Comparison of different inspiratory triggering settings in automated ventilators during cardiopulmonary resuscitation in a porcine model

  • Dingyu Tan ,

    Contributed equally to this work with: Dingyu Tan, Jun Xu, Shihuan Shao

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Jun Xu ,

    Contributed equally to this work with: Dingyu Tan, Jun Xu, Shihuan Shao

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Shihuan Shao ,

    Contributed equally to this work with: Dingyu Tan, Jun Xu, Shihuan Shao

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Yangyang Fu,

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Feng Sun,

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Yazhi Zhang,

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Yingying Hu,

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Joseph Walline,

    Affiliation Division of Emergency Medicine, Department of Surgery, Saint Louis University Hospital, Saint Louis, Missouri

  • Huadong Zhu ,

    zhuhuadong1970@126.com (ZH); yxzpumch@126.com (XY)

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

  • Xuezhong Yu

    zhuhuadong1970@126.com (ZH); yxzpumch@126.com (XY)

    Affiliation Department of Emergency, Peking Union Medical College Hospital, Chinese Academy of Medical sciences, Beijing, China

Comparison of different inspiratory triggering settings in automated ventilators during cardiopulmonary resuscitation in a porcine model

  • Dingyu Tan, 
  • Jun Xu, 
  • Shihuan Shao, 
  • Yangyang Fu, 
  • Feng Sun, 
  • Yazhi Zhang, 
  • Yingying Hu, 
  • Joseph Walline, 
  • Huadong Zhu, 
  • Xuezhong Yu
PLOS
x

Abstract

Background

Mechanical ventilation via automated in-hospital ventilators is quite common during cardiopulmonary resuscitation. It is not known whether different inspiratory triggering sensitivity settings of ordinary ventilators have different effects on actual ventilation, gas exchange and hemodynamics during resuscitation.

Methods

18 pigs enrolled in this study were anaesthetized and intubated. Continuous chest compressions and mechanical ventilation (volume-controlled mode, 100% O2, respiratory rate 10/min, and tidal volumes 10ml/kg) were performed after 3 minutes of ventricular fibrillation. Group trig-4, trig-10 and trig-20 (six pigs each) were characterized by triggering sensitivities of 4, 10 and 20 (cmH2O for pressure-triggering and L/min for flow-triggering), respectively. Additionally, each pig in each group was mechanically ventilated using three types of inspiratory triggering (pressure-triggering, flow-triggering and turned-off triggering) of 5 minutes duration each, and each animal matched with one of six random assortments of the three different triggering settings. Blood gas samples, respiratory and hemodynamic parameters for each period were all collected and analyzed.

Results

In each group, significantly lower actual respiratory rate, minute ventilation volume, mean airway pressure, arterial pH, PaO2, and higher end-tidal carbon dioxide, aortic blood pressure, coronary perfusion pressure, PaCO2 and venous oxygen saturation were observed in the ventilation periods with a turned-off triggering setting compared to those with pressure- or flow- triggering (all P<0.05), except when compared with pressure-triggering of 20 cmH2O (respiratory rate 10.5[10/11.3]/min vs 12.5[10.8/13.3]/min, P = 0.07; coronary perfusion pressure 30.3[24.5/31.6] mmHg vs 27.4[23.7/29] mmHg, P = 0.173; venous oxygen saturation 46.5[32/56.8]% vs 41.5[33.5/48.5]%, P = 0.575).

Conclusions

Ventilation with pressure- or flow-triggering tends to induce hyperventilation and deteriorating gas exchange and hemodynamics during CPR. A turned-off patient triggering or a pressure-triggering of 20 cmH2O is preferred for ventilation when an ordinary inpatient hospital ventilator is used during resuscitation.

