Influence of intraocular lens subsurface nanoglistenings on functional visual acuity

Correction

17 Apr 2017: The PLOS ONE Staff (2017) Correction: Influence of intraocular lens subsurface nanoglistenings on functional visual acuity. PLOS ONE 12(4): e0176318. https://doi.org/10.1371/journal.pone.0176318 View correction

Abstract

To investigate the influence of intraocular lens subsurface nanoglistenings (SSNGs) on functional visual acuity (FVA), thirty-nine eyes of 29 patients were examined in this study. The SSNG group comprised 19 eyes of 14 patients (75.7± 5.4 years, mean ± standard deviation), and the control group comprised 20 eyes of 15 patients (73.6 ± 6.5 years). The SSNGs were diagnosed on the basis of the typical whitish IOL appearance upon slit-lamp examination and results of densitometry regarding surface light scattering using Scheimpflug images. The FVA measurement system (AS-28; Kowa, Aichi, Japan) was used to examine changes in continuous visual acuity (VA) over time, and visual function parameters such as FVA, visual maintenance ratio (VMR), maximum VA, minimum VA, standard deviation of VA, and number of blinks were assessed. The results were compared between the SSNG and control groups, and correlations of FVA parameters with the intensity of surface light scattering, time after surgery, and age were also evaluated. There were significant differences in VMR (P = 0.035) and standard deviation of VAs (P = 0.031) between the two groups, although no significant differences were found in baseline VA, FVA, maximum VA, minimum VA, and number of blinks. None of the FVA parameters showed any significant correlations with the intensity of surface light scattering, time after surgery, or age. There is a possibility that VA is unstable during a continuous gazing task in patients with SSNGs.

Introduction

Over time, particular types of hydrophobic acrylic intraocular lens (IOL) develop a white, opaque appearance because of an increase in light scattering on their surface [1, 2]. Using scanning electron microscopy, Ong et al. [3] observed hydration-related vacuoles up to 120 μm from the surface of these IOLs; the vacuoles have diameters less than 200 nm and are known as subsurface nanoglistenings (SSNGs). The main cause of these phenomena is thought to be water aggregation in the surface layer, which in turn results from phase separation [4, 5]. The intensity of surface light scattering reportedly increases during the years following the IOL implantation [2, 6]; thus, clinicians fear that patients with this particular condition may suffer decreased visual function.

However, results differ as to whether the surface light scattering due to SSNGs degrades visual function. Several studies have suggested that SSNGs have no negative influence on visual acuity (VA) and/or contrast sensitivity [2, 4, 7], whereas various others have now indicated a risk of decline in visual function [6, 810]. For instance, Miyata et al. [6] showed that corrected distance VA tended to decrease in patients with surface light scattering of IOLs when the scattering intensity exceeded 50 computer-compatible tape (CCT) steps on a Scheimpflug image. Furthermore, Yoshida et al. [8] reported a patient with decreased VA and contrast sensitivity resulting from excessive surface light scattering in an acrylic IOL; they extracted the IOL and replaced it with a new one, thus improving the degraded visual function. Similarly, Matsushima et al. [9] reported a case series of five patients with decreased vision owing to severe SSNGs and glistening formation in IOLs. The same group also demonstrated that light transmittance was decreased in all the extracted IOLs; they suggested that this had caused the deterioration in visual function [8, 9]. Beheregaray et al. [10] showed that SSNG-associated increases in forward light scattering (FLS) were significantly correlated with a reduction in VA and in contrast sensitivity. The examples given indicate that SSNG-induced surface light scattering affects visual function somewhat, and that this effect depends on the severity of the SSNGs. The impact of these artifacts on quality of vision needs to be further examined from various perspectives.

Visual impairment is usually assessed using conventional VA testing with high-contrast optotypes. Contrast sensitivity and glare testing provide important additional information regarding visual function. However, the results of these examinations are expressed as one value; therefore, they do not reflect the entire spectrum of results. Recently, much attention has been given to the sequential changes in visual function, because daily tasks generally involve continuous, not static, acquisition of visual information. Functional visual acuity (FVA) testing was developed to assess dynamic changes in visual functioning [1114]; the technique has proven quite useful in detecting masked impairment of visual function in patients with dry eye [1113, 15, 16], mild cataract [17], posterior capsule opacification after cataract surgery [18], early presbyopia [19], epiretinal membrane [20], and age-related macular degeneration [21]. In addition, FVA testing detected slight visual deterioration induced by viscous eyedrops [22], eye ointment [23], and soft contact lenses (SCLs) [24], which was not identified using conventional VA testing.

