Pre-analytical error for point of care venous blood testing in acute ambulatory settings : a mixed methods service evaluation

PLOS ONE

Dear Dr Fanshawe,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

In addition to the few issues raised by the reviewers, please reconsider revising the title to clearly state that the paper is on the effect of pre-analytic handling  on three POCT instrument errors. One of the most common pro-analytic errors insufficient patient ID.

We would appreciate receiving your revised manuscript by Feb 07 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Pal Bela Szecsi, M.D. D.M.Sci.

Academic Editor

PLOS ONE

Journal Requirements:

1. When submitting your revision, we need you to address these additional requirements.

Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2.  Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

3. Thank you for stating the following in the Financial Disclosure section:

This research was funded by the National Institute for Health Research (NIHR) Community Healthcare MedTech and In Vitro Diagnostics Co-operative at Oxford Health NHS Foundation Trust. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care. The project was also supported by a grant from Becton, Dickinson and Company. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

We note that you received funding from a commercial source: Becton, Dickinson and Company.

Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc.

Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include your amended Competing Interests Statement within your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #2: N/A

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The study consider the pre—analytical errors recorded with the use of three specific devices, both in an hospital setting and community in ambulatory care. The authors consider the errors registered by the instruments and in particular the errors attributable to the operators.

The errors in the pre-analytical phase include many aspects, in addition to those that can be registered by the analytical system: the choice of the right timing with respect to the clinical conditions (for example for blood gas, you could quote Auvet doi: 10.1186/s13613-016-0152-6 ), the site of the sampling, the instruments used for venipuncture, the any dilution or contamination of the blood drawn, the sample hemolysis. Despite this, the study provides new data on the safety of POCT systems. This limitation must be discussed.

Another limitation of the study consists in the typology of errors that are strictly related to the specific type of instrument considered. These results cannot be reported to other types of analyzers.

In the qualitative study the secondary analysis is based on limited number of cases.

The "Staff awareness of factors", “staff actions” and “staff perception” chapters (lines 329 - 512) present individual anecdotal cases that could be placed in a section of additional materials rather than in the text of the paper. Alternatively, the proposed reports could be summarized.

In the conclusions a positive judgment is expressed regarding the frequency of errors attributable to the operators. But it is not known what the impact of these errors on care was, as the authors themselves claim, therefore the statement appears arbitrary and must be more prudent. In absolute terms the frequency is actually high, for example if we estimate it with the sigma metric.

Figure 3 shows little relevant data that can be briefly presented in the text.

Reviewer #2: This is a well written manuscript with focus on a topic for which there are very few published studies.

I only have one comment. The authors found that there was no difference in error rates between locations performing high and low volume POCT. I would have expected the error rate to be lower in areas performing high volume testing. Could the authors speculate or comment on why they saw no difference?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Response to reviewers

*Editor:

*In addition to the few issues raised by the reviewers, please reconsider revising the title to clearly state that the paper is on the effect of pre-analytic handling on three POCT instrument errors.

We agree and have modified the title to reflect this:

Pre-analytical error for three point of care venous blood testing platforms in acute ambulatory settings : a mixed methods service evaluation

*One of the most common pro-analytic errors insufficient patient ID.

This was true of the Afinion but this was likely to be only a technical result in accessing the device (quickly resolved by the operator) and so not likely to be an operator error. As such, although we thought it appropriate to report this in results tables we do not feel that it warrants more prominence than currently.

*Reviewer #1: The study consider the pre—analytical errors recorded with the use of three specific devices, both in an hospital setting and community in ambulatory care. The authors consider the errors registered by the instruments and in particular the errors attributable to the operators.

The errors in the pre-analytical phase include many aspects, in addition to those that can be registered by the analytical system: the choice of the right timing with respect to the clinical conditions (for example for blood gas, you could quote Auvet doi: 10.1186/s13613-016-0152-6 ), the site of the sampling, the instruments used for venipuncture, the any dilution or contamination of the blood drawn, the sample hemolysis. Despite this, the study provides new data on the safety of POCT systems. This limitation must be discussed.

Thank you – we agree that this should be mentioned and we have added to the relevant part of the Discussion: “It was not possible to assess whether an incorrectly filled cartridge or poor quality sample might have given an inaccurate and therefore clinically misleading test result if the device did not initiate an error message, and this may be affected by the site of sampling or the timing when the test was performed [28]” (citing the paper suggested).

*Another limitation of the study consists in the typology of errors that are strictly related to the specific type of instrument considered. These results cannot be reported to other types of analyzers.

We have added a sentence to the limitations in the Discussion: “Our results only relate to the POCT devices already used in the clinical environment considered in our study, and cannot be extrapolated to other analyzers.”

