Peer Review History

Original SubmissionSeptember 29, 2020
Decision Letter - Stephane Gasman, Editor

PONE-D-20-30666

Correlative fluorescence microscopy, transmission electron microscopy and secondary ion mass spectrometry (CLEM-SIMS) for cellular imaging

PLOS ONE

Dear Silvio,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the minor points raised by one of the reviewer. Espacially, the methods section needs to be better developed as it lacks some information.

Please submit your revised manuscript by december 15. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Stephane Gasman

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: I Don't Know

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The Authors are presenting a protocol that enables the correlation between three imaging techniques. Starting from a fully established CLEM protocol based on sequential imaging with Light Microscopy (LM) and Electron Microscopy (EM) of thin resin sections of vitrified/freeze substituted resin embedded material, the Authors succeeded in analyzing the same sections with secondary ion mass spectroscopy (SIMS). The resulting method, CLEM-SIMS, offer the possibility to overlay to the EM ultrastructure the compositional information gained by nanoSIMS having also the chance to target specific cells or organelles due to the correlation with LM.

As proof of principle the Authors imaged in a correlative manner cultured cells stained with MitoTracker or mito-mCitrine transfected. CLEM-SIMS approach showed that non-functional mitochondria (i.e. MitoTracker negative) despite having a similar morphology, have a molecular composition different from functional ones. This clearly shows the advantage of correlating analytical information with a CLEM approach over an independent correlation between SIMS and LM or EM alone.

I agree with the authors that their protocol will be extremely useful to correlating SIMS signal to specific cellular structure and potentially to study turnover of proteins and molecule.

I have only few technical concerns.

1)Uranyl acetate fluorescence

The Authors claimed that the fluorescent signal observed in the DAPI channel is the signal from uranyl acetate ( see lines 122-123; 130 and 435). In a recent publication Tuijtel, M. W. et al. (DOI: 10.1038/s41598-017-10905-x) used indeed uranyl acetate (UA) fluorescence in order to correlate CLEM images. Instead of what the Authors are claiming Tuijtel, M. W. et al. reported that the fluorescent signal coming from UA is visible using filter set for GFP (excitation 470/40, emission 525/50) at cryogenic temperature, and far less intense at 21°C. Moreover the fluorescent signal arising from UA revealed the outline of the cells, nucleoli and mitochondria. In figure 2 of this manuscript instead the DAPI channel is highlighting heterochromatic regions. Could the authors comment on these discrepancies?

2)Material and Methods

The authors should reported the excitation and emission parameter of the fluorescent filter sets and the imaging parameter used. Information of the type of transmission electron microscope used, camera and pixel sizes are missing.

Image registration

In the image registration chapter, the author should explain in more detail the procedure of alignment and ideally give an estimation of the alignment accuracy over the 3 imaging modalities. As suggestion the Authors could draw a line on structures visible in the 3 modalities (i.e mitochondria) calculate the full-with-half maximum of the intensity over the line for EM, LM and nanoSIMS in order to obtain an estimation of the alignment accuracy. Such images with their respective plots calculated from the signal of the 3 images modalities could be added as supplementary data.

Minor comments.

High pressure freezing/freeze substitution (HPF/FS).

The authors correctly decide to avoid chemical fixation and process their sample with HPF/FS as they explained in the Discussion chapter. Nevertheless, in order to process the samples in HPF planchettes, the cells were detached using trypsin. In my opinion for further studies the Author should consider to use sapphire disc in order to vitrify cultured cells without the need of tripisinization.

NanoSIMS compare to other analytical methods.

Other analytical methods have been recently developed in the contest of a correlative approach. The author did a good effort in highlighting potentiality and pitfall of the technology they developed. For a reader would be useful also to compare this new methods with other correlative approached based on other analytical techniques (i.e. energy dispersive X-ray (EDX) or electron energy loss spectroscopy (EELS)).

