Concerning the matching of magnetic susceptibility differences for the compensation of background gradients in anisotropic diffusion fibre phantoms

  • Ezequiel Farrher ,

    Contributed equally to this work with: Ezequiel Farrher, Johannes Lindemeyer

    e.farrher@fz-juelich.de

    Affiliation Institute of Neuroscience and Medicine – 4, Forschungszentrum Jülich GmbH, Jülich, Germany

    ORCID http://orcid.org/0000-0003-2902-8667

  • Johannes Lindemeyer ,

    Contributed equally to this work with: Ezequiel Farrher, Johannes Lindemeyer

    Affiliation Institute of Neuroscience and Medicine – 4, Forschungszentrum Jülich GmbH, Jülich, Germany

  • Farida Grinberg,

    Affiliations Institute of Neuroscience and Medicine – 4, Forschungszentrum Jülich GmbH, Jülich, Germany, Department of Neurology, Faculty of Medicine, RWTH Aachen University, Aachen, Germany

  • Ana-Maria Oros-Peusquens,

    Affiliation Institute of Neuroscience and Medicine – 4, Forschungszentrum Jülich GmbH, Jülich, Germany

  • N. Jon Shah

    Affiliations Institute of Neuroscience and Medicine – 4, Forschungszentrum Jülich GmbH, Jülich, Germany, Department of Neurology, Faculty of Medicine, RWTH Aachen University, Aachen, Germany, JARA – BRAIN – Translational Medicine, RWTH Aachen University, Aachen, Germany, Institute of Neuroscience and Medicine – 11, Forschungszentrum Jülich GmbH, Jülich, Germany, Department of Electrical and Computer Systems Engineering, and Monash Biomedical Imaging, School of Psychological Sciences, Monash University, Melbourne, Victoria, Australia

Concerning the matching of magnetic susceptibility differences for the compensation of background gradients in anisotropic diffusion fibre phantoms

  • Ezequiel Farrher, 
  • Johannes Lindemeyer, 
  • Farida Grinberg, 
  • Ana-Maria Oros-Peusquens, 
  • N. Jon Shah
PLOS
x

Abstract

Artificial, anisotropic fibre phantoms are nowadays increasingly used in the field of diffusion-weighted MRI. Such phantoms represent useful tools for, among others, the calibration of pulse sequences and validation of diffusion models since they can mimic well-known structural features of brain tissue on the one hand, but exhibit a reduced complexity, on the other. Among all materials, polyethylene fibres have been widely used due to their excellent properties regarding the restriction of water diffusion and surface relaxation properties. Yet the magnetic susceptibility of polyethylene can be distinctly lower than that of distilled water. This difference produces strong microscopic, background field gradients in the vicinity of fibre bundles which are not parallel to the static magnetic field. This, in turn, modulates the MRI signal behaviour. In the present work we investigate an approach to reduce the susceptibility-induced background gradients via reducing the heterogeneity in the internal magnetic susceptibility. An aqueous solution of magnesium chloride hexahydrate (MgCl2·6H2O) is used for this purpose. Its performance is demonstrated in dedicated anisotropic fibre phantoms with different geometrical configurations.

Introduction

Diffusion phantoms are highly demanded devices for a broad set of applications in diffusion-weighted (DW) MRI. These applications include the design and calibration of DW MRI experiments [15], validation and optimisation of high angular resolution diffusion imaging (HARDI) methods [610], tractography algorithms [1113], validation of diffusion models [14,15] and results from multi-centre studies [16,17]. Artificial, anisotropic diffusion phantoms found in the literature are made of a large variety of materials, e.g., rayon [6,18], hemp [18], polyester [1,5], acrylic fibres [13], glass capillaries [19], plastic capillaries [3,9], hollow fibres produced by co-electrospinning [2,12] and polyamide fibres [4,10]. One of the most frequently used materials is Dyneema® [14,15,18,20,21] (Dyneema® SK75 dtex1760, DSM, Geleen, The Netherlands). These rod-like, hydrophobic, polyethylene fibres have several advantages concerning physical properties compared to other synthetic materials such as nylon and fibre glass, particularly its lower surface relaxivity that results in overall longer relaxation times [20]. This enables diffusion measurements to be performed with higher signal-to-noise ratio (SNR) or longer echo times, which is especially important in the case of strong diffusion weightings and time-dependent studies. Mechanical flexibility is also an advantage when compared to, for instance, rigid capillaries, given that this flexibility facilitates manufacturing of phantoms with a wide range of geometrical configurations. Another advantage of Dyneema® fibres is that they are available with radii below 10 microns, which makes them well-suited for producing phantoms with sufficiently dense diffusion barriers, comparable to that of cellular structures.

Magnetostatic theory predicts that two media of different magnetic susceptibility, χ, placed in an external magnetic field introduce distortions to the field distribution in the vicinity of their interface [22]. These distortions, i.e., magnetic field gradients, depend on the difference in magnetic susceptibility, Δχ = χ2 - χ1, as well as on the geometric configuration of the media (curvature radii, interface orientation relative to the magnetic field, etc.). Thus, microscopic, susceptibility-induced magnetic field gradients (further on referred to as background gradients) are ubiquitous in NMR and MRI experiments involving magnetically heterogeneous media such as, among others, porous materials [2325] and biological tissues [2629]. In particular, background gradients appear as a common feature in anisotropic diffusion phantoms where the magnetic susceptibility of the diffusing liquid is a priori different from that of the diffusion-restricting material [4,20,21,30].

As initially demonstrated by Stejskal and Tanner [31], the DW NMR signal in an infinite medium as measured using conventional pulsed gradient spin-echo (PGSE) sequences in the presence of constant background gradients, is influenced by three distinct terms: i) a term dependent on the externally applied field gradient g, ii) a term purely dependent on the background gradients g0, and iii) an interference, cross-term resulting from the interaction between the background and the externally applied gradients. In SE experiments without externally applied field gradients, background gradients act, in the interplay with the diffusion process, as an extra mechanism reducing the measured transverse relaxation time, T2 [24]. This feature has been successfully used to provide biomarkers of trabecular bone density [27,32,33] or muscle microstructure [34], for example. Recently, computer simulations were used [35] to study the orientation-dependence of T2 due to background gradients in an array of cylinders under different configurations of Δχ, cylinder packing density and order.

In conventional PGSE, experiments are carried out by measuring the echo attenuation for several strengths of the diffusion-sensitising field gradient g, while keeping the sequence timing constant [36]. As a consequence, the signal attenuation due to g0 becomes merely a multiplicative factor. Special care must be taken though when the strength of g0 is so large that the signal-to-noise ratio (SNR) becomes critically low for a given echo-time. It has been shown that low values of the SNR in the magnitude signal can lead to a strong underestimation of the apparent diffusion coefficient [37]. On the other hand, the cross-term can lead to more complex implications. It is a priori spatially dependent, and therefore it may be cancelled in some regions whereas it may be greatly enhanced in others, even for small g0 strengths. The effect of the cross-term has previously been measured and empirically modelled by Zhong et al. [38,39] in artificial samples, post mortem and in rat tissue in vivo. Their results demonstrate an increase in the echo amplitude for increasing g0 strengths. This was interpreted as an effect of the regions where g was compensated by g0. The same empirical model was later used by Clark et al. [28] for the analysis of in vivo human brain data, showing similar results. In this context, the same increase in the echo amplitude due to background gradients introduced by the microvessels in the human brain was predicted by Kiselev [40], employing a more realistic model for brain microvasculature.