Given that many acrylic IOLs have now been implanted, and that SSNG-associated surface light scattering gradually increases over many years [2, 6], it is crucial that investigators evaluate the impact of SSNGs on continuous visual functioning. For this reason, we conducted the current prospective study to investigate the influence of SSNGs on the FVA by comparing the results of the FVA test between eyes with and without SSNGs.

Subjects and methods

Participants

We conducted a comparative prospective study at the Miyata Eye Hospital from April 2014 to March 2016. To examine the influence of SSNG on FVA, we established two groups—those with SSNGs and those without—and recruited candidates separately for each group. The eligibility criteria are shown in Table 1. The study was approved by the institutional review board of Miyata Eye Hospital, and adhered to the tenets of the Declaration of Helsinki. After explaining the nature of the study, each patient signed a written consent form before being enrolled. This consent procedure was approved by the IRB.

thumbnail
Table 1. Eligibility criteria for SSNG and control groups.

https://doi.org/10.1371/journal.pone.0173574.t001

The SSNGs were diagnosed by at least two ophthalmologists on the basis of the typical whitish IOL appearance upon slit-lamp examination. In addition, increased light scattering on the anterior IOL surface was quantitatively evaluated using a Scheimpflug system (EAS-1000; Nidek Co. Ltd., Aichi, Japan). Specifically, we obtained a Scheimpflug slit image of the IOL at the 0-degree meridian and measured the average scattering light intensity of the central 1.00 × 0.25-mm area of the anterior optic surface using the axial densitometry of the computer. The scattering light intensity was expressed in CCT steps ranging from 0 (minimum) to 255 (maximum) [2, 25]. Only eyes with the surface light scattering higher than 50 CCT steps were included in the SSNG group.

Functional visual acuity measurement system

An FVA measurement system (AS-28; Kowa, Aichi, Japan) was used to examine changes in continuous VA over time. This system has been described in detail elsewhere [1114]. Briefly, testing was performed monocularly, with best spectacle correction under photopic conditions, and measurements were initiated at the established baseline VA for each patient. Subjects delineated an automatically presented Landolt ring orientation using a joystick. Optotype size was changed in single steps depending on the patient’s responses: the optotype was enlarged when the response was incorrect, and reduced when the response was correct. If there was no response within two seconds, the answer was recorded as an error and the optotype was enlarged. This testing was continuously performed for 60 seconds under spontaneous blinking. The FVA measurements included several evaluation parameters: FVA, visual maintenance ratio (VMR), maximum VA, minimum VA, standard deviation of VAs, and number of blinks. FVA was defined as the average of all visual acuity values measured over time, because this average value may reflect daily vision more efficiently than the visual acuity measured at a specific time point. VMR was defined as FVA divided by baseline VA [13]. Using this index, it was possible to compare groups with different baseline VAs [13, 14]. The standard deviation of VA was used to indicate visual stability (or instability) during the 60-second testing period. Maximum and minimum VAs were defined as the best and worst VAs recorded during the examination, respectively (Fig 1) [26].

thumbnail
Fig 1. Parameters of functional visual acuity testing.

Continuous red line shows sequential visual acuities measured over a 60-second measurement session. Green line denotes functional visual acuity which is calculated as the average of all visual acuity values. Pink circle represents baseline visual acuity. Visual maintenance ratio refers to area beneath time-wise change in visual acuity (red oblique line area) divided by area beneath baseline visual acuity (purple square area). Maximum and minimum visual acuities (blue and orange circles) imply the best and worst values of visual acuity over the testing period.

https://doi.org/10.1371/journal.pone.0173574.g001

Statistical methods

Regarding the obtained data, normally distributed data were compared between the SSNG and control groups using the Student’s t-test, and data that were not normally distributed were compared using the Mann-Whitney U test. In addition, correlations between the various FVA parameters and (1) the intensity of surface light scattering, (2) time after surgery, and (3) age were evaluated using the Pearson correlation test and Spearman correlation test. A P-value < 0.05 was considered statistically significant.

