Strategies for intrapartum foetal surveillance in low- and middle-income countries: A systematic review

  • Natasha Housseine ,

    Roles Conceptualization, Data curation, Formal analysis, Methodology, Supervision, Writing – original draft, Writing – review & editing

    natasha.housseine@outlook.com, N.Housseine@umcutrecht.nl

    ‡ These authors are joint first authors on this work.

    Affiliations Department of Obstetrics and Gynaecology, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands, Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands, Department of Obstetrics and Gynaecology, Mnazi Mmoja Hospital, Zanzibar, Tanzania

  • Marieke C. Punt ,

    Roles Conceptualization, Data curation, Formal analysis, Methodology, Writing – original draft, Writing – review & editing

    ‡ These authors are joint first authors on this work.

    Affiliations Department of Obstetrics and Gynaecology, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands, Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands

  • Joyce L. Browne,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliation Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands

  • Tarek Meguid,

    Roles Writing – review & editing

    Affiliations Department of Obstetrics and Gynaecology, Mnazi Mmoja Hospital, Zanzibar, Tanzania, School of Health and Medical Science, State University of Zanzibar (SUZA), Zanzibar, Tanzania

  • Kerstin Klipstein-Grobusch,

    Roles Conceptualization, Writing – review & editing

    Affiliations Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands, Division of Epidemiology & Biostatistics, School of Public Health, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, South Africa

  • Barbara E. Kwast,

    Roles Writing – review & editing

    Affiliation International Consultant Maternal Health and Safe Motherhood, Leusden, The Netherlands

  • Arie Franx,

    Roles Writing – review & editing

    Affiliation Department of Obstetrics and Gynaecology, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands

  • Diederick E. Grobbee,

    Roles Conceptualization, Writing – review & editing

    Affiliation Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands

  • Marcus J. Rijken

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliations Department of Obstetrics and Gynaecology, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands, Julius Global Health, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht University, Utrecht, The Netherlands

Abstract

Background

The majority of the five million perinatal deaths worldwide take place in low-resource settings. In contrast to high-resource settings, almost 50% of stillbirths occur intrapartum. The aim of this study was to synthesise available evidence of strategies for foetal surveillance in low-resource settings and associated neonatal and maternal outcomes, including barriers to their implementation.

Methods and findings

The review was registered with Prospero (CRD42016038679). Five databases were searched up to May 1st, 2016 for studies related to intrapartum foetal monitoring strategies and neonatal outcomes in low-resource settings.

Two authors extracted data and assessed the risk of bias for each study. The outcomes were narratively synthesised. Strengths, weaknesses, opportunities and threats analysis (SWOT) was conducted for each monitoring technique to analyse their implementation.

There were 37 studies included: five intervention and 32 observational studies. Use of the partograph improved perinatal outcomes. Intermittent auscultation with Pinard was associated with lowest rates of caesarean sections (10–15%) but with comparable perinatal outcomes to hand-held Doppler and Cardiotocography (CTG). CTG was associated with the highest rates of caesarean sections (28–34%) without proven benefits for perinatal outcome. Several tests on admission (admission tests) and adjunctive tests including foetal stimulation tests improved the accuracy of foetal heart rate monitoring in predicting adverse perinatal outcomes.

Conclusions

From the available evidence, the partograph is associated with improved perinatal outcomes and is recommended for use with intermittent auscultation for intrapartum monitoring in low resource settings. CTG is associated with higher caesarean section rates without proven benefits for perinatal outcomes, and should not be recommended in low-resource settings. High-quality evidence considering implementation barriers and enablers is needed to determine the optimal foetal monitoring strategy in low-resource settings.

Introduction

Over two million stillbirths are estimated to occur yearly worldwide, of which >98% are in low-resource settings [1,2]. Almost half of the number of stillbirths in low- and middle-income countries (LMICs) occur during labour, whereas most stillbirths in high-income countries (HICs) take place during the antenatal period [3,4]. The time of labour and delivery is a challenging period for the foetus and can result in foetal asphyxia and associated irreversible organ damage and mortality [58]. Intrapartum foetal monitoring allows for prompt and effective intervention when needed, and avoids unnecessary interventions like caesarean sections (CS) by offering confirmation of a favourable foetal condition [9]. Methods of foetal surveillance include foetal heart rate (FHR) monitoring by intermittent auscultation (IA), cardiotocography (CTG) with foetal blood sampling and foetal electrocardiogram with ST-wave analysis [10,11]. Nearly all methods are considered to be high-tech, complex in operation, and require significant financial resources [12,13].

Although global consensus exists that some form of foetal monitoring should be used during labour to improve maternal and neonatal outcomes, there is no evidence for an ideal foetal monitoring system [8,11,14]. Studies on foetal monitoring have been primarily conducted in HICs and, based on variable level of evidence, consensus-based guidelines were developed for foetal surveillance, which may not be readily applicable to LMICs due to context-specific factors [11,1517]. Thus, in many low resource settings, low-cost and low-tech methods such as IA by Pinard stethoscope or hand-held Doppler, are the only accessible methods [18]. A review on intrapartum foetal surveillance (implementation) strategies for LMICs is not available. Therefore, the aim of this systematic review was to synthesize the available evidence for intrapartum foetal surveillance in low resource settings and a SWOT analysis was applied to analyse the implementation.