While background gradients can be exploited to obtain additional structural information on tissue microstructure and orientation [4145], they can lead to misinterpretation of the results, if they are not properly taken into account in the context of most DW MRI measurements [28,3840]. Therefore, in many DW MRI experiments, background gradients represent unwanted entities which need to be cancelled. A reduction of the effect of background gradients has been achieved via the design of advanced pulse sequences, as an alternative to the conventional Stejskal-Tanner PGSE sequence [31]. Some of the sequences attempt to reduce the effect of the g0 term [46,47], whereas others are meant to suppress the effect of the cross-term [46,48]. Nevertheless, a strong underlying assumption in these methods is that the background gradients change slowly throughout space, meaning that each diffusing molecule experiences the same constant background gradient during the observation time. Ref. [46] provides a comprehensive review on this topic.

A different approach for the suppression of background gradients is susceptibility matching. Its aim is to reduce, or ideally entirely eliminate, the difference in magnetic susceptibilities within the sample. This approach has been previously used in NMR spectroscopy studies [49] to reduce the line broadening in NMR spectra. More recently susceptibility matching was carried out in a polyamide fibre phantom. In their work, Laun et al. [4] used a solution of 83 grams of sodium chloride (NaCl) per kilogram of distilled water, in order to match the susceptibility of the solution to that of the polyamide fibres, which is 0.37 ppm more diamagnetic than distilled water.

The aim of this work is to design a strategy to suppress unwanted background gradients by matching the susceptibility of the liquid to that of the fibre material. In order to achieve this, the effects of background gradients on diffusion metrics in anisotropic phantoms constructed with Dyneema® fibres and different fibre configurations are assessed. We investigate the influence of background gradients on diffusion tensor imaging (DTI) invariant metrics, as well as on two methods for the analysis of HARDI data, namely constrained spherical deconvolution (CSD) [50] and q-ball imaging (QBI) [51]. Three types of dedicated phantoms were used for this purpose: i) bulk phantoms, i.e., test vials containing the solution of interest embedded in a cylindrical phantom; ii) parallel-fibre phantoms, i.e., showing single-modal diffusion profiles; iii) a phantom with two fibre populations crossing at the right angle, i.e., showing multi-modal diffusion profiles.

Theory

The PGSE signal attenuation for an infinite, homogeneous and anisotropic medium with uniform background gradients g0 = g0n0, in the presence of a diffusion sensitising, pulsed field gradient g = gn, is given by [24,36] (1) where S0 is the DW signal in the absence of diffusion weighting, D is the symmetric, positive-definite diffusion tensor, and n and n0 are the unit vectors along the diffusion sensitising and background gradients, respectively. The b-value due to the applied gradient is given by , whereas the b-value related to the cross-term can be shown to be , with g(t) and g0(t) denoting the gradients waveforms, TE the echo time and γ the gyromagnetic ratio. For the conventional Stejskal-Tanner PGSE sequence b = γ2g2δ2(Δ − δ/3) and bc = γ2gg0δ(Δ − δ/3)[TE − (Δ − δ)/2], where δ and Δ are the time duration and separation of the field gradient pulses, respectively [40]. Eq (1) corresponds to the conventional case in which the DW signal is acquired for several strengths of the field gradient while keeping the sequence timing constant. Thus, S0 includes a diffusion weighting term that is purely dependent on the background gradients, which effectively acts as an extra mechanism inducing transverse relaxation.

In the framework of DTI analysis, estimation of tensor elements is performed by regressing (2) to the measured DW signal. In Eq (2) the apparent diffusion tensor, Dapp, is introduced to emphasize the possible bias in D arising as a consequence of neglecting the cross term and the dependence on the pulse sequence and timing parameters employed. It is worth noticing that the diffusion weighting term related to background gradients only will not significantly bias the estimation, provided that the SNR is high enough. In case S0 is close to the noise floor, special care must be taken in order to avoid bias in the tensor elements. Conventionally, regression of Eq (2) to the experimental data is performed via a least-squares minimisation approach. However, for low SNR experiments the maximum likelihood estimator is preferred in order to ensure unbiased estimation [52].

System with cylindrical geometry

For a system of perfectly aligned cylinders, the background gradients only have components in the radial direction, i.e., perpendicular to the axis of symmetry [40]. Therefore, the DW signals acquired utilising gradient directions parallel (n||) and perpendicular (n) to the axis of symmetry can be calculated using Eq (1), yielding (3) where D|| and D denote the diffusivities in the direction parallel and perpendicular to the axis of symmetry, respectively. While the estimation of D|| using Eq (2) would remain unbiased, the dot product can take positive (parallel), negative (anti-parallel) or zero values. In such systems, this term may also vary throughout space. More importantly, for densely packed systems the gradient may be non-linear in the length-scale of the molecular root mean square displacement. Hence, the estimation of D via Eq (2) can potentially be biased.

Material and methods

Selection of the appropriate compound

Pure water is diamagnetic with a magnetic susceptibility of about χwater = -9.05ppm [53]. The susceptibility values of typical compounds range from strong paramagnetism (positive values) to comparably small diamagnetism (negative values). In particular, materials soluble in water with a magnetic susceptibility lower than that of water are rare. The susceptibility of the fibre material Dyneema® is approximately -10ppm [20], which is about 1ppm below water. Promising candidates for lowering the susceptibility value of aqueous solutions are salts since they are typically good solutes in water and many of them are strongly diamagnetic. Unfortunately, numerous salts also exhibit high electrical conductivity [54]. Hence compounds such as NaCl or KCl are poor candidates for approaching high susceptibility differences due to their high electrical conductivity and their consequently high radiofrequency absorption. Magnesium Chloride (MgCl2) is diamagnetic with a susceptibility of χ = 4π · (ρ/M) · χm(cgs) = −14.6ppm, χm(cgs) = -47.4cm3/mol [55], where ρ is the density and M represents the molar mass. It is also highly soluble in water with up to 552 g/l at 20°C [55] and its conductivity at a mass percentage of 25% is roughly half that of e.g. NaCl [54]. The hydrated form, magnesium chloride hexahydrate (MgCl2·6H2O) can be easily mixed with water. Accurate values of the magnetic susceptibility of different MgCl2·6H2O concentrations in distilled water are to our knowledge not reported in literature.

Bulk experiments

All MRI experiments were carried out in a whole-body 3T Siemens MAGNETOM Trio scanner (Siemens Medical Systems, Erlangen, Germany). The body coil was used for radiofrequency transmission and a 12-channel phased-array head coil was used for signal reception. All experiments were performed at room temperature. The gradient system provided a maximal gradient strength of 40 mT/m.

A set of 9 solutions was prepared with increasing MgCl2·6H2O concentrations, cv (v = 0,…,8), in distilled water in order to assess bulk NMR properties (Table 1). Test vials were filled with these solutions and embedded, one at a time, into a cylindrical polyethylene terephthalate (PET) plastic phantom (10 cm diameter, 14 cm height) filled with distilled water (Fig 1a).

thumbnail
Fig 1. Schematic representation of the three phantom types used in this study.