Reviewer #2: This is a very interesting and useful manuscript in which the authors present a novel and insightful methodology to correlate the ultrastructure of cells with the molecular composition of cell compartments. The technique is based in previous methodological advances achieved by the group of Dr. Silvio Rizzoli who leads this study. The novelty is the combination of electron microscopy and fluorescence microscopy with secondary ion mass spectrometry (SIMS) to visualize the localization of spots enriched in different types of natural isotopes, which is useful to, for example, detect specific organelles (such as mitochnodria o difference in chromatin organization). The group is pionneering here the implementation of high pressure freezing which is an advantage to investigate cellular ultrastructure under conditions that are better to preserve structures compared to conventional fixation methods. This is a methodological and useful study in which the conclusions are supported by high quality data. The authors explain and discuss the methods in detail. The paper is very nicely and clearly written and opens new perspectives to investigate specific aspects of the biology of the cell such as the aging of organelles.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Reply to Reviewers

Reviewer #1:

The Authors are presenting a protocol that enables the correlation between three imaging techniques. Starting from a fully established CLEM protocol based on sequential imaging with Light Microscopy (LM) and Electron Microscopy (EM) of thin resin sections of vitrified/freeze substituted resin embedded material, the Authors succeeded in analyzing the same sections with secondary ion mass spectroscopy (SIMS). The resulting method, CLEM-SIMS, offer the possibility to overlay to the EM ultrastructure the compositional information gained by nanoSIMS having also the chance to target specific cells or organelles due to the correlation with LM.

As proof of principle the Authors imaged in a correlative manner cultured cells stained with MitoTracker or mito-mCitrine transfected. CLEM-SIMS approach showed that non-functional mitochondria (i.e. MitoTracker negative) despite having a similar morphology, have a molecular composition different from functional ones. This clearly shows the advantage of correlating analytical information with a CLEM approach over an independent correlation between SIMS and LM or EM alone.

I agree with the authors that their protocol will be extremely useful to correlating SIMS signal to specific cellular structure and potentially to study turnover of proteins and molecule.

I have only few technical concerns.

We thank the Reviewer for the comments.

1)Uranyl acetate fluorescence

The Authors claimed that the fluorescent signal observed in the DAPI channel is the signal from uranyl acetate ( see lines 122-123; 130 and 435). In a recent publication Tuijtel, M. W. et al. (DOI: 10.1038/s41598-017-10905-x) used indeed uranyl acetate (UA) fluorescence in order to correlate CLEM images. Instead of what the Authors are claiming Tuijtel, M. W. et al. reported that the fluorescent signal coming from UA is visible using filter set for GFP (excitation 470/40, emission 525/50) at cryogenic temperature, and far less intense at 21°C. Moreover the fluorescent signal arising from UA revealed the outline of the cells, nucleoli and mitochondria. In figure 2 of this manuscript instead the DAPI channel is highlighting heterochromatic regions. Could the authors comment on these discrepancies?

As the Reviewer points out, the work from Tuijtel and collaborators employed uranyl acetate fluorescence at cryogenic temperatures. This procedure not only enhances the fluorescence of uranyl acetate, but also changes its emission spectrum (Jones et al., Chem Sci, 2015, 6:513). At room temperature a peak appears in the spectrum covered by the DAPI channel, which is measurable using a conventional microscope. This is more easily measured in our work than in the work of Tuijtel and colleagues, since our sections are substantially thicker.

The fluorescence of the uranyl acetate in the green channel (the one employed by Tuijtel and collaborator) should still be measurable at room temperature, according to Jones et al., Chem Sci, 2015, 6:513. A small signal can be detected, as suggested by Tuijtel and collaborators, but it is so low as to be negligible in comparison to specific GFP signals (please see Figure 1 of our manuscript).

The localization of the uranyl acetate fluorescence is not different, in our opinion, between the two studies. We also find nucleoli to be well-labeled, in all cells, and the heterochromatin seems to be well labeled in the work of Tuijtel and collaborators (please see their Figure 3). Mitochondria also are labeled in our work, albeit the signal is dim. As an example, please find below a view in which we do not show the MitoTracker or GFP channel, and we use the same colormap as Tuijtel and colleagues. The resulting image is very similar to the published work of Tuijtel and colleagues.