Results

In total, 39 eyes of 29 patients were enrolled in this study. The SSNG group comprised 19 eyes of 14 patients (two men, twelve women) with a mean age of 75.7 ± 5.4 (SD) years (range 66–84 years). The control group comprised 20 eyes of 15 patients (three men, twelve women) with a mean age of 73.6 ± 6.5 years (range 66–88 years). There was no significant difference in age between the groups (P = 0.2868). Five different models of the AcrySof IOL (Alcon, Inc., Fort Worth, TX) were used in the implants (MA60BM, MA60AC, MA30BA, SA60AT, and SN60WF). The intensity of surface light scattering was 137.5 ± 43.1 CCT in the SSNG group and 17.0 ± 2.6 CCT in the control group, showing a significant difference (P < 0.0001).

Table 2 shows the results of the FVA testing. There were significant differences in VMR (P = 0.035) and standard deviation of VAs (P = 0.031) between the SSNG and control groups, although no significant differences were found in baseline VA, FVA, maximum VA, minimum VA, and number of blinks. In addition, none of the FVA parameters showed any significant correlations with the intensity of surface light scattering, time after surgery, or age in the SSNG group (Table 3).

thumbnail
Table 2. Comparison of FVA testing results between SSNG and control groups.

https://doi.org/10.1371/journal.pone.0173574.t002

thumbnail
Table 3. Relationship of FVA parameters with the intensity of surface light scattering, time after surgery, and age in the SSNG group.

https://doi.org/10.1371/journal.pone.0173574.t003

Discussion

As shown in the results, the SSNG group had a lower VMR and a larger VA standard deviation than the control group, although the baseline VA did not differ between the groups. The lower VMR implies that patients with SSNGs had difficulty maintaining their baseline VA, and the larger VA standard deviation suggests that the VA fluctuated more widely during the continuous visual task. Together, these findings show that the VA is unstable in eyes with SSNGs. This was the first study to investigate and clarify the influence of SSNG on the stability of vision over time.

In this study, there were no significant correlations between the degree of surface light scattering and visual function parameters, probably because the index of surface light scattering—examined using a Scheimpflug system in this study—reflects backward light scattering (BLS): the light scattered out of the eye towards the light source that can be observed during slit-lamp examinations. In contrast, FLS comprises the light scattered towards the retina, hence it can reduce retinal image contrast, induce glare, and affect visual function [27]. Beheregaray et al. [10] examined both BLS and FLS in patients with SSNGs and showed no association between the two parameters. They also investigated the relationships of BLS and FLS with visual function, finding that increases in FLS, which were evaluated using an optical approach, were significantly correlated with reductions in VA and in contrast sensitivity, whereas BLS was not correlated with visual function in any way [10]. In light of these findings, it stands to reason that there were no significant relationships between the degree of surface light scattering and FVA parameters in this study. Unfortunately, we did not evaluate FLS in the current study; thus, further studies are necessary to clarify the association between FLS and FVA parameters.

We do not have enough data to definitely explain how SSNG-associated surface light scattering causes VA instability; however, several reasons can be assumed. Previous studies revealed that light transmittance is decreased in extracted IOLs with SSNGs [8, 9]. Naturally, lower incident light on the retina reduces visual function. In fact, we previously showed that almost all FVA parameters decreased under low lighting conditions in healthy eyes [28]. Intraocular scattering also degrades retinal image quality [29], so it is unsurprising that patients take longer to judge the orientation of optotypes, even when their decline in VA is not detected using conventional VA testing. The unique feature of the FVA testing system is that a time element is added to VA assessment—sequential measurements are made over a period of time. In addition, the optotype presentation time is ≤ 2 seconds for each measurement; hence, subjects need to rapidly respond to continuously changing optotypes throughout the examination. During conventional VA testing, it often takes time to elicit a response especially in elderly people; however, the acquired result does not reflect this delay in response—only a single VA value is given. In this regard, FVA testing differs from other visual examinations such as conventional VA, contrast sensitivity, and glare disability testing: FVA testing assesses another aspect of visual functioning. This property may be advantageous in that early or small SSNG-induced changes in visual function can be detected. Indeed, this kind of delayed response may also affect reading speed, which was regrettably not examined in the current study. Further studies should be conducted to elucidate this point.

There were some limitations to this study. Firstly, we did not examine FLS. As mentioned, there were no direct associations between the degree of surface light scattering, which corresponds to BLS, and visual parameter degradation in the present study. However, FLS is likely to be more closely related to retinal image quality and visual function [10, 27]. Thus, in future studies, the correlation between FLS and FVA parameters should be investigated in eyes with SSNGs; FLS can be measured using the double-pass or compensation-comparison methods [3032]. Especially, subjective FLS, which can be evaluated by the latter method through psychophysical straylight measurements, closely predicts functional light scattering [33], and hence it should be examined in relation to visual function in further researches regarding SSNGs.

Secondly, all implanted IOLs were made from hydrophobic acrylic materials by the same company, but light-filtering range (blue-light or ultraviolet-only filtering) and optical design (aspheric or spherical surface) were somewhat different among the IOL models. With regard to light filtering properties, many studies have shown that visual function in eyes with blue-light filtering IOLs is comparable to that in eyes with ultraviolet-only filtering IOLs [3437]. On the other hand, as for the influence of surface asphericity on visual performance, previous studies showed conflicting results [3841]. Although several researchers found no difference in visual performance including contrast sensitivity between aspheric and spherical IOLs [3840], the results of a Meta-Analysis showed that contrast sensitivity in eyes with aspheric IOLs was slightly better than that in eyes with spherical IOLs especially under mesopic conditions, although there was no significant difference in best-corrected visual acuity between the IOLs [41]. In our study, the FVA testing was performed under photopic conditions with high-contrast optotypes, and thus the influence of the difference in IOL asphericity on the study outcomes seems small. In addition, the distribution of the implanted IOL models was similar between the SSNG and control groups (S1 Table). Based on the above, we believe that the study outcomes were not greatly influenced by the difference of IOL models.

Another weakness of our study was the small sample size. Therefore, the current results should be confirmed in a larger study population with a unified IOL model.

In summary, this study investigated the influence of SSNGs on FVA; it found that SSNGs caused instability in VA during a gazing task although they did not seem to induce any decline in standard VA. Furthermore, patients with SSNGs have some difficulty judging or responding quickly to presented optotypes. In the ever-progressing developments of the information society, continuous-gazing tasks, such as reading, work at visual display terminals, and smartphone use are required in many situations of daily life. Driving also requires optimal continuous vision to avoid traffic accidents—especially at night and in bad weather conditions. In addition, the intensity of SSNG surface scattering continuously increases with time [2, 6]. Therefore, the current results need to be corroborated by broader-based studies. Finally, FVA testing detected slight changes in visual function in the current study; this may help physicians identify and understand unknown visual complaints in patients with SSNGs.

Supporting information

S1 Table. Distribution of implanted IOL models in each group.

https://doi.org/10.1371/journal.pone.0173574.s001

(DOCX)

Author Contributions

  1. Conceptualization: T. Hiraoka K. Miyata TO.
  2. Data curation: T. Hiraoka MI KT.
  3. Formal analysis: T. Hayashidera KT K. Minami.
  4. Funding acquisition: T. Hiraoka.
  5. Investigation: T. Hayashidera MI KT K. Minami.
  6. Methodology: T. Hiraoka K. Miyata.
  7. Project administration: K. Miyata TO.
  8. Resources: K. Miyata.
  9. Software: K. Miyata.
  10. Supervision: K. Miyata TO.
  11. Validation: K. Miyata TO.
  12. Visualization: K. Miyata TO.
  13. Writing – original draft: T. Hiraoka TO.
  14. Writing – review & editing: K. Miyata TO.

