Rotatable Small Permanent Magnet Array for Ultra-Low Field Nuclear Magnetic Resonance Instrumentation: A Concept Study

Abstract

Object

We studied the feasibility of generating the variable magnetic fields required for ultra-low field nuclear magnetic resonance relaxometry with dynamically adjustable permanent magnets. Our motivation was to substitute traditional electromagnets by distributed permanent magnets, increasing system portability.

Materials and Methods

The finite element method (COMSOL®) was employed for the numerical study of a small permanent magnet array to calculate achievable magnetic field strength, homogeneity, switching time and magnetic forces. A manually operated prototype was simulated and constructed to validate the numerical approach and to verify the generated magnetic field.

Results

A concentric small permanent magnet array can be used to generate strong sample pre-polarisation and variable measurement fields for ultra-low field relaxometry via simple prescribed magnet rotations. Using the array, it is possible to achieve a pre-polarisation field strength above 100 mT and variable measurement fields ranging from 20–50 μT with 200 ppm absolute field homogeneity within a field-of-view of 5 x 5 x 5 cubic centimetres.

Conclusions

A dynamic small permanent magnet array can generate multiple highly homogeneous magnetic fields required in ultra-low field nuclear magnetic resonance (NMR) and magnetic resonance imaging (MRI) instruments. This design can significantly reduce the volume and energy requirements of traditional systems based on electromagnets, improving portability considerably.

Introduction

Nuclear magnetic resonance (NMR) spectroscopy and magnetic resonance imaging (MRI) are non-invasive and non-destructive investigative tools that can provide information from the molecular to the macroscopic scale. These techniques harness the phenomenon of magnetic resonance due to the interaction, within a magnetic field, between precessing nuclear magnetic moments (nuclear spin systems) and electromagnetic radiation. NMR/MRI have a wide range of applications in materials science, structural biology, chemistry and medical imaging [13].

Conventional MRI instruments comprise three main components: a superconducting magnet to align the nuclear spins and generate net sample magnetisation; a transmitter/receiver coil system that radiates electromagnetic energy to the nuclear spin system and detects the NMR signal; and gradient coils that enable the encoding of spatial information allowing the generation of three dimensional images [2].

The signal-to-noise ratio (SNR) achieved in NMR/MRI is proportional to the magnitude of net sample magnetisation. Hence, the quality of NMR/MRI data is dependent on the strength and homogeneity of the main magnetic field (commonly referred to as the B0 field). Superconducting magnets have been utilised to increase field strength. These increase the bulk and cost of purchase, operation and maintenance of NMR/MRI instruments.

Partly in response to these drawbacks, over the last decade there has been growing interest in ultra-low magnetic field (ULF) NMR/MRI, which uses a main magnet field strength of less than 10 mT [414]. Potential advantages of ULF over high field NMR/MRI instruments include greater absolute magnetic field homogeneity, simple and low cost instrumentation and low power consumption [15]. ULF NMR/MRI offers the possibility of important new applications such as the ability to image in the presence of metal, for example in trauma, disaster and battlefield applications. At ULF, the Larmor frequency overlaps with a range of molecular and physiological processes such as protein folding, slow diffusion, molecular tumbling and enzyme catalysis which are difficult to observe at high field because of the large frequency mismatch [13]. This raises the possibility of new imaging paradigms sensitised to these processes. In addition, because superconducting magnets are not required, the instruments may be more portable, allowing ULF instruments to be used in remote locations [13].

Although based on the same fundamental principles of magnetic resonance as high field NMR/MRI, ULF instruments are set up differently. Prior to the measurement, sample magnetisation is generated by a pulsed magnetic field approximately three orders of magnitude higher (~0.05–0.1 T) than the Earth’s field. This technique is known as sample pre-polarisation and is a key strategy in ULF research to overcome low SNR which still severely restricts ULF-NMR/MRI applications [5, 9, 13]. Highly sensitive magnetometers are also used to increase SNR. Often, excitation pulses are not used to trigger the ULF-NMR/MRI signal. Instead, the ULF-NMR/MRI signal is generated and detected in the presence of a second magnetic field, the measurement field, that is perpendicular to the pre-polarisation field.

In most instruments, the magnetic fields in ULF-NMR/MRI instruments are generated using resistive coils, which have high power consumption and heat production [1113, 16]. Moreover, the presence of highly conductive materials in resistive coils contributes to signal loss due to sample heating effects, residual coil noise, transients and eddy currents, and destructive interference effects [12, 13, 17]. Vesanen et al. addressed these problems by using superconducting coils [18]. Here we propose a permanent magnet solution using Halbach arrays,

Halbach arrays are a versatile arrangement of permanent magnets that can be used to generate strong, highly homogeneous magnetic fields in a field of view (FOV) that is small compared to overall array volume [1921]. Halbach array technology has led to the development of new generations of benchtop and handheld NMR instruments, with field strength above 3T and inhomogeneity of less than 0.01% (100 ppm) [19, 2224].

