Surface radiation dose comparison of a dedicated extremity cone beam computed tomography (CBCT) device and a multidetector computed tomography (MDCT) machine in pediatric ankle and wrist phantoms

  • Sebastian Tschauner ,

    Contributed equally to this work with: Sebastian Tschauner, Robert Marterer

    sebastian.tschauner@medunigraz.at

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Robert Marterer ,

    Contributed equally to this work with: Sebastian Tschauner, Robert Marterer

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Eszter Nagy,

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Georg Apfaltrer,

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Michael Riccabona,

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Georg Singer,

    Affiliation Division of General Pediatric and Adolescence Surgery, Medical University of Graz, Graz, Austria

  • Georg Stücklschweiger,

    Affiliation Competence Centre for Medical Physics and Radiation Protection, University Hospital Graz, Graz, Austria

  • Helmuth Guss,

    Affiliation Competence Centre for Medical Physics and Radiation Protection, University Hospital Graz, Graz, Austria

  • Erich Sorantin

    Affiliation Division of Pediatric Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

Abstract

Objectives

To evaluate and compare surface doses of a cone beam computed tomography (CBCT) and a multidetector computed tomography (MDCT) device in pediatric ankle and wrist phantoms.

Methods

Thermoluminescent dosimeters (TLD) were used to measure and compare surface doses between CBCT and MDCT in a left ankle and a right wrist pediatric phantom. In both modalities adapted pediatric dose protocols were utilized to achieve realistic imaging conditions. All measurements were repeated three times to prove test-retest reliability. Additionally, objective and subjective image quality parameters were assessed.

Results

Average surface doses were 3.8 ±2.1 mGy for the ankle, and 2.2 ±1.3 mGy for the wrist in CBCT. The corresponding surface doses in optimized MDCT were 4.5 ±1.3 mGy for the ankle, and 3.4 ±0.7 mGy for the wrist. Overall, mean surface dose was significantly lower in CBCT (3.0 ±1.9 mGy vs. 3.9 ±1.2 mGy, p<0.001). Subjectively rated general image quality was not significantly different between the study protocols (p = 0.421), whereas objectively measured image quality parameters were in favor of CBCT (p<0.001).

Conclusions

Adapted extremity CBCT imaging protocols have the potential to fall below optimized pediatric ankle and wrist MDCT doses at comparable image qualities. These possible dose savings warrant further development and research in pediatric extremity CBCT applications.

Background

Extremity cone beam computed tomography (CBCT) represents an imaging modality that produces cross-sectional studies from volumetric acquisitions. It employs an X-ray tube and a flat-panel detector rotating opposite to each other in a gantry [13]. The main difference to modern multidetector computed tomography (MDCT) scanners operated in volumetric mode is the use of a flat panel detector and the absence of pre-detector collimation [2, 3]. The resulting simplicity of the CBCT construction facilitates small and mobile machines [4] and weight-bearing applications [511].

CBCT as an imaging method is not a new concept, and it is widely used in dental imaging for instance [1218]. Previous studies on extremity CBCT indicated comparable diagnostic value compared to MDCT and characterized it as a possible alternative in adult orthopedic imaging [1922]. Recently, initial experiences of pediatric CBCT applications were published as well [23].

Extraordinary dose-saving potentials in adults have been described in the knee and ankle region [24, 25]. These low-dose capabilities drew attention of pediatric radiology research on the new modality lately. Pugmire et al. published lower extremity CBCT doses in children compared to MDCT [23]. Conversely, Lang et al. and Neubauer et al. did not substantiate these findings and reported comparable doses of CBCT and MDCT examinations, when taking image quality into account [20, 22]. As a consequence, actual CBCT dose savings compared to MDCT remain a legitimate subject for debate and research.

The purpose of the current study was to compare surface radiation doses of an established and optimized MDCT and a novel extremity CBCT scanner, both depicted in Fig 1a, in pediatric anthropomorphic ankle and wrist phantoms. Adapted pediatric imaging protocols on both machines were used to achieve realistic results at our dedicated pediatric radiology division. Surface doses were assessed by employing thermoluminescence dosimeters (TLDs) [26].

thumbnail
Fig 1. Images of measurement setup.

a) CBCT (front) and MDCT (background) scanners used for dosimetry, b) pediatric ankle phantom in the CBCT gantry, c) pediatric wrist phantom in the CBCT gantry.

https://doi.org/10.1371/journal.pone.0178747.g001

Methods

Dosimetric studies were conducted on a left lower leg and a right forearm of a pediatric whole body phantom PBU-70 (Kyoto Kagaku Co. Ltd, Kyoto, Japan) [27]. The anthropomorphic phantoms were modelled after a 4-year-old child of 105 cm height and 20 kg weight. The used extremities embedded cortical and cancellous bone in soft tissue equivalent material (Fig 1b and 1c).