(Find figure enclosed in "Response to Reviewers" file)

Figure for reviewers. Uranyl Acetate in DAPI channel. a) Uranyl acetate fluorescence in 160 nm thick resin sections of HeLa cells imaged in the DAPI channel, using a green colormap. b) The same image of the DAPI channel displayed in grayscale with inverted look up table (LUT). c) Transmission electron micrograph of the same region on the resin section imaged at 120 kV with 3500x original magnification. Scale bars: 5µm.

2)Material and Methods

The authors should reported the excitation and emission parameter of the fluorescent filter sets and the imaging parameter used. Information of the type of transmission electron microscope used, camera and pixel sizes are missing.

We have now added these missing parameters.

Image registration

In the image registration chapter, the author should explain in more detail the procedure of alignment and ideally give an estimation of the alignment accuracy over the 3 imaging modalities. As suggestion the Authors could draw a line on structures visible in the 3 modalities (i.e mitochondria) calculate the full-with-half maximum of the intensity over the line for EM, LM and nanoSIMS in order to obtain an estimation of the alignment accuracy. Such images with their respective plots calculated from the signal of the 3 images modalities could be added as supplementary data.

We have now performed this type of analysis, by measuring the full-with-half-maximum (FWHM) for several mitochondria that are labeled in all channels. The resulting values are similar, in spite of differences in the resolutions of the three techniques. We have now added the respective figure as Supplementary Data.

Minor comments.

High pressure freezing/freeze substitution (HPF/FS).

The authors correctly decide to avoid chemical fixation and process their sample with HPF/FS as they explained in the Discussion chapter. Nevertheless, in order to process the samples in HPF planchettes, the cells were detached using trypsin. In my opinion for further studies the Author should consider to use sapphire disc in order to vitrify cultured cells without the need of tripisinization.

We will use this technique in the future. We have now made a comment on this approach in our Discussion.

NanoSIMS compare to other analytical methods.

Other analytical methods have been recently developed in the contest of a correlative approach. The author did a good effort in highlighting potentiality and pitfall of the technology they developed. For a reader would be useful also to compare this new methods with other correlative approached based on other analytical techniques (i.e. energy dispersive X-ray (EDX) or electron energy loss spectroscopy (EELS)).

We have added a paragraph on these important techniques in the Discussion section.

Reviewer #2:

This is a very interesting and useful manuscript in which the authors present a novel and insightful methodology to correlate the ultrastructure of cells with the molecular composition of cell compartments. The technique is based in previous methodological advances achieved by the group of Dr. Silvio Rizzoli who leads this study. The novelty is the combination of electron microscopy and fluorescence microscopy with secondary ion mass spectrometry (SIMS) to visualize the localization of spots enriched in different types of natural isotopes, which is useful to, for example, detect specific organelles (such as mitochnodria o difference in chromatin organization). The group is pionneering here the implementation of high pressure freezing which is an advantage to investigate cellular ultrastructure under conditions that are better to preserve structures compared to conventional fixation methods. This is a methodological and useful study in which the conclusions are supported by high quality data. The authors explain and discuss the methods in detail. The paper is very nicely and clearly written and opens new perspectives to investigate specific aspects of the biology of the cell such as the aging of organelles.

We thank the Reviewer for the comments.

Attachments
Attachment
Submitted filename: Reply to Reviewers.docx
Decision Letter - Stephane Gasman, Editor

Correlative fluorescence microscopy, transmission electron microscopy and secondary ion mass spectrometry (CLEM-SIMS) for cellular imaging

PONE-D-20-30666R1

Dear Silvio,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Stephane Gasman

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: I Don't Know

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Formally Accepted
Acceptance Letter - Stephane Gasman, Editor

PONE-D-20-30666R1

Correlative fluorescence microscopy, transmission electron microscopy and secondary ion mass spectrometry (CLEM-SIMS) for cellular imaging

Dear Dr. Rizzoli:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff