Detection of intravenous infiltration using impedance parameters in patients in a long-term care hospital

  • Ihn Sook Jeong ,

    Contributed equally to this work with: Ihn Sook Jeong, Eun-Joo Lee

    Roles Conceptualization, Funding acquisition, Project administration, Writing – original draft, Writing – review & editing

    Affiliation College of Nursing, Pusan National University, Yangsan, Republic of Korea

  • Eun-Joo Lee ,

    Contributed equally to this work with: Ihn Sook Jeong, Eun-Joo Lee

    Roles Data curation, Investigation, Methodology, Writing – review & editing

    Affiliation Dept, of Nursing, Dong-Eui University, Busan, South Korea

  • Jae Hyung Kim ,

    Roles Formal analysis, Writing – original draft

    jhkim@inje.ac.kr

    ‡ These authors also contributed equally to this work.

    Affiliation Research Institute of Nursing Science, Pusan National University, Yangsan, Republic of Korea

  • Gun Ho Kim ,

    Roles Data curation, Methodology

    ‡ These authors also contributed equally to this work.

    Affiliation Dept. of Medical Science, School of Medicine, Pusan National University, Yangsan, Republic of Korea

  • Young Jun Hwang ,

    Roles Data curation, Formal analysis

    ‡ These authors also contributed equally to this work.

    Affiliation Dept. of Medical Science, School of Medicine, Pusan National University, Yangsan, Republic of Korea

  • Gye Rok Jeon

    Roles Conceptualization, Formal analysis, Investigation, Methodology, Supervision, Writing – original draft

    ‡ These authors also contributed equally to this work.

    Affiliation Dept. of Biomedical Engineering, School of Medicine, Pusan National University, Yangsan, Republic of Korea

Abstract

This study was aimed to evaluate the changes of impedance parameters of patients who were admitted to a long-term care hospital by measuring bioelectrical impedance. The subjects were 18 patients who had infusion therapy through peripheral intravenous (IV) catheters and had at least an infiltration. The impedance parameters were measured with a multi-channel impedance measuring instrument (Vector Impedance Meter) twice; at starting IV infusion after catheter insertion and infiltration detected. As results, the resistance (R) after infiltration significantly decreased compared to the initial resistance. At 50 kHz, the resistances were 498.2±79.3 [Ω] before infiltration and 369.4±85.6 [Ω] after infiltration. The magnitude of the reactance (XC) decreased after infiltration. At 50 kHz, the measured reactance was -31.1±8.3 [Ω] before infiltration and -24.5±5.9 [Ω] after infiltration. The data points plotted in the R-XC graph shifted from the first quadrant before infiltration to third quadrant after infiltration. Our findings suggest that bioelectrical impedance is an effective method for detection of infiltration in a noninvasive and quantitative manner.

Introduction

Infiltration is one of the most common health complications in infusion therapy involving a peripheral intravenous (PIV) catheter [1]. Patients with infiltration often experience pain and skin damage, and can even a persistent disability of the arm [2]. Thus, infiltration can significantly impact the patient’s wellbeing and must be carefully monitored at an early stage [3]. Infiltration is known to be difficult to detect particularly during its initial stages. To date, infiltration has relied primarily on clinical and visual methods to examine skin and tissues surrounding the IV catheter site for factors such as tissue pressure, color, edema, swelling and temperature. However, these methods are disadvantageous because infiltration can only be confirmed well after infiltration has progressed considerably or tissue damage has occurred in subcutaneous tissue. As recognition of early signs and symptoms of infiltration can effectively limit the amount of fluid that leaks from the vein into the subcutaneous tissue [4], several infiltration detection systems have been introduced.

Optical methods near infrared (NIR) light have been developed to facilitate noninvasive monitoring of IV sites [5,6]. High-resolution variable frequency ultrasound imaging is increasingly being used in the noninvasive evaluation of various cutaneous diseases [7]. Ultrasound could detect small volumes of fluids such as cosmetic fillers and subcutaneous injections subcutaneous injections and thus provides a potential reference standard for the future evaluation of IV monitoring device [8]

The infiltration detection system using bioelectrical impedance analysis (BIA) can be another option [9]. BIA has been employed to diagnose diseases [10] and to assess the hydration status, body composition, muscle–fat ratio, obesity degree, lean balance, edema, and nutritional status of the patients [11,12]. As increase in local conductive IV fluids is known to reduce the measured impedance [13], fluid leakage due to infiltration can increase the skin firmness which leads to change in bioelectrical impedance around the IV catheter site [9]. Therefore, we have investigated changes of impedance parameters such as impedance, resistance, reactance, and capacitance between pre and post infiltration, as the IV solution accumulates in the subcutaneous tissues and cell membrane functions in rabbits [14] and pigs [15]. When IV solution was properly infused into the vein on the back of rabbit, the impedance parameters (impedance, resistance, reactance, and capacitance) measured at five injection sites showed almost similar behavior with very slight standard deviations. On the other hand, the impedance parameters were significantly different before and after infiltration [14]. In addition, when infiltration was intentionally induced in the vein of a pig's posterior ear, impedance parameters (resistance, reactance, and capacitance) showed significant differences before and after infiltration [15].

