Ultra-high resolution, 3-dimensional magnetic resonance imaging of the atherosclerotic vessel wall at clinical 7T

  • Martin J. Willemink,

    Roles Data curation, Formal analysis, Investigation, Writing – original draft, Writing – review & editing

    Affiliations Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands, Department of Radiology, Stanford University School of Medicine, Stanford, California, United States of America

  • Bram F. Coolen,

    Roles Supervision, Writing – review & editing

    Affiliations Department of Biomedical Engineering and Physics, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands, Department of Radiology and Nuclear Medicine, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands

  • Hadrien Dyvorne,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Philip M. Robson,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Ilda Bander,

    Roles Investigation, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Seigo Ishino,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Alison Pruzan,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Arthi Sridhar,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Bei Zhang,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Priti Balchandani,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Venkatesh Mani,

    Roles Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Gustav J. Strijkers,

    Roles Supervision, Writing – review & editing

    Affiliation Department of Biomedical Engineering and Physics, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands

  • Aart J. Nederveen,

    Roles Supervision, Writing – review & editing

    Affiliation Department of Radiology and Nuclear Medicine, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands

  • Tim Leiner,

    Roles Supervision, Writing – review & editing

    Affiliation Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands

  • Zahi A. Fayad,

    Roles Conceptualization, Funding acquisition, Methodology, Resources, Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • Willem J. M. Mulder,

    Roles Conceptualization, Funding acquisition, Methodology, Resources, Supervision, Writing – review & editing

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Medical Biochemistry, Academic Medical Centre, University of Amsterdam, Amsterdam, The Netherlands

  •  [ ... ],
  • Claudia Calcagno

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Resources, Supervision, Writing – original draft, Writing – review & editing

    claudia.calcagno@mssm.edu

    Affiliations BioMedical Engineering and Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America, Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America

  • [ view all ]
  • [ view less ]

Abstract

Accurate quantification and characterization of atherosclerotic plaques with MRI requires high spatial resolution acquisitions with excellent image quality. The intrinsically better signal-to-noise ratio (SNR) at high-field clinical 7T compared to the widely employed lower field strengths of 1.5 and 3T may yield significant improvements to vascular MRI. However, 7T atherosclerosis imaging also presents specific challenges, related to local transmit coils and B1 field inhomogeneities, which may overshadow these theoretical gains. We present the development and evaluation of 3D, black-blood, ultra-high resolution vascular MRI on clinical high-field 7T in comparison lower-field 3T. These protocols were applied for in vivo imaging of atherosclerotic rabbits, which are often used for development, testing, and validation of translatable cardiovascular MR protocols. Eight atherosclerotic New Zealand White rabbits were imaged on clinical 7T and 3T MRI scanners using 3D, isotropic, high (0.63 mm3) and ultra-high (0.43 mm3) spatial resolution, black-blood MR sequences with extensive spatial coverage. Following imaging, rabbits were sacrificed for validation using fluorescence imaging and histology. Image quality parameters such as SNR and contrast-to-noise ratio (CNR), as well as morphological and functional plaque measurements (plaque area and permeability) were evaluated at both field strengths. Using the same or comparable imaging parameters, SNR and CNR were in general higher at 7T compared to 3T, with a median (interquartiles) SNR gain of +40.3 (35.3–80.1)%, and a median CNR gain of +68.1 (38.5–95.2)%. Morphological and functional parameters, such as vessel wall area and permeability, were reliably acquired at 7T and correlated significantly with corresponding, widely validated 3T vessel wall MRI measurements. In conclusion, we successfully developed 3D, black-blood, ultra-high spatial resolution vessel wall MRI protocols on a 7T clinical scanner. 7T imaging was in general superior to 3T with respect to image quality, and comparable in terms of plaque area and permeability measurements.

Introduction

Cardiovascular disease due to atherosclerosis is the main cause of morbidity and mortality worldwide [1]. In the imaging community, there has been much focus in developing and validating non-invasive in vivo techniques to characterize arterial vessel wall atherosclerosis and to measure atherosclerotic disease burden non-invasively [24]. Specifically, vessel wall measurements by quantitative magnetic resonance imaging (MRI), such as vessel wall area, have already been demonstrated to be good predictors of future cardiovascular events [58]. These parameters have also been successfully used as surrogate imaging markers of treatment efficacy for approved and novel anti-atherosclerotic compounds [914]. In the long term, non-invasive vascular MR imaging may aid in more efficiently diagnosing and stratifying individuals at high risk for cardiovascular events, as well as monitoring patient treatment response [15, 16].

At present, vessel wall MRI still faces significant challenges [912, 1719] when acquired at the most common clinical field strengths of 1.5 and 3T. MRI of the small arterial wall requires imaging with very high-spatial resolution, for accurate quantification of plaque size and to correctly identify relevant high-risk plaque components, such as lipid rich necrotic core and fibrous cap [16]. Ideally, vascular MRI acquisitions should be 3-dimensional (3D), use isotropic voxels and allow extensive spatial coverage along a whole vessel, while avoiding partial volume artifacts in tortuous, disease-prone areas [20, 21]. In addition, efficient “black blood” preparations, which suppress the bright signal from flowing blood, and perivascular fat signal suppression are needed to ensure optimal contrast and delineation of the arterial wall with respect to the vessel lumen and other neighboring structures [20, 22, 23]. To satisfy all these requirements while maintaining high scan time efficiency, signal-to-noise ratio (SNR) and spatial resolution often need to be compromised when imaging on 1.5 and 3T scanners. This significantly affects the accuracy of atherosclerosis burden measurements and the characterization of plaque composition, especially when using advanced quantitative MR techniques [24]. Particularly, the measurement of functional parameters relevant to high-risk plaque phenotypes, such as enhanced vessel wall permeability and neovascularization using high temporal resolution dynamic contrast enhanced (DCE) MRI [25], may be significantly impacted by low SNR measurements.

The advent of high-field 7T clinical MRI scanners offers new opportunities to surpass these challenges and improve the characterization of at-risk atherosclerotic plaques [26, 27]. In principle, the intrinsically higher SNR at 7T may allow 3D vessel wall imaging with isotropic high spatial resolution, improved image quality and within imaging times suitable for clinical examinations. However, arterial vessel wall MRI at 7T also presents specific issues, for example related to the use of local transmit coils and B1 field inhomogeneities [28], which may overshadow these theoretical advantages over lower field 1.5 and 3T clinical scanners.

In this study, we aimed to assess the advantages and disadvantages of high-field clinical 7T vessel wall MRI for quantification of plaque burden and permeability, in comparison with 3T field strength. 3D vessel-wall black blood MRI protocols developed on a clinical 7T scanner were applied for high and ultra-high spatial resolution in vivo imaging of the atherosclerotic rabbit abdominal aorta and benchmarked against equivalent acquisitions at clinical 3T [18, 29]. The rabbit aorta is similar in size to the human coronary arteries [30], and rabbit aortic plaques recapitulate many phenotypical characteristics of human plaques [31]. Since rabbits can be imaged on clinical MR scanners, this relevant animal model of atherosclerosis has been extensively used in the past by our group and others for robust development, testing and validation of cardiovascular MR imaging protocols before translation into humans [17, 3239].

Methods

Ethics statement

This study was conducted after review and approval by the Institutional Animal Care and Use Committee (IACUC) of the Icahn School of Medicine at Mount Sinai.

Animal model and study design

Atherosclerosis was induced in 8 male New Zealand White (NZW) rabbits using a previously validated approach [9, 10, 1719, 29, 4042], consisting of a combination of high-fat diet and double balloon injury of the infrarenal abdominal aorta. All rabbits were imaged between 5 and 7 months of high-fat diet containing 0.3% cholesterol and 4.7% coconut oil for 2 months and 0.15% and 4.7% coconut oil thereafter. Animals were sacrificed after the last imaging time-point (Fig 1). Animals were imaged twice, once on a 7T clinical scanner (Magnetom, Siemens Healthineers, Erlangen, Germany), and once on a 3T clinical scanner (mMR, Siemens Healthineers, Erlangen, Germany).

thumbnail
Fig 1. Study design.

(A) Atherosclerosis was induced in 8 male New Zealand White rabbits. Rabbits were imaged 5–7 months after high-fat diet initiation. 3D imaging sequences were acquired from just below the left renal artery until the aortic bifurcation (B).

https://doi.org/10.1371/journal.pone.0241779.g001

7T rabbit coil design

At 7T, a custom-made coil was specifically designed to achieve optimal transmit/receive efficiency along the entire rabbit abdominal aorta for high SNR imaging of the arterial vessel wall. This coil consists of a transmit-only quadrature-driven high-pass birdcage and 12 receive-only loop elements. The transmit birdcage is located on a cylinder with a diameter of 20.3 cm, while the receive elements are located on a cylinder with a diameter of 13.3 cm. The two cylinders are concentric. The 12-channel receive-only array is arranged into two rows in the z direction, with 6 elements in each row (Fig 2E). For the 3T scans a Tx/Rx 15-channel knee coil was used for imaging (Quality Electrodynamics (QED), Mayfield, OH).

thumbnail
Fig 2. Image quality results.

Images of the aorta acquired with 3D T2 weighted SPACE at high resolution (0.63 mm3, HR) and ultra-high resolution (0.43 mm3, UHR) on a 3T and 7T MRI scanner (A). Boxplots indicating SNR measured in the vessel wall (B) and lumen (C); and CNR, D) for T2w SPACE, and pre-contrast and post-contrast T1w MERGE sequences. The 12-channel receive-only coil is displayed (E). * indicates significantly different based on Wilcoxon-signed ranks tests; SPACE, Sampling Perfection with Application optimized Contrasts using different flip angle Evolution; MERGE, Motion sensitized driven Equilibrium prepared Rapid Gradient Echo; Pre, pre-contrast administration; Post, post-contrast administration.

https://doi.org/10.1371/journal.pone.0241779.g002

MR image acquisition

On both scanners, localizer scans and time-of-flight (TOF) angiography were used to clearly identify the rabbit abdominal aorta [18, 29]. Subsequently, the following sequences were acquired:

  1. 3D T2w SPACE [20, 22] (Sampling Perfection with Application optimized Contrasts using different flip angle Evolution),
  2. Pre-contrast 3D T1w MERGE [21] (Motion Sensitized Driven Equilibrium (MSDE) [23] prepared Rapid Gradient Echo),
  3. Dynamic contrast enhanced (DCE) MRI using 3D T1w MERGE [18], and
  4. Post-contrast 3D T1w MERGE [21].

3D T2w SPACE was used to quantify atherosclerotic plaque burden (defined as vessel wall area), while DCE-MRI and pre/post-contrast 3D T1w MERGE were used to quantify plaque neovascularization/permeability. DCE-MRI was acquired before, during and after manual injection of the T1-shortening MR contrast agent gadolinium (Gd) diethylenetriamine penta-acetic acid (DTPA), at the dose of 0.1 mmol/Kg body weight. At both field strengths, all sequences, with the exception of DCE-MRI, were acquired using 2 different spatial resolutions: 0.6x0.6x0.6 mm3 (High Resolution, ‘HR’) and 0.4x0.4x0.4 mm3 (Ultra-High Resolution, ‘UHR’). 3T and 7T acquisition times were 3:50 min for T2SPACE HR and 11:06 min for T2SPACE UHR, while acquisition times of pre- and post-contrast 3D T1w MERGE were 2:39 min for 3T and 7T HR, and 5:46 min for 3T and 7T UHR, respectively. DCE-MRI was imaged only at a resolution of 0.6x0.6x0.6 mm3 (HR), since only one contrast agent injection was performed per imaging session, and to allow imaging with a sufficiently high temporal resolution for the quantification of plaque permeability [18, 29]. For all sequences, imaging parameters were kept as consistent as possible across field strengths to allow for direct comparison of signal-to-noise (SNR) measurements. No parallel imaging was used. Tables detailing imaging parameters for all sequences at each field strength can be found in the S1 Table.

MR image analysis

After image acquisition, data was transferred to a workstation with Osirix software for image analysis (http://www.osirix-viewer.com). All images were reformatted in the axial plane before analysis. For all sequences, analyses were performed in axial slices from the left renal artery, down to the aortic bifurcation.

Signal to noise (SNR) and contrast to noise (CNR) ratio measurements.

Vessel wall signal intensity was calculated by averaging the MR signal in the area defined by the inner and outer vessel wall contours. Noise regions of interest (ROIs) were traced on the 3 middle slices of each sequence, in the native acquisition plane (sagittal). Care was taken to place noise regions in the middle section of the image, where transmit and receive efficiency are higher, and in areas with no artifacts. Standard deviation of noise ROIs, std(noise), was calculated for each slice, averaged across the 3 slices, and then used to calculate SNR assuming Rician distributed noise as follows [43]: (1)

Contrast-to-noise ratio (CNR) between vessel wall and lumen was calculated as the normalized difference between vessel wall and lumen SNR. SNR and CNR were calculated for 3D T2w SPACE, and pre and post-contrast 3D T1w MERGE sequences, at both field strengths and spatial resolutions. Slice-by-slice SNR and CNR were averaged for each rabbit from the renal arteries to the iliac bifurcation, and these values were used for statistical analysis. All ROIs were manually placed.

Vessel wall area measurements.

For each traced axial slice, vessel wall area was calculated from 3D T2w SPACE acquisitions by tracing and subtracting the areas defined by the outer and inner vessel wall contours. Slice-by-slice area measurements were averaged for each rabbit from the renal arteries to the iliac bifurcation, and these values were used for statistical analysis.

Dynamic Contrast Enhanced (DCE) MRI.

Enhanced vessel wall permeability and increased neovascularization are hallmarks of atherosclerotic plaques at high-risk for causing severe clinical events, such as myocardial infarction and stroke [44]. In the past, our group and others have demonstrated that these features can be reliably quantified using in vivo DCE-MRI at 1.5T and 3T, both in patients [4549] and large animal models [17, 18, 32] of atherosclerosis. This technique consists in the rapid acquisition of images before, during and after the injection of a gadolinium (Gd) based, T1 shortening contrast agent. Here, for DCE-MRI analysis inner and outer vessel wall contours were traced on post-contrast 3D T1w MERGE images. In several rabbit studies we have demonstrated that the area under the contrast agent uptake curve (AUC) from in vivo vessel wall DCE-MRI is a good surrogate measure of atherosclerotic plaque permeability [17, 18, 32], as validated by histology [17, 32], and by measuring vessel wall permeability to Evans Blue (EB) dye using near infra-red fluorescence (NIRF) [18]. In line with these historical data, plaque contrast agent uptake was measured by calculating the area-under-the-curve (AUC) of the MRI signal intensity over time, up to 2 minutes after contrast agent injection, averaged in each vessel wall ROI [18]. For statistical analysis, slice-by-slice AUC values were averaged across all analyzed slices, or in segments corresponding to ex vivo Evans Blue ROIs (see below for more details).

Ex vivo Evans Blue (EB) Near Infrared Fluorescence (NIRF) and histology

Within 48 hours from the second imaging session, rabbits were euthanized using an intravenous injection of sodium pentobarbital (100 mg/kg), and immediately perfused using 1L of saline solution. Before euthanasia animals were injected with 6 mL of 0.5% Evans Blue (EB, Sigma Aldrich, St. Louis, MO) in 1x PBS through a 22G catheter placed in a marginal ear vein. For each session, one animal not injected with EB served as control. EB was allowed to circulate for 30 minutes to allow for binding to serum albumin and extravasation in the vessel wall. After euthanasia, rabbit abdominal aortas were harvested, measured in length and imaged with a Xenogen IVIS-200 optical imaging system (Perkin Elmer Inc., Shelton, CT, USA). The excitation and emission filters for Evans Blue were set to 605 and 680 nm, respectively, with an exposure time of 4 s, binning at 8, FOV 12.2 cm. After fluorescence imaging, the abdominal aorta was divided into 12 segments, which were fixed using a 10% buffered formalin solution. The first 4 segments encompassed a 1.6 cm segment (0.4 cm per segment) right below the left renal artery, where formation of atherosclerotic plaques was more prominent. The other 8 segments were equally spaced from the bottom of the 4th section, up to the iliac bifurcation. Within 24 hours from fixation, specimens were embedded into paraffin blocks. Each paraffin block was sectioned (5 μm) onto glass slides and stained using Masson’s trichrome stain for vessel wall area measurements. NIRF images were analyzed by placing square 0.5x0.5 cm2 ROIs in the rabbit abdominal aortas from the left renal artery, down to the iliac bifurcation. An ROI of the same size was also placed in a portion of skeletal muscle harvested from each rabbit, acquired within the same FOV and with the same imaging parameters as a control tissue. Average and total radiant efficiency [(p/s/cm2/sr)/(μW/cm2)] for each ROI were recorded as a measure of EB deposition in the aorta. Outer and inner vessel wall contours were traced on digitized histology slides using Panoramic Viewer software (http://www.3dhistech.com/pannoramic_viewer). Vessel wall area was calculated by subtracting inner from outer vessel wall contour and averaged for each rabbit across the whole aorta (segments 1–12).

Statistical analysis

Statistical analyses were performed using IBM SPSS version 24.0 (SPSS Inc, Chicago, Illinois, USA) and MedCalc version 16.8.4 (Mariakerke, Belgium). Unless otherwise stated, data are listed as medians (first interquartile–third interquartile). Non-parametric tests were used since all quantified parameter values were not normally distributed. The Wilcoxon signed ranks test was applied to compare dependent data (vessel wall area measurements using imaging and image quality in terms of SNR and CNR) and independent data were assessed with the independent samples Mann-Whitney U test (average and total radiant efficiency on Evans Blue Near Infra-Red Fluorescence and vessel wall area measurements using histology). Correlations between vessel wall area measurements performed on imaging and histology were evaluated using the Spearman’s correlation coefficient with 95% confidence intervals (95% CI). The Spearman’s correlation coefficient was also used for comparison between DCE measurements and EB results. Correlations between all DCE measurements combined for both time points at 3T and 7T were evaluated using the intra-class correlation coefficient (ICC) with 95% CI. These ICC values were calculated for all individual segment measurements as well as for averaged measurements per animal. ICC analyses were also used to compare VWA measured at 3T with 7T. A p-value < 0.05 was considered statistically significant.

Results

Image quality: SNR and CNR

At the same spatial resolution, we found SNR to be significantly higher (p<0.05) at 7T compared to 3T for almost all acquisitions, with the exception of vessel wall pre-contrast UHR 3D T1w MERGE (p = 0.093) and lumen UHR 3D T2w SPACE (p = 0.069), where the SNR increase at 7T was not statistically significant. Median SNR gain at 7T with respect to 3T was +40.3 (35.3–80.1)%.

In vessel wall imaging, the CNR indicates the ability to discern the vessel wall from the lumen and the surrounding tissues. CNR was also found to be significantly higher at 7T as compared to 3T, except for vessel wall pre-contrast UHR 3D T1w MERGE (p = 0.889). Median CNR gain was +68.1 (38.5–95.2)%.

Irrespective of field strength, aortic vessel wall MRI SNR and CNR were higher at HR (0.6x0.6x0.6 mm3 voxel size) as compared to UHR (0.4x0.4x0.4 mm3) for all acquisitions (p = 0.012 for all comparisons). Our SNR and CNR results are in line with previously published data of the human carotid arteries [26, 27].

Detailed results of image quality measurements SNR and CNR are listed in Table 1 and Fig 2.

thumbnail
Table 1. Image quality in terms of signal-to-noise and contrast-to-noise ratios.

https://doi.org/10.1371/journal.pone.0241779.t001

Vessel wall area measurements by 3D T2 weighted SPACE MRI and histology

There were no absolute differences between VWA measurements from HR and UHR 3D T2w SPACE acquired at either 3T or 7T (p = 0.779 and p = 0.069, respectively). These results are in concordance with data obtained in clinical studies on the carotid arteries using 2D T1w fast gradient echo (FGE) and T2w turbo spin echo (TSE) imaging [27]. As shown by previous studies, the absolute value of quantitative plaque burden measurements depends on the acquired spatial resolution, with lower resolutions resulting in higher measurements [50]. In line with these results, in our study vessel wall area (VWA) measurements were significantly higher for HR as compared to UHR 3D T2w SPACE, both at 3T and 7T (p = 0.012). Intra-class correlation coefficients between VWA measured with 3T and 7T were 0.873 (95% CI 0.493–0.973) for HR and 0.966 (95% CI 0.840–0.993) for UHR images.

Correlations between VWA calculated from imaging and histology were positive but not significant for all spatial resolutions and field strengths with Spearman’s correlation coefficients ranging from 0.405 (95% CI -0.420–0.863, p = 0.320) for UHR 3D T2w SPACE acquired at 7T to 0.476 (95% CI -0.344–0.884, p = 0.233) for HR 3D T2 SPACE acquired at 7T. This result may be due to the fact that MRI and histology were correlated after averaging across the whole vessel, because of intrinsic difficulties in accurately matching corresponding imaging and histological slices. Results of the vessel wall area measurements are listed in Table 2 and displayed in Fig 3.

thumbnail
Fig 3. Vessel wall area results.

Results of vessel wall area imaging measurements for 3T and 7T (A). Spearman’s correlations between imaging and histology were positive (however not significant) (B). Comparisons of high resolution 3T and 7T images and histology (C) for a plaque with concentric thickening (left) and a larger plaque with eccentric thickening (right). HR, high resolution; UHR, ultra-high resolution; VWA, vessel wall area.

https://doi.org/10.1371/journal.pone.0241779.g003

Vessel wall permeability by in vivo DCE-MRI and ex vivo NIRF imaging

In our study, AUC measurements from 3T and 7T DCE-MRI were positively and significantly correlated, with an ICC of 0.201 [95% CI 0.037–0.354]) (Fig 4), indicating modest concordance between measurements taken at the two different field strengths. In fact, AUC measurements differed significantly between 3T (2.546 [1.810–3.571) and 7T (2.011 [1.552–2.622], p<0.001). However, AUC measurements were positively and significantly correlated with average and total radiant efficiency of ex vivo Evans Blue NIRF for both 3T and 7T DCE (Table 3 and Fig 4). Correlations were stronger for 7T (ρ = 0.385 (95% CI 0.146–0.582) for average radiant efficiency and ρ = 0.392 (95% CI 0.153–0.587) for total radiant efficiency) compared to 3T (ρ = 0.321 (95% CI 0.075–0.530) for average radiant efficiency and ρ = 0.291 (95% CI 0.043–0.506) for total radiant efficiency). Average and total radiant efficiency, averaged across all ROIs in a vessel, and across animals, were substantially higher in animals injected with EB (total radiant efficiency 6.43 (5.15–9.24) (p/s/cm2/sr)/(μW/cm2)) with respect to non-injected, control animals (0.56 (0.55–0.56) (p/s/cm2/sr)/(μW/cm2).

thumbnail
Fig 4. Correlations between 3T and 7T MRI displayed for Dynamic Contrast Enhanced images (DCE) data.

(A). Absolute comparison of DCE measurements between 3T and 7T (B). Correlation between Evans blue measurements and DCE measurements (C). Infra-red fluorescence images of two rabbits, one Evans blue injected rabbit and one control rabbit without Evans blue injection (D).

https://doi.org/10.1371/journal.pone.0241779.g004

thumbnail
Table 3. Correlations between Evans blue and dynamic contrast enhanced imaging for all measurements.

https://doi.org/10.1371/journal.pone.0241779.t003

Discussion

In this manuscript, we describe the development of a 3D black blood atherosclerosis imaging protocol with ultra-high spatial resolution, high SNR and extensive spatial coverage, on a clinical high-field 7T MR scanner.

As a first step, we have chosen to deploy our 3D vessel wall imaging protocol in atherosclerotic rabbits. Assessing the performance of 7T vessel wall imaging in this validated animal model before translation into humans has allowed us to more easily control for possible confounding factors that may render the comparison between different field strengths more challenging. Some of these potential issues are i) bulk motion, which can be avoided by the use of anesthesia in the animal model; ii) the prominent field inhomogeneities apparent in certain relevant human anatomies, such as the neck, where the carotid arteries are located, compared to the rabbit abdomen; and iii) the presence of tortuous vessels in plaque prone areas in humans, such as the carotid bulb, or the intracranial circulation, in comparison with the rabbit infra-renal abdominal aorta.

Because of its specific features, the rabbit model has been extensively used in the past by our groups and others for developing, testing and validating cardiovascular MR imaging protocols before translation into humans [17, 3239]. Atherosclerotic rabbits are a validated model of cardiovascular disease: the rabbit abdominal aorta is comparable in terms of vessel wall thickness and area to the human coronary arteries [30], and rabbit aortic plaques present many of the characteristic features of high-risk human lesions, such as the presence of active plaque inflammation, abundant plaque macrophages, as well as neovascularization and enhanced vessel wall permeability [51]. While being a small animal model, rabbits can be imaged on clinical MR scanners, and therefore represent an attractive and cost-effective alternative to larger animals for the robust development of protocols that can be readily translated into humans.

Our 7T vascular imaging protocol relies on 3D T2w SPACE [20] and 3D T1w MERGE [21], two validated MR sequences commonly employed for vessel wall imaging to quantify plaque burden and neovascularization/permeability. These sequences were evaluated at the isotropic resolution of 0.6x0.6x0.6 mm3 (referred to as “High Resolution”, HR), commonly used in our rabbit imaging protocols at 3T [18, 29, 42], and at the higher resolution of 0.4x0.4x0.4 mm3 (referred to as “Ultra-High Resolution”, UHR). This setup allowed us to evaluate both the effects of spatial resolution and field strength on vessel wall SNR, vessel wall/lumen CNR, and, ultimately, on the quantification of vessel wall area, and plaque neovascularization/permeability.

In our study, we found vessel wall SNR to be overall significantly higher at 7T compared to 3T, with a median SNR gain of +40.3%. This value is considerably lower than the expected theoretical gain of 133% when assuming direct proportionality between SNR and field strength increase. Previous studies employing 2D imaging have also found that vessel wall SNR at 7T is consistently higher than 3T [26, 27], although a higher variability of SNR across subjects at the higher 7T field strength was observed. These and our results may be attributed to the use of different transmit/receive arrays at 3T and 7T, which may significantly impact SNR and image quality regardless of field strength. Other studies, specifically investigating the use of ‘Delay Alternating with Nutation for Tailored Excitation' (DANTE) black blood preparation pulses, found only a modest increase in SNR and CNR at 7T [52]. In this case [52] more prominent aliasing and motion artifacts were observed at 7T, as a result of the prolonged scan time necessary to avoid exceeding specific absorption rate (SAR) limits, which may also have contributed to the observed marginal gain in SNR at this field strength. A lower than theoretical SNR gain at 7T may be also attributable to increased T1 and reduced T2 at this higher field strength, leading to a decrease in SNR at equal acquisition times. While imaging parameters (particularly TR and TE) could be optimized to maximize SNR efficiency, the choice for these parameters is not just dictated by relaxation parameters but also by limitations in total scan time and design of the 3D spin/gradient echo read-out train. In addition, not much literature is available on vessel wall relaxation parameters at 7T. We therefore feel that in keeping the sequence parameters similar between 3T and 7T, the best possible comparison between field strengths has been made.

In concordance with clinical studies on the carotid arteries [27], we found good agreement between vessel wall area measured by T2w SPACE at 3T and 7T, for both HR and UHR images. As previously shown [50], we found that vessel wall area was higher in HR with respect to UHR images. Vessel wall measurements by MRI showed a positive correlation with vessel wall area measured by histology, which was, however, not statistically significant. This may be attributed to the fact that vessel wall area by MRI and histological measurements cannot be perfectly slice-matched and therefore were correlated after averaging across the whole vessel. Moreover, we found that VWA measured by MRI was overestimated as compared to histology. This well-known discrepancy is due to the much higher resolution provided by microscopic histological images (μm2), compared to the much coarser resolution in vivo MRI (mm2). Therefore, while histological images allow for a very accurate quantification of vessel wall area, quantifying plaque burden by MRI carries a margin of error which is directly proportional to image resolution. In fact, as it can be observed in Fig 3, panel B overestimation of vessel wall area by MRI is more prominent at in HR versus UHR images.

While others have reported the use of contrast enhancement imaging for improved detection of vessel wall lesions in patients with posterior circulation ischemia [53] and of peripheral intracranial segments [54] at 7T, to our knowledge we are the first to report the use of dynamic contrast enhanced (DCE) MRI to quantify neovascularization and permeability in the atherosclerotic vessel wall at this higher field strength. We found good concordance between DCE-MRI results at 3T and 7T, and confirmed a correlation between AUC by DCE-MRI and vessel wall permeability evaluated by ex vivo Evans Blue, as previously validated at 3T [18]. Differences in AUC values between the two field strengths may be attributed to expected differences in contrast agent relaxivity at 3 and 7T, and to the method used to convert the MR signal to contrast agent concentration prior to AUC calculation. As previously described and validated, contrast agent concentration was estimated in this study as the relative signal enhancement at each point in time after contrast agent injection [18]. A more accurate quantification of contrast agent concentration, which may eliminate these discrepancies, requires accurate quantification of vessel wall pre-contrast T1 relaxation time with the same spatial coverage and resolution of the sequence used for DCE-MRI [55]. While such a method may certainly lead to a more accurate quantification of plaque enhancement, neovascularization and permeability, 3D T1 mapping of the arterial vessel wall is currently a topic of active investigation and its application in this context was beyond the scope of our study [56, 57].

Conclusions

In conclusion, we demonstrate the successful development of a 3D black blood MRI protocol to assess plaque burden and neovascularization/permeability in a rabbit animal model of atherosclerosis on a high-field clinical 7T scanner, in comparison with 3T. This approach allows imaging with high spatial resolution and extensive spatial coverage of the rabbit arterial vessel wall, comparable in size to the human coronary arteries, with high SNR and CNR. We find that measures of plaque burden and neovascularization/permeability are comparable between the two field strengths, with SNR and vessel wall/lumen CNR being consistently higher at 7T with respect to 3T.

Acknowledgments

The authors would like to thank Chang Ho Wessel and Samantha Baxter.

References

  1. 1. Writing Group M, Mozaffarian D, Benjamin EJ, Go AS, Arnett DK, Blaha MJ, et al. Heart Disease and Stroke Statistics-2016 Update: A Report From the American Heart Association. Circulation. 2016;133(4):e38–360. pmid:26673558
  2. 2. Ripa RS, Pedersen SF, Kjaer A. PET/MR Imaging in Vascular Disease: Atherosclerosis and Inflammation. PET Clin. 2016;11(4):479–88. pmid:27593251
  3. 3. Dweck MR, Williams MC, Moss AJ, Newby DE, Fayad ZA. Computed Tomography and Cardiac Magnetic Resonance in Ischemic Heart Disease. J Am Coll Cardiol. 2016;68(20):2201–16. pmid:27855810
  4. 4. Patel AK, Suri HS, Singh J, Kumar D, Shafique S, Nicolaides A, et al. A Review on Atherosclerotic Biology, Wall Stiffness, Physics of Elasticity, and Its Ultrasound-Based Measurement. Curr Atheroscler Rep. 2016;18(12):83. pmid:27830569
  5. 5. McDermott MM, Kramer CM, Tian L, Carr J, Guralnik JM, Polonsky T, et al. Plaque Composition in the Proximal Superficial Femoral Artery and Peripheral Artery Disease Events. JACC Cardiovasc Imaging. 2016;10(9):1003–12. pmid:27838307
  6. 6. Papini GD, Di Leo G, Bandirali M, Cotticelli B, Flor N, Restivo P, et al. Is Carotid Plaque Contrast Enhancement on MRI Predictive for Cerebral or Cardiovascular Events? A Prospective Cohort Study. J Comput Assist Tomogr. 2016;41(2):321–6.
  7. 7. Xu D, Hippe DS, Underhill HR, Oikawa-Wakayama M, Dong L, Yamada K, et al. Prediction of high-risk plaque development and plaque progression with the carotid atherosclerosis score. JACC Cardiovasc Imaging. 2014;7(4):366–73. pmid:24631510
  8. 8. Mani V, Muntner P, Gidding SS, Aguiar SH, El Aidi H, Weinshelbaum KB, et al. Cardiovascular magnetic resonance parameters of atherosclerotic plaque burden improve discrimination of prior major adverse cardiovascular events. J Cardiovasc Magn Reson. 2009;11:10. pmid:19393089
  9. 9. Lobatto ME, Calcagno C, Metselaar JM, Storm G, Stroes ES, Fayad ZA, et al. Imaging the efficacy of anti-inflammatory liposomes in a rabbit model of atherosclerosis by non-invasive imaging. Methods Enzymol. 2012;508:211–28. pmid:22449928
  10. 10. Lobatto ME, Calcagno C, Otten MJ, Millon A, Ramachandran S, Paridaans MP, et al. Pharmaceutical development and preclinical evaluation of a GMP-grade anti-inflammatory nanotherapy. Nanomedicine. 2015;11(5):1133–40. pmid:25791805
  11. 11. Vucic E, Calcagno C, Dickson SD, Rudd JH, Hayashi K, Bucerius J, et al. Regression of inflammation in atherosclerosis by the LXR agonist R211945: a noninvasive assessment and comparison with atorvastatin. JACC Cardiovasc Imaging. 2012;5(8):819–28. pmid:22897996
  12. 12. Vucic E, Dickson SD, Calcagno C, Rudd JH, Moshier E, Hayashi K, et al. Pioglitazone modulates vascular inflammation in atherosclerotic rabbits noninvasive assessment with FDG-PET-CT and dynamic contrast-enhanced MR imaging. JACC Cardiovasc Imaging. 2011;4(10):1100–9. pmid:21999870
  13. 13. Fayad ZA, Mani V, Woodward M, Kallend D, Abt M, Burgess T, et al. Safety and efficacy of dalcetrapib on atherosclerotic disease using novel non-invasive multimodality imaging (dal-PLAQUE): a randomised clinical trial. Lancet. 2011;378(9802):1547–59. pmid:21908036
  14. 14. Tang TY, Howarth SP, Miller SR, Graves MJ, Patterson AJ, JM UK-I, et al. The ATHEROMA (Atorvastatin Therapy: Effects on Reduction of Macrophage Activity) Study. Evaluation using ultrasmall superparamagnetic iron oxide-enhanced magnetic resonance imaging in carotid disease. J Am Coll Cardiol. 2009;53(22):2039–50. pmid:19477353
  15. 15. Dweck MR, Aikawa E, Newby DE, Tarkin JM, Rudd JH, Narula J, et al. Noninvasive Molecular Imaging of Disease Activity in Atherosclerosis. Circ Res. 2016;119(2):330–40. pmid:27390335
  16. 16. Coolen BF, Calcagno C, van Ooij P, Fayad ZA, Strijkers GJ, Nederveen AJ. Vessel wall characterization using quantitative MRI: what's in a number? Magn Reson Mater Phy. 2018;31(1):201–22. pmid:28808823
  17. 17. Calcagno C, Cornily JC, Hyafil F, Rudd JH, Briley-Saebo KC, Mani V, et al. Detection of neovessels in atherosclerotic plaques of rabbits using dynamic contrast enhanced MRI and 18F-FDG PET. Arterioscler Thromb Vasc Biol. 2008;28(7):1311–7. pmid:18467641
  18. 18. Calcagno C, Lobatto ME, Dyvorne H, Robson PM, Millon A, Senders ML, et al. Three-dimensional dynamic contrast-enhanced MRI for the accurate, extensive quantification of microvascular permeability in atherosclerotic plaques. NMR Biomed. 2015;28(10):1304–14. pmid:26332103
  19. 19. Cornily JC, Hyafil F, Calcagno C, Briley-Saebo KC, Tunstead J, Aguinaldo JG, et al. Evaluation of neovessels in atherosclerotic plaques of rabbits using an albumin-binding intravascular contrast agent and MRI. J Magn Reson Imaging. 2008;27(6):1406–11. pmid:18504763
  20. 20. Fan Z, Zhang Z, Chung YC, Weale P, Zuehlsdorff S, Carr J, et al. Carotid arterial wall MRI at 3T using 3D variable-flip-angle turbo spin-echo (TSE) with flow-sensitive dephasing (FSD). J Magn Reson Imaging. 2010;31(3):645–54. pmid:20187208
  21. 21. Balu N, Yarnykh VL, Chu B, Wang J, Hatsukami T, Yuan C. Carotid plaque assessment using fast 3D isotropic resolution black-blood MRI. Magn Reson Med. 2011;65(3):627–37. pmid:20941742
  22. 22. Mihai G, Winner MW, Raman SV, Rajagopalan S, Simonetti OP, Chung YC. Assessment of carotid stenosis using three-dimensional T2-weighted dark blood imaging: Initial experience. J Magn Reson Imaging. 2012;35(2):449–55. pmid:22147541
  23. 23. Wang J, Yarnykh VL, Hatsukami T, Chu B, Balu N, Yuan C. Improved suppression of plaque-mimicking artifacts in black-blood carotid atherosclerosis imaging using a multislice motion-sensitized driven-equilibrium (MSDE) turbo spin-echo (TSE) sequence. Magn Reson Med. 2007;58(5):973–81. pmid:17969103
  24. 24. Alam SR, Stirrat C, Richards J, Mirsadraee S, Semple SI, Tse G, et al. Vascular and plaque imaging with ultrasmall superparamagnetic particles of iron oxide. J Cardiovasc Magn Reson. 2015;17:83. pmid:26381872
  25. 25. van Hoof RH, Heeneman S, Wildberger JE, Kooi ME. Dynamic Contrast-Enhanced MRI to Study Atherosclerotic Plaque Microvasculature. Curr Atheroscler Rep. 2016;18(6):33. pmid:27115144
  26. 26. Koning W, de Rotte AA, Bluemink JJ, van der Velden TA, Luijten PR, Klomp DW, et al. MRI of the carotid artery at 7 Tesla: quantitative comparison with 3 Tesla. J Magn Reson Imaging. 2015;41(3):773–80. pmid:24578311
  27. 27. Kroner ES, van Schinkel LD, Versluis MJ, Brouwer NJ, van den Boogaard PJ, van der Wall EE, et al. Ultrahigh-field 7-T magnetic resonance carotid vessel wall imaging: initial experience in comparison with 3-T field strength. Invest Radiol. 2012;47(12):697–704. pmid:22996317
  28. 28. Kraff O, Quick HH. 7T: Physics, safety, and potential clinical applications. J Magn Reson Imaging. 2017;46(6):1573–89. pmid:28370675
  29. 29. Kim Y, Lobatto ME, Kawahara T, Lee Chung B, Mieszawska AJ, Sanchez-Gaytan BL, et al. Probing nanoparticle translocation across the permeable endothelium in experimental atherosclerosis. Proc Natl Acad Sci U S A. 2014;111(3):1078–83. pmid:24395808
  30. 30. Kim WY, Stuber M, Bornert P, Kissinger KV, Manning WJ, Botnar RM. Three-dimensional black-blood cardiac magnetic resonance coronary vessel wall imaging detects positive arterial remodeling in patients with nonsignificant coronary artery disease. Circulation. 2002;106(3):296–9. pmid:12119242
  31. 31. Phinikaridou A, Hallock KJ, Qiao Y, Hamilton JA. A robust rabbit model of human atherosclerosis and atherothrombosis. J Lipid Res. 2009;50(5):787–97. pmid:19141434
  32. 32. Calcagno C, Lairez O, Hawkins J, Kerr SW, Dugas MS, Simpson T, et al. Combined PET/DCE-MRI in a Rabbit Model of Atherosclerosis: Integrated Quantification of Plaque Inflammation, Permeability, and Burden During Treatment With a Leukotriene A4 Hydrolase Inhibitor. JACC Cardiovasc Imaging. 2018;11(2 Pt 2):291–301.
  33. 33. Senders ML, Lobatto ME, Soler R, Lairez O, Perez-Medina C, Calcagno C, et al. Development and Multiparametric Evaluation of Experimental Atherosclerosis in Rabbits. Methods Mol Biol. 2018;1816:385–400. pmid:29987836
  34. 34. Wang YX, Kuribayashi H, Wagberg M, Holmes AP, Tessier JJ, Waterton JC. Gradient echo MRI characterization of development of atherosclerosis in the abdominal aorta in Watanabe Heritable Hyperlipidemic rabbits. Cardiovasc Intervent Radiol. 2006;29(4):605–12. pmid:16729232
  35. 35. Millon A, Canet-Soulas E, Boussel L, Fayad Z, Douek P. Animal models of atherosclerosis and magnetic resonance imaging for monitoring plaque progression. Vascular. 2014;22(3):221–37. pmid:24907292
  36. 36. Chen H, Ricks J, Rosenfeld M, Kerwin WS. Progression of experimental lesions of atherosclerosis: assessment by kinetic modeling of black-blood dynamic contrast-enhanced MRI. Magn Reson Med. 2013;69(6):1712–20. pmid:22829477