Permanent magnets do not require electric current flow to generate magnets fields. Hence, sample heating due to energy dissipation in resistive material is avoided, cooling devices obviated and power consumption significantly reduced compared to resistive coil technology. Moreover, the conductivity of the material used in magnets is much lower than materials like copper used in resistive coils. Hence, eddy current effects from rapid changes in magnetic field, which can lead to signal artefacts and noise, are reduced.

A Halbach array has recently been introduced in a prototype portable MRI scanner suitable for human brains (FOV = 16 cm) [25]. The array was used to generate a static field (B0 = 77 mT) within the volume which was quite inhomogeneous (1% inhomogeneity; ~10000 ppm) due to presence of fringe fields originating from both ends of the Halbach array which has a length and diameter of ~ 36 cm. The B0 inhomogeneity was exploited to encode spatial information and rotation of the Halbach array about the sample was required to generate a 2D image [25].

In all current designs of NMR/MRI instruments using Halbach arrays, resistive coils and excitation pulses are employed to generate signals and gradients. Our new approach is an extension of the concept of the Halbach array that obviates the need for resistive coils for low and ULF NMR/MRI by introducing a dynamic adjustable small permanent magnet array (SPMA) to generate and switch between multiple magnetic fields. This is achieved by prescribed rotations of individual magnets, allowing adjustments of magnetic field magnitudes and orientations.

In this study, we examined the potential of SPMAs for ULF-NMR relaxometry. By applying the finite element method (FEM) to Maxwell’s equation, we determined the magnetic fields generated by the dynamic SPMA and analysed it in terms of achievable field strength and field homogeneity. In addition, a manually operated SPMA was simulated and built to demonstrate the ability to generate the magnetic fields required for ULF relaxometry. An ULF relaxometry instrument for measuring the longitudinal (T1) and transversal relaxation (T2) time in a sample, requires two perpendicular and dynamic switchable magnetic fields: the pre-polarisation field Bp and the measurement field Bm. A schematic representation of the application of Bp and Bm to perform a basic ULF relaxometry measurement is shown in Fig 1: The pulsed Bp (Fig 1A) with magnitudes typically ranging from 30–70 milliteslas (mT) generates the net sample magnetisation M, according to Curie’s law (Fig 1B). The orientation and the magnitude of Bm defines the axis of precession of M and the Larmor frequency, respectively (Fig 1C). After Bp is switched off, the precession of M about Bm generates the sample signal known as free induction decay (FID) and is detected by the sensor (S) (Fig 1D). The duration and measurability of the FID depends on the demagnetisation characteristics of the sample and sensor location and orientation [17].

thumbnail
Fig 1. Schematic presentation of ULF relaxometry measurement.

(a) Pre-polarisation coil is switched on to generate Bp. (b) The net magnetization M reaches its maximum right before the pre-polarisation coil is switched off. (c) The measurement field Bm, perpendicular to Bp is switched on. (d) The net magnetisation vector M precesses about Bm and decays, the sample demagnetises. The localised magnetic field sensor (S) detects the sample signal (FID) during demagnetisation

https://doi.org/10.1371/journal.pone.0157040.g001

Materials and Methods

SPMA design

The dynamic SPMA for ULF relaxometry is shown in Fig 2. It consists of cylindrical magnets of finite length, transversely magnetised (i.e. in the x-y plane) arranged in three concentric cylindrical arrays as indicated in Fig 2A. Each magnet of array A was assumed to be individually pivot-mounted about its own axis in the z-coordinate direction to allow the generation of different magnetic field configurations by prescribed rotations of each magnet. In contrast, the orientation of each magnet within array B and C was fixed but the two arrays were able to rotate about the z-axis (see Fig 2B).

thumbnail
Fig 2. Setup of the dynamic SPMA model for ULF relaxometry.

(a) The SPMA model consists of three concentric cylindrical arrays with transversely (x-y plane) magnetised rods. Array A, required for pre-polarisation, consists of 24 magnets; Arrays B and C, required for generating the variable measurement field, consist of 12 magnets each. The origin of the coordinate system was positioned at the centre of the arrays and the axes of the arrays were aligned along the z-axis. In the Halbach configuration, each magnet has a specific orientation different to the neighbouring magnets. To study the magnetic forces, the magnets of array A were numbered counter-clockwise starting at the right-hand-side. (b) Side view indicating the concentric SPMA setup and sizes. Array A is fixed but each magnet rotates individually along its own axis in the z-coordinate direction (small red circular arrow). Arrays B and C (with fixed magnet orientation) rotate about the z-axis, indicated by the large red circular arrow

https://doi.org/10.1371/journal.pone.0157040.g002

Our SPMA design is based on strong, highly homogeneous magnetic fields generated by a Halbach dipole cylinder or array [23]. The three concentric cylindrical arrays, A, B and C generate the two mutually perpendicular magnetic fields required for ULF relaxometry measurements: Bp (array A) and Bm (array B and C). Throughout this paper it is assumed that Bp is oriented along the x-axis and Bm along the y-axis. Each cylindrical magnet within the SPMA is transversely magnetised (x-y plane) with remanent magnetisation Br. The following parameters were used when designing the array: field of view (FOV) within the centre of the SPMA 5 x 5 x 5 cm3; Bp magnitude >100 mT; Bm magnitude 20–50 μT (852–2130 Hz). The FOV was chosen to be sufficient for both ULF-NMR measurements and for a small ULF imaging device. Bm corresponds to the proton (1H) Larmor frequency ωL as determined by the Larmor equation (1) with the units for ωL being rad/sec and fL being Hz. γ is the gyromagnetic ratio, which is 267.5 x 106 rad/sec/T for protons.