References

  1. 1. Nishihara H, Yaguchi S, Onishi T, Chida M, Ayaki M (2003) Surface scattering in implanted hydrophobic intraocular lenses. J Cataract Refract Surg 29:1385–1388. pmid:12900250
  2. 2. Miyata K, Otani S, Nejima R, Miyai T, Samejima T, Honbo M, et al. (2009) Comparison of postoperative surface light scattering of different intraocular lenses. Br J Ophthalmol 93:684–687. pmid:19276100
  3. 3. Ong MD, Callaghan TA, Pei R, Karakelle M (2012) Etiology of surface light scattering on hydrophobic acrylic intraocular lenses. J Cataract Refract Surg 38:1833–1844. pmid:22999603
  4. 4. Mönestam E, Behndig A (2011) Impact on visual function from light scattering and glistenings in intraocular lenses, a long-term study. Acta Ophthalmol 89:724–728. pmid:20064111
  5. 5. Matsushima H, Mukai K, Nagata M, Gotoh N, Matsui E, Senoo T (2009) Analysis of surface whitening of extracted hydrophobic acrylic intraocular lenses. J Cataract Refract Surg 35:1927–1934. pmid:19878825
  6. 6. Miyata K, Honbo M, Otani S, Nejima R, Minami K. Minami (2012) Effect on visual acuity of increased surface light scattering in intraocular lenses. J Cataract Refract Surg 38:221–226. pmid:22322162
  7. 7. Hayashi K, Hirata A, Yoshida M, Yoshimura K, Hayashi H (2012) Long-term effect of surface light scattering and glistenings of intraocular lenses on visual function. Am J Ophthalmol 154:240–251. pmid:22633349
  8. 8. Yoshida S, Matsushima H, Nagata M, Senoo T, Ota I, Miyake K (2011) Decreased visual function due to high-level light scattering in a hydrophobic acrylic intraocular lens. Jpn J Ophthalmol 55:62–66. pmid:21331695
  9. 9. Matsushima H, Nagata M, Katsuki Y, Ota I, Miyake K, Beiko GH, et al. (2015) Decreased visual acuity resulting from glistening and sub-surface nano-glistening formation in intraocular lenses: A retrospective analysis of 5 cases. Saudi J Ophthalmol 29:259–263. pmid:26586975
  10. 10. Beheregaray S, Yamamoto T, Hiraoka T, Oshika T (2014) Influence on visual function of forward light scattering associated with subsurface nanoglistenings in intraocular lenses. J Cataract Refract Surg 40:1147–1154. pmid:24874769
  11. 11. Goto E, Yagi Y, Matsumoto Y, Tsubota K (2002) Impaired functional visual acuity of dry eye patients. Am J Ophthalmol 133:181–186. pmid:11812420
  12. 12. Goto E, Yagi Y, Kaido M, Matsumoto Y, Konomi K, Tsubota K (2003) Improved functional visual acuity after punctal occlusion in dry eye patients. Am J Ophthalmol 135:704–705. pmid:12719079
  13. 13. Kaido M, Dogru M, Yamada M, Sotozono C, Kinoshita S, Shimazaki J, et al. (2006) Functional visual acuity in Stevens-Johnson syndrome. Am J Ophthalmol 142:917–922. pmid:17157576
  14. 14. Kaido M, Dogru M, Ishida R, Tsubota K (2007) Concept of functional visual acuity and its applications. Cornea 26(9 Suppl 1):S29–35.
  15. 15. Ishida R, Kojima T, Dogru M, Kaido M, Matsumoto Y, Tanaka M, et al. (2005) The application of a new continuous functional visual acuity measurement system in dry eye syndromes. Am J Ophthalmol 139:253–258. pmid:15733985
  16. 16. Kaido M, Ishida R, Dogru M, Tamaoki T, Tsubota K (2008) Efficacy of punctum plug treatment in short break-up time dry eye. Optom Vis Sci 85:758–763. pmid:18677228
  17. 17. Yamaguchi T, Negishi K, Dogru M, Saiki M, Tsubota K (2009) Improvement of functional visual acuity after cataract surgery in patients with good pre- and postoperative spectacle-corrected visual acuity. J Refract Surg 25:410–415. pmid:19507792
  18. 18. Wakamatsu TH, Yamaguchi T, Negishi K, Kaido M, Matsumoto Y, Ishida R, et al. (2011) Functional visual acuity after neodymium:YAG laser capsulotomy in patients with posterior capsule opacification and good visual acuity preoperatively. J Cataract Refract Surg 37:258–264. pmid:21241907
  19. 19. Katada Y, Negishi K, Watanabe K, Shigeno Y, Saiki M, Torii H, et al. (2016) Functional visual acuity of early presbyopia. PLoS One 11:e0151094. pmid:26959362
  20. 20. Nishi Y, Shinoda H, Uchida A, Koto T, Mochimaru H, Nagai N, et al. (2013) Detection of early visual impairment in patients with epiretinal membrane. Acta Ophthalmol 91:e353–357. pmid:23387556