Based on the preclinical studies, this study was aimed to evaluate the impedance parameters of patients at a long-term care hospital by measuring bioelectrical impedance. This is a preliminary study aimed to develop a point-of-care device and an automated impedance device that can detect infiltration in the nursing field.

Materials and methods

Study subjects

The subjects of this prospective observational study were 18. who met the following inclusion criteria; patients who had infusion therapy through PIV catheters from July 18th to September 7th, 2017, had at least an infiltration, and gave written informed consent by themselves or their legally authorized representatives (LARs). There were 3 males and 15 females, and their mean age was 79.5 years, ranged 57–87 years. Mean diameter of infiltration was 2.3cm, ranged 0.4-7cm. Table 1 shows the more characteristics of subjects.

Study instrument

Multi-channel impedance measuring instrument (Vector Impedance Meter) developed by Kim et al. for detection of infiltration is as follows [16]. A weak alternating current (800 μA) and multiple frequencies (10 kHz~10 MHz) are applied to the human body to measure impedance parameters (resistance: R, reactance: XC, phase angle: θ) in the live body. Radio frequency (RF) weak current is applied to the electrode located in the periphery of the injection site and the impedance and phase change of the arbitrary points among the multiple channels or specified channels.

Study procedure

The study protocol was approved by the IRB committee of Pusan National University Yangsan Hospital (IRB No. 03-2016-017). When the patients needed IV infusion therapy, the staff nurses inserted 22gauge IV catheter, and explained the purpose and method to the patients or LARs. When they agreed to participate in the study, we obtained written informed consents. And then, we attached electrodes near the IV catheter site at both sides to apply the alternating current and collecting the voltage, and took initial measurement one impedance parameter (Z, R, XC, Cm) as a function of frequency during infusing an IV solution at the rate of around 60 drops per min. While infusing the IV solution into the subject’s vein, an AC was applied to the electrodes with varying frequencies. To reduce the measurement error, we marked the electrode sites after taking initial measurement and removing the electrodes.

The staff nurses examined the tissue surrounding IV catheter site every shift to identify the signs and symptoms of infiltration such as swelling, redness, and/or discomfort [17] according to their practice manual. When the infiltration was suspected, they quickly reported to a study assistant staying in the hospital to make additional measurement before removing the IV catheter at the same electrode site as at the initial measurement. The impedance parameters measured were kept in the multi-channel impedance measuring instrument automatically and transferred to a PC in text file for data storage and analysis.

Fig 1 shows the electrode placement for impedance measurements in the subject’s ankle.

thumbnail
Fig 1. Electrode placement for impedance measurement on the patient's ankle.

https://doi.org/10.1371/journal.pone.0213585.g001

Data analysis

Data processing was performed using Excel 2016 (Microsoft Corporation, Redmond, Wash., USA) and the data were graphically analyzed using OriginPro 9.0.0 (OriginLab Corporation, Northampton, Massachusetts, USA).

Results

Resistance (R) as a function of frequency (f)

Fig 2 shows the mean and standard deviation of the resistance (R) as a function of frequency before and after infiltration. Resistance decreased almost inversely with frequency, with resistances showing a significant difference before and after infiltration. While the difference in resistance before and after infiltration was observed clearly, the standard deviation was relatively large. For example, at 20 kHz, the resistances were 643.6±124.1 [Ω] before infiltration and 481.2±114.3 [Ω] after infiltration which corresponded to a 25.2% decrease in resistance. At 50 kHz, the resistances were 498.2±79.3 [Ω] before infiltration and 369.4±85.6 [Ω] after infiltration which corresponded to a 25.8% decrease in resistance.

thumbnail
Fig 2. Resistance as a function of frequency during infusion IV solution into the vein.