CBCT dosimetry was performed on a Planmed Verity® scanner (Planmed Oy, Helsinki, Finland). Aided by laser position markers, the phantoms were centered in the middle of the gantry. Both extremities were positioned proximo-distally with the intra-articular spaces of either the talocrural or the radioulnar joint at the center (compare Fig 1b and 1c). The field of view (FOV) had a diameter of 16 cm and an extension of 12 cm proximo-distally. CBCT tube rotation angle was 210°.

Correspondingly, multidetector computed tomography (MDCT) dosimetric studies were conducted on a Toshiba Aquilion One® (Toshiba Medical Systems Corporation, Otawara-shi, Japan), a 320-row scanner. Equally to the CBCT, ankle and wrist phantoms were centered in the gantry. Proximo-distal scan length was 12 cm. A “small” FOV (diameter of 9.94 cm) was used for the ankle, and a “double small” FOV (diameter of 9.28 cm) for the wrist. All MDCT volumetric acquisitions were completed in a single tube rotation of 360° without table incrementation (no pitch).

Three study protocols were measured both in the left ankle and the right wrist phantom:

  1. CBCT, study protocol
  2. MDCT (routine), protocol used in routine imaging at the authors’ institution
  3. MDCT (CTDI equivalent), protocol adapted to match the CBCT’s CT dose index volume (CTDIvol) values

Each of these three study protocols consisted of three exposure levels:

  • low”, suitable for pre-schoolers younger than 6 years
  • medium”, suitable for schoolers aged 6 to 12 years
  • high”, suitable for teenagers older than 12 years

In CBCT, the “high” protocol corresponded to the manufacturer-lowered pre-set for children. Prior to the dosimetric analyses, the study authors consensually agreed on the two other lowered CBCT exposure settings “medium” and “low”, with the aim of a barely sufficient image quality to securely make a diagnosis. The settings were based on initial experiences with phantoms, animal cadavers, and patients. The first MDCT protocol set (MDCT routine) consisted of exposure settings actually used in clinical routine at the authors’ institution. The second set [MDCT (CTDI equivalent)] was adapted to match the CTDIvol (16 cm phantom) values of the respective CBCT protocol, taking the underlying diverging CTDIvol phantoms in both machines into account. This exposure set was meant to show dose conformity between the modalities. All described protocols and exposure settings are summarized in Table 1.

thumbnail
Table 1. Exposure parameters for CBCT and MDCT protocols and phantoms.

https://doi.org/10.1371/journal.pone.0178747.t001

Lithium fluoride (LiF) thermoluminescence dosimeters TLD-100 square rods with diameters of 1×6 mm (Thermo Fisher Scientific Inc., Waltham, Massachusetts, USA) were used. TLDs were calibrated to the X-ray quality of both devices (CBCT and MDCT) individually. Dose results are based on calibration and correction factors determined at the Competence Center of Medical Physics and Radiation Protection, University Hospital of Graz, Austria. TLD glow-curves were read out on a Harshaw TLD Model 5500 reader with planchet heating system and WinREMS readout software (Thermo Fisher Scientific Inc., Waltham, Massachusetts, USA). Surface doses were measured in 6 circular positions around the examined joints. For each measurement a pair of TLDs was attached in the same way at every 1, 3, 5, 7, 9, and 11 o’clock position, seen in the direction of the extremity as shown in Fig 2. All dosimeters were consecutively irradiated 10 times per study protocol in order to ensure that a sufficient amount of radiation had been applied. Additionally, all protocols were repeated 3 times to prove the measurement test-retest reliability. After exposure TLDs were read out by a medical physicist with longstanding experience. A total of 322 (of 324) successful TLD readouts were performed, while 2 TLD readouts failed due to material fatigue.

thumbnail
Fig 2. Positioning of TLDs around the ankle and wrist.

Dosimeter pairs were attached to the phantoms in clockwise position at 1, 3, 5, 7, 9, and 11 o’clock as displayed. a) ankle phantom. b) wrist phantom.

https://doi.org/10.1371/journal.pone.0178747.g002

After the respective image acquisitions, axial slices with thicknesses of 1.4 millimeters were reconstructed relative to the examined body part. The reconstruction kernel was “Standard” without iterative reconstruction in CBCT, and FC30 “Bone sharp” plus “AIDR” iterative reconstruction in MDCT.

FIJI 1.49v [20], an ImageJ distribution (open source image processing software, http://rsbweb.nih.gov/ij/) was used to measure mean pixel values and noise of air, cancellous bone, and soft tissue. Ellipsoid regions of interest (ROI) were placed in matching axial slices and positions and the resulting Hounsfield units (HU) were assessed. These measurements were repeated in 10 different matching image positions across all studies. Maximum and mean HU values of cortical bone and the respective standard deviations were read out in all series too. This information was used to calculate a normalization factor (NF = maximum cortical bone intensity of single study / maximum cortical bone intensity of all studies) applied to contrast-to-noise ratios (CNR), with the aim to extenuate device specific effects of differences in HU display [28, 29], and of various kVp settings [30, 31]. NF was 1.0 for the study containing the highest maximum cortical bone pixel value, and lower for all other exposure levels (compare S2 Dataset, listing all objective and subjective image measurements). Combined mean image noise (of air, cancellous bone, and soft tissue), normalized CNR [(mean cortical bone – mean air) / SD air * NF] and signal-to-noise ratios (SNR = mean [material] / SD [material]) were calculated for every CBCT and MDCT protocol as parameters for objective image quality.

Subjective image quality was rated consensually on a five grade Likert scale (1 = excellent, 2 = good, 3 = average, 4 = fair, 5 = poor) by three radiologists with 4, 6 and 28 years of experience in musculoskeletal CT. The assessed parameters included overall quality, contrast, sharpness, noise, beam hardening, aliasing, and ring artifacts.

Collected data was imported and analyzed in SPSS Statistics Version 21 software (IBM Corp., Armonk, NY). Descriptive statistics were used to explore the measured values. Mean values were compared with independent samples t-tests. Pearson correlations were used to assess relations between dose and image quality parameters. Intraclass correlation [two-way mixed average measures, ICC(3,k)] coefficients and a Bland-Altman plot were calculated to prove dose measurements’ test-retest reliability. P values less than 0.05 were assumed to be statistically significant.

An ethics committee approval was not needed for this phantom study.

Results

In CBCT the mean surface dose was 3.0 ±1.9 mGy, averaged over low, medium, and high ankle and wrist protocols. Significantly higher surface doses were measured for MDCT (routine) with 3.9 ±1.1 (p<0.001). As expected, differences between the CBCT and the CTDI-matched MDCT (CTDI equivalent) protocol were not statistically significant (mean 3.0 ±1.9 mGy vs. 3.0 ±1.4 mGy, p = 0.903).

Significant linear correlations between surface doses and both CTDIvol (r = 0.957, p<0.001) and DLP (0.950, p<0.001) were found. Surface doses in CBCT were significantly lower compared to MDCT (routine) in the majority of the exposure levels apart from “high” in general (p = 0.633), “high” in the ankle phantom (p = 0.131), and overall in the ankle phantom (p = 0.053). Respective findings are summarized in Table 2.

thumbnail
Table 2. Independent samples t-test results between CBCT and MDCT (routine).

https://doi.org/10.1371/journal.pone.0178747.t002

In CBCT the posterior TLDs at 5 and 7 o’clock positions showed significantly lower surface doses than the anterior positions at 1 and 11 o’clock (mean 2.5 ±1.5 mGy vs. mean 3.5 ±2.0 mGy, p = 0.014), accounted for by the incomplete CBCT tube rotation of 210 degrees. This is graphically depicted in Fig 3c. In MDCT, no significant differences between the surface doses at the different dosimeter positions were detected. Mean values of all measured surface doses are listed in Table 3; respective raw data is available in S1 Dataset.

thumbnail
Table 3. Measured surface doses at the six positions around the ankle and wrist.

https://doi.org/10.1371/journal.pone.0178747.t003

thumbnail
Fig 3. Box plots displaying surface doses in mGy for both examined devices.

a) ankle and wrist, and b) for the different exposure settings “low”, “medium”, and “high”. c) Mean surface doses at different positions around the joints in CBCT and MDCT.

https://doi.org/10.1371/journal.pone.0178747.g003

The objectively measured image parameters were significantly in favor of CBCT (all p<0.001), even though the noise reduction algorithm of iterative reconstruction had been enabled in MDCT: Mean noise was 33.3 ±17.8 HU in CBCT and 63.2 ±15.6 HU in MDCT. Normalized CNR was 55.4 ±28.5 HU in CBCT, and 39.4 ±5.1 HU in MDCT. SNR was 28.2 ±15.1 HU in CBCT and 8.8 ±1.8 HU in MDCT. Noise, CNR and SNR are graphically depicted in Fig 4.

thumbnail
Fig 4. Objective image quality assessment.

CNR, SNR and noise in a) ankle and b) wrist phantom across all exposure settings.

https://doi.org/10.1371/journal.pone.0178747.g004

Subjective overall image quality decreased with lower exposure settings (R = -0.594, p = 0.009), and showed significant negative linear correlations with surface dose in both devices (R = -0.816, p = 0.048 in CBCT; and R = -0.693, p = 0.012 in MDCT). Overall image quality was not significantly different between both devices (CBCT 2.7 ±0.8 and MDCT 3.0 ±1.0 points, p = 0.456) and the study protocols [CBCT 2.7 ±0.8 and MDCT (routine) 2.3 ±0.5 points, p = 0.421]; neither was sharpness or contrast. Objectively measured and subjectively rated noise were significantly correlating (R = 0.809, p<0.001). Comparative examples of CBCT and MDCT images are given in Fig 5a and 5b. In contrast to CBCT, the raters did not detect any beam hardening artifacts in MDCT (example depicted in Fig 5c). On the other hand aliasing and noise were rated stronger in MDCT (p<0.001), exemplarily shown in Fig 5d. All subjective image ratings are listed in Table 4.

thumbnail
Fig 5. Examples of images quality at different dose levels and image artifacts.

a) Side-by-side comparison of matching slices in CBCT (left) and MDCT (right) ankle examinations rated with good image quality. b) Wrist phantom scanned in CBCT (left) and MDCT (right) rated poor image quality. c) Beam hardening artifacts in CBCT. d) Aliasing artifacts and image noise in MDCT.

https://doi.org/10.1371/journal.pone.0178747.g005

Excellent test-retest reliability was shown for the TLD surface dose measurements by an overall ICC(3,k) of 0.953 (p<0.001). Split up between the modalities, the ICC was 0.946 (p<0.001) in CBCT and 0.922 (p<0.001) in MDCT. The measurement repeatability is depicted in Fig 6.

thumbnail
Fig 6. A Bland-Altman plot shows test-retest reliability of all TLD surface dose measurements among each other.

Mean 0.003 (SD ±0.66) mGy shown as solid horizontal line, and upper and lower 1.96 SD intervals presented as dotted horizontal lines at 1.288 and -1.282 mGy respectively. All valid measurements from all six TLD positions were included in the plot.

https://doi.org/10.1371/journal.pone.0178747.g006

Discussion

In this study surface doses applied by an extremity CBCT and a MDCT scanner in pediatric ankle and wrist phantoms were assessed. Adapted pediatric imaging protocols with age-related exposure settings were used to compare both scanners as realistically as possible.

To the best of our knowledge, this is the first study assessing and systematically comparing extremity CBCT doses in pediatric ankle and wrist phantoms. In adults, exceptional dose-saving potentials of CBCT compared to MDCT have been published for the knee and ankle region by Koivisto et al. [24, 25]. In contrast, recent studies by Lang et al. and Neubauer et al. did not confirm these findings and the authors reported comparable doses of extremity CBCT and MDCT examinations, when taking image quality into account [20, 22]. The currently available clinical pediatric extremity CBCT studies report doses that usually are lower than MDCT [23, 32]. As a result, dose optimization and possible dose savings still remain subject of debate and warrant future studies, especially in the radio-susceptible pediatric population.