(a) Bulk phantoms, consisting of a plastic vial containing the concentration of interest, embedded in a cylindrical PET phantom filled with distilled water. Field distortions around the vial where used to assess the magnetic susceptibility. (b) Parallel-fibres phantoms containing concentrations c0, c2, c4, c6 and c8 in the interstitial space. Shrinking tubes were used to compress the fibres and keep them parallel. (c) Crossing-fibre phantom, with fibres wound around a Perspex platform and crossing at right angle.

https://doi.org/10.1371/journal.pone.0176192.g001

Magnetic susceptibility is assessed via the field distortions generated by the cylindrical tube in the surrounding water as in quantitative susceptibility imaging (QSM, [56,57]). The strongest distortions can be achieved with the cylinder orientation perpendicular to the static magnetic field [58]. The field distribution inside the phantom is estimated by linear regression on the phase data acquired with a multiple-echo, gradient-echo (GRE) sequence at 1mm isotropic resolution; matrix-size, 128×128×160; flip-angle, 14°; TE = 3ms (8 echoes, ΔTE = 4ms; bandwidth, BW = 501 Hz/pixel); repetition time, TR = 60ms; using monopolar readout. Automated threshold- and morphology-based segmentation is applied to the signal magnitude in order to generate masks for the vial content, mv, and the water volume, mw. An evaluation area surrounding the vial, me, is defined by expanding the vial mask. Our in-house software, MUBAFIRE [59], is used to correct for field distortions originating from sources outside of the phantom. The susceptibility distribution χv inside the vial is assumed to be constant as the liquid is homogeneous. Furthermore, setting χ to 0 within the water mask defines χwater as the reference offset. The difference between the measured field and the one generated by dipole convolution [60] of χv within a cylindrical water mask, me, around the test vial is evaluated (Fig 2) and minimised for χv: (4) where d is the appropriate dipole kernel [61]. The internal stray field of the vial can bias the background field correction. In order to compensate for this effect, background correction and susceptibility estimation are iterated three times. In each iteration, the estimated vial field of the previous iteration is subtracted during background correction. The whole procedure is illustrated in Fig 2. The accuracy of the susceptibility estimate is indicated by the relative difference between background-corrected field and the field generated by the determined susceptibility χv, evaluated within me. Error values are defined as the mean-corrected field difference divided by the mean-corrected measured field.

thumbnail
Fig 2. Data processing workflow for the assessment of the bulk magnetic susceptibility values.

Starting from the top left, the raw phase is spatially unwrapped, the echoes are realigned and the field map is estimated by linear regression; background fields are removed using MUBAFIRE and a reduced ROI for evaluation is selected; setting a constant susceptibility value inside of the vial volume, the difference between the hereby generated and the measured field distortion is minimised; the process is iterated three times in order to take the estimated vial field into account in the background correction.

https://doi.org/10.1371/journal.pone.0176192.g002

For the measurement of the bulk T1 relaxation times all vials were placed in the so-called “revolver phantom”. Measurements were performed with the TAPIR sequence [62]. Protocol parameters were: TR = 25 ms; flip-angle, 20°, inversion-times, 20 ≤ TI ≤ 4180 ms; BW = 320 Hz/pixel; voxel size, 1×1×4 mm3; matrix-size, 200×200×1. Data were processed using in-house Matlab scripts (Matlab 2015a, The MathWorks, Natick, MA, USA).

The bulk T2 and proton density (PD) for all vials are estimated with the help of a 2D spin-echo multi-contrast sequence provided by the manufacturer, using the following protocol parameters: 32 contrasts with an inter-echo time spacing of 50 ms; TR = 104 ms; number of averages, AVGs = 3; BW = 781 Hz/pixel; voxel size, 2×2×10 mm3 and matrix-size, 96×128×1. The echo attenuation, S(TE), is assumed to be monoexponential (5) where S0 is the signal at TE = 0. S0 and T2 are estimated voxel-wise via non-linear least-squares using the Nelder-Mead algorithm with in-house Matlab scripts. Finally the PD for each concentration cv is evaluated as PDv = S0,v/S0,0, with S0,v being the corresponding signal at TE = 0 [14,21].

Measurement of the bulk diffusion coefficient, D, for all concentrations cv, was carried out using a 2D twice-refocused SE (TRSE) EPI sequence with bipolar diffusion weighting gradients provided by the manufacturer [63]. Protocol parameters were: TR/TE = 8000/112 ms; AVGs = 16; BW = 1628 Hz/pixel; voxel-size, 2×2×10 mm3; matrix-size, 96×128×1; 16 b-values, [0:0.2:3.0] ms/μm2 along a single gradient direction; GRAPPA acceleration factor 2, with 24 reference lines. All DW MRI experiments in this work are performed such that different b-values are achieved by varying the gradient strength g while keeping the timing parameters δ and Δ constant. Considering the isotropic diffusion of this case, S0 and D are estimated voxel-wise via non-linear maximisation of the log-likelihood function as described in the section Maximum likelihood estimation of diffusion parameters.

Parallel-fibre phantom

All fibre phantoms were constructed using Dyneema® fibres (rod-like fibres with a radius of approximately 8 μm) [14,20,21]. Five parallel-fibre phantoms were built using bundles of Dyneema® fibres aligned in parallel. Each bundle was placed inside a shrinking tube approximately 5 centimetres wide [14,20] and afterwards heated to shrink it to a final diameter of about 2.5 centimetres. The fibre density for each phantom was 0.64 ± 0.02. Each phantom was then embedded in plastic vials that were filled with MgCl2·6H2O concentrations c0, c2, c4, c6 and c8 (Table 1), in such a way that the interstitial space between the fibres was filled via capillary force. Finally, all vials were assembled in a cylindrical PET plastic container (10 cm diameter, 14 cm height) filled with distilled water for the subsequent MRI experiments (Fig 1b).

Experiments were carried out for five different orientations of the fibre bundles with respect to the static field B0 at θ = 0°, 22.5°, 45°, 67.5° and 90°. The TRSE sequence was used. Protocol parameters were: TR/TE = 4200/82 ms; AVGs = 10; BW = 1563 Hz/pixel; voxel-size, 2×2×2 mm3; matrix-size, 80×128×36; 5 b-values, [0:0.25:1.0] ms/μm2 along 30 gradient directions; GRAPPA acceleration factor 2, with 24 reference lines; phase-encoding direction right-to-left. An extra non-DW volume with opposite phase-encoding direction was acquired for the correction of EPI distortions.

Eddy-current and EPI distortion correction is applied to all volumes with the help of the EDDY tool available in FSL [6466]. Subsequently, the diffusion weighting directions are reoriented according to the transformation matrices obtained in the former step [67]. Spatial smoothing with a Gaussian convolution kernel (full-width-half-maximum 2.3 mm) is applied to the images. Given the anisotropic nature of diffusion in this case, S0 and the apparent diffusion tensor Dapp are estimated as described in the section Maximum likelihood estimation of diffusion parameters.

Crossing-fibre phantom

The crossing-fibre phantom was built by tightly winding the fibres around a Perspex support (Plexiglas®) in order to keep the geometrical shape (Fig 1c). Several layers of fibres were stacked in perpendicularly alternating directions in such a way that the resulting thickness was approximately 10 mm [21]. The fibre density for this phantom was 0.68 ± 0.02. The whole setup was immersed in a PET plastic cylindrical container (10 cm diameter, 4 cm height). In order to assess the effect of internal gradients, two different cases were considered:

  1. Unmatched case. The phantom was first filled with distilled water and the whole set of MRI experiments was carried out.
  2. Matched case. The phantom was drained and refilled with the matching MgCl2·6H2O concentration (see section Results: c3 = 1.87 mol/l with χ3 ≈ -1.01 ppm), repeating the same MRI experiments.

In both cases the phantom was placed in a vacuum chamber for four hours in order to remove remaining air bubbles.