Introduction

The goal of ventilation during cardiopulmonary resuscitation (CPR) is to provide sufficient oxygenation and removal of carbon dioxide (CO2) to improve tissue oxygenation and acidosis. Apart from high-quality chest compressions, hyperventilation has gained increasing attention in recent years because it may negatively affect outcomes in cardiac arrest [13]. Manual ventilation with respiratory rates greater than 20 breaths/ min is common during CPR from cardiac arrest [46]. An automated transport ventilator (ATV) was found to be as effective as a bag valve mask to deliver ventilation during CPR once the airway is secured [7, 8], and it is recommend by the current guidelines for prolonged resuscitation [9]. In addition, using of ATV may help to avoid uncontrolled ventilation and improve the quality of chest compressions [10, 11]. However, mechanical ventilation by an ordinary automated ventilator rather than an ATV is quite common during resuscitation in actually clinical practice especially in inpatient hospital settings.

Different from the totally time-triggered mandatory breaths of Intermittent Positive Pressure Ventilation (IPPV) in ATV, the IPPV mode in most hospital-based ventilators (such as volume controlled mode) works with time- and patient-triggering simultaneously in order to increase the patient-ventilator synchrony [12]. Patient triggering systems such as pressure- and flow-triggering cannot be turned off in most of the commonly used modern ventilators. Except for respiratory rate (RR), using other ventilatory parameters (e.g. minute ventilation, peak pressure limit) during advanced life support is not recommend under current guidelines [13]. We observed in clinical practice that the changes in airway pressure and airflow induced by chest compressions during CPR may lead to abnormal inspiratory triggering and cause remarkably high ventilation rates. The effects of different inspiratory triggering sensitivity settings of ventilators on actual ventilation during CPR was seldom reported. Additionally, we wanted to see if no triggering was of any benefit in comparison to pressure- or flow- triggering. In this study we investigated the influence of different inspiratory triggering settings using typical hospital-based ventilators on actual ventilation, gas exchange and hemodynamics in a porcine model of cardiac arrest.

Materials and methods

Animal preparation

With approval of the ethics committee for animal experiments at Peking Union Medical College Hospital, the study was performed on a total of 18 domestic pigs (weight, 31.1±1.8 kg).

The animals were fasted overnight and premedicated with one dose of intramuscular pentobarbital sodium (30 mg/kg), diazepam (0.3 mg/kg) and atropine (0.02 mg/kg). The animals were then placed supine in a U-shaped fixing frame. An intravenous catheter was inserted into a lateral ear vein and followed by propofol infusion (2–3 mg/kg/hr) to maintain anesthesia. Each pig was intubated with a size 6.5 endotracheal tube and then mechanically ventilated with room air, using volume-control mode (Hamilton-G5, Hamilton Medical AG; Bonaduz, Switzerland), with a tidal volume of 10 mL/kg and a RR adjusted to maintain partial pressure of end-tidal carbon dioxide (PETCO2) at 35~40 mmHg and pulse oxygen saturation > 95%. Arterial blood gases (Gem 3000; Instrumentation Laboratory, Bedford, MA) were analyzed to verify the baseline conditions. A warming blanket was used to maintain the animal’s temperature between 36–38°C.

The right internal jugular vein and left common carotid artery were surgically exposed and cannulated for right atrial pressure (RAP) and central aortic blood pressure (AoP) measurements, as well as blood sampling. The left external jugular vein was also fitted with a catheter which provided access for ventricular fibrillation induction. Normal saline solution at 5ml/kg/hr was infused prior to the induction of ventricular fibrillation to maintain RAP between 3 and 5 mm Hg. Electrocardiograms, AoP and RAP were continuously monitored and recorded by a T8 Mindray monitor (Mindray Biological Medical Electronic Co, Ltd, Shenzhen, China). Coronary perfusion pressure (CPP) was calculated as the gradient between AoP and RAP during the decompression phase of chest compressions. RR, minute ventilation volume and airway pressure were continuously recorded by a BIOPAC MP-150 system via a piezometric tube and a flow sensor linked to the endotracheal tube.

Experimental protocol

Ventricular fibrillation was induced by 24V/50Hz AC current with a right ventricular internal-pacing electrode. After 3 minutes of untreated ventricular fibrillation without any ventilation, continuous chest compressions via a mechanical CPR device (WISH-SL-FS-A, Wuhan, China) were started at a rate of 100 compressions/min and a depth of 30% of the anteroposterior chest diameter. Mechanical ventilation (Hamilton-G5) was performed simultaneously for three periods of 5 minutes each, using a volume-controlled mode with a constant flow of 30L/min, zero end-expiratory pressure, FiO2 1.0, RR 10/min, tidal volume of 10 mL/kg, I:E 1:2, and the upper airway pressure limit was set to 60 cmH2O (Fig 1). According to the different inspiratory triggering sensitivity settings, the animals were divided into three groups. In the first group of six pigs (Group trig-4), three periods (5 minutes each) of mechanical ventilation were characterized by a pressure-triggering sensitivity of 4 cmH2O, a flow-triggering sensitivity of 4 L/min, and turned-off triggering respectively. Similar to Group trig-4, six pigs in Group trig-10 were ventilated with a pressure-triggering of 10 cmH2O, a flow-triggering of 10 L/min and turned-off triggering, while six pigs in Group trig-20 were ventilated with a pressure-triggering of 20 cmH2O, a flow-triggering of 20 L/min and turned-off triggering. Six sequences in each group were generated due to random assortment of the three different triggering settings, and each animal in each group was ventilated in accordance with one of the six sequences. Arterial and central venous blood gas samples were drawn at baseline and at the endpoint of each 5 minute ventilation period. The animals were finally sacrificed by infusion of potassium chloride at the end of the experiment. Respiratory and hemodynamic parameters at last minute of each period were collected, and the mean values were analyzed.

  1. No.1 Turned-off ---- Pressure-triggering ---- Flow-triggering
  2. No.2 Turned-off ---- Flow-triggering ---- Pressure-triggering
  3. No.3 Pressure-triggering ---- Turned-off ---- Flow-triggering
  4. No.4 Pressure-triggering ---- Flow-triggering ---- Turned-off
  5. No.5 Flow-triggering ---- Turned-off ---- Pressure-triggering
  6. No.6 Flow-triggering ---- Pressure-triggering ---- Turned-off

Statistical analysis

Non-parametric tests were applied in this study due to the small sample size. Statistical analysis was performed with SPSS 17.0 for Windows (SPSS, Inc.), using the Mann-Whitney U test or Kruskal-Wallis H test. Continuous variables are expressed as median (25th/75th percentiles) except the animals’ weight. Statistical significance was fixed at a P value of less than 0.05.

Results

Data from all 18 animals were included in the analysis. The results of respiratory parameters, hemodynamic parameters, and blood gases analyses were similar in the three groups at baseline (Table 1).

In each group, turned-off triggering provided a RR identical to the set value, while the pressure- and flow-triggering could induce significantly higher actual RR with the exception of a pressure-trigger of 20 cmH2O (all P<0.05, Table 2). Except for the pressure-trigger of 20 cmH2O, significantly higher minute ventilation volumes, mean airway pressures, and lower PETCO2, AoP and CPP were also induced by the pressure- and flow-triggering settings compared to turned-off triggering (all P<0.05, Fig 2).

thumbnail
Fig 2. Aortic blood pressure, right atrial pressure and mean airway pressure during the inspiratory phase of an animal in Group trig-20.