In our study, we explicitly did not try to assess effective radiation doses, because different factors and unknown variables influence their correct calculation in pediatric extremities [33]. Mainly, the distribution of radiosensitive red bone marrow and relatively radio-insensitive yellow bone marrow are known to vary at different ages and to decrease until adulthood [3436]. Additionally, inter-individual variety of marrow distribution and conversion may occur [37], rendering it impossible to predict the actual individual patient risk of an examination expressed by a specific effective dose value. Moreover, effective dose is believed not to be able to serve as a valid parameter of patient risk per se [3841], and stochastic radiation induced damage may be underestimated [42, 43]. Effective dose was not intended to be applied to patients or children [38, 44], and appropriate pediatric tissue weighting factors remain a subject of ongoing debate and research [4547]. In respect of the stated shortcomings in regard to effective dose in children, we have chosen to compare both modalities based on surface doses only.

In the current study image noise, normalized CNR, and SNR were examined as three basic, objectively measurable image quality parameters. In both CBCT and MDCT altering dose settings directly influenced the resulting image quality, as expected. Due to the missing pre-detector collimation in CBCT, scatter radiation additionally decreased the image quality through image artifacts [20, 22, 48, 49]. Therefore, scatter radiation is normally corrected by mathematical algorithms [50, 51], which may not equally correct for scatter induced artifacts in every situation. On the other hand, iterative reconstruction algorithms of the MDCT decrease image noise and alter the image impression [52, 53], which was not available or applied in CBCT. Due to the reasons given above, a direct comparison between different CBCT devices or CBCT and MDCT machines is further complicated regarding objective, and especially subjective image quality. We found quantitative image quality in favor of CBCT at comparative dose levels, whereas subjectively no significant difference was noted. However, alike Neubauer et al. and Lang et al. [20, 22] we detected more image-degrading beam hardening artifacts in CBCT.

In contrast to previous related studies by Koivisto et al. [24, 25] we have chosen TLDs over MOSFET dosimeters. The main reason was the better availability of, and longer experience with TLDs in our institution. Prior studies have not shown significant differences between both methods of dose measurements [5456].

The following study limitations need to be mentioned: The most important limitation is the fact that phantoms were used instead of pediatric patients or cadaveric specimens, which we had favored due to reasons of availability, longevity and consistency. These phantoms of the ankle and wrist may not be completely anthropomorphic with different material compositions and X-ray absorption. Another important limitation is the choice of appropriate and matchable exposure settings. The general differences between the modalities complicate their comparison. The exposure parameters in the analyzed MDCT had been optimized and validated in clinical routine for many years, and this amount of dose titration was certainly not possible in the CBCT machine. We therefore decided to run two sets of protocols in MDCT, one actually used in clinical routine, and one that provided matching CTDIvol values with the CBCT scanner.

Conclusions

Compared to an optimized MDCT, CBCT surface doses were significantly lower in a majority of the examined study protocols with superior objective, but only equivalent subjective image quality. Even though the previously reported extraordinary CBCT dose savings could not be substantiated in the examined pediatric ankle and wrist phantoms, the study results raise hope that CBCT could be a valuable low-dose alternative in pediatric extremity trauma imaging.

Supporting information

S1 Dataset. Raw data of surface dose measurements.

https://doi.org/10.1371/journal.pone.0178747.s001

(XLSX)

S2 Dataset. Raw data of objective and subjective image quality measurements.

https://doi.org/10.1371/journal.pone.0178747.s002

(XLSX)

Author Contributions

  1. Conceptualization: ST RM EN G. Singer G. Stücklschweiger ES.
  2. Data curation: ST RM EN GA MR G. Singer G. Stücklschweiger ES HG.
  3. Formal analysis: ST G. Stücklschweiger ES HG.
  4. Funding acquisition: ES.
  5. Investigation: ST RM EN GA MR.
  6. Methodology: G. Stücklschweiger HG ES.
  7. Project administration: TS ES G. Singer.
  8. Resources: G. Singer G. Stücklschweiger ES.
  9. Supervision: ES MR G. Singer.
  10. Validation: ST ES.
  11. Visualization: ST.
  12. Writing – original draft: ST.
  13. Writing – review & editing: ST RM EN GA MR G. Singer G. Stücklschweiger ES HG.