Experiments were carried out for three orientations of the phantom such that the angle, θ, between one of the fibre populations and B0 was θ = 0°, 22.5° and 45°. Thus, the angle between the remaining fibre population and B0 was 90°–θ = 90°, 67.5° and 45°, respectively. Protocol parameters were: TR/TE = 5000/113 ms; AVGs = 8; BW = 1563 Hz/pixel; voxel size, 2×2×2 mm3; matrix size, 88×128×14; GRAPPA acceleration factor 2, with 24 reference lines; phase-encoding direction right-to-left. An extra non-DW volume with the opposite phase-encoding direction was acquired for the correction of EPI distortions. The applied b-values were: b = 0, 1.0 ms/μm2 (unmatched) and b = 0, 2.0 ms/μm2 (matched) along 64 gradient directions. The different b-values for the unmatched and the matched cases were chosen so as to maintain the comparability of the products (bD)unmatched and (bD)matched. The non-DW signal, S0, and the apparent diffusion tensor Dapp are estimated voxel-wise as described in the section Maximum likelihood estimation of diffusion parameters.

An evaluation of the performance of the CSD and QBI methods in assessing fibre directionality is also carried out. The normalised fibre orientation distribution (FOD) defined in the framework of CSD and the orientation density function (ODF) in the case of QBI are obtained voxel-wise using the toolbox ExploreDTI [68]. Results are generated using maximum spherical-harmonic order lmax = 6. Fibre orientations are assessed via the local maxima of the FOD and ODF, using a Newton optimisation algorithm [69]. A threshold equal to 10% of the maximum peak is used to reject spurious peaks. The deviation angle, Δα, between the predicted and the underlying, mean fibre direction is evaluated as (6) where p = 1, 2 denotes the peak number, N is the number of voxels in the ROI, mp,i is the voxel-wise unit vector pointing in the pth fibre direction and Mp is a vector pointing along the average pth fibre direction, .

Maximum likelihood estimation of diffusion parameters

The maximum likelihood estimator (ML) is used in order to avoid bias in the diffusion parameters due to the low SNR observed in some of the DW experiments described above [52]. The ML estimator is written as: (7) where S(β) is the signal model (Eq (2)), β is a vector containing all parameters of interest, X is a vector containing the measured signals xi, px = ip(xi|S(β), σ) denotes the joint probability density function for xi, and p(xi|S(β), σ) is the Rice distribution with background noise parameter σ [52]. Here, the parameter σ is estimated following the method proposed by Aja-Fernández et al. [70]. For each case, maximisation of Eq (7) is performed using the Nelder-Mead algorithm with in-house Matlab scripts.

Further evaluation of the DTI invariant scalar maps, namely axial (AD), representing D||, radial (RD), representing D, and mean (MD) apparent diffusivities as well as fractional anisotropy (FA), is performed as described elsewhere [36]. Furthermore, a ROI analysis is carried out for all metrics. The brackets 〈A〉 denote the average of a given parameter A over a ROI, whereas the standard deviation is calculated as , where N is the number of voxels in the ROI.

Results

Bulk experiments

Fig 3 illustrates the bulk magnetic susceptibility difference with respect to distilled water, χ (a), the relative proton density PD (b), the relaxation times T1 and T2 (c), and the bulk diffusion coefficient D (d), for the 9 concentrations of MgCl2·6H2O (Table 1). The susceptibility estimation shows high precision with a relative field error of around 2%. Increasing the concentration of MgCl2·6H2O leads to a monotonous decrease of all parameters. In particular, for the concentration c3 (1.87 mol/l) the magnetic susceptibility is χ3 ≈ -1.01 ppm, which practically matches the susceptibility of the Dyneema fibres (Fig 3a) [20]. Concomitant with that change there is a reduction in the proton density, in both relaxation times and in the diffusion coefficient. For the concentration c3 the following values are observed: PD = (0.83±0.03), T1 = (1700±100) ms, T2 = (1040±20) ms and D = (1.13±0.02) μm2/ms.

thumbnail
Fig 3. Concentration dependence of bulk NMR quantities.

Magnetic susceptibility (a), proton density (b), relaxation times T1 and T2 (c) and diffusion coefficient (d).

https://doi.org/10.1371/journal.pone.0176192.g003

Parallel-fibre phantom

Fig 4 shows the maps of the DTI metrics AD (a), RD (b), MD (c) and FA (d), for all five vials and for the two extreme angles θ = 0° (left-hand side) and 90° (right-hand side). One can see that all parameters and all vials show a rather homogeneous behaviour for θ = 0°. Whereas for θ = 90° severe image artefacts can be seen for all parameters and all vials except for the concentrations c2 and c4.

thumbnail
Fig 4. Maps of the DTI scalar parameters.

AD (a), RD (b), MD (c) and FA (d) for the orientations θ = 0° (left-hand side) and θ = 90° (right-hand side). ROI-averaged values 〈AD〉 (e), 〈RD〉 (f), 〈MD〉 (g) and 〈FA〉 (h) and the normalised standard deviations σAD/〈AD〉 (i), σRD/〈RD〉 (j), σMD/〈MD〉 (k) and σFA/〈FA〉 (l). The bulk 〈AD〉 is shown as a reference by the dashed, grey line (e). The ROIs contain 1300±200 voxels.

https://doi.org/10.1371/journal.pone.0176192.g004

Fig 4e–4h demonstrate the angular dependence for the ROI-averaged values 〈AD〉, 〈RD〉, 〈MD〉 and 〈FA〉, respectively. The normalised standard deviations σAD/〈AD〉, σRD/〈RD〉, σMD/〈MD〉 and σFA/〈FA〉 are shown in Fig 4i–4l, respectively. It can be seen that for the concentration c0 all diffusion metrics are strongly underestimated when θ ≥ 45°. The rest of the concentrations show a qualitatively more stable behaviour for 0° ≤ θ ≤ 90°. However, Fig 4i–4l demonstrate that the normalised standard deviations for concentrations c0, c6 and c8, i.e., for susceptibility values far from the matching concentration, are strongly increased for θ ≥ 45°, whereas for c2 and c4 the change is virtually negligible in the whole angular range.

The change of the SNR values, conventionally calculated as SNR = S(b = 0)/σ, for 0° ≤ θ ≤ 90° and all concentrations is shown in Fig 5. A strong decrease is seen for c0, c6 and c8 and θ ≥ 45° (~60% to ~80%), whereas for concentrations c2 and c4 the reduction is much smaller (~25%).

thumbnail
Fig 5. SNR values in the parallel-fibre phantoms.

Calculated as SNR = S(b = 0)/σ, for 0° ≤ θ ≤ 90° in all concentrations.

https://doi.org/10.1371/journal.pone.0176192.g005

The results of the bulk measurements as well as of the parallel-fibre scenario clearly indicate c3 as the optimal concentration of the MgCl2·6H2O solution.

Crossing-fibre phantom

Fig 6 shows the maps of the diffusion tensor eigenvalues λi, i = 1,2,3 (a) and the metrics MD (b) and FA (c) for the unmatched and the matched cases. Two angular orientations are shown: θ = 0° and θ = 45°. All maps show differences between the experimental setups with the phantom positioned at θ = 0° and θ = 45° for the unmatched case, whereas for the matched case the differences virtually disappear.

thumbnail
Fig 6. Maps of the DTI parameters in the crossing-fibre phantom.

