induction heating of circular ferromagnetic plates. IEEE Trans Magn 1979;15:1174–81.
  4. 4. Green A. W. 10 kHz inductively coupled power transfer—concept and control. Proc 5th Int Conf Power Electron Var Drives 1994;1994:694–9.
  5. 5. Jang Y, Jovanović MM. A contactless electrical energy transmission system for portable-telephone battery chargers. IEEE Trans Ind Electron 2003;50:520–7.
  6. 6. Choi B, Nho J, Cha H, Ahn T, Choi S. Design and Implementation of Low-Profile Contactless Battery Charger Using Planar Printed Circuit Board Windings as Energy Transfer Device. IEEE Trans Ind Electron 2004;51:140–7.
  7. 7. Hui SYR, Ho WWC. A new generation of universal contactless battery charging platform for portable consumer electronic equipment. IEEE Trans Power Electron 2005;20:620–7.
  8. 8. Elliott G a. J, Covic G a., Kacprzak D, Boys JT. A New Concept: Asymmetrical Pick-Ups for Inductively Coupled Power Transfer Monorail Systems. IEEE Trans Magn 2006;42:3389–91.
  9. 9. Laakso I, Hirata A. Evaluation of the induced electric field and compliance procedure for a wireless power transfer system in an electrical vehicle. Phys Med Biol 2013;58:7583–93. pmid:24107570
  10. 10. Park SW, Wake K, Watanabe S. Incident electric field effect and numerical dosimetry for a wireless power transfer system using magnetically coupled resonances. IEEE Trans Microw Theory Tech 2013;61:3461–9.
  11. 11. Park SW. Dosimetry for Resonance-Based Wireless Power Transfer Charging of Electric Vehicles. J Electromagn Eng Sci 2015;15:129–33.
  12. 12. Herbertz J. Comment on the ICNIRP guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz). Health Phys 1998;75:535. pmid:9790564
  13. 13. International Commission on Non-Ionizing Radiation Protection. Guidelines for limiting exposure to time-varying electric and magnetic fields (1 Hz to 100 kHz). Health Phys 2010;99:818–36. pmid:21068601
  14. 14. Ieee C95.1–2005. IEEE Standard for Safety Levels With Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz. vol. 2005. 2006.
  15. 15. IEC62311. Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0 Hz–300 GHz) 2007.
  16. 16. Nagaoka T, Watanabe S, Sakurai K, Kunieda E, Watanabe S, Taki M, et al. Development of realistic high-resolution whole-body voxel models of Japanese adult males and females of average height and weight, and application of models to radio-frequency electromagnetic-field dosimetry. Phys Med Biol 2004;49:1–15. pmid:14971769
  17. 17. Gabriel C. Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies. Environ Heal 1996; Report No.:21. Report N.AL/OE-TR- 1996–0037.
  18. 18. Kunz K, Luebbers R. The finite difference time domain method for electromagnetics. 1993.
  19. 19. Park SW, Wake K, Watanabe S. Calculation errors of the electric field induced in a human body under quasi-static approximation conditions. IEEE Trans Microw Theory Tech 2013;61:2153–60.