Methods

This review was registered with the PROSPERO registry for systematic reviews (CRD42016038679). It adhered to PRISMA guidelines (S1 File) [19] and was conducted according to the Cochrane methodology [20].

Research questions

This review aimed to answer two research questions: (1) what is the available evidence for strategies of intrapartum foetal surveillance in low- and middle-income countries and their associated neonatal and maternal outcomes? (2) what are the strengths, weaknesses, opportunities, and threats (SWOT) associated with the implementation of these intrapartum foetal surveillance strategies?

Eligibility criteria

Observational or intervention studies concerning women receiving intrapartum foetal surveillance with reported neonatal outcomes in low resource settings were eligible for inclusion. These included studies on admission tests, which were defined as tests performed to determine foetal wellbeing upon arrival in labour in a birth facility. Low resource settings were defined as low-income, lower-middle- and upper-middle income countries (LICs, L-MICs, and UMICs respectively), according to the World Bank classification [21]. Conference abstracts, reports, editorials, presentations, and project protocols were excluded.

Information sources and search

The search was conducted in the following electronic databases: Pubmed/MEDLINE, The Cochrane Library, EMBASE, POPLINE and Global Health Library to include all articles up to May 1st, 2016. For every database, a search string was developed with the support of a librarian specialised in medical sciences, using pre-defined search (Title/Abstract) and MeSH/Emtree terms when applicable. References were manually searched for additional studies. Only for the Global Health Library, limits were used (humans/English). The full search strings are available in Appendix A in S2 File.

Study selection

Mendeley reference software was used to remove duplicates. Subsequently, two reviewers (MCP and NH) independently screened articles based on title and abstract, after which full-text screening was performed. In case of disagreement, a third reviewer (MJR) was consulted. Authors were contacted once in case of inaccessible full-texts, and a study excluded if no reply was received.

Data collection process

Data extraction of the included studies was conducted by one reviewer (MCP) and double-checked for accuracy by a second reviewer (NH). A standardised data extraction sheet was created (Appendix B in S2 File) SWOT analysis was applied to the methods, results and discussion sections of the selected articles whenever mentioned and recorded in the same extraction sheet as all other outcomes. Outcome measurements were noted as percentages and calculated when possible in case of different reporting strategy. Sensitivity, specificity, positive and negative predictive values (PPV and NPV respectively) were collected when available. The corresponding author or organisation was emailed once in case of incomplete data. In case of disagreements during the extraction process, other members of the review team were contacted (JB, MJR).

Risk of bias assessment

The level of bias was assessed for each study using the Cochrane Risk of Bias Tool (S3 File) and the Newcastle-Ottawa Quality Assessment Scale for intervention and observational studies, respectively (S4 and S5 File) [20,22,23]. Colour coding of the table was assigned as red, green and yellow for high, low and unclear (Cochrane) or intermediate (Newcastle-Ottawa) risk respectively. Judgement of bias was determined (MCP) and double-checked for accuracy (NH). Any disagreement during this process was resolved by contacting other members of the review team (JB, MJR).

Data synthesis

Due to heterogeneity in domains, determinants, study designs and reported outcomes, a senior statistician from the Cochrane Collaboration advised not to conduct a meta-analysis. This review, therefore, consists of a narrative analysis of strategies for intrapartum foetal surveillance and their corresponding outcomes. The quantitative results of all studies were summarised according to study design: intervention and descriptive studies. For each method of foetal monitoring, SWOT findings were summarised according to each component.

Results

A total of 10,195 articles were obtained after removal of duplicates and including nine articles from cross-referencing (Fig 1). After title- and abstract screening, 518 articles were screened in full-text, of which 38 were included. Two publications reported on the same study [24,25]. The final 37 included studies consisted of five (13.5%) intervention studies (three randomised controlled trials (RCT)[2628] and two clustered RCT [24,25,29]), and 32 (86.5%) observational studies (23 cohort studies, six cross-sectional studies, and three case-control studies) [3060]. The studies were conducted in Africa (n = 16), Asia (n = 21) and Europe (n = 1) and were from LICs (n = 6), L-MICs (n = 21) and UMICs (n = 11). Many studies were from urban settings (urban: n = 16, rural: n = 1, both: 2), for 18 studies this could not be determined. Studies were on: admission tests and early intrapartum (CTG, n = 7, IA: 1, other methods, n = 6) [3235,44,48,51,56,57,59,61], ongoing intrapartum FHR monitoring (IA, n = 8; CTG, n = 11) [2628,30,31,3639,41,43,4547,50,53,58], adjunctive tests (n = 9) [39,4143,45,46,49,52,58] and partograph (n = 5) [24,25,29,40,54,55] (Tables 13).

thumbnail
Table 1. Quality assessment of randomised controlled trials (n = 5).

https://doi.org/10.1371/journal.pone.0206295.t001

thumbnail
Table 2. Quality assessment of the observational studies (n = 32).

https://doi.org/10.1371/journal.pone.0206295.t002

thumbnail
Table 3. Foetal monitoring methods as predictors of birth outcomes.

https://doi.org/10.1371/journal.pone.0206295.t003

Risk of bias of studies

A summary of quality assessment for intervention studies is provided in Table 1 and for observational studies in Table 2. Study performance of the five intervention studies was overall moderate, however, blinding of the participants or researchers was not done (5/5 high risk) and confounders were often not considered (1/5 high risk; 3/5 unclear risk and 1/5 low risk). Quality of observational studies was low to moderate; classified as low risk in: 81.3% for selection process, in 25% for comparability and in 25% outcome/exposure of studies.