Eigenvalues λi, i = 1,2,3 (a) and metrics MD (b) and FA (c) for the unmatched case (top) and the matched case (bottom). Two angular orientations are shown: θ = 0° on the left-hand side, and θ = 45° on the right-hand side. The angular dependence for the ROI-averaged values for 〈λi〉 (d), 〈MD〉 (e) and 〈FA〉 (f). The corresponding normalised standard deviation values (g), σMD/〈MD〉 (h) and σFA/〈FA〉 (i), for the unmatched (black-solid lines) as well as for the matched (red-dashed lines) cases.

https://doi.org/10.1371/journal.pone.0176192.g006

The angular dependence for the ROI-averaged values is shown in Fig 6d–6f for 〈λi〉, 〈MD〉 and 〈FA〉, respectively. The corresponding normalised standard deviation values , σMD/〈MD〉 and σFA/〈FA〉 are illustrated in Fig 6g–6i, respectively, for the unmatched as well as for the matched cases. For the unmatched case one finds λ1>λ2λ3 for θ = 0°, whereas λ1λ2>λ3 for θ = 45°, denoting a strong angular dependence of the shape of the diffusion tensor ellipsoid with regard to B0. On the contrary, for the matched case one observes that λ1λ2>λ3 for both orientations. Moreover, the normalised standard deviations for all parameters show higher values for the unmatched case than for the matched case.

The performance of the CSD and QBI methods is depicted in Fig 7. The unmatched case is shown in the panel on the left-hand side for three phantom orientations (Fig 7a–7c). Similarly the matched case is shown in the panel on the right-hand side (Fig 7d–7f). In the unmatched case and θ = 0° only the fibre population parallel to B0 is resolved, whereas the fibre population perpendicular to B0 is not visible (Fig 7a). Only for the orientation θ = 45° both fibre populations can be resolved (Fig 7c). Yet, for the matched case both fibre populations are clearly resolved, regardless of the phantom orientation with respect to B0 (Fig 7d–7f).

thumbnail
Fig 7. Performance of the CSD and QBI methods in the assessment of fibre orientations.

The unmatched case is shown in the panel on the left-hand side for three phantom orientations (a-c). Similarly the matched case is shown in the panel on the right-hand side (d-f). The background image in each case corresponds to the colour-coded FA from conventional DTI.

https://doi.org/10.1371/journal.pone.0176192.g007

The dependence of the mean deviation angle, Δα, with regard to θ as estimated using Eq (6), is depicted in Fig 8a for CSD and Fig 8b for QBI. For the unmatched case, one can clearly observe that Δα1> Δα2, for both CSD and QBI, whilst for the matched case, Δα1 Δα2. Both mean deviation values are lower than for the unmatched case.

thumbnail
Fig 8. Dependence of the mean deviation angle Δα with regard to θ.

Δα is evaluated for CSD (a) and for QBI (b) using Eq (6). Black lines correspond to the unmatched case, whereas red lines denote the matched case. Circles denote the main fibre direction and squares refer to the secondary fibre direction.

https://doi.org/10.1371/journal.pone.0176192.g008

Discussion

Bulk experiments

The characterisation of the bulk water NMR properties with increasing solute concentration reveals that the reduction in the bulk magnetic susceptibility entails other side effects, namely a decrease of the bulk longitudinal and transverse relaxation times, of the bulk diffusivity and of the proton density.

A decrease in T1 is beneficial as it allows for a reduction in repetition time without significantly affecting the SNR. On the other hand, a reduction in PD and T2 a priori gives rise to negative effects in DW MRI, since they contribute to diminish the SNR for a given TE. A decrease of ~10% in PD for c3 clearly implies a direct reduction in the SNR by the same amount. On the other hand, the observed decrease in T2 from 2600 ms for c0 to around 1040 ms for c3 introduces a minor reduction of the SNR values by ~6%, for an echo-time of 100 ms (typically used in DW MRI).

A decrease in the bulk diffusivity values from 2.2 μm2/ms for c0 to 1.13 μm2/ms for c3 was also measured. The bulk water diffusivity for increasing concentrations of MgCl2·6H2O has previously been investigated for the range of c = [0.24, 5.43] mol/l at 25°C using NMR [71]. The authors interpret such reduction in terms of the so-called two-phase model. Under this approach, water molecules are considered to be in fast exchange, on a nanosecond timescale, between two distinguishable water pools, namely the hydration layer at the ion Mg2+ and the bulk water. The hydration layer, composed by six H2O molecules, diffuses as a whole unit with a slower diffusivity, , whereas the remaining pool, referring to the water beyond the hydration layer, is characterised by a faster diffusivity, Dbulk. Thus, in the fast exchange limit, the mean diffusivity as measured with a PGSE sequence is given by , with f(c) being the mole fraction of the hydration layer [71]. In our case, the validity of the fast exchange limit is particularly important given that the DW signal attenuation will remain monoexponential, which is the desired behaviour for the bulk liquid used in diffusion phantoms. This behaviour was observed for all concentrations. Furthermore, the bulk diffusivity for c3 more closely resembles the free diffusion inside axonal bodies, i.e., ~1.4 μm2/ms [72].

Parallel-fibre phantoms

According to Eq (3), the apparent RD can a priori be biased if the cross term is not taken into account in the signal model and the magnetic susceptibilities are not matched. However, given that the cross term can take negative and positive values, one needs to consider its behaviour in the whole volume of interest in order to predict the actual behaviour.

Our results show that there is an increasing underestimation of RD for increasing strength of the background gradients, i.e., increasing the angle of the fibre with respect to B0. Here, we refer to the works by Zhong et al. [38,39], in which a symmetric distribution for the cross term around zero in a heterogeneous, isotropic system was assumed. In their work, Zhong et al. postulate that the background gradients show a slow spatial change, i.e., each diffusing molecule experiences a constant background gradient during the time course of the experiment. Their results showed that the overall effect of the cross term is a decrease in the estimated diffusivity with increasing strength of the background gradients. In our parallel-fibre phantoms, the diffusion length ld, i.e., the root mean square displacements during TE = 82 ms, was approximately 12 μm for c0 and 7.5 μm for c8 in the radial direction. On the other hand, considering the observed mean volume fibre fraction of ~0.64, one can estimate the structural length ls, i.e., the distances between the fibres in the interstitial space, to be in the order of 3 μm (assuming a simple hexagonal arrangement of the lattice as an example). Therefore given that ls is shorter, the spins cross the pores (structural lengths) many times before dephasing and the magnetic field inhomogeneities are motionally averaged [73,74]. In other words, the water molecules experience a mean background gradient with slow spatial change and therefore, a similar argument to the one utilised by Zhong et al. can be used to explain the observed underestimation in RD in our work.

The apparent AD should instead ideally remain independent of the orientation as predicted by Eq (3). This can be observed for concentrations c2-c8. However, AD exhibits a strong underestimation for c0 at θ > 45°, as shown in Fig 4a and 4e. A possible explanation for that could be the fact that as a result of the strong reduction in T2 due to the strong background gradients, the SNR reaches very low values (Fig 5). As a consequence, the relative standard deviation σAD/〈AD〉 reaches values of the order of 0.6 to 0.8 (Fig 4i), turning the estimation unreliable. However, given that all tensor elements were assessed using the asymptotically unbiased ML estimator, one can expect only a marginal influence of the low SNR in AD. On the other hand, a stronger influence is expected by the likely presence of some imperfections in the intra-voxel fibre alignment, which would result in an axial component of the background gradients. Moreover, the difference in magnetic susceptibility between the liquid inside the vial and the surrounding distilled water (see Fig 1, middle phantom), can create long-range field gradients that can a priori affect the estimation of AD.

Despite the aforementioned results, all parameters remained practically unchanged for the concentrations c2 and c4. Moreover, as a consequence of the minor reduction in the SNR values, the relative standard deviation for all parameters remains low across the whole angular range.