Magnetisation pattern

Four different magnetisation patterns were considered in this study. They were created by prescribed rotations of each cylindrical magnet in array A (Fig 3), with Br of each magnet indicated by white arrows: Halbach (Fig 3A), reverse Halbach (Fig 3B), tangential (Fig 3C) and radial (Fig 3D). The Halbach pattern is known to achieve a strong homogeneous magnetic field in the centre of the array A, while the lowest field strength or field cancellation (indicated by irregular field distribution in the centre) is achieved with the reverse Halbach, tangential or radial magnetisation patterns. Different magnetisation patterns and numbers of permanent magnets in the array lead to different magnetic field distributions, field strength and homogeneity in the centre of the array. All of these variations affect suitability for ULF relaxometry measurements. A study on the impact of such effects is provided in the following sections.

thumbnail
Fig 3. Definition and visualisation of the magnetisation patterns considered in this study.

(a) Halbach, (b) reverse Halbach, (c) tangential and (d) radial. Shown as vector plots are the magnet remanent magnetisation (thick white arrows) and the normalized magnetic field distribution (thin white arrows) in the centre of the SPMA. The Halbach pattern leads to a strong, highly directional magnetic field, while the other patterns lead to non-directional, weak fields.

https://doi.org/10.1371/journal.pone.0157040.g003

Simulation environment

The complexity of the fields associated with different conformations of the SPMA precluded an analytical approach. Hence we undertook a rigorous numerical analysis using COMSOL® (version 4.3b, AC/DC module, Magnetostatic), a commercial finite element method (FEM) simulation environment with a computer-aided design interface for 3D model design. Simulations were carried out using an x64-based 16 core PC (Dell Precision T7600) with 128 GB of RAM. In the FEM simulation, the SPMA model was discretised in 3D-tetrahedral meshes using a predetermined mesh distribution and density that had been optimised using COMSOL. Mesh density was manually increased to achieve sub-millimetre spatial resolution in the centre of the array. The number of mesh points generally ranged between 40–50 million which ensured convergent and accurate results within reasonable time frames. The size of the cylindrically shaped computational window (diameter 1.2 m, length 2.1 m) was set to be sufficiently large to be able to model the SPMA (diameter 0.4 m, length 0.7 m) and to minimise numerical errors due to discontinuities. The standard COMSOL Neumann boundary condition was used to represent magnetic shielding implemented in a previously described ULF instrument developed at the Centre for Advanced Imaging [17], with thickness d = 12 cm and permeability μr = 5500. The relative permeability of the material in the magnets was set to 1.05 and for the surrounding environment (air) it was 1 (see S1 File).

Pre-polarisation array (array A)

The outer Array A which generates Bp, had an assumed fixed radius RA = 11cm and array length L = 70 cm. It consisted of n identical cylindrical magnets each with a remanent magnetic field strength (Br) of 1 Tesla (T). Notably, other commercially available magnet cross-sections lead to equivalent qualitative results provided that magnet size is small compared to the distance to the centre of the array. An ideal Halbach array is characterized by a continuous change in the azimuthal direction of the magnetisation vector, which at present cannot be achieved. Hence, cylindrical Halbach arrays are discretised by identical magnets with constant Br and the approximation improved by progressively increasing the number of magnets along the circumference. In this study n = 12, 16 and 24 magnets were considered.

The array radius RA was fixed to ensure that the SPMA fits within the magnetic shielding device. To allow the performance of SPMAs with different numbers of magnets to be compared while keeping the overall size constant, a fill factor was introduced. This factor quantifies the ratio of magnetic material to air gap (dm to da) along the circumference (see Fig 4). A fill factor of 0.75, for instance, corresponds to 75% occupancy by magnets (dm) and 25% by air (da) along the circumference. Magnet diameter dm is thus dependent on the total number of magnets n and is approximately: (2)

thumbnail
Fig 4. Section of an array with radius RA.