The period indicated by dotted box is the decompression phase of chest compressions. Coronary perfusion (shadow area) is indicated by the area between red and blue line in the dotted box. (A), Turned-off triggering. (B), A pressure-triggering of 20 cmH2O. (C), A flow-triggering of 20 L/min.

https://doi.org/10.1371/journal.pone.0171869.g002

In each group, significantly lower arterial pH, PaO2, and higher PaCO2 were observed in the periods with turned-off triggering compared to those with pressure- or flow- triggering (all P<0.05), except in the comparison with pressure triggering in Group trig-20 (Table 3). The venous blood gas analyses showed that most of the differences of venous pH, PvO2 and PvCO2 between the different triggering settings were not statistically significant. Turned-off triggering was related to higher venous oxygen saturation (SvO2) than those of pressure- or flow- triggering settings in each group (all P<0.05), with the exception of a pressure-triggering of 20 cmH2O (46.5[32/56.8] % vs 41.5[33.5/48.5] %, P = 0.575).

In Group trig-20, the venous-arterial CO2 gradients for turned-off triggering, pressure-triggering and flow-triggering were 8(-0.25/10) mmHg, 15(2.5/26) mmHg and 31.5(29.8/35.3) mmHg respectively (Fig 3). The venous-arterial CO2 gradients for flow-triggering were significantly higher than those of turned-off triggering (P = 0.006). The comparison between turned-off triggering and pressure-triggering did not show statistical significance (P = 0.148).

thumbnail
Fig 3. Venous-arterial CO2 gradients in Group trig-20 (median, 25/75% percentiles, min/max) [mmHg].

https://doi.org/10.1371/journal.pone.0171869.g003

Discussion

This study investigated ventilation via a typical hospital ventilator using different inspiratory triggering settings during CPR. Compared to the turned-off patient triggering, most settings of pressure- or flow-triggering induced hyperventilation through a high RR, with the exception of pressure-triggering at 20 cmH2O. Though a higher PaO2 was observed, elevated mean airway pressures and deteriorating CPP, PETCO2 and SvO2 seen with these settings indicated adverse prognoses. A pressure-triggering of 20 cmH2O had similar ventilation volume, gas exchange and hemodynamics as compared to turned-off triggering, and could be an alternative choice if an ATV or turned-off patient triggering was not available in clinical practice.

Excessive ventilation is strongly suggested to be avoided in the current guidelines, due to its adverse effects increasing intrathoracic pressure, decreasing venous return, coronary perfusion, and diminishing cardiac output and survival [1, 13]. Avoidance of hyperventilation during CPR is a broad consensus, though clinical studies confirming the association between hyperventilation and cardiac arrest outcomes are limited. However, hyperventilation commonly occurs during CPR with RR exceeding 10 breaths/min 63% of the time and exceeding 20 breaths/min 20% of the time, despite guideline recommendations [6]. The precipitating factors of hyperventilation are usually multifactorial including an adrenaline-driven arousal response, resuscitator inexperience, and CPR delivered at off-hours [6, 14, 15]. RR by manual ventilation could even reach 37 breaths/min with advanced airways, and providers often fail to ventilate at recommended rates even after retraining [1, 4, 16].

The use of a mechanical ventilator during CPR can theoretically eliminate hyperventilation by totally controlling ventilation rates and tidal volumes without patient triggering. An ATV is recommended in the guidelines as one of the ventilation alternatives for prolonged resuscitation efforts to provide adequate ventilation and allow resuscitators to perform other tasks [9], while the automatic mode of the oxygen-powered, flow-limited ventilator should not be used due to its adverse effects on venous return and forward blood flow [17]. The mechanical properties of typical hospital-based ventilators are usually superior to transport ventilators, and these ordinary ventilators are often involved in CPR due to their availability and convenience in our clinical practice. However, patient-triggering cannot be turned off in most modern ventilators (e.g. Puritan Bennette 7200ae & 840, MAQUET servo-i, Drager infinity C500, and GE Engstrom carestation, among others). The current guidelines do not give any specific ventilatory recommendations except for the RR with a secured airway, and, to the best of our knowledge, there are no studies focusing on different inspiratory triggering settings during CPR until now.