Narrative synthesis of quantitative results

A summary of the FHR monitoring strategies and their outcomes is provided in Table 3. Detailed results of each intervention and descriptive study are presented in S1S3 Table.

Admission tests

Neonatal outcomes.

We identified only observational studies for admission tests. The study of IA on admission (n = 1) showed that absent FHR by hand-held Doppler was associated with a much higher perinatal mortality (938/1000 deliveries) compared to when FHR was present (13/1000 deliveries) [59].

Admission CTG, a 20-minute recording, was assessed in seven studies in Asia. Studies were on low risk (n = 3) [44,57,60], high risk (n = 3) [34,48,56], and mixed-risk pregnancies(n = 1) [32]. Abnormal CTG traces were associated with intrapartum FHR abnormalities (foetal distress) [32,34,44,48,57], meconium-stained liquor [32,57]; low Apgar scores at 5 minutes [32,44,48,57,60], perinatal deaths [32,44,48], and admission to neonatal intensive care unit (NICU) [32,44,48,57]. Test performance of admission CTG varied across studies: PPV of 19% to 88%, while the NPV was between 88.6% to 100% [32,48,56,57,60].

Maternal perception of sound-provoked foetal movement (i.e. foetal acoustic stimulation, FAST, n = 2) performed well in predicting foetal distress, perinatal death, Apgar score <7 at 5 minutes, admission to NICU [60,61]. It also improved the test performance of CTG in two studies (PPV: 45.2% to 65.5%, 19% to 73.6%, NPV: 94.2% to 96%, 100% to 100%) [56,60]. Admission amniotic fluid index (aAFI) performed worse than admission CTG (specificity: 64% and 92% respectively, n = 326) [35]. In one study (n = 330) rapid Biophysical profile (rBPP i.e. combination of sound-provoked ultrasound-detected foetal movement and AFI) had PPV (50%) and NPV (99.1%) for poor neonatal outcomes [51]. Umbilical artery Doppler velocity (UADV) in labour did not predict neonatal outcomes in two studies [33,49].

Maternal outcomes.

Only studies on admission CTG reported mode of delivery. Abnormal traces increased CS rates compared to reactive CTG traces (between 42.7% to 100% and 20.7% to 30%, respectively, p<0.05) [32,34,44,57].

Ongoing intrapartum foetal surveillance

Neonatal outcomes.

There were three RCTs comparing IA and CTG: Uganda (n = 1971) [27], Zimbabwe (n = 1255) [26], India (n = 100) [28]. In Uganda, hand-held Doppler and Pinard stethoscope were compared [27]. The RCT in Zimbabwe had four arms: 1) intermittent CTG traces(n = 318) 2) hand-held Doppler (n = 312), 3) Pinard (n = 310)and 4) routine monitoring with Pinard (n = 315). In the first three groups, research midwives ensured they assessed FHR every 30 minutes for 10 minutes per study protocol and caregiving midwives were supposed to adhere to the same frequency by following hospital protocol [26]. Continuous CTG monitoring (n = 50) was compared to IA (n = 50) in women with a history of CS in India [28]. In these studies, detection of FHR abnormalities was significantly different in Pinard, Doppler and CTG groups (Table S1and S2). However, no significant changes in perinatal deaths, low Apgar scores at 1 and 5 minutes and admission to NICU were observed [2628].

The study in Zimbabwe reported fewer cases of neonatal seizures and hypoxic-ischaemic encephalopathy (HIE) in the hand-held Doppler group compared to the Pinard groups (zero vs 15; and one vs 17 respectively) [26]. Although foetal distress was diagnosed in the three treatment groups, protocol violations, delays or unavailable operative deliveries led to the majority of perinatal deaths [26].

One observational study (Tanzania, n = 10271) showed that detection of an absent or abnormal FHR with foetal stethoscope was strongly associated with fresh stillbirths, neonatal deaths, low Apgar score and neonatal resuscitation [36]. In observational studies, pathological CTG traces were associated with low Apgar score at one minute [30,39], umbilical cord indices [30,50] and HIE [53] as compared to normal traces. However, contrasting findings were seen for umbilical cord indices (PPV 11.6% vs 100%) [43,50] and five minutes Apgar scores [30,37].

Several studies identified adjunctive tests for FHR monitoring. Foetal pulse oximetry had a comparable test performance compared to foetal blood sampling (n = 150) [42]. Meconium was mostly effective in predicting neonatal outcomes when combined with abnormal FHR [39,41,43,46]. Foetal scalp stimulation test (FSST) combined with IA were good predictors of perinatal outcomes: umbilical cord pH, Apgar score at one and five minutes, neonatal death and NICU admission [45].

The multi-centre partograph-intervention study in Southeast Asia which included 35 484 women showed a significant reduction in intrapartum stillbirths (0.50% to 0.31%, p = 0.024), but no significant reduction in Apgar scores, neonatal deaths, NICU, and resuscitation [24,25]. Training midwives to use the partograph reduced low Apgar scores at 1 minute but no improvement in other perinatal outcomes [29]. Observational studies showed that crossing the alert and action lines on the partograph was associated with a higher incidence of neonatal resuscitation and fresh stillbirths [54,55]. Substandard use of partograph was associated with low Apgar score [40].

Maternal outcomes.

The RCT in Zimbabwe showed that CTG and hand-held Doppler significantly increased CS rates due to foetal distress compared to Pinard. (63%, 67% and 41% respectively) [26]. The RCT in India showed a trend towards increasing CS rate in the CTG group due to foetal distress compared to IA (47% vs 18%) [28]. The Uganda RCT showed no difference in overall CS rates between hand-held Doppler and Pinard [27]. No clear difference was observed for operative vaginal delivery [26,28]. Duration of labour [26]. postpartum haemorrhage, maternal fever, ruptured uterus and maternal death [28] were similar. Meconium was associated with increased CS rates in India (clear liquor 17% vs meconium 33%) [39,58]. Nonreactive FSST detected by IA was associated with a significant increase in operative vaginal deliveries and CS rates [45]. Two clustered RCT on the partograph showed that training and the use of partograph led to significant reduction in length of labour and obstructed labour and oxytocin use but no changes in CS rate or maternal mortality [24,25,29]. There was no increased CS rate due to foetal distress. There was a reduction in vaginal examinations but no change in postpartum haemorrhage and maternal sepsis.27,33 The partograph significantly increased the number of referrals of women in labour to higher level centres [29].

Narrative synthesis of SWOT analysis

Detailed SWOT results of the given foetal monitoring methods are provided in Table 4. Admission CTG were recommended for triaging labours and resource allocation when resources are scarce [32,48,57]. The Pinard, hand-held Doppler and partograph were strategies reported as simple and low-cost [2527,36,54,59]. IA allowed for greater mobility of the women than CTG and was easily accessible, but difficult to carry out in busy maternity wards [43]. The hand-held Doppler may be more mother- and user-friendly than the Pinard [27,59] but required consumables [36]. Some of these challenges were eliminated when using the wind-up Doppler. The use of CTG required a high level of skills, resources, and costs [26,31]. Combining FHR monitoring with simpler adjunctive tests such as meconium, FAST, FSST, and FPO may provide non-invasive and reliable ways to confirm foetal wellbeing, avoiding unnecessary interventions [42,45,56,60,61].

thumbnail
Table 4. SWOT analysis of methods of intrapartum foetal monitoring.

https://doi.org/10.1371/journal.pone.0206295.t004

Strengths of the partograph were its low-cost, pictorial overview of labour allowing timely recognition for complications and action [25,29,40,54,55]. A major threat was an underuse of partograph due to a shortage of staff, lack of knowledge, training, and guidelines, unavailability of copies and hesitant attitudes of staff [29,40,54]. Opportunities to increase partograph use lie in providing partograph copies, training, and appropriate management guidelines [25,29,40,54]. A major threat to all intrapartum foetal surveillance studies was limited or unavailability of intervention including timely operative deliveries [26,27,31,36].

Discussion

Main findings

This systematic review and SWOT analysis provide an overview of the evidence of intrapartum foetal monitoring strategies in low-resource settings on perinatal and maternal outcomes. The use of CTG increased the rates of CS but had no effect on adverse perinatal outcomes compared to IA [26,28]. IA and the partograph is the preferred method in low-resource settings for FHR monitoring.

The observational studies in this review suggest that admission tests (including CTG, IA or FAST) can predict adverse outcomes in LMICs, and mode of delivery in both low and high-risk pregnancies [32,34,48,56,57,60,61]. We suggest that admission tests might have a much better use in low resource settings because of: 1) the incidence of intrapartum stillbirths could modify the predictive test results [11], 2) inadequate risk assessment and stratification during antenatal care, making admission tests a good screening tool to identify high-risk foetuses and 3) a triaging tool for better allocation of resources in settings with heavy workload and scarce (human) resources [32,34,48,57].

The overall evidence shows that CTG does not improve outcomes but increases the number of CS compared to IA. It is unclear whether hand-held Doppler improves neonatal outcomes, and it may increase CS rate. Similar findings on CTG and hand-held Doppler are reported in the Cochrane meta-analyses [11,62]. A study in South Africa showed pregnant women preferred hand-held Doppler over Pinard or CTG [63]. However, the number of CS presents real concerns for maternal safety in low resource settings [6468]. Foetal heart monitoring may have false positivity for foetal distress leading to unnecessary intervention. The current review identified simple and cheap strategies to strengthen the test performance of intrapartum FHR monitoring including foetal stimulation tests (FAST and FSST) and meconium. However, their effectiveness is not known and should be tested in future studies. Contrary to a Cochrane review, which did not include the large study in South East Asia [69], the partograph was useful for monitoring and decision-making for the intrapartum care of the mother, foetus and labour progress, and was associated with reduced intrapartum stillbirths in low-resource settings [25,29,40,54,55]. The BOLD initiative and WHO guidelines stress the importance of supportive, person-centred care during labour and childbirth rather than focus on cervical dilatation only [7072].

Challenges exist in up-scaling effective interventions in low-resource settings [18,73]. Given the resource constraints, the SWOT analysis shows that the ideal method of intrapartum foetal monitoring should be: simple, affordable, robust, safe, reliable and sustainable [18,74]. Yet, most monitoring systems require maintenance and adequate staffing who need to be trained and supervised. For example, although IA and partographs are low-tech and -cost technology, they highly depend on human resources. A strong commitment to investing in high quality research of existing and new strategies of real-life implementation for intrapartum foetal monitoring is required. These may include new ways to monitor foetal well-being, context-appropriate guidelines, and healthcare workforce strengthening [15,75]. A substantial time-lag between recognition of foetal compromise and delivery as a major cause of severe asphyxia and death was identified in this review [26,27,31,36]. Importantly, emergency obstetric and newborn care including operative vaginal deliveries and neonatal resuscitation should be readily available to ensure both prompt diagnosis and successive intervention.

Strengths and limitations

A strength of this review is the systematic assessment of neonatal and maternal outcomes and SWOT analysis. Although an extensive and inclusive search in five international databases was conducted, studies performed in low-resource settings and published in national journals might not have been indexed in the searched databases. Limitations are also inherent in the reviewed articles and include the quality of the evidence, the lack of detailed reporting of implementation factors and relevant outcomes such as contraction monitoring, maternal morbidity and mortality, CS rates, professional and maternal opinion. RCTs did not guarantee appropriate and timely interventions which confounded the results. We intended to evaluate evidence for all intrapartum foetal monitoring strategies in low-resource settings using a meta-analysis, however, due to heterogeneity in designs and outcomes, only a narrative review could be performed.

Conclusion

Of the foetal monitoring strategies that have been studied in LMICs, the partograph and intermittent auscultation is the preferred strategy for intrapartum foetal surveillance in low-resource settings because of reduced intrapartum stillbirths (partograph), lower caesarean section rates (Pinard) and easier implementation. CTG is associated with higher caesarean section rates without proven benefits for perinatal outcomes, and should not be recommended in low-resource settings until new research delivers evidence for better perinatal outcomes. The benefit and harms of admission tests, adjunctive tests and hand-held Doppler on perinatal and maternal outcomes should be determined in future studies in low resource settings. High-quality RCT studies of foetal monitoring should include clear management protocols with timely interventions. Moreover, there is a need to harmonise core outcomes in foetal monitoring studies. Consideration of implementation factors will also be essential to determine the real-world optimal foetal monitoring approach.

Supporting information

S2 File. Full search strategy for five databases and data extraction sheet.

https://doi.org/10.1371/journal.pone.0206295.s002

(DOCX)

S4 File. New Castle-Ottawa scale for cohort and case-control studies.

https://doi.org/10.1371/journal.pone.0206295.s004

(PDF)

S5 File. New Castle-Ottawa scale cross-sectional studies.

https://doi.org/10.1371/journal.pone.0206295.s005

(DOC)

S2 Table. Characteristics and quantitative results of included randomised controlled trials.

https://doi.org/10.1371/journal.pone.0206295.s007

(DOCX)

S3 Table. Characteristics and quantitative results of included observational studies.

https://doi.org/10.1371/journal.pone.0206295.s008

(DOCX)

Acknowledgments

We would like to acknowledge the valuable support of the UMC Utrecht Librarians (Paulien Wiersma and Tessa Pronk) regarding the search string for the databases and the Dutch Cochrane Centre (Rob Scholten) for consultation in data analysis.

References

  1. 1. GBD 2015 Child Mortality Collaborators. Global, regional, national, and selected subnational levels of stillbirths, neonatal, infant, and under-5 mortality, 1980–2015: a systematic analysis for the Global Burden of Disease Study 2015. Lancet 2016;388.
  2. 2. Blencowe H, Cousens S, Jassir FB, Say L, Chou D, Mathers C, et al. National, regional, and worldwide estimates of stillbirth rates in 2015, with trends from 2000: a systematic analysis. Lancet Glob Heal 2016;4:e98–108.
  3. 3. Lawn JE, Blencowe H, Waiswa P, Amouzou A, Mathers C, Hogan D, et al. Stillbirths: rates, risk factors, and acceleration towards 2030. Lancet 2016;387:587–603. pmid:26794078
  4. 4. Goldenberg RL, McClure EM, Bann CM. The relationship of intrapartum and antepartum stillbirth rates to measures of obstetric care in developed and developing countries. Acta Obstet Gynecol Scand 2007;86:1303–9. pmid:17963057
  5. 5. Velaphi S, Pattinson R. Avoidable factors and causes of neonatal deaths from perinatal asphyxia-hypoxia in South Africa: National perinatal survey. Ann Trop Paediatr 2007;27:99–106. pmid:17565806
  6. 6. Goldenberg RL, Saleem S, Pasha O, Harrison MS, McClure EM. Reducing stillbirths in low-income countries. Acta Obs Gynecol Scand 2016;95:135–43.
  7. 7. Pattinson R. Reducing intrapartum stillbirths—Experience in South Africa. Int J Gynecol Obstet 2009;107:S66–7.
  8. 8. Ayres-De-Campos D, Arulkumaran S. FIGO consensus guidelines on intrapartum fetal monitoring: Physiology of fetal oxygenation and the main goals of intrapartum fetal monitoring. Int J Gynecol Obstet 2015;131:5–8.
  9. 9. Lewis D, Downe S. FIGO consensus guidelines on intrapartum fetal monitoring: Intermittent auscultation. Int J Gynecol Obstet 2015;131:9–12.
  10. 10. Lewis D, Downe S, Ayres-De-Campos D, Arulkumaran S, Visser GH, Ayres-De-Campos D, et al. FIGO consensus guidelines on intrapartum fetal monitoring: Adjunctive technologies. Int J Gynecol Obstet 2015;131:25–9.
  11. 11. Alfirevic Z, Devane D, Gml G, Cuthbert A. Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour (Review). Cochrane Database Syst Rev 2017:Art. No.: CD006066.
  12. 12. Vijgen SMC, Westerhuis MEMH, Opmeer BC, Visser GHA, Moons KGM, Porath MM, et al. Cost-effectiveness of cardiotocography plus ST-analysis of the fetal electrocardiogram compared to cardiotocography only in the prevention of cerebral palsy. Am J Obstet Gynecol 2009;201:S192.
  13. 13. The Royal College of Obstetrics and Gynaecology. The use of electronic monitoring: The use and interpretation of cardiotocography in intrapartum fetal surveillance. Clinical Guideline number 8. 2001.
  14. 14. Ayres-De-Campos D, Arulkumaran S. FIGO consensus guidelines on intrapartum fetal monitoring: Introduction. Int J Gynecol Obstet 2015;131:3–4.
  15. 15. Maaløe N, Housseine N, Roosmalen J Van, Bygbjerg IC, Tersbøl BP, Khamis RS, et al. Labour management guidelines for a Tanzanian referral hospital: The participatory development process and birth attendants ‘ perceptions. BMC Pregnancy Childbirth 2017;17:1–11.
  16. 16. Pitchforth E, Lilford RJ, Kebede Y, Asres G, Stanford C, Frost J. Social Science & Medicine Assessing and understanding quality of care in a labour ward: A pilot study combining clinical and social science perspectives in Gondar, Ethiopia. Soc Sci Med 2010;71:1739–48. pmid:20855142
  17. 17. Kruk ME, Leslie HH, Verguet S, Mbaruku GM, Adanu RMK, Langer A. Quality of basic maternal care functions in health facilities of fi ve African countries: an analysis of national health system surveys. Lancet 2016;4:e845–55. pmid:27670090
  18. 18. Justus Hofmeyr G, Haws RA, Bergstr??m S, Lee ACC, Okong P, Darmstadt GL, et al. Obstetric care in low-resource settings: What, who, and how to overcome challenges to scale up? Int J Gynecol Obstet 2009;107:S21–45.
  19. 19. Moher D, Liberati A, Tetzlaff J, Altman DG, Group TP, Oxman A, et al. Preferred Reporting Items for Systematic Reviews and Meta-Analyses: The PRISMA Statement. PLoS Med 2009;6:e1000097. pmid:19621072
  20. 20. Higgings JP E. Cochrane Handbook for Systematic Reviews of Interventions Version 5.1.0 [updated March 2011]. The Cochrane Collaboration, 2011. Available from www.handbook.cochrane.org.
  21. 21. The World Bank. World Bank Group. World Bank Country and Lending Groups. [cited 2016 Jul 25]. Available from: https://datahelpdesk.worldbank.org/knowledgebase/articles/906519-world-bank-country-and-lending-groups
  22. 22. Wells G, Shea B, O’Connell D, Peterson J, Welch V, Losos M, et al. The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomised studies in meta-analyses. Ottawa Research Institute; c2016 [cited 2016 Jul 25]. Available from: http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp
  23. 23. Herzog R, Álvarez-pasquin MJ, Díaz C, Luis J, Barrio D, Estrada JM, et al. BMC Public Health; c2016 [cited 2016 Jul 25]. Are healthcare workers’ intentions to vaccinate related to their knowledge, beliefs and attitudes? a systematic review. [Additional file 3] Available on: https://bmcpublichealth.biomedcentral.com/articles/10.1186/1471-2458-13-154