Crossing-fibre phantoms

DTI analysis demonstrates that the ellipsoid representation of the diffusion tensor can appear strongly changed in the presence of background gradients. For the unmatched case and the orientation for θ = 0°, the tensor eigenvalues follow the relation λ1>λ2λ3 with λ1 being the diffusivity along the fibre population parallel to B0 (Fig 6a). This relation corresponds to the prolate shape of the ellipsoid, observed in systems with cylindrical geometry. In other words, only the fibre population parallel to B0 is observable. This result can be explained by the fact that the remaining fibre population suffers from a strong reduction in the SNR leading to a strong underestimation in the diffusivities, similar to that observed in the parallel-fibre phantom for θ = 90° (Fig 4e and 4f). Although the expected oblate shape of the ellipsoid for such fibre configuration, i.e., λ1λ2>λ3, is recovered for the orientation θ = 45°, the absolute values of λi are still biased by the background gradients. The effect on the estimation of the fibre populations can be more clearly observed in the CSD and QBI analysis in Fig 7a–7c, where both fibre populations are recovered only for θ = 45°.

On the other hand, the dependence of the tensor eigenvalues on the orientation with respect to B0 is completely suppressed in the matched case, as shown in Fig 6. Similarly, both fibre populations are clearly identified using CSD and QBI, independently of the phantom orientation. This is even more emphasised by Fig 8, showing that the angular deviations for both populations are similar and practically independent of the phantom orientation.

Notwithstanding the advantages of the improved design of the phantom, one recognises several limitations of the fibres themselves regarding their capability for mimicking white matter tissue properties. One of their intrinsic limitations is that, being rod-like fibres, they resemble the diffusion properties of only the extracellular space in the white matter tissue. Moreover, white matter fibres are not perfect cylinders, but their profile is modulated by the Ranvier nodes, a fact that is not reflected in Dyneema fibres.

Conclusions

In this work we have assessed one of the crucial points in the construction of anisotropic diffusion fibre phantom, namely the necessary independence of the diffusion metrics on the orientation of the phantom in the static magnetic field. We showed that susceptibility differences between fibre and liquid are responsible for the angular dependence of the diffusion response signal and consequently of the diffusion parameters and the reconstruction of fibre tracts.

It is demonstrated that an experimentally determined concentration of 1.87 mol/l MgCl2·6H2O dissolved in distilled water removes the susceptibility difference between the liquid and the fibres, and hence eliminates the microscopic gradient fields in the interstitial space. Susceptibility matching leads to complete angular independence of the diffusion metrics in complex anisotropic diffusion phantoms. This, in turn, allows one to conduct DW MRI experiments without the need to consider the geometric alignment of the involved structures with respect to the direction of static magnetic field. Moreover, our findings are of further interest for the investigation of diffusion in complex scenarios involving susceptibility differences and local microscopic field gradients.

Supporting information

S1 File. Archive containing results.

Datasets are indexed and described in CONTENTS.txt.

https://doi.org/10.1371/journal.pone.0176192.s001

(ZIP)

Acknowledgments

The authors thank Mr Michael Schöneck for assistance in the preparation of solutions and Dr. Nazim Lechea for his help in data processing.

Author Contributions

  1. Conceptualization: EF JL FG AMOP NJS.
  2. Formal analysis: EF JL.
  3. Funding acquisition: NJS.
  4. Investigation: EF JL.
  5. Methodology: EF JL FG AMOP NJS.
  6. Project administration: EF JL.
  7. Resources: EF JL FG NJS.
  8. Software: EF JL.
  9. Supervision: EF JL NJS.
  10. Visualization: EF JL.
  11. Writing – original draft: EF JL.
  12. Writing – review & editing: EF JL FG AMOP NJS.