*In the qualitative study the secondary analysis is based on limited number of cases. The "Staff awareness of factors", “staff actions” and “staff perception” chapters (lines 329 - 512) present individual anecdotal cases that could be placed in a section of additional materials rather than in the text of the paper. Alternatively, the proposed reports could be summarized.

In qualitative research, emphasis is placed on gaining rich, deep information from a relatively small number of cases, where each has substantial experience of, and involvement in, the phenomenon under investigation (in this case, POC testing). The secondary analysis strand is based on data collected in a qualitative case study of one ambulatory care unit published elsewhere (Jones C, Glogowska M, Locock L, Lasserson D. Embedding new technologies in practice – a normalization process theory study of point of care testing. BMC Health Surv Res. 2016;16:591) which encompassed 14 episodes of ethnographic observation (approximately 60 hours) and 7 semi-structured interviews with staff (6 of which were audio-recorded and transcribed). In a similar way, the prospective ethnographic strand (which took place in another ambulatory care unit), represented 15 hours of observation.

For a qualitative study, this represents a considerable dataset, which was able to provide both a description of the context in which the POC testing had been implemented and a detailed interpretation of how usual practice around POC testing took place.

As regards the “Staff awareness of factors”, “staff actions” and “staff perception” sections, the material here are rigorously collected and robustly analysed data from the secondary analysis and the prospective ethnographic strands of the qualitative component. The qualitative component was part of the original mixed methods design which aimed at producing a more comprehensive assessment of preanalytical error in POC testing in this setting. Where quotations are included, it is because they are reflective of the views of more than just the one participant to whom they are attributed and are not simply ‘anecdotal cases’. Thus, we regard the qualitative material as playing an important role in aiding our understanding, and providing some explanation of, the quantitative findings and do not think it should be moved into additional materials.

*In the conclusions a positive judgment is expressed regarding the frequency of errors attributable to the operators. But it is not known what the impact of these errors on care was, as the authors themselves claim, therefore the statement appears arbitrary and must be more prudent. In absolute terms the frequency is actually high, for example if we estimate it with the sigma metric.

We agree that our study was unable to assess the impact on patient care (this was not the objective). We have amended the Conclusions of the Abstract as a result: “In the observed settings, the rate of errors attributable to operators of the primary point of care device was less than 1 in 40. In some cases, errors may lead to a small increase in resource use or time required so adequate staff training is necessary to prevent adverse impact on patient care.”

*Figure 3 shows little relevant data that can be briefly presented in the text.

We have moved this figure to a Supporting Information file (S2 Graph) and shortened the text.

*Reviewer #2: This is a well written manuscript with focus on a topic for which there are very few published studies.

I only have one comment. The authors found that there was no difference in error rates between locations performing high and low volume POCT. I would have expected the error rate to be lower in areas performing high volume testing. Could the authors speculate or comment on why they saw no difference?

Thank you – we had also anticipated that the error rate might be lower in areas performing high volume testing but were limited in the amount of location-specific information that could be collected to determine factors that might influence error rates at different locations. We have amended the relevant part of the Discussion to now read: “Error rates in different locations did not appear to be strongly associated with total test use. This was unexpected but could be explained by location specific factors that we were unable to measure, such as the experience of individual operators. An interesting extra research question would be to investigate whether new operators have higher error rates or improve with practice.”

Attachments
Attachment
Submitted filename: Response to reviewers.docx
Decision Letter - Pal Bela Szecsi, Editor

Pre-analytical error for three point of care venous blood testing platforms in acute ambulatory settings : a mixed methods service evaluation

PONE-D-19-29304R1

Dear Dr. Fanshawe,

We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.

Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.

Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

With kind regards,

Pal Bela Szecsi, M.D. D.M.Sci.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #2: N/A

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: (No Response)

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The paper was corrected as requested. However, I believe that research can be deepened with a new study with respect to clinical risk (for example through interviews, questionnaires, event analysis) and with respect to patient safety. Also the process of patient identification sometimes is a critical issue. But this is only a my suggestion.

Reviewer #2: The authors have sufficiently addressed the concerns I raised. This manuscript is well written and addresses an important topic.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Formally Accepted
Acceptance Letter - Pal Bela Szecsi, Editor

PONE-D-19-29304R1

Pre-analytical error for three point of care venous blood testing platforms in acute ambulatory settings : a mixed methods service evaluation

Dear Dr. Fanshawe:

I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

For any other questions or concerns, please email plosone@plos.org.

Thank you for submitting your work to PLOS ONE.

With kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Pal Bela Szecsi

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .