Validated LC-MS/MS assay for quantification of the newly approved tyrosine kinase inhibitor, dacomitinib, and application to investigating its metabolic stability

  • Ali S. Abdelhameed ,

    Roles Formal analysis, Funding acquisition, Resources, Validation, Writing – review & editing

    asaber@ksu.edu.sa (ASA); mzeidan@ksu.edu.sa (MWA)

    Affiliation Department of Pharmaceutical Chemistry, College of Pharmacy, King Saud University, Riyadh, Kingdom of Saudi Arabia

  • Adnan A. Kadi,

    Roles Formal analysis, Investigation, Resources, Software, Writing – review & editing

    Affiliation Department of Pharmaceutical Chemistry, College of Pharmacy, King Saud University, Riyadh, Kingdom of Saudi Arabia

  • Mohamed W. Attwa ,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Resources, Supervision, Writing – original draft

    asaber@ksu.edu.sa (ASA); mzeidan@ksu.edu.sa (MWA)

    Affiliations Department of Pharmaceutical Chemistry, College of Pharmacy, King Saud University, Riyadh, Kingdom of Saudi Arabia, Students’ University Hospital, Mansoura University, Mansoura, Egypt

  • Haitham AlRabiah

    Roles Data curation, Investigation, Supervision, Visualization, Writing – review & editing

    Affiliation Department of Pharmaceutical Chemistry, College of Pharmacy, King Saud University, Riyadh, Kingdom of Saudi Arabia

Abstract

Dacomitinib (DMB) is a second-generation irreversible tyrosine kinase inhibitor (TKI) that is claimed to overcome the disadvantages of the resistance reported for first-line epidermal growth factor receptor (EGFR) TKIs. Towards the end of 2018, the US Food and Drug Administration approved DMB in the form of VIZIMPRO tablets. In the current study, a validated LC-MS/MS assay was established for DMB quantification in rat liver microsomes (RLMs) with application to the drug metabolic stability assessment. Chromatographic resolution of DMB and lapatinib (internal standard) was achieved using an isocratic mobile phase and a reversed-phase C18 column. The linearity of the established LC-MS/MS assay ranged from 2 to 500 ng/mL with r2 ≥ 0.9999. The limit of detection (LOD) and limit of quantification (LOQ) were 0.35 and 1.1 ng/mL, respectively. The precision and accuracy (both intra-day and inter-day) were 0.84–3.58% and 92.2–100.32%, respectively. The metabolic stability of DMB in the RLM matrix was estimated by calculating two parameters, in vitro t1/2 (0.97 mL/min/kg) and intrinsic clearance (157.5 min). Such values infer that DMB would be excreted very slowly from the human body, which might lead to possible bioaccumulation. To the best of our knowledge, this is the first method for DMB analysis in RLMs with metabolic stability estimation.

Introduction

Lung cancer is the leading cause of death among all cancer types, in particular, non-small cell lung cancer (NSCLC) is considered the most widespread [15], with an incidence of approximately 90%. The epidermal growth factor receptor (EGFR) signaling pathway has gained importance in the last few years as a therapeutic target for NSCLC [6]. Tyrosine kinase inhibitors (TKIs) that control EGFR are very efficient in the treatment of cancers possessing EGFR mutations, with a characteristic therapeutic window. First-line TKIs controlling EGFR (e.g., erlotinib and gefitinib) have good initial responses against these mutations [7, 8]. Unfortunately, acquired resistance in ~60% of patients and toxicities that occur during treatment [9, 10] decrease their therapeutic efficacies [11, 12]. This has led scientists to develop second-generation, irreversible EGFR TKIs (e.g., dacomitinib (DMB) and avitinib) [13, 14].

DMB (Fig 1) overcomes the acquired resistance observed with first-line EGFR TKIs [1315]. It was shown to improve progression-free survival when compared with that of gefitinib in the treatment of NSCLC patients with positive EGFR mutations. This represents a new achievement for the treatment of these patients [16]. On September 27, 2018, the Food and Drug Administration (FDA) approved DMB in the form of VIZIMPRO tablets for the first-line treatment of patients with metastatic NSCLC harboring EGFR exon 19 deletions or exon 21 L858R substitution mutations [17]. In addition, a DMB marketing authorization application was accepted by the European Medicines Agency (EMA) for the same indication [18].