  21. 21. Tomita Y, Nagai N, Suzuki M, Shinoda H, Uchida A, Mochimaru H, et al. (2016) Functional visual acuity in age-related macular degeneration. Optom Vis Sci 93:70–76. pmid:26583795
  22. 22. Ishioka M, Kato N, Takano Y, Shimazaki J, Tsubota K (2009) The quantitative detection of blurring of vision after eyedrop instillation using a functional visual acuity system. Acta Ophthalmol 87: 574–575. pmid:18507722
  23. 23. Hiraoka T, Yamamoto T, Okamoto F, Oshika T (2013) Changes in functional visual acuity and ocular wavefront aberration after administration of eye ointment. J Ocul Pharmacol Ther 29:770–775. pmid:23808568
  24. 24. Watanabe K, Kaido M, Ishida R, Dogru M, Negishi K, Tsubota K (2014) The effect of tinted soft contact lens wear on functional visual acuity and higher-order aberrations. Cont Lens Anterior Eye 37:203–208. pmid:24721637
  25. 25. Tanaka Y, Kato S, Miyata K, Honbo M, Nejima R, Kitano S, et al. (2004) Limitation of Scheimpflug videophotography system in quantifying posterior capsule opacification after intraocular lens implantation. Am J Ophthalmol 137:732–735. pmid:15059713
  26. 26. Kaido M, Toda I, Ishida R, Konagai M, Dogru M, Tsubota K (2011) Age-related changes in functional visual acuity in healthy individuals. Jpn J Ophthalmol 55:183–189. pmid:21584727
  27. 27. Fujikado T, Kuroda T, Maeda N, Ninomiya S, Goto H, Tano Y, et al. (2004) Light scattering and optical aberrations as objective parameters to predict visual deterioration in eyes with cataracts. J Cataract Refract Surg 30:1198–1208. pmid:15177593
  28. 28. Hiraoka T, Hoshi S, Okamoto Y, Okamoto F, Oshika T (2015) Mesopic Functional Visual Acuity in Normal Subjects. PLoS One 10:e0134505. pmid:26218066
  29. 29. Artal P, Ferro M, Miranda I, Navarro R (1993) Effects of aging in retinal image quality. J Opt Soc Am A 10:1656–1662. pmid:8350155
  30. 30. Saad A, Saab M, Gatinel D (2010) Repeatability of measurements with a double-pass system. J Cataract Refract Surg 36:28–33. pmid:20117702
  31. 31. Artal P, Benito A, Pérez GM, Alcón E, De Casas A, Pujol J, et al. (2011) An objective scatter index based on double-pass retinal images of a point source to classify cataracts. PLoS One 6:e16823. pmid:21326868
  32. 32. Franssen L, Coppens JE, van den Berg TJ (2006) Compensation comparison method for assessment of retinal straylight. Invest Ophthalmol Vis Sci 47:768–776. pmid:16431978
  33. 33. Henriksen BS, Kinard K, Olson RJ (2015) Effect of intraocular lens glistening size on visual quality. J Cataract Refract Surg 41:1190–1198. pmid:26189379
  34. 34. Rodríguez-Galietero A, Montés-Micó R, Muñoz G, Albarrán-Diego C (2005) Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation. J Cataract Refract Surg 31:1736–1740. pmid:16246777
  35. 35. Muñoz G, Belda-Salmerón L, Albarrán-Diego C, Ferrer-Blasco T, Fernández-Porrero A (2012) Contrast sensitivity and color perception with orange and yellow intraocular lenses. Eur J Ophthalmol 22:769–775. pmid:22287168
  36. 36. Lavric A, Pompe MT (2014) Do blue-light filtering intraocular lenses affect visual function? Optom Vis Sci 91:1348–1354. pmid:25216318
  37. 37. Li X, Kelly D, Nolan JM, Dennison JL, Beatty S (2016) The evidence informing the surgeon's selection of intraocular lens on the basis of light transmittance properties. Eye (Lond). [Epub ahead of print]
  38. 38. Muñoz G, Albarrán-Diego C, Montés-Micó R, Rodríguez-Galietero A, Alió JL (2006) Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg 32:1320–1327. pmid:16863968
  39. 39. Wang L, Tang X (2010) Meta-analysis of clinical randomized controlled trials on aspherical intraocular lens. Zhonghua Yan Ke Za Zh 46:254–257.
  40. 40. Muñoz G, Albarrán-Diego C, Galotto MÁ, Pascual J, Ferrer-Blasco T (2011) Lack of effect of intraocular lens asphericity on visual performance with acrylic intraocular lenses. Eur J Ophthalmol 21:723–731. pmid:21319139
  41. 41. Liu J, Zhao J, Ma L, Liu G, Wu D, Zhang J (2013) Contrast sensitivity and spherical aberration in eyes implanted with AcrySof IQ and AcrySof Natural intraocular lens: the results of a meta-analysis. PLoS One 8:e77860. pmid:24205001