Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

< Back to Article

Influence of Depth of Interaction upon the Performance of Scintillator Detectors

Figure 7

Energy resolution against DOI for 30 mm LYSO:Ce scintillator crystals.

The energy resolution (%) versus the depth of interaction plotted for two Proteus 2×2×30 mm3 LYSO:Ce scintillator crystals in the unwrapped and wrapped configurations. For a DOI greater than 5 mm, the mean energy resolutions are 16.52±0.02% and 13.92±0.01% for the unwrapped and wrapped configurations respectively.