Circle (O) represents the resistance measured before infiltration (BI) and cross (×) represents the resistance measured after infiltration (AI).

https://doi.org/10.1371/journal.pone.0213585.g002

Reactance (XC) as a function of frequency (f)

Fig 3 shows the reactance (XC) of the cell membrane as a function of frequency before and after infiltration. The absolute magnitude of the reactance decreased in the applied frequency range. In addition, the absolute magnitude of the reactance after infiltration becomes smaller than before infiltration. For example, at 20 kHz, the measured reactance was -65.5±18.5 [Ω] before infiltration and -52.4±13.7 [Ω] after infiltration, corresponding to a 20.0% decrease in reactance. At 50 kHz, the measured reactance was -31.1±8.3 [Ω] before infiltration and -24.5±5.9 [Ω] after infiltration, corresponding to a 21.2% decrease in reactance.

thumbnail
Fig 3. Reactance as a function of frequency before and after infiltration.

https://doi.org/10.1371/journal.pone.0213585.g003

Resistance (R) versus reactance (XC)

Fig 4 shows the relationship between absolute magnitude of resistance (R) and reactance (XC) before and after infiltration at 50 kHz. Compared to the values measured before infiltration, both R and Xc decreased after infiltration. The positions of R and Xc mainly distributed in the first quadrant before infiltration have shifted to the third quadrant.

thumbnail
Fig 4. Reactance versus resistance before and after infiltration.

https://doi.org/10.1371/journal.pone.0213585.g004

Capacitance (Cm) of cell membrane as a function of frequency (f)

Fig 5 shows the capacitance of the cell membrane as a function of frequency. The capacitance of the cell membrane decreases almost inversely as frequency increases. The capacitance of the cell membrane after infiltration increases in the 20–500 kHz frequency range. At 50 kHz, the capacitance of the cell membrane after infiltration increased by 28.0% compared to the capacitance before infiltration.

thumbnail
Fig 5. Capacitance of cell membrane before and after infiltration.

https://doi.org/10.1371/journal.pone.0213585.g005

Discussion

Impedance indicators have been proposed to detect detection of infiltration of IV solutions that accumulate in the extracellular fluid (ECF) and affect the cell membrane function. In previous experiments that studied infiltration in the rabbit's ears and the human’s forearms, impedance parameters showed a significant difference within 2–3 minutes after infiltration in rabbits, whereas impedance parameters gradually changed over 15 minutes after infiltration in humans [18]. In this human study, impedance parameters before and after infiltration in hospitalized subjects were analyzed using BIA, and infiltration phenomena were quantitatively explained by using infiltration mechanism and equivalent circuit of human cells. Impedance parameters (R, XC, Cm) showed significant differences in the impedance parameters before and after infiltration.

After infiltration, a reduction in resistance was significantly observed at all frequencies. This indicates that the IV solution flowing from the vein due to infiltration accumulates in the ECF (including interstitial fluid) [19,20]. The large standard deviations shown in Fig 2 are thought to be caused not only by the locations of the forearms, upper arms, heels, hands, legs, feet and thighs (listed in decreasing order of frequency of infiltration) but also the size of infiltration. When AC with a frequency of 20 kHz was used, the resistance was relatively high since AC of frequency of 20 kHz does not pass through the cell membrane, and decreased with increasing frequency.

For reactance, the absolute magnitude of the reactance decreased in the applied frequency range. This is caused by the blood components released from the venous blood vessels in the infiltration, which trigger the cell membranes to aggregate in a linear/ parallel manner, so that the capacitance of the cell membrane increases, and thus the absolute magnitude of the reactance decreases, approaching zero [21]. The IV solution and blood components were absorbed into the cell membrane due to infiltration, critically reducing the cell membrane’s ability to slow down the electric current. Since change in reactance reflects the function of the cell membrane, it is possible to detect a variation in skin or subcutaneous tissue (or necrosis) due to side effects of drug-induced infiltration. Important information about tissue necrosis that occur in neonates with thin blood vessels or in patients experiencing frequent infiltration can be obtained using this phenomenon.

For resistance versus reactance, the results of this measurement are in good agreement with Nescolarde’s study that measured bioelectrical impedance for injuries caused in soccer players' lower extremity muscles [22]. As one recovers from damage caused by infiltration, resistance and reactance are expected to move toward the upper right direction. For instance, localized bio-impedance was used to assess injuries to lower body muscles by measuring soft tissue hydration and cell membrane integrity. Impedance measurements of edema in the calf of the athlete showed that the position of R-Xc shifted to the lower left according to the severity of injury and to the upper right during the recovery [22]. Compared to non-injury values, R, and Xc decreased with increasing severity of injury: grade III (23.1% and 45.1%), grade II (20.6% and 31.6%) and grade I (11.9% and 23.5%). These results indicate that a decrease in R reflects local body fluid accumulation, and a reduction in Xc indicates significant side effects due to compromised cell membrane integrity [23].