Each cylindrical magnet, numbered in a counter-clockwise direction and with diameter dm, is evenly arranged along the circumference to ensure equidistant air gaps. The fill factor is defined as the ratio between dm and da. In this example dm equals da and the fill factor is 0.5

https://doi.org/10.1371/journal.pone.0157040.g004

Measurement array (arrays B and C)

The measurement field Bm was generated by superimposing two magnetic fields generated by arrays B (BB) and C (BC) with radii RB = 8 cm and RC = 9 cm, respectively, each with the fixed Halbach pattern. As with the pre-polarisation array A, this arrangement ensured a highly directional and homogeneous magnetic field along the y-axis. Magnetic fields from the two arrays cancel within the FOV when (a) the magnetic field magnitudes BB and BC are matched and (b) their directions are opposite (Fig 5A). By rotating arrays B and C by the same angle but in opposite directions about the z-axis, the total magnetic field remains parallel to the y-axis and comprises the measurement field Bm (Fig 5B).

thumbnail
Fig 5. Principle of generating Bm simulated with COMSOL.

(a) Two concentric arrays B and C (RB = 9 cm and RC = 8 cm), each with Halbach magnetisation pattern (see Fig 3A), generate opposing magnetic fields BB and BC. If their magnitude is matched, the field is nearly cancelled in the centre. (b) By rotating array B and C respectively clockwise and counter-clockwise about the axis of symmetry of the SPMA (see red arrows), the x-component of the magnetic field remains cancelled but a y-component, Bm, is generated

https://doi.org/10.1371/journal.pone.0157040.g005

The most important magnetic field properties for NMR experiments are field strength, directionality and homogeneity [21]. In the case of the ULF relaxometer proposed here the qualitative targets to optimize the SNR were:

  • uniform field direction and maximum magnitude of Bp during sample pre-polarisation (switched on).
  • minimum magnitude for Bp during the measurement period.
  • uniform field direction and homogeneous variable measurement field Bm (arrays B and C).

Manual SPMA

A manually operated SPMA was built to demonstrate the generation, cancellation and regulation of Bp and Bm. Like the SPMA for ULF relaxometry in the numerical study, the prototype consisted of three concentric arrays to generate the pre-polarisation field, Bp (array D), and the measurement field, Bm (array E and F), see Fig 6. The manually adjustable arrays were composed of ferrite permanent magnets of rectangular cross-section (ferrite grade Y30BH, Br = 0.39 T, AMF Magnetics, Australia) chosen for their cost-effectiveness and availability and for simplicity of construction. Array D (Fig 6A, array I and II) had a radius RD of 15 cm and comprises 12 magnets (15L x 2.5W x 2.54H cm) equally spaced around the circumference achieving a fill factor of 0.32. Twelve magnets (15L x 1.2W x 0.6H cm) are used in array E (RE = 10.5 cm, Fig 6A, array III) and 6 magnets (15L x 1.2W x 0.6H cm) in array F (RF = 7.5 cm, Fig 6A, array IV), leading to fill factors of 0.16 and 0.11, respectively. For each array, the magnets were pressure-fitted in pairs of medium density fibreboard (MDF) rings that hold the ends of the bar magnets. The moderate magnetisation of the ferrite magnets allowed safe manual rotations of the arrays. MDF rings of different diameter fitted within each other keeping the relative position of each array in place (Fig 6B). Two manually interchangeable frames for array D were built to hold the magnets for two different configurations: tangential (to generate Bm) and Halbach (to generate Bp); see Fig 6A, arrays I and II.

thumbnail
Fig 6. SPMA prototype.

(a) Elements of the SPMA prototype shown separately: Array D with Halbach pattern (I) and tangential pattern (II), array E (III) array F (IV) each with Halbach pattern. Array D magnets are fitted in the MDF frame I or II to achieve Bp ON or Bp OFF configurations, respectively. (b) Arrays E (III) and F (IV) fitted inside array D with tangential pattern (II). Bm magnitude control is achieved by rotating arrays E (III) and F (IV) in opposite directions with prescribed angles α. The white arrows indicate the magnetisation direction of each magnet

https://doi.org/10.1371/journal.pone.0157040.g006

A Gaussmeter (F.W. Bell, Model 5080, Milwaukee, USA) mounted on a custom built 3-axis adjustable Cartesian holder was used to obtain magnetic field measurements in a grid (5 x 5 x 3 points in, x, y, z) covering the FOV (5 x 5 x 5 cm3) for 2 different angular settings (α = 0°, and 10°, see Fig 6B).

Results

I. SPMA for ULF relaxometry–Numerical simulation

Array A during pre-polarisation (‘switched on’).

The magnetic flux density distribution of Array A during pre-polarisation (Bp) generated with the Halbach pattern (see Fig 3A), with magnet properties L = 70 cm, dm = 2.16 cm, Br = 1 T, is presented as x-y cross section plot at the centre of the array (z = 0) in Fig 7A. In the same figure, an arrow plot shows the local field direction of Bp parallel to the x-axis, and the FOV is visualised by light grey disks. In this simulation, arrays B and C are not included since Bp produced by array A is more than 1000 times larger than Bm. The magnetic field distortion due to differences in electric permeability can be ignored because the relative permeability of the permanent magnets is very similar to that of the air (1.05 versus 1.00).

thumbnail
Fig 7. Magnitude and direction plots of magnetic flux density of a SPMA.