  24. 24. Lennox CE, Kwast BE, Farley TM. Breech labor on the WHO partograph. Int J Gynaecol Obs 1998;62:117–27.
  25. 25. World Health Organization partograph in management of labour. World Health Organization Maternal Health and Safe Motherhood Programme. Lancet (London, England) 1994;343:1399–404.
  26. 26. Mahomed K, Nyoni R, Mulambo T, Kasule J, Jacobus E. Randomised controlled trial of intrapartum fetal heart rate monitoring. BMJ 1994;308:497–500. pmid:8136665
  27. 27. Byaruhanga R, Bassani DG, Jagau A, Muwanguzi P, Montgomery AL, Lawn JE. Use of wind-up fetal Doppler versus Pinard for fetal heart rate intermittent monitoring in labour: a randomised clinical trial. BMJ Open 2015;5:e006867. pmid:25636792
  28. 28. Madaan M, Trivedi SS. Intrapartum electronic fetal monitoring vs. intermittent auscultation in postcesarean pregnancies. Int J Gynaecol Obs 2006;94:123–5.
  29. 29. Fahdhy M, Chongsuvivatwong V, V FMC. Evaluation of World Health Organization partograph implementation by midwives for maternity home birth in Medan, Indonesia. Midwifery 2005;21:301–10. pmid:16076515
  30. 30. Aboulghar WM, Ibrahim MA, Allam IS, Hosny W, Otify M. Validity of cardiotocography in the diagnosis of acute fetal hypoxia in low resources settings. Internet J Gynecol Obstet 2013;17.
  31. 31. Adanikin AI, Awoleke JO. Clinical suspicion, management and outcome of intrapartum fetal distress in a public hospital with limited advanced fetal surveillance. J Matern Fetal Neonatal Med 2016:1–20.
  32. 32. Rahman H, Renjhen P, Dutta S. Reliability of admission cardiotocography for intrapartum monitoring in low resource setting. Niger Med J 2012;53:145–9. pmid:23293415
  33. 33. Howarth GR, Pattinson RC, Kirsten G, Truter H, Odendaal HJ. Umbilical artery Doppler velocimetry in the prediction of intrapartum fetal compromise. S Afr Med J 1992;81:248–50. pmid:1542814
  34. 34. Kulkarni AA, Shrotri AN. Admission test: a predictive test for fetal distress in high risk labour. J Obs Gynaecol Res 1998;24:255–9.
  35. 35. Kushtagi P, Deepika KS. Amniotic fluid index at admission in labour as predictor of intrapartum fetal status. J Obs Gynaecol 2011;31:393–5.
  36. 36. Langli Ersdal H, Mduma E, Svensen E, Sundby J, Perlman J. Intermittent detection of fetal heart rate abnormalities identify infants at greatest risk for fresh stillbirths, birth asphyxia, neonatal resuscitation, and early neonatal deaths in a limited-resource setting: a prospective descriptive observational stud. Neonatology 2012;102:235–42. pmid:22907583
  37. 37. Odendaal HJ, Sandenbergh HA. Acceleration patterns of the fetal heart rate before and during labour. S Afr Med J 1977;52:473–5. pmid:905907
  38. 38. Odendaal HJ, Steyn W, Theron GB, Norman K, Kirsten GF. Does a nonreactive fetal heart rate pattern really mean fetal distress? Am J Perinatol 1994;11:194–8. pmid:8048984
  39. 39. Odongo BE, Ndavi PM, Gachuno OW, Sequeira E. Cardiotocography and perinatal outcome in women with and without meconium stained liquor. East Afr Med J 2010;87:199–204. pmid:23057282
  40. 40. Ogwang S, Karyabakabo Z, Rutebemberwa E, OSKZR E. Assessment of partogram use during labour in Rujumbura Health Sub District, Rukungiri District, Uganda. Afr Heal Sci 2009;9 Suppl 1:S27–34.
  41. 41. Oladapo OT, Sotimehin SA, Ayoola-Sotubo O. Predictors of severe neonatal compromise following caesarean section for clinically diagnosed foetal distress. West Afr J Med 2009;28:327–32. pmid:20383839
  42. 42. Bakr AF, Al-Abd M, Karkour T. Fetal pulse oximetry and neonatal outcome: a study in a developing country. J Perinatol 2005;25:759–62. pmid:16281048
  43. 43. Parveen S. Umbilical cord arterial blood base excess as gold standard for foetal well being screening test validity at term delivery. J Pak Med Assoc 2010;60:347–50. pmid:20527603
  44. 44. Raouf S, Sheikhan F, Hassanpour S, Bani S, Torabi R, Shamsalizadeh N. Diagnostic value of non stress test in latent phase of labor and maternal and fetal outcomes. Glob J Heal Sci 2015;7:177–82.
  45. 45. Rathore AM, Ramji S, Devi CB, Saini S, Manaktala U, Batra S. Fetal scalp stimulation test: an adjunct to intermittent auscultation in non-reassuring fetal status during labor. J Obs Gynaecol Res 2011;37:819–24.
  46. 46. Rotich SK, Ndavi MP, Rukaria-Kaumbutho R, Kigondu CS. Early perinatal outcome in cases delivered through Caesarian section following clinical diagnosis of severe foetal distress at Kenyatta National Hospital. East Afr Med J 2006;83:250–8. pmid:16866219
  47. 47. Roy KK, Baruah J, Kumar S, Deorari AK, Sharma JB, Karmakar D. Cesarean section for suspected fetal distress, continuous fetal heart monitoring and decision to delivery time. Indian J Pediatr 2008;75:1249–52. pmid:19190880
  48. 48. Sandhu GS, Raju R, Bhattacharyya TK, Shaktivardhan . Admission cardiotocography screening of high risk obstetric patients. Med J Armed Forces India 2008;64:43–5. pmid:27408079