The capacitance of the cell membrane after infiltration increased, which may be due to the absorption of IV solution and blood components (red blood cells, white blood cells, platelets, etc.) from the vein into the surrounding cell membranes. In our previous experiments [14], the capacitance of the cell membrane was measured as a function of time up to 25minutes post initiation of infiltration into the veins of the rabbit's ear. The capacitance of cell membranes in rabbit’s ear increased markedly during infiltration and remained almost constant after infiltration. Capacitance of the cell membrane decreases almost inversely as frequency increases. Body capacitance represents the absolute amount of energy storage in the body due to intact cellular membranes. High capacitance indicates that the human body stores energy efficiently, while low capacitance indicates that the cell is inefficient. When a low-frequency current is applied, the cell membrane mainly acts as a resistor. Conversely, when a high-frequency current is applied, the cell membrane behaves primarily as a capacitor [21].

To our knowledge, this study is one of the few studies to identify the relationship between impedance parameters and infiltration. And, it is meaningful to suggest bioelectrical impedance as an effective method for detection of infiltration. That is, we can assume that infiltration may occurred when the resistance (R) and the absolute magnitude of the reactance (XC) decrease, and body capacitance (Cm) increase compared to baseline measure.

However, careful interpretation of the results from the study is required as this study has the following several limitations. First, there is limitation in generalizability because of small sample size with old ages in a single long-term care hospital. Second, the staff nurses inspected the occurrence of infiltration in a regular manner, and measured the parameters at infiltration as quickly as possible. However, there might have a gap between real and detected time of infiltration occurrence, and the parameters at infiltration may not represent them right after the infiltration occurrence. Third, we did not measure the subject’s body composition such as weight and body mass index. However, it may not influence the results because we measured the parameters for the same subjects repeatedly and their body composition is hardly changed for short time between before and after measurement (mean IV day was 1.47 days). Fourth, there were changes in the type of infusate from Combiflex inj. 1000ml (JW Pharmaceutical, Korea) to normal saline or Hartmann’s dextrose at initial and additional measurement for two subjects. It can make a difference for impedance parameters if a conductive (e.g., saline, balanced crystalloid solution) or a non-conductive (e.g., aqua, glucose) solution. As Combiflex inj. 1000ml contains amino acids, electrolytes, emulsion and glucose, changes in the type of infusate may not have a big influence on the results. Fifth, the distance between 2 electrodes is not controlled and it may affect the impedance measurement because the magnitude of the impedance values is inversely proportional to the distance between the electrodes. However, we marked the electrode site at initial measurement, and use the same site for additional measurement. As this study compared the changes in impedance parameters before and after infiltration, the fail to control the distance between electrodes may not influence on the results. Lastly, as we adopted observational study not clinical trial due to ethical issue. we could not control the IV catheter sites and the size of infiltration, which may influence the large standard deviation of the resistance with frequency.

Conclusions

The resistance (R) after infiltration significantly decreased compared to the initial resistance, the absolute magnitude of the reactance (XC) decreased after infiltration, and the data points plotted in the R-XC graph were distributed on the upper right before infiltration but on the lower left after infiltration. And, body capacitance of the cell membrane decreased almost inversely as frequency increased after infiltration. Our findings suggest that bioelectrical impedance is an effective method for detection of infiltration in a noninvasive and quantitative manner. Repeated studies with different age groups of subjects are recommended to increase the generalizability of the results. And we also recommend further studies to identify the smallest infiltration size that impedance parameters can detect the infiltration, which needs to confirmed whether impedance parameter can use a method for early detection of infiltration.

Acknowledgments

This research was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT and Future Planning (2015R1A2A2A04003415).