Magnetic flux (a) during pre-polarisation (Bp) and (d) measurement state (Bm). Cross-sectional plots through the point of origin along the x-axis (b) and y-axis (c) of the ratio Bpy/Bp (red) and Bpz/Bp (blue). Bpy and Bpz are the y- and z-components of the pre-polarisation field Bp (= Bpx). Bpy and Bpz are at least six orders of magnitude smaller in all directions within the FOV. Plots along the z-axis were omitted since all ratios are well below 10−9. (e) Cross-sections of Bmx/Bm and Bmz/Bm along the x-axis (blue), y-axis (red) and z-axis (black) demonstrating the x and z-component of the resultant magnetic field, generated by arrays A, B and C, are at least three orders of magnitude smaller than Bm (f). Percent deviation from measurement field magnitude of Bm, plotted along the x-axis (curve 1, blue), y-axis (curve 2, red) and z-axis (curve 3, black). Arrays B and C were rotated by ~4.5° to achieve a magnitude of 40 μT

https://doi.org/10.1371/journal.pone.0157040.g007

The directionality and homogeneity of Bp was assessed by the ratio of the minor field component (Bpy, Bpz, as the y and z components of the Bp field, respectively) to the main component of Bp (along the x-axis). Fig 7B and 7C show cross-sectional plots of the magnetic flux density ratios Bpy/Bp (red) and Bpz/Bp (blue) along x and y, demonstrating that Bpy and Bpz were at least six orders of magnitude smaller than Bp, in keeping with the high directionality of the latter along the x-axis. Plots along the z-axis are omitted since the ratios and the minor components are even smaller than Bpy and Bpz.

Twelve, 16 and 24 magnets with a fill factor of 0.75, 12 magnets with a fill factor of 0.375 and 16 magnets with a fill factor of 0.5 were analysed to study the effect of these parameters on the pre-polarisation field. The main results are summarized in Table 1 and the field magnitude along the x-axis is shown in Fig 8. The solid line plots correspond to 12 magnets, dashed line plots to 16 magnets and dash-dotted plots to 24 magnets. For a constant fill factor of 0.75, the achievable field strength increases with decreasing magnet number, as illustrated in Fig 8A, because of the greater magnet surface and volume, according to Eq 2. In contrast, by reducing the fill factor or the number of magnets, the pre-polarisation field strength decreased if magnet size was kept constant, as shown in Fig 8B. However, in all cases the magnetic field variation of Bp remained below 0.023% (230 ppm), as demonstrated in Fig 8C and 8D. The high field homogeneity is due to the combination of a small FOV compared to the large volume of the SPMA. Fig 9 illustrates the relative variation in magnitude of Bp in three dimensions for n = 24 magnets (fill factor 0.75) with respect to the magnitude at the centre plotted along the z-axis in 2 cm steps on the x-z (Fig 9A) and y-z planes (Fig 9B) demonstrating the high field homogeneity within the whole volume of the FOV.

thumbnail
Fig 8. 2D cross-sectional plots of pre-polarisation field Bp along the x-axis (switched on).

For array A with constant fill factor, curve 1 (solid line) corresponds to 12 magnets, curve 2 (dashed line) to 16 and curve 3 (dash-dotted line) to 24 magnets. For array A with constant magnet dimensions (L = 70 cm, dm = 2.16 cm), curve 4 (dash-dotted line) corresponds to 24 magnets, curve 5 (dashed line) to 16 magnets and curve 6 (solid line) to 12 magnets. (a) In array A with constant fill factor 0.75, the field strength within the field of view (FOV) decreases with the number of magnets, since magnet volume and surface area increase. (b) For array A with constant magnet size, the field strength decreases with decreasing numbers of magnets. (c) Within the FOV the field inhomogeneity decreases slightly with decreasing numbers of magnets for both constant fill factor (c) and magnet size (d). In all cases, the field inhomogeneity within the FOV was well below 0.02% (200 ppm)

https://doi.org/10.1371/journal.pone.0157040.g008

thumbnail
Fig 9. Relative magnitude variation of the pre-polarisation field, Bp, generated by array A with 24 magnets.

Field inhomogeneity shown as line plots in z = 2 cm steps along the x axis (a) and y axis (b). Plotted are percent magnitude deviations from the magnitude of Bp at the centre of the array. Within the chosen FOV of 5 x 5 x 5 cm3, the inhomogeneity is less than 0.02% in all cases

https://doi.org/10.1371/journal.pone.0157040.g009

thumbnail
Table 1. Achievable magnetic field strength at the centre of array A and field inhomogeneity within the field of view (FOV) during pre-polarisation for varying number of magnets and fill factors calculated with COMSOL.

https://doi.org/10.1371/journal.pone.0157040.t001

Array A after pre-polarisation (‘switched off’).

During the measurement period, the magnetic field within the centre of the SPMA produced by array A must be minimised to enable precession and relaxation of the magnetisation vector of the sample. This is achieved by rotating each magnet of array A from the Halbach pattern (Fig 3A) to one of the three possible magnetisation patterns considered in this study, the reverse Halbach (Fig 3B), tangential (Fig 3C) or radial (Fig 3D) pattern. Fig 10 illustrates cross-sectional plots of the magnetic flux density generated by each pattern with the FOV indicated by the grey shaded area. Only the tangential pattern enabled field cancellation to magnitudes below 1 μT (fill factor = 0.75) within the FOV (Fig 10C). The residual magnetic field magnitude generated by array A is low enough (< 10−8 T for 16 or 24 magnets, see Fig 10C) not to interfere with the measurement field Bm (20–50 μT). Arrays B and C were excluded from the simulations in this section. The magnitude of the magnetic flux density generated by array A was below 1 μT in a larger area with a larger number of magnets (see Fig 10C). This is expected since the SPMA conforms better to an ideal Halbach array as the number of magnets increases. A similar effect was also observed during pre-polarisation with the Halbach pattern (Fig 8). From Fig 10C, it can be concluded that 16 (curve 2) or 24 cylindrical magnets (curve 3) are suitable to ensure minimum interference from array A during the measurement period.

thumbnail
Fig 10. Magnetic flux density plots of array A along the x-axis for different magnetisation patterns, corresponding to Fig 4.

In each array, 12 (solid line 1), 16 (dashed line 2) and 24 (dash-dotted line 3) magnets (L = 70 cm, dm = 2.16 cm, Br = 1 T, fill factor = 0.75) are considered for the (a) reverse Halbach, (b) radial and (c) tangential patterns. The FOV is indicated by the shaded area. Only the tangential pattern (c) is able to cancel the magnetic field within the FOV to magnitudes below 1 μT

https://doi.org/10.1371/journal.pone.0157040.g010

The inaccuracy in the angular position of the permanent magnets of array A in the tangential configuration induces a field inhomogeneity of about 90μT/degree. This indicates that high precision actuators of less than 40 arcsecond tolerance would be required to keep the Bp field below 2μT.

Energy considerations for array A.

Within the array, strong permanent magnets may generate considerable repulsive and attractive forces. We examined the total magnetic energy within array A as an index of how much energy was needed to change each permanent magnet in array A from pre-polarisation (‘switched on’) to measurement (‘switched off’) configurations. This parameter also relates to the mechanical stability of the array and may influence switching time. Since arrays B and C were of much lower volume and remanent magnetisation of the component magnets, their contribution was not considered.

Table 2 lists the total magnetic energy contained in array A for different magnetisation patterns and numbers of magnets. The total magnetic energy increased with as the number of magnets decreased due to the higher magnet volume and surface area. The highest total magnetic energy for a given number of magnets was achieved with the radial pattern and the lowest with the tangential pattern. With the radial pattern, each magnet experienced repelling forces only (magnetisation vector parallel) whereas only attracting forces were present with the tangential pattern (magnetisation vector anti-parallel). The latter is similar to two adjacent bar magnets which are allowed to rotate freely. Their opposite poles will attract each other (magnetisation vector anti-parallel) and form a stable configuration in the lowest total energy state.

thumbnail
Table 2. Total magnetic energy contained within array A for different magnetisation pattern and varying numbers of magnets.

https://doi.org/10.1371/journal.pone.0157040.t002

Due to the complex arrangement of magnetisation vectors in the Halbach pattern, each magnet has different stored total magnetic energy, as highlighted in Fig 11A for 12, 16 and 24 magnets. Consequently, the energy required to rotate each magnet varies when array A switches from the Halbach pattern (pre-polarisation field) to the tangential pattern (measurement state), which is illustrated in Fig 11B. As an example, the total magnetic energy difference for magnet number 7 of array A with 12 magnets (see Fig 2 for numbering) was -117 Joules (J) see Fig 11B, curve 1. The negative sign indicates energy release when magnet number 7 rotates from pre-polarisation to measurement states and +117 J is required to rotate it from the measurement to the pre-polarisation state. The findings favour rapid switching from the pre-polarisation to the measurement state because the magnets of array A tend naturally towards the tangential magnetisation pattern due to its lower total magnetic energy. Moreover, since the transition time from measurement to pre-polarisation is not a critical factor, it can be chosen to minimise mechanical vibrations, a potential source of measurement error.

thumbnail
Fig 11. Total stored magnetic energy for each magnet in array A with different number of magnets.

The solid line (curve 1) corresponds to 12 magnets, the dashed line (curve 2) to 16 magnets and the dash-dotted line (curve 3) to 24 magnets in array A. Magnet numbering follows Fig 2A. (a) Magnetic energy for each magnet in array A during pre-polarisation with Halbach magnetization pattern (see Fig 3A). (b) Magnetic energy difference between pre-polarisation and measurement with tangential magnetisation pattern. Negative values in Fig 11B indicate that all the magnets move to a lower magnetic energy state.

https://doi.org/10.1371/journal.pone.0157040.g011

The switching time from the pre-polarisation to the measurement state was estimated by assuming that the difference in magnetic energy is predominantly released as rotational energy. The rotational energy as a function of angular velocity (ω) is given by: (3) where I = 1/2 mmagrm2 is the moment of inertia of the cylindrical magnet about the axis of rotation, and mmag and rm denote the mass and radius of the magnet, respectively.