  49. 49. Stuart IP, Lindow SW, van der Elst CW. Fetal acidosis and Doppler velocimetry of the umbilical arteries in labor. Ultrasound Obs Gynecol 1993;3:256–9.
  50. 50. Tasnim N, Mahmud G, Akram S. Predictive accuracy of intrapartum cardiotocography in terms of fetal acid base status at birth. J Coll Physicians Surg Pak 2009;19:632–5. pmid:19811714
  51. 51. Tongprasert F, Jinpala S, Srisupandit K, Tongsong T, TFJSSKT T, Tongprasert F, et al. The rapid biophysical profile for early intrapartum fetal well-being assessment. Int J Gynaecol Obs 2006;95:14–7.
  52. 52. Gupta V, Bhatia BD, Mishra OP. Meconium stained amniotic fluid: antenatal, intrapartum and neonatal attributes. Indian Pediatr 1996;33:293–7. pmid:8772903
  53. 53. Bogdanovic G, Babovic A, Rizvanovic M, Ljuca D, Grgic G, Djuranovic-Milicic J. Cardiotocography in the prognosis of perinatal outcome. Med Arch 2014;68:102–5. pmid:24937932
  54. 54. Dujardin B, De Schampheleire I, Sene H, Ndiaye F. Value of the alert and action lines on the partogram. Lancet 1992;339:1336–8. pmid:1350000
  55. 55. Bolbol-Haghighi N, Keshavarz M, Delvarianzadeh M, Molzami S. Evaluation of the alert line of partogram in recognizing the need for neonatal resuscitation. Iran J Nurs Midwifery Res 2015;20:560–4. pmid:26457092
  56. 56. Chittacharoen A, Chaitum A, Suthutvoravut S, Herabutya Y. Fetal acoustic stimulation for early intrapartum assessment of fetal well-being. Int J Gynaecol Obs 2000;69:275–7.
  57. 57. David B, Saraswathi K. Role of admission CTG as a screening test to predict fetal outcome and mode of delivery. Res J Pharm Biol Chem Sci 2014;5:295–9.
  58. 58. Duhan N, Paul A, Duhan U, Anjali . Meconium staining of amniotic fluid- a poor indicator of fetal compromise. JK Sci 2010;12:184–6.
  59. 59. Goldenberg RL, McClure EM, Kodkany B, Wembodinga G, Pasha O, Esamai F, et al. A multi-country study of the “intrapartum stillbirth and early neonatal death indicator” in hospitals in low-resource settings. Int J Gynaecol Obstet 2013;122:230–3. pmid:23796259
  60. 60. Goonewardene M, Hanwellage K. Fetal acoustic stimulation test for early intrapartum fetal monitoring. Ceylon Med J 2011;56:14–8. pmid:21542428
  61. 61. Chittacharoen A, Srijhantuik K, Suthutvoravut S, Herabutya Y. Maternal perception of sound-provoked fetal movement in the early intrapartum period. Int J Gynaecol Obs 1997;56:129–33.
  62. 62. Martis R, Emilia O, Nurdiati Detty S. Intermittent auscultation (IA) of fetal heart rate in labour for fetal well-being. Cochrane Database Syst Rev 2010.
  63. 63. Mangesi L, Hofmeyr G, Woods D. Assessing the preference of women for different methods of monitoring the fetal heart in labour. S Afr J Obstet Gynaecol 2009;15:58–9.
  64. 64. Briand V, Dumont A, Abrahamowicz M, Sow A, Traore M, Rozenberg P, et al. Maternal and Perinatal Outcomes by Mode of Delivery in Senegal and Mali: A Cross-Sectional Epidemiological Survey. PLoS ONE 7(10) 2012;7.
  65. 65. Gebhardt GS, Chb MB, Sci M, Mmed O, Sa F, Fawcus S, et al. Maternal death and caesarean section in South Africa: Results from the 2011–2013 Saving Mothers Report of the National Committee for Confidential Enquiries into Maternal Deaths. S Afr Med J 2015;105:287–91. pmid:26294872
  66. 66. Harrison MS, Goldenberg RL. Cesarean section in sub-Saharan Africa. Matern Heal Neonatol Perinatol 2016:1–10.
  67. 67. Iyoke CA, Ugwu GO, Ezugwu FO, Lawani OL, Onah HE. Risks associated with subsequent pregnancy after one caesarean section: A prospective cohort study in a Nigerian obstetric population. Niger J Clin Pract 2014;17.
  68. 68. Litorp H, Kidanto HL, Rööst M, Abeid M, Nyström L, Essén B. Maternal near-miss and death and their association with caesarean section complications: a cross-sectional study at a university hospital and a regional hospital in Tanzania. BMC Pregnancy Childbirth 2014;14.
  69. 69. Lavender T, Hart A, Smyth RM. Effect of partogram use on outcomes for women in spontaneous labour at term. Cochrane Database Syst Rev 2013;7:CD005461.
  70. 70. WHO recommendations: intrapartum care for a positive childbirth experience. Geneva: World Health Organization; 2018. Licence: CC BY-NC-SA 3.0 IGO. n.d.
  71. 71. Oladapo OT, Souza JP, Fawole B, Mugerwa K, Alves D, Souza H, et al. Progression of the first stage of spontaneous labour: A prospective cohort study in two sub- Saharan African countries. PLoS Med 2018;15
  72. 72. Souza J, Oladapo O, Fawole B, Mugerwa K, Reis R, Barbosa-Junior F, et al. Cervical dilatation over time is a poor predictor of severe adverse birth outcomes: a diagnostic accuracy study. BJOG 2018.
  73. 73. Dickson KE, Simen-kapeu A, Kinney M V, Huicho L, Vesel L, Lackritz E, et al. Every Newborn 4 Every Newborn: health-systems bottlenecks and strategies to accelerate scale-up in countries 2014;384.
  74. 74. Wyatt J. Appropriate medical technology for perinatal care in low-resource countries. Ann Trop Paediatr 2008;28:243–51. pmid:19021939
  75. 75. de Francisco Shapovalova N, Meguid T, Campbell J, Al. E. Health-care workers as agents of sustainable development. Lancet Glob Heal 2015;3:e249–50.