  37. 37. Phinikaridou A, Ruberg FL, Hallock KJ, Qiao Y, Hua N, Viereck J, et al. In vivo detection of vulnerable atherosclerotic plaque by MRI in a rabbit model. Circ Cardiovasc Imaging. 2010;3(3):323–32. pmid:20194634
  38. 38. Molony D, Park J, Zhou L, Fleischer C, Sun HY, Hu X, et al. Bulk Flow and Near Wall Hemodynamics of the Rabbit Aortic Arch: A 4D PC-MRI Derived CFD Study. J Biomech Eng. 2018. pmid:30140921
  39. 39. Cuadrado I, Saura M, Castejon B, Martin AM, Herruzo I, Balatsos N, et al. Preclinical models of atherosclerosis. The future of Hybrid PET/MR technology for the early detection of vulnerable plaque. Expert Rev Mol Med. 2016;18:e6. pmid:27056676
  40. 40. van der Valk FM, van Wijk DF, Lobatto ME, Verberne HJ, Storm G, Willems MC, et al. Prednisolone-containing liposomes accumulate in human atherosclerotic macrophages upon intravenous administration. Nanomedicine. 2015;11(5):1039–46. pmid:25791806
  41. 41. Lobatto ME, Fayad ZA, Silvera S, Vucic E, Calcagno C, Mani V, et al. Multimodal clinical imaging to longitudinally assess a nanomedical anti-inflammatory treatment in experimental atherosclerosis. Mol Pharm. 2010;7(6):2020–9. pmid:21028895
  42. 42. Lobatto ME, Calcagno C, Millon A, Senders ML, Fay F, Robson PM, et al. Atherosclerotic plaque targeting mechanism of long-circulating nanoparticles established by multimodal imaging. ACS Nano. 2015;9(2):1837–47. pmid:25619964
  43. 43. Brown RW, Cheng Y-CN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 2nd Edition: Wiley-Blackwell; 2014 June 2014. 1008 p.
  44. 44. Kolodgie FD, Yahagi K, Mori H, Romero ME, Trout HHR, Finn AV, et al. High-risk carotid plaque: lessons learned from histopathology. Semin Vasc Surg. 2017;30(1):31–43. pmid:28818257
  45. 45. Kerwin WS, Oikawa M, Yuan C, Jarvik GP, Hatsukami TS. MR imaging of adventitial vasa vasorum in carotid atherosclerosis. Magn Reson Med. 2008;59(3):507–14. pmid:18306402
  46. 46. Kerwin WS, O'Brien KD, Ferguson MS, Polissar N, Hatsukami TS, Yuan C. Inflammation in carotid atherosclerotic plaque: a dynamic contrast-enhanced MR imaging study. Radiology. 2006;241(2):459–68. pmid:16966482
  47. 47. Kerwin W, Hooker A, Spilker M, Vicini P, Ferguson M, Hatsukami T, et al. Quantitative magnetic resonance imaging analysis of neovasculature volume in carotid atherosclerotic plaque. Circulation. 2003;107(6):851–6. pmid:12591755
  48. 48. Dong L, Kerwin WS, Chen H, Chu B, Underhill HR, Neradilek MB, et al. Carotid artery atherosclerosis: effect of intensive lipid therapy on the vasa vasorum—evaluation by using dynamic contrast-enhanced MR imaging. Radiology. 2011;260(1):224–31. pmid:21493792
  49. 49. Chen H, Wu T, Kerwin WS, Yuan C. Atherosclerotic plaque inflammation quantification using dynamic contrast-enhanced (DCE) MRI. Quant Imaging Med Surg. 2013;3(6):298–301. pmid:24404443
  50. 50. Harteveld AA, Denswil NP, Van Hecke W, Kuijf HJ, Vink A, Spliet WGM, et al. Ex vivo vessel wall thickness measurements of the human circle of Willis using 7T MRI. Atherosclerosis. 2018;273:106–14. pmid:29715587
  51. 51. Spagnoli LG, Orlandi A, Mauriello A, Santeusanio G, de Angelis C, Lucreziotti R, et al. Aging and atherosclerosis in the rabbit. 1. Distribution, prevalence and morphology of atherosclerotic lesions. Atherosclerosis. 1991;89(1):11–24. pmid:1772469
  52. 52. Papoutsis K, Li L, Near J, Payne S, Jezzard P. A purpose-built neck coil for black-blood DANTE-prepared carotid artery imaging at 7T. Magn Reson Imaging. 2017;40:53–61. pmid:28438710
  53. 53. Harteveld AA, van der Kolk AG, van der Worp HB, Dieleman N, Zwanenburg JJM, Luijten PR, et al. Detecting Intracranial Vessel Wall Lesions With 7T-Magnetic Resonance Imaging: Patients With Posterior Circulation Ischemia Versus Healthy Controls. Stroke. 2017;48(9):2601–4. pmid:28701579
  54. 54. Umutlu L, Theysohn N, Maderwald S, Johst S, Lauenstein TC, Moenninghoff C, et al. 7 Tesla MPRAGE imaging of the intracranial arterial vasculature: nonenhanced versus contrast-enhanced. Acad Radiol. 2013;20(5):628–34. pmid:23473725
  55. 55. Tofts PS, Brix G, Buckley DL, Evelhoch JL, Henderson E, Knopp MV, et al. Estimating kinetic parameters from dynamic contrast-enhanced T(1)-weighted MRI of a diffusable tracer: standardized quantities and symbols. J Magn Reson Imaging. 1999;10(3):223–32. pmid:10508281
  56. 56. Qi H, Sun J, Qiao H, Zhao X, Guo R, Balu N, et al. Simultaneous T1 and T2 mapping of the carotid plaque (SIMPLE) with T2 and inversion recovery prepared 3D radial imaging. Magn Reson Med. 2018. pmid:29802629
  57. 57. Coolen BF, Poot DH, Liem MI, Smits LP, Gao S, Kotek G, et al. Three-dimensional quantitative T1 and T2 mapping of the carotid artery: Sequence design and in vivo feasibility. Magn Reson Med. 2016;75(3):1008–17. pmid:25920036