References

  1. 1. Siewerdsen JH, Jaffray DA. Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med Phys. 2001;28(2):220–31. pmid:11243347.
  2. 2. Shaw CC. Cone Beam Computed Tomography. Philadelphia: CRC Press; 2014. 1 online resource (266 pages) p.
  3. 3. Jaju PP. Cone beam computed tomography: a clinician's guide to 3D imaging. New Delhi: Jaypee Brothers Medical Publishers (P) Ltd; 2015. xvii, 113 pages: illustrations; 28 cm. p.
  4. 4. Shaw CC. Cone beam computed tomography. Boca Raton: CRC Press, Taylor & Francis Group; 2014. xviii, 251 pages, 4 unnumbered pages of plates: illustrations (some colour); 29 cm. p.
  5. 5. Collan L, Kankare JA, Mattila K. The biomechanics of the first metatarsal bone in hallux valgus: a preliminary study utilizing a weight bearing extremity CT. Foot Ankle Surg. 2013;19(3):155–61. pmid:23830162.
  6. 6. Tuominen EK, Kankare J, Koskinen SK, Mattila KT. Weight-bearing CT imaging of the lower extremity. AJR American journal of roentgenology. 2013;200(1):146–8. pmid:23255755.
  7. 7. Hirschmann A, Pfirrmann CW, Klammer G, Espinosa N, Buck FM. Upright cone CT of the hindfoot: comparison of the non-weight-bearing with the upright weight-bearing position. European radiology. 2014;24(3):553–8. pmid:24071992.
  8. 8. Colin F, Horn Lang T, Zwicky L, Hintermann B, Knupp M. Subtalar joint configuration on weightbearing CT scan. Foot Ankle Int. 2014;35(10):1057–62. pmid:25015393.
  9. 9. Halonen KS, Mononen ME, Jurvelin JS, Toyras J, Salo J, Korhonen RK. Deformation of articular cartilage during static loading of a knee joint--experimental and finite element analysis. Journal of biomechanics. 2014;47(10):2467–74. pmid:24813824.
  10. 10. Carrino JA, Al Muhit A, Zbijewski W, Thawait GK, Stayman JW, Packard N, et al. Dedicated cone-beam CT system for extremity imaging. Radiology. 2014;270(3):816–24. pmid:24475803
  11. 11. Thawait GK, Demehri S, AlMuhit A, Zbijweski W, Yorkston J, Del Grande F, et al. Extremity cone-beam CT for evaluation of medial tibiofemoral osteoarthritis: Initial experience in imaging of the weight-bearing and non-weight-bearing knee. European journal of radiology. 2015;84(12):2564–70. pmid:26388464.
  12. 12. Weber MT, Stratz N, Fleiner J, Schulze D, Hannig C. Possibilities and limits of imaging endodontic structures with CBCT. Swiss Dent J. 2015;125(3):293–311. pmid:26168687.
  13. 13. Hol C, Hellen-Halme K, Torgersen G, Nilsson M, Moystad A. How do dentists use CBCT in dental clinics? A Norwegian nationwide survey. Acta Odontol Scand. 2015;73(3):195–201. pmid:25415368.
  14. 14. Machado GL. CBCT imaging—A boon to orthodontics. Saudi Dent J. 2015;27(1):12–21. pmid:25544810
  15. 15. Kapila SD, Nervina JM. CBCT in orthodontics: assessment of treatment outcomes and indications for its use. Dentomaxillofac Radiol. 2015;44(1):20140282. pmid:25358833
  16. 16. Kiljunen T, Kaasalainen T, Suomalainen A, Kortesniemi M. Dental cone beam CT: A review. Phys Med. 2015;31(8):844–60. pmid:26481816.
  17. 17. Agrawal JM, Agrawal MS, Nanjannawar LG, Parushetti AD. CBCT in orthodontics: the wave of future. J Contemp Dent Pract. 2013;14(1):153–7. pmid:23579915.
  18. 18. Alamri HM, Sadrameli M, Alshalhoob MA, Sadrameli M, Alshehri MA. Applications of CBCT in dental practice: a review of the literature. Gen Dent. 2012;60(5):390–400; quiz 1–2. pmid:23032226.
  19. 19. Faccioli N, Foti G, Barillari M, Atzei A, Mucelli RP. Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography. Skeletal radiology. 2010;39(11):1087–95. pmid:20224906.
  20. 20. Lang H, Neubauer J, Fritz B, Spira EM, Strube J, Langer M, et al. A retrospective, semi-quantitative image quality analysis of cone beam computed tomography (CBCT) and MSCT in the diagnosis of distal radius fractures. European radiology. 2016. pmid:27003138.