thumbnail
Fig 1. Chemical structures of dacomitinib and lapatinib (IS).

https://doi.org/10.1371/journal.pone.0214598.g001

To the best of our knowledge, a single LC/MS-MS assay was lately published reporting the analysis of DMB in rat plasma [19]. The purpose of the present study was to establish a validated LC-MS/MS assay to quantify DMB in rat liver microsomes (RLMs) as a different biological matrix to the drug and to allow the application of this assay to investigate the DMB metabolic stability by calculating two important parameters (i.e., intrinsic clearance and in vitro half-life (t1/2)). These parameters could then be utilized for in vivo t1/2, hepatic clearance, and bioavailability calculations. Bioavailability is important because it provides information about the metabolism of the investigated compound; if the compound is rapidly metabolized, it will exhibit low bioavailability in vivo [20].

Experimental

Reagents and chemicals

All chemicals and solvents were of analytical grade. Dacomitinib (DMB) and lapatinib (internal standard; LTP; IS) were purchased from Med Chem Express (Princeton, NJ, USA). Rat liver microsomes (RLMs), Acetonitrile (ACN), ammonium formate (NH4COOH), and formic acid (HCOOH) were purchased from Sigma Aldrich (St. Louis, MO, USA). HPLC-grade water (H2O) was obtained from the Milli-Q plus filtration system (Millipore, Billerica, MA, USA).

LC-MS/MS methodology

All LC-MS/MS parameters were optimized to achieve the best chromatographic resolution of DMB and IS with good separation. LTP was chosen as the IS in the DMB analysis because the same extraction procedure can be applied efficiently with a great success for both compounds (DMB and LTP recoveries were 97.91±3.74% and 97.2 ± 1.3%, respectively in the RLM matrix) and the elution time of LTP is comparable to that of DMB. The proposed procedure is rapid with 4 min run time. Both LTP and DMB are TKIs and are not co-administered to patients, so this assay might be applied for pharmacokinetics or therapeutic drug monitoring (TDM) for subjects treated with DMB.

Agilent eclipse plus C18 column (100 mm in length, 2.1 mm in internal diameter and 1.8 μm particle size) was used for chromatographic resolution of analytes. The column temperature was adjusted at 22±1 °C. A triple quadrupole (QqQ) mass spectrometer (Agilent Technologies, CA, USA). with an electrospray ionization source interface (ESI), running in the positive mode, was used for detection. Low purity nitrogen (11 L/min) was utilized as the drying gas in the ESI source and high purity nitrogen (55 psi) was employed as the collision gas. The values of capillary voltage (V) and ESI temperature (T) were set at 4000 V and 350°C, respectively. The instruments and data acquisition were controlled using the Mass Hunter software (Agilent Technologies, CA, USA). DMB was quantified using the multiple reaction monitoring (MRM) scanning mode, for the mass reactions (parent to fragment ions) from 470.1→319 and 470.1→124 for DMB and 581→365 and 581→365 for LTP (Scheme 1). The fragmentor voltage (FV) was set at 140 V with collision energy (CE) of 30 eV for DMB, and FV of 140 V and 145 V with CE of 30 eV and 32 eV for LTP. MRM transitions were used for determination of DMB to eliminate any interference caused by the RLM matrix components and enhance the sensitivity of the proposed LC-MS/MS analytical method (Fig 2).

thumbnail
Fig 2. MRM mass spectrum transitions of DMB (A) and lapatinib (IS) (B).

https://doi.org/10.1371/journal.pone.0214598.g002

Preparation of DMB calibration standards

Both DMB and LTP are soluble in dimethyl sulfoxide (DMSO). DMB was dissolved in DMSO at 2 mg/mL. Afterwards, this stock solution was diluted 10 times with mobile phase to prepare DMB solution 1 (200 μg/mL), which was then diluted 10 times with mobile phase to prepare DMB solution 2 (20 μg/mL). LTP stock solution (100 μg/mL) was prepared in DMSO and then diluted fifty times with the mobile phase to prepare LTP (IS) solution 3 (2 μg/mL). DMB solution 2 (20 μg/mL) was diluted with specific RLM matrix (40 μL in 1 mL of phosphate buffer at pH: 7.4) to generate fifteen calibration standards: 2, 5, 10, 15, 20, 30, 40, 50, 80, 100, 150, 200, 300, 400, and 500 ng/mL. Four standards (2, 15, 150, and 400 ng/mL) were selected as quality controls, namely the lower limit quality control (LLQC), low quality control (LQC), medium quality control (MQC), and high quality control (HQC), respectively.