The pre-polarisation field strength generated by array A with 24 magnets exceeded 100 mT, set as a target for our study (Fig 8A, Table 1). Hence, the study of magnetic forces in array A was performed for this configuration. The average total magnetic energy for array A was approximately +/- 20 J (curve 3, Fig 11B). With a magnet diameter rm = 1.08 cm (refer to Eq 2, fill factor of 0.75) and using the average density of rare-earth magnets (7400 Kg/m3), the angular speed was estimated to be around 600 rad/sec i.e. 10 ms for one revolution (100 Hz). Therefore, in principle, rapid switching of the pre-polarisation field can be achieved using SPMAs.

The mechanical force required to rotate each magnet from the measurement state to the pre-polarisation state was calculated by relating torque (τ) to magnetic energy (E): (4) with θ being the amount of rotation. Assuming that the force F was applied tangentially to the magnet (τ = F∙rm), the average mechanical force required was estimated to be ~300 N. Hence, as each magnet experiences about τ = 3.3 Nm, rotation of individual magnets could be achieved by commercially available rotary actuators [26]. Rapid switching in the presence of large forces requires careful control of magnet rotation and an engineering design that minimizes mechanical (torsional) vibrations both in individual magnets and in the entire array.

Arrays B and C (measurement array).

The measurement field Bm was generated by two concentric cylindrical arrays comprising 12 magnets, each with the Halbach pattern. The simulations accounted for the presence of array A in the tangential configuration. Fig 7D shows an arrow plot of the magnetic field in the x-y plane at z = 0 generated by array A with 24 magnets “switched off” i.e. in the tangential pattern (see Fig 3C) and by the Halbach pattern of arrays B and C. The local field direction of Bm in the FOV was in the y-coordinate direction. The measurement field Bm = 40 μT was generated by rotating arrays B and C in opposite directions, each by an angle of α = 4.5°. The directionality and homogeneity of Bm was evaluated by assessing the ratio of the minor field components (Bmx, Bmz along x- and z) to the main component Bm (along the y-coordinate direction), plotted in Fig 7E, along the x- (solid line), y- (dash-dotted line) and z- (dashed line) axes. In all cases the minor components of Bm in the FOV were at least three orders of magnitude smaller than the main component.

Fig 7F shows detailed line plots of Bm along the x- (curve 1), y- (curve 2) and z-axes (curve 3) through the centre of the SPMA. Relative magnitude varied by less than 0.02% (200 ppm) within the FOV. The absolute magnetic field variation was around 8 nT (Bm = 40 μT) corresponding to a line broadening in the NMR spectrum due to this field inhomogeneity of less than 0.2 Hz, according to Eq 1.

For the chosen magnet characteristics (Ø = 0.3 cm, L = 70 cm and Br = 0.2 T), rotation angles between 0 and 5° led to Bm values ranging between 0–50 μT. Precise rotations of arrays B and C are thus necessary to control the magnitude and direction of Bm. A small mismatch of the rotation angles of arrays B and C could lead to a tilt from the defined axis of precession (y-axis) and the creation of an additional x-component in Bm. This would result in a measurement field that is not perpendicular to sample magnetisation, resulting in a decrease in signal strength. Since the angles are relatively small, this effect is likely to be negligible. A misalignment of 0.5° of one array with the other arrays being correctly aligned resulted in a change in the magnitude of Bm from 55.4 μT to 52.6 μT corresponding to a proton Larmor frequency change of 118 Hz. Field inhomogeneity was increased from 0.02% (200 ppm) to around 0.05% (500 ppm) on average in all three directions. During an ULF experiment switching of Bm is not strictly necessary, since Bp is at least three orders of magnitude larger than Bm. Therefore, arrays B and C can be carefully adjusted prior to experiments to minimize effects due to misalignment.

Magnetic fields of two arrays with different radii but the same number of identical magnets cannot be matched [23]. Hence, in arrays B and C field matching was achieved by reducing Br from 0.2 T to 0.16 T for the magnets in array C. This is practical as rare earth magnets have a large range of standard remanent magnetisations, as detailed in MMPA 0100 –Standard Specifications for Permanent Magnet Materials.

II. Manual SPMA measurement

Bp and Bm in the manual SPMA prototype are presented for the pre-polarisation (Fig 12A) and the measurement state (Fig 12B). The direction of Bp and Bm within the FOV is qualitatively visualized by a custom-built array of small pivot-mounted needles. For comparison, a corresponding COMSOL model (see S2 File) was designed to simulate the magnetic fields within the FOV shown within the figure insets in blue circles. The visualised magnetic fields of the prototype are shown in the figure insets in red circles.

thumbnail
Fig 12. Comparison of Bp and Bm generated by a SPMA prototype (left hand side) with numerical simulation (right hand side).