In general, the maximum adjustable range of pressure- or flow-triggering of most modern ventilators is 20 cmH2O or L/min. This study demonstrated that a pressure-triggering of 20 cmH2O rather than a flow-triggering of 20 L/min could effectively control hyperventilation. Because of design characteristics, flow-triggering provides more sensitive triggering and relatively shorter time delay than pressure-triggering [18, 19]. It’s reasonable to surmise from our study that airway pressures are usually between -10 and -20 cmH2O, while airflow is more than 20 L/min in most cases during the decompression phase of chest compressions. Adverse effects of hyperventilation were verified in this fundamental research again. High ventilation rates greater than 20 breaths/min induced high minute ventilation volume, increased mean airway pressure, and decreased CPP in this study.

However, contradictory viewpoints and research challenging the adverse effects of hyperventilation or high airway pressures during CPR have emerged in recent years [2022]. Higher mean airway pressure was found to be associated with higher ROSC in clinical settings [22]. A novel ventilation mode (Chest Compression Synchronized Ventilation, CCSV) which increases cyclic intrathoracic pressure differences and avoids negative airway pressure shows better effects on oxygenation and hemodynamics compared to IPPV [23]. Meanwhile, the need for artificial ventilation in the first minutes of CPR has also been challenged in the past several years [24]. Passive ventilation such as continuous positive airway pressure (CPAP) with pure oxygen via a Boussignac device or CPAP plus pressure support ventilation, was found to an effective alternative to mechanical ventilation with similar or superior gas exchange [25, 26]. Nevertheless, positive-pressure ventilation is still the “gold standard” of ventilation during CPR, though further investigations are needed to clarify the specific effects of ventilation volume and airway pressure on the outcomes of CPR.

As regards gas exchange in this study, high RR caused by pressure- and flow-triggering were correlated with decreased PETCO2, alkalosis and better arterial oxygenation. Arterial blood gas values were often used as measurable indicators of resuscitation effects in may studies of CPR [2729]. However, PaCO2 and PaO2 are simultaneously influenced by ventilation and pulmonary blood flow [30]. With significantly decreasing pulmonary blood flow (increasing ventilation/perfusion ratio), PaO2 increases and PaCO2 decreases, approaching the composition of inspired gas, which is opposite to conventional practice [30]. Similar to the current study, increased PaO2 and decreased PaCO2 during low blood flow states of CPR, may indicate not only adequate or excessive ventilation but also the negative effects of deteriorating blood flow. In general, oxygen delivery to the heart and brain is limited by blood flow rather than by arterial oxygen content during CPR [9, 31, 32].

Venous blood gases are thought to be more useful than arterial blood gases to assess perfusion during resuscitation, because they are less affected by ventilation changes [29, 30]. Actually, similar values of venous blood gases analysis were observed in most groups in our study, except that controlled ventilation rates with turned-off triggering or pressure-triggering of 20 cmH2O showed better SvO2. However, the same phenomenon was also observed in human CPR [33]. Subjects with restoration of spontaneous circulation (ROSC) demonstrated significantly different values of venous PH, PO2, and PCO2 during resuscitation, while a higher SvO2 indicated high rate of ROSC from cardiac arrest.

In addition to decreased CPP and SvO2, flow-triggering also correlated with increased venous-arterial CO2 gradients. Low blood flow states such as during chest compressions or shock, PaCO2 positively correlates with blood flow while venous PCO2 negatively correlates with blood flow [29, 34]. The increased venous-arterial CO2 gradients in this study indicated decreased blood flow during CPR, and suggest again that hyperventilation due to flow-triggering deteriorates cardiac output and tissue blood flow.