Extraction of DMB from RLMs matrix

Analytes extractions were performed using ACN protein precipitation, a standard technique for metabolic stability experiments [21]. Specifically, 2 mL ACN was added to each mL of calibration standard followed by centrifugation at 14000 rpm (12 min at 4 °C) to deproteinize by precipitation [22]. All supernatants were then filtered using syringe filters (0.22 μm pore size). Fifty microliters of IS solution 3 was added to 1 mL of each filtered samples and transferred to 1.5 mL vials. Two microliters of each sample was then injected into the LC-MS/MS for analysis. Similarly, blank samples were prepared by using the stated phosphate buffer without RLM matrix to confirm that RLM components did not interfere with the elution time of DMB and IS. A calibration curve was established by plotting the peak area ratio of DMB to IS (y axis) against the nominal values (x axis). A linear regression equation was used to validate the linearity of the established assay. Slope, intercept, and coefficient of determination (r2) values were computed.

Method validation

The established LC-MS/MS assay was validated for sensitivity, assay recovery, linearity, reproducibility, specificity, limit of quantification (LOQ), limit of detection (LOD). and stability according to the US Food and Drug Administration (FDA) guidelines [23]. The validation parameters of the LC-MS/MS assay that was developed to quantify DMB were described in more detail previously [22, 24]. These parameters included the least squares statistical method was utilized to calculate the calibration curve equations (y = ax + b). The linear fit was verified using the r2 value.

Metabolic stability of DMB

The metabolic stability study for DMB was performed by assessing the decrease in DMB concentration after incubation with RLMs. One micromolar DMB was incubated with RLMs (1 mg microsomal protein / 1 mL of phosphate buffer) in triplicate. The medium for metabolic reaction was phosphate buffer (pH 7.4) containing 3.3 mM MgCl2. The mixture was pre-incubated for 10 min in a 37 °C water bath. The metabolic reaction was initiated by adding NADPH (1 mM) and was terminated by adding 2 mL ACN at specific time intervals (0, 2.5, 5, 7.5, 10, 15, 20, 40, and 50 min). ACN was used for stopping the metabolic reaction and protein precipitation method that was used for extraction of DMB from RLMs incubation. All incubates were centrifuged at 14000 rpm (12 min at 4 °C). All supernatants were then filtered using syringe filters (0.22 μm pore size). Fifty μL of IS solution 3 was added to 1 mL of each filtered samples and transferred to 1.5 mL vials. Two microliters of each sample was then injected into the LC-MS/MS for analysis. The metabolic stability curve for DMB was then created.

Results and discussion

HPLC–MS/MS methodology

Optimization of chromatographic parameters, including the mobile phase pH, mobile phase constituents, and C18 column, was performed. The pH of the aqueous portion (10 mM NH4COOH) was adjusted to 4.2 with formic acid. A pH above this value caused peak tailing and an unnecessary increase in retention time. The ratio of the aqueous portion to the organic portion (ACN) of the mobile phase was adjusted to 30%: 70%; higher volume of ACN resulted in overlapping of the chromatographic peaks with bad peak resolution while reducing the ACN volume <70% resulted in longer run times. RRLC columns with different types of stationary phases (e.g., HILIC columns) were tested, but both DMB and IS were not retained on such columns; the unsurpassed results were obtained using C18 column. DMB was quantified using MRM from 470.1→319 and 470.1→124 for DMB and 581→365 and 581→365 for IS. The MRM detection mode of the mass spectrometer was utilized for the quantification of DMB ions to avoid potential interference from the RLM matrix constituents and improve the sensitivity of the established LC-MS/MS assay (Fig 2).

The chromatographic separation of DMB and IS was achieved in 4 min. DMB and IS chromatographic peaks eluted at 1.5 min. and 2.6 min, respectively. The analytes peaks were well resolved, with no carryover to the blank samples (i.e., RLM matrix samples or RLM plus IS samples). Fig 3 shows the overlaid MRM chromatograms of 15 DMB calibration standards, showing concentration-dependency and consistency in their profiles.

thumbnail
Fig 3. Overlaid MRM chromatograms of fifteen calibration standards of DMB (2–500 ng/mL) and IS (100 ng/mL).

https://doi.org/10.1371/journal.pone.0214598.g003

Validation of the developed LC-MS/MS assay

Specificity.