  21. 21. Huang AJ, Chang CY, Thomas BJ, MacMahon PJ, Palmer WE. Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects. Skeletal radiology. 2015;44(6):797–809. pmid:25652734.
  22. 22. Neubauer J, Neubauer C, Gerstmair A, Krauss T, Reising K, Zajonc H, et al. Comparison of the Radiation Dose from Cone Beam Computed Tomography and Multidetector Computed Tomography in Examinations of the Hand. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2016. pmid:26981913.
  23. 23. Pugmire BS, Shailam R, Sagar P, Liu B, Li X, Palmer WE, et al. Initial Clinical Experience With Extremity Cone-Beam CT of the Foot and Ankle in Pediatric Patients. AJR American journal of roentgenology. 2016;206(2):431–5. pmid:26797374.
  24. 24. Koivisto J, Kiljunen T, Kadesjo N, Shi XQ, Wolff J. Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region. J Foot Ankle Res. 2015;8:8. pmid:25788986
  25. 25. Koivisto J, Kiljunen T, Wolff J, Kortesniemi M. Assessment of effective radiation dose of an extremity CBCT, MSCT and conventional X ray for knee area using MOSFET dosemeters. Radiation protection dosimetry. 2013;157(4):515–24. pmid:23825221.
  26. 26. Burch WM. Thermoluminescence: a simple clinical dosimeter. Australas Radiol. 1967;11(3):280–5. pmid:6063522.
  27. 27. Kagaku Kyoto. Pediatric Whole Body Phantom "PBU-70" Kitanekoya-cho Fushimi-ku Kyoto, Japan: Kyoto Kagaku Co., Ltd. 2016 [2016-05-07]. https://kyotokagaku.com/products/detail03/ph-2c.html.
  28. 28. Varshowsaz M, Goorang S, Ehsani S, Azizi Z, Rahimian S. Comparison of Tissue Density in Hounsfield Units in Computed Tomography and Cone Beam Computed Tomography. J Dent (Tehran). 2016;13(2):108–15. pmid:27928239
  29. 29. Silva IM, Freitas DQ, Ambrosano GM, Boscolo FN, Almeida SM. Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography. Braz Oral Res. 2012;26(6):550–6. pmid:23184166.
  30. 30. Hardie AD, Perry JD, Bradshaw ML, Picard MM. Increased renal cyst density on abdominal CT at 100-kVp compared with 120-kVp: a preliminary evaluation. Clin Imaging. 2015;39(4):642–5. pmid:25794849.
  31. 31. Cho ES, Chung TS, Oh DK, Choi HS, Suh SH, Lee HK, et al. Cerebral computed tomography angiography using a low tube voltage (80 kVp) and a moderate concentration of iodine contrast material: a quantitative and qualitative comparison with conventional computed tomography angiography. Investigative radiology. 2012;47(2):142–7. Epub 2011/11/23. pmid:22104960.
  32. 32. Ludlow JB, Ivanovic M. Weightbearing CBCT, MDCT, and 2D imaging dosimetry of the foot and ankle. International Journal of Diagnostic Imaging. 2014;1(2):1–9.
  33. 33. Strauss KJ. Dose indices: everybody wants a number. Pediatric radiology. 2014;44 Suppl 3:450–9. pmid:25304704.
  34. 34. Taccone A, Oddone M, Dell'Acqua AD, Occhi M, Ciccone MA. MRI "road-map" of normal age-related bone marrow. II. Thorax, pelvis and extremities. Pediatric radiology. 1995;25(8):596–606. pmid:8570312.
  35. 35. Zawin JK, Jaramillo D. Conversion of bone marrow in the humerus, sternum, and clavicle: changes with age on MR images. Radiology. 1993;188(1):159–64. pmid:8511291.
  36. 36. Waitches G, Zawin JK, Poznanski AK. Sequence and rate of bone marrow conversion in the femora of children as seen on MR imaging: are accepted standards accurate? AJR American journal of roentgenology. 1994;162(6):1399–406. pmid:8192007.
  37. 37. Guillerman RP. Marrow: red, yellow and bad. Pediatric radiology. 2013;43 Suppl 1:S181–92. pmid:23478934.
  38. 38. Brenner DJ. Effective dose: a flawed concept that could and should be replaced. The British journal of radiology. 2008;81(967):521–3. pmid:18443016.
  39. 39. Li X, Samei E, Segars WP, Sturgeon GM, Colsher JG, Frush DP. Patient-specific radiation dose and cancer risk for pediatric chest CT. Radiology. 2011;259(3):862–74. pmid:21467251