References

  1. 1. Bach M, Fritzsche KH, Stieltjes B, Laun FB. Investigation of resolution effects using a specialized diffusion tensor phantom. Magn Reson Med. 2014;71: 1108–1116. pmid:23657980
  2. 2. Hubbard PL, Zhou F-L, Eichhorn SJ, Parker GJM. Biomimetic phantom for the validation of diffusion magnetic resonance imaging. Magn Reson Med. 2015;73: 299–305. pmid:24469863
  3. 3. Lin C-P, Wedeen VJ, Chen J-H, Yao C, Tseng W-YI. Validation of diffusion spectrum magnetic resonance imaging with manganese-enhanced rat optic tracts and ex vivo phantoms. Neuroimage. 2003;19: 482–495. http://dx.doi.org/10.1016/S1053-8119(03)00154-X pmid:12880782
  4. 4. Laun FB, Huff S, Stieltjes B. On the effects of dephasing due to local gradients in diffusion tensor imaging experiments: relevance for diffusion tensor imaging fiber phantoms. Magn Reson Imaging. 2009;27: 541–548. http://dx.doi.org/10.1016/j.mri.2008.08.011 pmid:18977104
  5. 5. Watanabe M, Aoki S, Masutani Y, Abe O, Hayashi N, Masumoto T, et al. Flexible ex vivo phantoms for validation of diffusion tensor tractography on a clinical scanner. Radiat Med. 2006;24: 605–609. pmid:17111268
  6. 6. Perrin M, Poupon C, Rieul B, Leroux P, Constantinesco A, Mangin J-F, et al. Validation of q-ball imaging with a diffusion fibre-crossing phantom on a clinical scanner. Philos Trans R Soc London B Biol Sci. The Royal Society; 2005;360: 881–891.
  7. 7. Descoteaux M, Deriche R, Le Bihan D, Mangin J-F, Poupon C. Multiple q-shell diffusion propagator imaging. Med Image Anal. 2011;15: 603–621. http://dx.doi.org/10.1016/j.media.2010.07.001 pmid:20685153
  8. 8. Poupon C, Rieul B, Kezele I, Perrin M, Poupon F, Mangin J-F. New diffusion phantoms dedicated to the study and validation of high-angular-resolution diffusion imaging (HARDI) models. Magn Reson Med. 2008;60: 1276–1283. pmid:19030160
  9. 9. Tournier J-D, Yeh C-H, Calamante F, Cho K-H, Connelly A, Lin C-P. Resolving crossing fibres using constrained spherical deconvolution: Validation using diffusion-weighted imaging phantom data. Neuroimage. 2008;42: 617–625. http://dx.doi.org/10.1016/j.neuroimage.2008.05.002 pmid:18583153
  10. 10. Moussavi-Biugui A, Stieltjes B, Fritzsche K, Semmler W, Laun FB. Novel spherical phantoms for Q-ball imaging under in vivo conditions. Magn Reson Med. 2011;65: 190–194. pmid:20740652
  11. 11. Wilkins B, Lee N, Gajawelli N, Law M, Leporé N. Fiber estimation and tractography in diffusion MRI: Development of simulated brain images and comparison of multi-fiber analysis methods at clinical b-values. Neuroimage. 2015;109: 341–356. http://dx.doi.org/10.1016/j.neuroimage.2014.12.060 pmid:25555998
  12. 12. Hubbard PL, Parker GJM. Validation of Tractography. In: Johansen-Berg H, Behrens TEJ, editors. Diffusion MRI. Second Edi. San Diego: Academic Press; 2014. pp. 453–480. http://dx.doi.org/10.1016/B978-0-12-396460-1.00020-2
  13. 13. Fillard P, Descoteaux M, Goh A, Gouttard S, Jeurissen B, Malcolm J, et al. Quantitative evaluation of 10 tractography algorithms on a realistic diffusion MR phantom. Neuroimage. 2011;56: 220–234. http://dx.doi.org/10.1016/j.neuroimage.2011.01.032 pmid:21256221
  14. 14. Fieremans E, De Deene Y, Delputte S, Özdemir MS, D’Asseler Y, Vlassenbroeck J, et al. Simulation and experimental verification of the diffusion in an anisotropic fiber phantom. J Magn Reson. 2008;190: 189–199. http://dx.doi.org/10.1016/j.jmr.2007.10.014 pmid:18023218
  15. 15. Reischauer C, Staempfli P, Jaermann T, Boesiger P. Construction of a temperature-controlled diffusion phantom for quality control of diffusion measurements. J Magn Reson Imaging. 2009;29: 692–698. pmid:19243053
  16. 16. Zhu T, Hu R, Qiu X, Taylor M, Tso Y, Yiannoutsos C, et al. Quantification of accuracy and precision of multi-center DTI measurements: A diffusion phantom and human brain study. Neuroimage. 2011;56: 1398–1411. http://dx.doi.org/10.1016/j.neuroimage.2011.02.010 pmid:21316471
  17. 17. Chenevert TL, Galbán CJ, Ivancevic MK, Rohrer SE, Londy FJ, Kwee TC, et al. Diffusion coefficient measurement using a temperature-controlled fluid for quality control in multicenter studies. J Magn Reson Imaging. 2011;34: 983–987. pmid:21928310
  18. 18. Lorenz R, Bellemann EM, Hennig J, Il’yasov AK. Anisotropic Phantoms for Quantitative Diffusion Tensor Imaging and Fiber-Tracking Validation. Appl Magn Reson. 2008;33: 419–429.
  19. 19. Yanasak N, Allison J. Use of capillaries in the construction of an MRI phantom for the assessment of diffusion tensor imaging: demonstration of performance. Magn Reson Imaging. 2006;24: 1349–1361. http://dx.doi.org/10.1016/j.mri.2006.08.001 pmid:17145407
  20. 20. Fieremans E, De Deene Y, Delputte S, Özdemir MS, Achten E, Lemahieu I. The design of anisotropic diffusion phantoms for the validation of diffusion weighted magnetic resonance imaging. Phys Med Biol. 2008;53: 5405. pmid:18765890
  21. 21. Farrher E, Kaffanke J, Celik AA, Stöcker T, Grinberg F, Shah NJ. Novel multisection design of anisotropic diffusion phantoms. Magn Reson Imaging. 2012;30: 518–526. pmid:22285876
  22. 22. Jackson JD. Classical electrodynamics. 3rd ed. New York, NY: Wiley; 1999.
  23. 23. Majumdar S, Gore JC. Studies of diffusion in random fields produced by variations in susceptibility. J Magn Reson. 1988;78: 41–55. http://dx.doi.org/10.1016/0022-2364(88)90155-2
  24. 24. Kärger J, Pfeifer H, Heink W. Principles and applications of self-diffusion measurements by nuclear magnetic resonance. In: Waugh J, editor. Advances in Magnetic Resonance. 1250 Sixth Avenue San Diego, California 92101: Academic Press, Inc.; 1988. pp. 1–89.
  25. 25. Goelman G, Prammer MG. The CPMG Pulse Sequence in Strong Magnetic Field Gradients with Applications to Oil-Well Logging. J Magn Reson Ser A. 1995;113: 11–18. http://dx.doi.org/10.1006/jmra.1995.1050
  26. 26. Beaulieu C, Allen PS. An in vitro evaluation of the effects of local magnetic-susceptibility-induced gradients on anisotropic water diffusion in nerve. Magn Reson Med. 1996;36: 39–44. pmid:8795018
  27. 27. De Santis S, Rebuzzi M, Di Pietro G, Fasano F, Maraviglia B, Capuani S. In vitro and in vivo MR evaluation of internal gradient to assess trabecular bone density. Phys Med Biol. 2010;55: 5767. pmid:20844335
  28. 28. Clark CA, Barker GJ, Tofts PS. An in Vivo Evaluation of the Effects of Local Magnetic Susceptibility-Induced Gradients on Water Diffusion Measurements in Human Brain. J Magn Reson. 1999;141: 52–61. http://dx.doi.org/10.1006/jmre.1999.1872 pmid:10527743
  29. 29. Palombo M, Gentili S, Bozzali M, Macaluso E, Capuani S. New insight into the contrast in diffusional kurtosis images: Does it depend on magnetic susceptibility? Magn Reson Med. 2015;73: 2015–2024. pmid:24894844
  30. 30. Lindemeyer J, Oros-Peusquens A-M, Farrher E, Grinberg F, Shah NJ. Orientation and Microstructure Effects on Susceptibility Reconstruction: a Diffusion Phantom Study. Proceedings of the International Society for Magnetic Resonance in Medicine. 2011. p. 4516.
  31. 31. Stejskal EO, Tanner JE. Spin Diffusion Measurements: Spin Echoes in the Presence of a Time Dependent Field Gradient. J Chem Phys. 1965;42.
  32. 32. Capuani S, Piccirilli E, Di Pietro G, Celi M, Tarantino U. Microstructural differences between osteoporotic and osteoarthritic femoral cancellous bone: an in vitro magnetic resonance micro-imaging investigation. Aging Clin Exp Res. 2013;25: 51–54.
  33. 33. Rebuzzi M, Vinicola V, Taggi F, Sabatini U, Wehrli FW, Capuani S. Potential diagnostic role of the MRI-derived internal magnetic field gradient in calcaneus cancellous bone for evaluating postmenopausal osteoporosis at 3 T. Bone. 2013;57: 155–163. http://dx.doi.org/10.1016/j.bone.2013.07.027 pmid:23899635
  34. 34. Packer KJ. The effects of diffusion through locally inhomogeneous magnetic fields on transverse nuclear spin relaxation in heterogeneous systems. Proton transverse relaxation in striated muscle tissue. J Magn Reson. 1973;9: 438–443. http://dx.doi.org/10.1016/0022-2364(73)90186-8