Fig 4 reveals good separation of the DMB and IS peaks and the absence of peaks with the blank RLM matrix at the corresponding retention times, which supports the specificity of the developed methodology. No carryover effect of DMB and IS was observed in the MS chromatograms.

thumbnail
Fig 4. MRM chromatograms of blank (A), blank + IS (B), and DMB 15 ng/mL LQC (C).

The blank RLM matrix revealed no matrix interference.

https://doi.org/10.1371/journal.pone.0214598.g004

Sensitivity and linearity.

The linear range and correlation coefficient (r2) for the proposed methodology were 2–500 ng/mL and ≥ 0.9999 in the RLM matrix, respectively. The regression equation of the DMB calibration curve was y = 0.4919x + 0.6092. The LOD and LOQ were equal to 0.35 and 1.1 ng/mL, respectively. The LLQC peak showed a very high signal-to-noise ratio (S/N) and good peak shape that validated the sensitivity of the developed LC-MS/MS assay (Fig 5).

Extraction recovery.

The RSD values of six repetitions for each concentration level in the calibration curve were less than 1.96% in the RLM matrix (Table 1). Back calculations of the fifteen samples of DMB in the RLM matrix (Calibration standards and QC samples) confirmed the performance of the developed methodology. The recovery of DMB detected in the spiked RLM matrix samples was 97.91 ± 3.74% with a RSD of less than 1.96%.

thumbnail
Table 1. Data of back-calculated DMB concentrations for the calibration standards from the RLM matrix.

https://doi.org/10.1371/journal.pone.0214598.t001

Precision and accuracy.

Considering both intra-day and inter-day values, the precision and accuracy ranged from 0.84 to 3.58% and 92.20 to 100.32%, respectively (Table 2). The mean percentage of DMB recovery was 97.91 ± 3.74% in the RLM matrix (Table 1). The intra- and inter-day accuracy and precision values are acceptable according to International Conference for Harmonization (ICH) guidelines [25, 26].

thumbnail
Table 2. Precision and accuracy (intra-day and inter-day) of the developed assays.

https://doi.org/10.1371/journal.pone.0214598.t002

Matrix effects.

The absence of an RLM matrix effect was confirmed by analyzing six different batches of RLM matrixes form six different rats, in which these batches were extracted and spiked with DMB (LLOQ (15 ng/mL), LQC (15 ng/mL), MQC (150 ng/mL) and HQC (400 ng/mL)) and IS. The abovementioned batches were named set 1. Set 2 was prepared in a similar way, except utilizing the mobile phase instead of the RLM matrix. Thus, the matrix effect was calculated from the following equation:

The studied RLM matrix that contained DMB had a matrix effect of 94.46 ± 2.94%. Accordingly, these results reveal that the influence of the RLM matrix on DMB and IS (IS) ionization was low (Table 3).

Stability

The stability of DMB in RLMs matrix was evaluated under all conditions that might have been encountered before the analysis. DMB displayed good stability in RLMs matrix samples after storage at −20 °C for 28 days as stability values were ranged from 98.89 to 98.9%. Detailed stability data for DMB is summarized in Table 4. There was no observed degradation of analytes under the tested conditions indicating that DMB displayed good stability in all analyses.

thumbnail
Table 4. Stability of dacomitinib (DMB) under different storage conditions.

https://doi.org/10.1371/journal.pone.0214598.t004

Metabolic stability

The DMB concentration in the RLM matrix was calculated by the displacement of the peak area ratios in the regression equation of the calibration curve. A metabolic stability curve was drawn by plotting the natural log (Ln) of the remaining percent of DMB on the y-axis against incubation time on the x-axis (Fig 6). The linear portion of the plotted curve was used to calculate in vitro t1/2 [27]. The regression equation for this linear region was y = -0.0044x + 4.6048 with r2 = 0.9989 (Table 5).

thumbnail
Fig 6. Metabolic stability curve of DMB treated with RLMs.

https://doi.org/10.1371/journal.pone.0214598.g006

thumbnail
Table 5. Metabolic stability parameters for DMB incubation with RLMs.

https://doi.org/10.1371/journal.pone.0214598.t005

By using the following equation, where the slope was 0.0044:

The intrinsic clearance (CLint) of DMB was calculated, according to the in vitro t1/2 method [20], using the following equation:

From these results, the metabolic stability of DMB was characterized by a very low CLint (0.97 mL/min/kg) and a very long in vitro t1/2 (157.5 min), which resulted in a very slow clearance of DMB from the blood by the liver. This probably resulted in a very high in vivo bioavailability that corroborated the high oral bioavailability previously reported and also indicated that DMB would possibly be bioaccumulated after multiple doses [28].