Our study has several limitations. First, the sample of animals was relatively small and outcomes of resuscitation efforts such as ROSC could not be detected and compared because of the cross-over design. Second, 5 minutes for each ventilation period was relatively short. That was chosen because the differences of respiratory parameters, gas exchange and hemodynamics between different triggering settings were significant after a ventilation duration of 5 minutes, and prolonged CPR would induce rapid deterioration of artificial circulation. Finally, an ATV was not involved in our study to serve as a point of comparison with our hospital-based ventilators. When the triggering of a ventilator is turned off, its ventilation process is basically identical to that of an ATV. Furthermore, the purpose of this study was to evaluate the feasibility of using different inspiratory triggering settings in an ordinary ventilator during CPR.

In conclusion, for ordinary automated ventilators found in inpatient hospital settings, pressure-triggering of 4 or 10 cmH2O and flow-triggering are improper for ventilation during CPR, due to their induction of hyperventilation leading to the deterioration in gas exchange and hemodynamics. Turned-off patient triggering is preferred, and a pressure-triggering of 20 cmH2O could be a viable alternative if turned-off triggering is not available.

Supporting information

S1 File. Relevant data underlying the findings described in manuscript.

https://doi.org/10.1371/journal.pone.0171869.s001

(XLSX)

Acknowledgments

We thank the Institute of Life Monitoring, Mindray Corporation for providing the T8 monitors used in this study. Mindray Corporation did not have any role in the study design, data collection, data analysis, preparation of the manuscript, or decision to publish.

Author Contributions

  1. Conceptualization: HZ XY DT JX JW.
  2. Formal analysis: DT JX SS FS YH YZ.
  3. Funding acquisition: XY DT.
  4. Investigation: DT SS YF FS YH YZ.
  5. Methodology: DT JX SS YF.
  6. Project administration: XY HZ.
  7. Resources: XY DT JX HZ JW.
  8. Writing – original draft: DT JX SS HZ XY.
  9. Writing – review & editing: DT JX SS HZ XY JW.