  35. 35. Knight MJ, Kauppinen RA. Diffusion-mediated nuclear spin phase decoherence in cylindrically porous materials. J Magn Reson. 2016;269: 1–12. http://dx.doi.org/10.1016/j.jmr.2016.05.007 pmid:27208416
  36. 36. Le Bihan D, Mangin J-F, Poupon C, Clark CA, Pappata S, Molko N, et al. Diffusion tensor imaging: Concepts and applications. J Magn Reson Imaging. John Wiley & Sons, Inc.; 2001;13: 534–546.
  37. 37. Miller AJ, Joseph PM. The use of power images to perform quantitative analysis on low SNR MR images. Magn Reson Imaging. 1993;11: 1051–1056. pmid:8231670
  38. 38. Zhong J, Gore JC. Studies of restricted diffusion in heterogeneous media containing variations in susceptibility. Magn Reson Med. 1991;19: 276–284. pmid:1881316
  39. 39. Zhong J, Kennan RP, Gore JC. Effects of susceptibility variations on NMR measurements of diffusion. J Magn Reson. 1991;95: 267–280. http://dx.doi.org/10.1016/0022-2364(91)90217-H
  40. 40. Kiselev VG. Effect of magnetic field gradients induced by microvasculature on NMR measurements of molecular self-diffusion in biological tissues. J Magn Reson. 2004;170: 228–235. http://dx.doi.org/10.1016/j.jmr.2004.07.004 pmid:15388085
  41. 41. Chen WC, Foxley S, Miller KL. Detecting microstructural properties of white matter based on compartmentalization of magnetic susceptibility. Neuroimage. Elsevier Inc.; 2013;70: 1–9.
  42. 42. Lee J, Shmueli K, Fukunaga M, van Gelderen P, Merkle H, Silva AC, et al. Sensitivity of MRI resonance frequency to the orientation of brain tissue microstructure. Proc Natl Acad Sci U S A. 2010;107: 5130–5135. pmid:20202922
  43. 43. Caporale A, Palombo M, Macaluso E, Guerreri M, Bozzali M, Capuani S. The γ-parameter of anomalous diffusion quantified in human brain by MRI depends on local magnetic susceptibility differences. Neuroimage. 2017;147: 619–631. http://dx.doi.org/10.1016/j.neuroimage.2016.12.051 pmid:28011255
  44. 44. Duyn JH, van Gelderen P, Li T-Q, de Zwart J a, Koretsky AP, Fukunaga M. High-field MRI of brain cortical substructure based on signal phase. Proc Natl Acad Sci U S A. 2007;104: 11796–11801. pmid:17586684
  45. 45. Liu C. Susceptibility tensor imaging. Magn Reson Med. 2010;63: 1471–1477. pmid:20512849
  46. 46. Zheng G, Price WS. Suppression of background gradients in (B0 gradient-based) NMR diffusion experiments. Concepts Magn Reson Part A. 2007;30A: 261–277.
  47. 47. Meiboom S, Gill D. Modified Spin Echo Method for Measuring Nuclear Relaxation Times. Rev Sci Instrum. 1958;29.
  48. 48. Sun PZ, Seland JG, Cory D. Background gradient suppression in pulsed gradient stimulated echo measurements. J Magn Reson. 2003;161: 168–173. pmid:12713966
  49. 49. Ballon D, Mahmood U, Jakubowski A, Koutcher JA. Resolution enhanced NMR spectroscopy in biological systems via magnetic susceptibility matched sample immersion chambers. Magn Reson Med. 1993;30: 754–758. pmid:8139459
  50. 50. Tournier J-D, Calamante F, Connelly A. Robust determination of the fibre orientation distribution in diffusion MRI: Non-negativity constrained super-resolved spherical deconvolution. Neuroimage. 2007;35: 1459–1472. http://dx.doi.org/10.1016/j.neuroimage.2007.02.016 pmid:17379540
  51. 51. Tuch DS. Q-ball imaging. Magn Reson Med. Wiley Subscription Services, Inc., A Wiley Company; 2004;52: 1358–1372.
  52. 52. Sijbers J, den Dekker AJ. Maximum likelihood estimation of signal amplitude and noise variance from MR data. Magn Reson Med. 2004;51: 586–594. pmid:15004801
  53. 53. Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys. 1996;23.
  54. 54. Haynes WM, editor. CRC Handbook of Chemistry and Physics—Electrical Conductivity of Aqueous Solutions. 70th ed. CRC Press, Boca Raton; 1989.
  55. 55. Lide DR, editor. CRC Handbook of Chemistry and Physics, Internet Version 2005 [Internet]. CRC Press, Boca Raton, FL; 2005. http://www.hbcpnetbase.com
  56. 56. Reichenbach JR. The future of susceptibility contrast for assessment of anatomy and function. Neuroimage. 2012;62: 1311–1315. http://dx.doi.org/10.1016/j.neuroimage.2012.01.004 pmid:22245644
  57. 57. Wang Y, Liu T. Quantitative susceptibility mapping (QSM): Decoding MRI data for a tissue magnetic biomarker. Magn Reson Med. 2015;73: 82–101. pmid:25044035
  58. 58. Haacke EM, Brown RW, Thompson MR, Venkatesan R, others. Magnetic resonance imaging: physical principles and sequence design. New York: Wiley; 1999.
  59. 59. Lindemeyer J, Oros-Peusquens A-M, Shah NJ. Multistage Background Field Removal (MUBAFIRE)—Compensating for B0 Distortions at Ultra-High Field. PLoS One. 2015;10: e0138325. pmid:26393515
  60. 60. Koch KM, Papademetris X, Rothman DL, de Graaf RA. Rapid calculations of susceptibility-induced magnetostatic field perturbations for in vivo magnetic resonance. Phys Med Biol. 2006;51: 6381. pmid:17148824
  61. 61. Marques JP, Bowtell R. Application of a Fourier-based method for rapid calculation of field inhomogeneity due to spatial variation of magnetic susceptibility. Concepts Magn Reson Part B Magn Reson Eng. 2005;25B: 65–78.
  62. 62. Shah NJ, Zaitsev M, Steinhoff S, Zilles K. A New Method for Fast Multislice T1 Mapping. Neuroimage. 2001;14: 1175–1185. http://dx.doi.org/10.1006/nimg.2001.0886 pmid:11697949
  63. 63. Reese TG, Heid O, Weisskoff RM, Wedeen VJ. Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo. Magn Reson Med. 2003;49: 177–182. pmid:12509835
  64. 64. Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TEJ, Johansen-Berg H, et al. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23, Supple: S208–S219. http://dx.doi.org/10.1016/j.neuroimage.2004.07.051
  65. 65. Woolrich MW, Jbabdi S, Patenaude B, Chappell M, Makni S, Behrens T, et al. Bayesian analysis of neuroimaging data in FSL. Neuroimage. 2009;45: S173–S186. http://dx.doi.org/10.1016/j.neuroimage.2008.10.055 pmid:19059349
  66. 66. Jenkinson M, Smith S. A global optimisation method for robust affine registration of brain images. Med Image Anal. 2001;5: 143–156. http://dx.doi.org/10.1016/S1361-8415(01)00036-6 pmid:11516708
  67. 67. Leemans A, Jones DK. The B-matrix must be rotated when correcting for subject motion in DTI data. Magn Reson Med. 2009;61: 1336–1349. pmid:19319973
  68. 68. Leemans A, Jeurissen B, Sijbers J, Jones DK. ExploreDTI: a graphical toolbox for processing, analyzing, and visualizing diffusion MR data. Proc Intl Soc Mag Reson Med. 2009. p. 3537.
  69. 69. Jeurissen B, Leemans A, Tournier J-D, Jones DK, Sijbers J. Investigating the prevalence of complex fiber configurations in white matter tissue with diffusion magnetic resonance imaging. Hum Brain Mapp. 2013;34: 2747–2766. pmid:22611035
  70. 70. Aja-Fernández S, Tristán-Vega A, Alberola-López C. Noise estimation in single- and multiple-coil magnetic resonance data based on statistical models. Magn Reson Imaging. 2009;27: 1397–1409. http://dx.doi.org/10.1016/j.mri.2009.05.025 pmid:19570640
  71. 71. Struis RPWJ, De Bleijser J, Leyte JC. An NMR contribution to the interpretation of the dynamical behavior of water molecules as a function of the magnesium chloride concentration at 25.degree.C. J Phys Chem. 1987;91: 6309–6315.
  72. 72. Barazany D, Basser PJ, Assaf Y. In vivo measurement of axon diameter distribution in the corpus callosum of rat brain. Brain. 2009;132: 1210–1220. Available: http://dx.doi.org/10.1093/brain/awp042 pmid:19403788
  73. 73. Mitchell J, Chandrasekera TC, Johns ML, Gladden LF, Fordham EJ. Nuclear magnetic resonance relaxation and diffusion in the presence of internal gradients: The effect of magnetic field strength. Phys Rev E. American Physical Society; 2010;81: 26101.
  74. 74. Di Pietro G, Palombo M, Capuani S. Internal Magnetic Field Gradients in Heterogeneous Porous Systems: Comparison Between Spin-Echo and Diffusion Decay Internal Field (DDIF) Method. Appl Magn Reson. 2014;45: 771–784.