Conclusions

A validated LC-MS/MS methodology was developed for quantifying the newly approved drug, DMB. The established assay is highly sensitive (LOD = 0.35 ng/mL), eco-friendly (small volume of ACN), fast (run time = 4 min.), accurate (92.2 to 100.32%), and has high recovery (97.91±3.74%). The LC-MS/MS assay was applied for DMB metabolic stability assessment with an RLM matrix, and the two parameters, in vitro t1/2 (157.5 min) and CLint (0.97 mL/min/kg), were calculated. Accordingly, with such high t1/2 and low CLint values, DMB can be further investigated for its drug plasma concentration and effect on kidney function due to possible drug bioaccumulation as it might have a very low extraction ratio with a very slow excretion.

Acknowledgments

The authors would like to extend their sincere appreciation to the Deanship of Scientific Research at the King Saud University for funding this work through the Research Group Project No. RG-1435-025.

References

  1. 1. Siegel RL, Miller KD, Jemal A. Cancer statistics, 2016. CA: A Cancer Journal for Clinicians. 2016;66(1):7–30.
  2. 2. Haghgoo SM, Allameh A, Mortaz E, Garssen J, Folkerts G, Barnes PJ, et al. Pharmacogenomics and targeted therapy of cancer: Focusing on non-small cell lung cancer. European Journal of Pharmacology. 2015;754:82–91. pmid:25725115
  3. 3. Abubakar I, Tillmann T, Banerjee A. Global, regional, and national age-sex specific all-cause and cause-specific mortality for 240 causes of death, 1990–2013: a systematic analysis for the Global Burden of Disease Study 2013. Lancet. 2015;385(9963):117–71. pmid:25530442
  4. 4. Ettinger DS, Akerley W, Bepler G, Blum MG, Chang A, Cheney RT, et al. Non–small cell lung cancer. Journal of the National Comprehensive Cancer Network. 2010;8(7):740–801. pmid:20679538
  5. 5. Larsen JE, Cascone T, Gerber DE, Heymach JV, Minna JD. Targeted therapies for lung cancer: clinical experience and novel agents. Cancer Journal. 2011;17(6):512–27.
  6. 6. Costa DB, Kobayashi SS. Whacking a mole-cule: clinical activity and mechanisms of resistance to third generation EGFR inhibitors in EGFR mutated lung cancers with EGFR-T790M. Translational Lung Cancer Research. 2015;4(6):809–15. pmid:26798593
  7. 7. Gridelli C, Rossi A, Carbone DP, Guarize J, Karachaliou N, Mok T, et al. Non-small-cell lung cancer. Nature Reviews Disease Primers. 2015;1:15009. pmid:27188576
  8. 8. Peters S, Zimmermann S, Adjei AA. Oral epidermal growth factor receptor tyrosine kinase inhibitors for the treatment of non-small cell lung cancer: comparative pharmacokinetics and drug–drug interactions. Cancer Treatment Reviews. 2014;40(8):917–26. pmid:25027951
  9. 9. Metro G, Crinò L. Advances on EGFR mutation for lung cancer. Translational Lung Cancer Research. 2012;1(1):5–13. pmid:25806150
  10. 10. Denis MG, Vallée A, Théoleyre S. EGFR T790M resistance mutation in non small-cell lung carcinoma. Clinica Chimica Acta. 2015;444:81–5.
  11. 11. Jorge S, Kobayashi S, Costa D. Epidermal growth factor receptor (EGFR) mutations in lung cancer: preclinical and clinical data. Brazilian Journal of Medical and Biological Research. 2014;47(11):929–39. pmid:25296354
  12. 12. Finlay MRV, Anderton M, Ashton S, Ballard P, Bethel PA, Box MR, et al. Discovery of a potent and selective EGFR inhibitor (AZD9291) of both sensitizing and T790M resistance mutations that spares the wild type form of the receptor. Journal of Medicinal Chemistry. 2014;57(20):8249–67. pmid:25271963
  13. 13. Tan C-S, Gilligan D, Pacey S. Treatment approaches for EGFR-inhibitor-resistant patients with non-small-cell lung cancer. Lancet oncology. 2015;16(9):e447–59. pmid:26370354
