Influence of Depth of Interaction upon the Performance of Scintillator Detectors

Figure 2

Photograph of the DOI timing coincidence apparatus.

The timing coincidence apparatus set up for DOI measurements. The reference detector, shown on the left of the image, is a 2×2×5 mm3 Agile Ca-co-doped LSO:Ce wrapped in PTFE tape coupled to a Hamamatsu MPPC S10931-050P SiPM using Rhodorsil 47 V optical grease. On the right, the source is shown close to the 3D-printed clamp holding the same photodetector coupled to the scintillator crystal under investigation.

Figure 2

doi: https://doi.org/10.1371/journal.pone.0098177.g002