  40. 40. Martin CJ. Effective dose: how should it be applied to medical exposures? The British journal of radiology. 2007;80(956):639–47. pmid:17646189.
  41. 41. Harrison J, Day P. Radiation doses and risks from internal emitters. J Radiol Prot. 2008;28(2):137–59. pmid:18495991.
  42. 42. Lee C, Flynn MJ, Judy PF, Cody DD, Bolch WE, Kruger RL. Body Size-Specific Organ and Effective Doses of Chest CT Screening Examinations of the National Lung Screening Trial. AJR American journal of roentgenology. 2017;208(5):1082–8. pmid:28267354.
  43. 43. Newman B, Ganguly A, Kim JE, Robinson T. Comparison of different methods of calculating CT radiation effective dose in children. AJR American journal of roentgenology. 2012;199(2):W232–9. pmid:22826426.
  44. 44. International Commission on Radiological Protection. ICRP publication 26: Recommendations of the International Commission on Radiological Protection: (adopted January 17, 1977). Oxford; New York: Published for the International Commission on Radiological Protection by Pergammon Press; 1977. 53 p.
  45. 45. Brady Z, Cain TM, Johnston PN. Comparison of organ dosimetry methods and effective dose calculation methods for paediatric CT. Australasian physical & engineering sciences in medicine / supported by the Australasian College of Physical Scientists in Medicine and the Australasian Association of Physical Sciences in Medicine. 2012;35(2):117–34. Epub 2012/04/11. pmid:22492218.
  46. 46. Seidenbusch MC, Harder D, Regulla DF, Schneider K. Conversion factors for determining organ doses received by paediatric patients in high-resolution single slice computed tomography with narrow collimation. Z Med Phys. 2014;24(2):123–37. Epub 2014/03/13. pmid:24630933.
  47. 47. Lee C, Kim KP, Bolch WE, Moroz BE, Folio L. NCICT: a computational solution to estimate organ doses for pediatric and adult patients undergoing CT scans. J Radiol Prot. 2015;35(4):891–909. Epub 2015/11/26. pmid:26609995.
  48. 48. Nagarajappa AK, Dwivedi N, Tiwari R. Artifacts: The downturn of CBCT image. J Int Soc Prev Community Dent. 2015;5(6):440–5. pmid:26759795
  49. 49. Makins SR. Artifacts interfering with interpretation of cone beam computed tomography images. Dent Clin North Am. 2014;58(3):485–95. pmid:24993920.
  50. 50. Jin JY, Ren L, Liu Q, Kim J, Wen N, Guan H, et al. Combining scatter reduction and correction to improve image quality in cone-beam computed tomography (CBCT). Med Phys. 2010;37(11):5634–44. pmid:21158275.
  51. 51. Li J, Yao W, Xiao Y, Yu Y. Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm. Journal of applied clinical medical physics / American College of Medical Physics. 2013;14(6):4346. pmid:24257276.
  52. 52. von Falck C, Bratanova V, Rodt T, Meyer B, Waldeck S, Wacker F, et al. Influence of sinogram affirmed iterative reconstruction of CT data on image noise characteristics and low-contrast detectability: an objective approach. PLoS One. 2013;8(2):e56875. pmid:23468886
  53. 53. Millon D, Vlassenbroek A, Van Maanen AG, Cambier SE, Coche EE. Low contrast detectability and spatial resolution with model-based Iterative reconstructions of MDCT images: a phantom and cadaveric study. European radiology. 2017;27(3):927–37. pmid:27300195.
  54. 54. Kinhikar RA, Murthy V, Goel V, Tambe CM, Dhote DS, Deshpande DD. Skin dose measurements using MOSFET and TLD for head and neck patients treated with tomotherapy. Appl Radiat Isot. 2009;67(9):1683–5. pmid:19369084.
  55. 55. Koivisto J, Schulze D, Wolff J, Rottke D. Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study. Dentomaxillofac Radiol. 2014;43(8):20140202. pmid:25143020
  56. 56. De Lin M, Toncheva G, Nguyen G, Kim S, Anderson-Evans C, Johnson GA, et al. Application of MOSFET detectors for dosimetry in small animal radiography using short exposure times. Radiation research. 2008;170(2):260–3. pmid:18666818