References

  1. 1. Aufderheide TP, Sigurdsson G, Pirrallo RG, Yannopoulos D, McKnite S, von Briesen C, et al. Hyperventilation-induced hypotension during cardiopulmonary resuscitation. Circulation. 2004;109(16):1960–5. pmid:15066941
  2. 2. Aufderheide TP, Lurie KG. Death by hyperventilation: a common and life-threatening problem during cardiopulmonary resuscitation. Crit Care Med. 2004;32(9 Suppl):S345–51. Epub 2004/10/29. pmid:15508657
  3. 3. Cheifetz IM, Craig DM, Quick G, McGovern JJ, Cannon ML, Ungerleider RM, et al. Increasing tidal volumes and pulmonary overdistention adversely affect pulmonary vascular mechanics and cardiac output in a pediatric swine model. Crit Care Med. 1998;26(4):710–6. Epub 1998/04/29. pmid:9559609
  4. 4. O'Neill JF, Deakin CD. Do we hyperventilate cardiac arrest patients? Resuscitation. 2007;73(1):82–5. pmid:17289248
  5. 5. Abella BS, Alvarado JP, Myklebust H, Edelson DP, Barry A, O'Hearn N, et al. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. JAMA. 2005;293(3):305–10. Epub 2005/01/20. pmid:15657323
  6. 6. McInnes AD, Sutton RM, Orioles A, Nishisaki A, Niles D, Abella BS, et al. The first quantitative report of ventilation rate during in-hospital resuscitation of older children and adolescents. Resuscitation. 2011;82(8):1025–9. Epub 2011/04/19. PubMed Central PMCID: PMCPMC3703745. pmid:21497007
  7. 7. Johannigman JA, Branson RD, Johnson DJ, Davis K Jr., Hurst JM. Out-of-hospital ventilation: bag—valve device vs transport ventilator. Acad Emerg Med. 1995;2(8):719–24. Epub 1995/08/01.
  8. 8. Salas N, Wisor B, Agazio J, Branson R, Austin PN. Comparison of ventilation and cardiac compressions using the Impact Model 730 automatic transport ventilator compared to a conventional bag valve with a facemask in a model of adult cardiopulmonary arrest. Resuscitation. 2007;74(1):94–101. Epub 2007/02/09. pmid:17287062
  9. 9. Link MS, Berkow LC, Kudenchuk PJ, Halperin HR, Hess EP, Moitra VK, et al. Part 7: Adult Advanced Cardiovascular Life Support: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18 Suppl 2):S444–64. Epub 2015/10/17.
  10. 10. Weiss SJ, Ernst AA, Jones R, Ong M, Filbrun T, Augustin C, et al. Automatic transport ventilator versus bag valve in the EMS setting: a prospective, randomized trial. South Med J. 2005;98(10):970–6. Epub 2005/11/22. pmid:16295811
  11. 11. Vt Hurst, West S, Austin P, Branson R, Beck G. Comparison of ventilation and chest compression performance by bystanders using the Impact Model 730 ventilator and a conventional bag valve with mask in a model of adult cardiopulmonary arrest. Resuscitation. 2007;73(1):123–30. Epub 2006/12/19. pmid:17175090
  12. 12. Sassoon CS, Gruer SE. Characteristics of the ventilator pressure- and flow-trigger variables. Intensive Care Med. 1995;21(2):159–68. Epub 1995/02/01. pmid:7775698
  13. 13. Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ, et al. Part 5: Adult Basic Life Support and Cardiopulmonary Resuscitation Quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18 Suppl 2):S414–35. Epub 2015/10/17.
  14. 14. Pitts S, Kellermann AL. Hyperventilation during cardiac arrest. Lancet. 2004;364(9431):313–5. Epub 2004/07/28. pmid:15276374
  15. 15. Park SO, Shin DH, Baek KJ, Hong DY, Kim EJ, Kim SC, et al. A clinical observational study analysing the factors associated with hyperventilation during actual cardiopulmonary resuscitation in the emergency department. Resuscitation. 2013;84(3):298–303. Epub 2012/08/14. pmid:22885095
  16. 16. Nikolla D, Lewandowski T, Carlson J. Mitigating hyperventilation during cardiopulmonary resuscitation. Am J Emerg Med. 2015. Epub 2016/01/08.
  17. 17. Brooks SC, Anderson ML, Bruder E, Daya MR, Gaffney A, Otto CW, et al. Part 6: Alternative Techniques and Ancillary Devices for Cardiopulmonary Resuscitation: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18 Suppl 2):S436–43. Epub 2015/10/17.
  18. 18. Sassoon CS. Mechanical ventilator design and function: the trigger variable. Respir Care. 1992;37(9):1056–69. Epub 1992/09/01. pmid:10145700
  19. 19. Dimitriou G, Greenough A, Laubscher B, Yamaguchi N. Comparison of airway pressure-triggered and airflow-triggered ventilation in very immature infants. Acta Paediatr. 1998;87(12):1256–60. Epub 1999/01/23. pmid:9894826
  20. 20. Gazmuri RJ, Ayoub IM, Radhakrishnan J, Motl J, Upadhyaya MP. Clinically plausible hyperventilation does not exert adverse hemodynamic effects during CPR but markedly reduces end-tidal PCO(2). Resuscitation. 2012;83(2):259–64. Epub 2011/08/23. pmid:21854734
  21. 21. Chalkias A, Xanthos T. Timing positive-pressure ventilation during chest compression: the key to improving the thoracic pump? Eur Heart J Acute Cardiovasc Care. 2015;4(1):24–7. Epub 2014/01/02. pmid:24381094
  22. 22. Chalkias A, Pavlopoulos F, Koutsovasilis A, d'Aloja E, Xanthos T. Airway pressure and outcome of out-of-hospital cardiac arrest: A prospective observational study. Resuscitation. 2017;110:101–6. Epub 2016/11/15. pmid:27840003
  23. 23. Kill C, Galbas M, Neuhaus C, Hahn O, Wallot P, Kesper K, et al. Chest Compression Synchronized Ventilation versus Intermitted Positive Pressure Ventilation during Cardiopulmonary Resuscitation in a Pig Model. PLoS One. 2015;10(5):e0127759. Epub 2015/05/27. PubMed Central PMCID: PMCPMC4444197. pmid:26011525
  24. 24. Hupfl M, Selig HF, Nagele P. Chest-compression-only versus standard cardiopulmonary resuscitation: a meta-analysis. Lancet. 2010;376(9752):1552–7. Epub 2010/10/19. PubMed Central PMCID: PMCPMC2987687. pmid:20951422
  25. 25. Bertrand C, Hemery F, Carli P, Goldstein P, Espesson C, Ruttimann M, et al. Constant flow insufflation of oxygen as the sole mode of ventilation during out-of-hospital cardiac arrest. Intensive Care Med. 2006;32(6):843–51. Epub 2006/05/23. pmid:16715326
  26. 26. Kleinsasser A, Lindner KH, Schaefer A, Loeckinger A. Decompression-triggered positive-pressure ventilation during cardiopulmonary resuscitation improves pulmonary gas exchange and oxygen uptake. Circulation. 2002;106(3):373–8. Epub 2002/07/18. pmid:12119256
  27. 27. Kill C, Hahn O, Dietz F, Neuhaus C, Schwarz S, Mahling R, et al. Mechanical ventilation during cardiopulmonary resuscitation with intermittent positive-pressure ventilation, bilevel ventilation, or chest compression synchronized ventilation in a pig model. Crit Care Med. 2014;42(2):e89–95. Epub 2013/10/26. pmid:24158168
  28. 28. Lurie KG, Yannopoulos D, McKnite SH, Herman ML, Idris AH, Nadkarni VM, et al. Comparison of a 10-breaths-per-minute versus a 2-breaths-per-minute strategy during cardiopulmonary resuscitation in a porcine model of cardiac arrest. Respir Care. 2008;53(7):862–70. Epub 2008/07/03. pmid:18593487
  29. 29. Cavus E, Meybohm P, Bein B, Steinfath M, Poppel A, Wenzel V, et al. Impact of different compression-ventilation ratios during basic life support cardiopulmonary resuscitation. Resuscitation. 2008;79(1):118–24. Epub 2008/07/01. pmid:18586375
  30. 30. Idris AH, Staples ED, O'Brien DJ, Melker RJ, Rush WJ, Del Duca KD, et al. Effect of ventilation on acid-base balance and oxygenation in low blood-flow states. Crit Care Med. 1994;22(11):1827–34. Epub 1994/11/01. pmid:7956288
  31. 31. Chardoli M, Heidari F, Rabiee H, Sharif-Alhoseini M, Shokoohi H, Rahimi-Movaghar V. Echocardiography integrated ACLS protocol versus conventional cardiopulmonary resuscitation in patients with pulseless electrical activity cardiac arrest. Chin J Traumatol. 2012;15(5):284–7. Epub 2012/10/17. pmid:23069099
  32. 32. Ornato JP, Garnett AR, Glauser FL. Relationship between cardiac output and the end-tidal carbon dioxide tension. Ann Emerg Med. 1990;19(10):1104–6. Epub 1990/10/01. pmid:2121075
  33. 33. Rivers EP, Martin GB, Smithline H, Rady MY, Schultz CH, Goetting MG, et al. The clinical implications of continuous central venous oxygen saturation during human CPR. Ann Emerg Med. 1992;21(9):1094–101. Epub 1992/09/01. pmid:1514720
  34. 34. Mallat J, Pepy F, Lemyze M, Gasan G, Vangrunderbeeck N, Tronchon L, et al. Central venous-to-arterial carbon dioxide partial pressure difference in early resuscitation from septic shock: a prospective observational study. Eur J Anaesthesiol. 2014;31(7):371–80. Epub 2014/03/15. pmid:24625464