  14. 14. Liao B-C, Lin C-C, Yang JC-H. Second and third-generation epidermal growth factor receptor tyrosine kinase inhibitors in advanced nonsmall cell lung cancer. Current Opinion in Oncology. 2015;27(2):94–101. pmid:25611025
  15. 15. Jänne PA, Yang JC-H, Kim D-W, Planchard D, Ohe Y, Ramalingam SS, et al. AZD9291 in EGFR inhibitor–resistant non–small-cell lung cancer. New England Journal of Medicine. 2015;372(18):1689–99. pmid:25923549
  16. 16. Wu YL, Cheng Y, Zhou X, Lee KH, Nakagawa K, Niho S, et al. Dacomitinib versus gefitinib as first-line treatment for patients with EGFR-mutation-positive non-small-cell lung cancer (ARCHER 1050): a randomised, open-label, phase 3 trial. Lancet Oncology. 2017;18(11):1454–66. pmid:28958502
  17. 17. U.S. Food and Drug Administration (FDA). Vizimpro® (dacomitinib) tablets. In: Center for Drug Evaluation and Research, editor. MD, USA 27th September 2018.
  18. 18. Barrón F, Zatarain-Barrón ZL, Cardona AF, Arrieta O. Extending the curve: survival of EGFR-mutated lung cancer patients in the 21st century. Journal of Thoracic Disease. 2018;10(3):1265–8. pmid:29708166
  19. 19. Qiu X, Lin Q, Ning Z, Qian X, Li P, Ye L, et al. Quantitative bioanalytical assay for the human epidermal growth factor receptor (HER) inhibitor dacomitinib in rat plasma by UPLC-MS/MS. Journal of Pharmaceutical and Biomedical Analysis. 2019;166:66–70. pmid:30612075
  20. 20. Baranczewski P, Stanczak A, Sundberg K, Svensson R, Wallin A, Jansson J, et al. Introduction to in vitro estimation of metabolic stability and drug interactions of new chemical entities in drug discovery and development. Pharmacological Reports. 2006;58(4):453–72. pmid:16963792
  21. 21. Kadi A.A., Abdelhameed A.S., Darwish H.W., Attwa M.W. and Al-Shakliah N.S. A highly efficient and sensitive LC-MS/MS method for the determination of afatinib in human plasma: application to a metabolic stability study. Biomedical Chromatography. 2016; 30(8), 1248–1255. pmid:26683307
  22. 22. Alrabiah H, Kadi AA, Attwa MW, Abdelhameed AS. A simple liquid chromatography-tandem mass spectrometry method to accurately determine the novel third-generation EGFR-TKI naquotinib with its applicability to metabolic stability assessment. RSC Advances. 2019;9(9):4862–69.
  23. 23. U.S. Food and Drug Administration (FDA). Guidance for industry; Analytical procedures and methods validation; Chemistry, manufacturing, and controls documentation. US Department of Health and Human Services. Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER) 2000.
  24. 24. Attwa MW, Kadi AA, Darwish HW, Abdelhameed AS. Investigation of the metabolic stability of olmutinib by validated LC-MS/MS: quantification in human plasma. RSC Advances. 2018;8(70):40387–94.
  25. 25. ICH F. Guidance for Industry: Q2B Validation of Analytical Procedures: Methodology. U.S. Department of Health and Human Services, Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER) 1996.
  26. 26. Abdelhameed AS, Attwa MW, Kadi AA. An LC–MS/MS method for rapid and sensitive high-throughput simultaneous determination of various protein kinase inhibitors in human plasma. Biomedical Chromatography. 2017;31(2):e3793.
  27. 27. Caldwell G, Yan Z. Optimization in drug discovery: in vitro methods: Springer Science & Business Media; 2014.
  28. 28. Peters S, Zimmermann S, Adjei AA. Oral epidermal growth factor receptor tyrosine kinase inhibitors for the treatment of non-small cell lung cancer: comparative pharmacokinetics and drug-drug interactions. Cancer Treatment Reviews. 2014;40(8):917–26. pmid:25027951