References

  1. 1. Rickard CM, Webster J, Wallis MC, Marsh N, McGrail MR, French V, et al. Routine versus clinically indicated replacement of peripheral intravenous catheters: a randomised controlled equivalence trial. Lancet. 2012;380(9847):1066–1074. pmid:22998716
  2. 2. Hadaway L. Infiltration and extravasation. Am J Nurs. 2007;107(8):64–72. pmid:17667395
  3. 3. Oya M, Takahashi T, Tanabe H, Oe M, Murayama R, Yabunaka K, et al. Low-temperature infiltration identified using infrared thermography in patients with subcutaneous edema revealed ultrasonographically: a case report. Drug Dis Ther. 2016;10(2):117–122. https://doi.org/10.5582/ddt.2016.01033.
  4. 4. Hadaway L. Protect patients from I.V. infiltration. Amer Nurs To. 2010;5(2): https://www.americannursetoday.com/protect-patients-from-i-v-infiltration-3/
  5. 5. Wintec LW. Optical detection of intravenous infiltration, US 7,826,890 B1, USA. http://www.google.com/patents/US7826890. 2005.
  6. 6. An optical device for detecting intravenous infiltration. http://www.ivteam.com/optical-iv.pdf. 2006.
  7. 7. Mandava A, Ravuri P.R. Konathan R, High-resolution ultrasound imaging of cutaneous lesions, Indian J Radiol Imaging. 2013;23(3):269–277.
  8. 8. Gottlieb M, Sundaram T, Holladay D, Nakitende D. Ultrasound-guided peripheral intravenous line placement: a narrative review of evidence-based best practices. West J Emerg Med. 2017;18(6):1047–1054. pmid:29085536
  9. 9. Jambulingam JA, McCrory R, West L, Inan OT. Non-invasive, multi-modal sensing of skin stretch and bioimpedance for detecting infiltration during intravenous therapy. Conf Proc IEEE Eng Med Biol Soc. 2016:4755–4758. pmid:28269333
  10. 10. Khalil SF, Mohktar MS, Ibrahim F. The theory and fundamentals of bioimpedance analysis in clinical status monitoring and diagnosis of diseases. Sensors. 2014;14(6):10895–10928. pmid:24949644
  11. 11. Berlit S, Brade J, Tuschy B, Foldi E, Walz-Eschenlohr U, Leweling H, et al. Whole-body versus segmental bioelectrical impedance analysis in patients with edema of the upper limb after breast cancer treatment. Anticancer Res. 2013;33(8):3403–3406. pmid:23898111
  12. 12. Buffa R, Mereu E, Comandini O, Ibanez ME, Marini E. Bioelectrical impedance vector analysis (BIVA) for the assessment of two-compartment body composition. Eur J Clin Nutr. 2014;68(11):1234–1240. pmid:25139557
  13. 13. Bicen AO, West LL, Cesar L, Inan OT. Toward non-invasive and automatic intravenous infiltration detection: evaluation of bioimpedance and skin strain in a pig model. IEEE J Transl Eng Health Med. 2018;6:4100207. pmid:29692956
  14. 14. Kim JH, Shin BJ, Lee MS, Kim YJ, Jeong IS, Jeon GR. Detection of intravenous infiltration in the posterior ear of the rabbit using bioelectrical impedance: pilot study. IJEAS. 2017;4(5):122–128.
  15. 15. Kim JH, Hwang YJ, Kim GH, Jeong IS, Jeon GR. Early Detection of infiltration induced in the veins of pig's ear and human's forearm by using bioimpedance: pilot study. J Korea Multimedia Soc. 2018;21(1):34–44.
  16. 16. Kim BC, Kim CM, Lee CH. Multi-channel impedance measuring method and multi-channel impedance measuring instrument. WO 2014/035040 A1, Patient PCT/KR2013/005779. 2013.
  17. 17. Wengström Y, Margulies A. European Oncology Nursing Society extravasation guidelines. Eur J Oncol Nurs. 2008;12(4):357–361. pmid:18765210
  18. 18. Kim JH, Hwang YJ, Kim GH, Shin BJ, Kim YJ, Lee EJ, Jeon GR. Comparison of infiltration induced in veins of rabbit’s ear and human’s forearm by using bioelectrical impedance: Pilot study. J Sensor Sci & Tech. 2017;26(5):1–8.
  19. 19. Kim JH, Kim SS, Kim SH, Baik SW, Jeon GR. Bioelectrical impedance analysis at popliteal regions of human body using BIMS. J Sen Sci Tech. 2016;25(1):1–7.
  20. 20. Hernández-Balaguera E, López-Dolado E, Polo JL. Obtaining electrical equivalent circuits of biological tissues using the current interruption method, circuit theory and fractional calculus. Roy So Chem. 2016; 6: 22312–22319.
  21. 21. Urone PP. College Physics (2nd Edition), Brooks/Cole, USA, pp. 484/584, 2001.
  22. 22. Nescolarde L, Yanguas J, Lukaski H, Alomar X, Rosell-Ferrer J, Rodas G. Localized bioimpedance to assess muscle injury. Physiol Meas. 2013;34(2):237–245. pmid:23354019
  23. 23. Nescolarde L, Yanguas J, Lukaski H, Alomar X, Rosell-Ferrer J, Rodas G. Effects of muscle injury severity on localized bioimpedance measurements. Physiol Meas. 2015;36(1):27–42. pmid:25500910