(a) Field direction of Bp indicated by array of needles (top inset, red circle) and surface plot of COMSOL (bottom inset, blue circle). (b) Field direction of Bm measured (top inset, red circle) and simulated (bottom inset, blue circle).

https://doi.org/10.1371/journal.pone.0157040.g012

Table 3 summarizes the measured and calculated main field parameters for Bp and Bm generated by arrays E and F with rotation angles (α) of 0, 5 and 10°. A non-zero magnitude of Bm = 0.4 mT is present for α = 0° because the magnetic field generated by each array individually (BE and BF) is not matched, hence the direction of Bm is not parallel to the y-axis. The dependence of Bm on the rotation angle α was correctly predicted by the simulations. However, a slight deviation was caused by the presence of a residual field generated by array D for the tangential pattern of 70 μT compared to the simulated 1 μT (Fig 6A, II). This may be due to uncompensated misalignment of individual magnets, variations in magnet dimensions or manufacturing imperfections which were not considered in the numerical model. Further optimization of the manual SPMA (for instance, with shimming [19, 25]) and detailed field homogeneity evaluation were not attempted since the primary purpose was to demonstrate the switching capabilities of Bp and the adjustability of Bm. Low quality magnets in the prototype setup were also likely to have a strong effect on field homogeneity, as shown in Table 3. Measured field inhomogeneity was higher in all configurations than for simulated fields. Discrepancies in Bm were higher than in Bp. Magnets producing Bm were closer to the FOV than magnets producing Bp, resulting in greater sensitivity to variations the former. Additionally, the precision of the Gaussmeter (30μT + 1% of absolute scale) was lower for fields in the amplitude range of Bm than those in the range for Bp.

thumbnail
Table 3. Comparison of simulated and measured magnetic fields generated by the SPMA prototype.

https://doi.org/10.1371/journal.pone.0157040.t003

The magnetic field maps of Bp and Bm generated by the manual SPMA prototype were obtained for pre-polarisation (Fig 13A) and measurement configurations (Fig 13B). Good qualitative agreement between simulation and measurement can be seen in Fig 13A. However, an offset of about 0.3 mT was present due to the higher remanent magnetisation of the magnets used compared to the manufacturer’s specifications which were used for the simulation. The measurements plotted in Fig 13B showed how Bm can be regulated in the setup. Here, the mismatch present between simulation and measurement is similar for both α = 0 and α = 10. This is mainly due to the leftover magnetic field produced by array D in the tangential configuration.

thumbnail
Fig 13.

Comparison of (a) Bp and (b) Bm generated by a SPMA prototype with results of numerical simulation. (a) The measured magnitude of Bp (yellow) shows an offset of 200 μT compared to the simulated value (blue), due to a higher remanent magnetisation than specifications provided by the manufacturer. (b) Arrow plot of Bm for two angular positions of arrays E and F. Dashed lines represent the measured field and continuous lines represent the simulated field for α = 00 and α = 100. The mismatch between simulation and measurement is produced predominantly by the leftover field produced by array D in the tangential configuration.

https://doi.org/10.1371/journal.pone.0157040.g013

Discussion

We introduced a small dynamic adjustable small permanent magnet array (SPMA) as a novel approach to generate multiple magnetic field configurations required for ULF NMR/MRI. As an advance on Halbach arrays, the SPMA enables magnetic fields to be generated through a combination of magnetisation patterns (see Fig 3) obtained by prescribed rotations of individual pivot-mounted permanent magnets and rotations of permanent magnets arrays. Two magnetisation patterns were implemented, the Halbach and tangential patterns, to generate and cancel the pre-polarisation field Bp. Two concentric arrays of permanent magnets were introduced to generate a variable measurement field Bm for ULF relaxometry at different frequencies. This would allow physical processes to be studied as a function of frequency.

Our simulation with 24 permanent magnets (L = 70 cm, dm = 2.16 cm and Br = 1 T) predicted pre-polarisation field magnitudes above 100 mT for the SPMA, higher than presently achieved in ULF instruments using resistive coil technology [13, 27, 28]. The simulation also predicted a magnetic field inhomogeneity of Bp less than 0.03% (300 ppm) within a field of FOV of 5 x 5 x 5 cm3. The SPMA field homogeneity is comparable to previously described Halbach array designs with stacked rectangular magnets for NMR [19, 29]. Higher and more homogeneous sample pre-polarisation should increase SNR [9, 10, 12].

Variable measurement fields ranging from near zero to 50 μT were generated by small rotations of two concentric cylindrical Halbach arrays B and C with the outer array A in the tangential pattern. Nominal magnitude deviations were below 0.02% or 200 ppm without shimming. This is equivalent to spectral line broadening of less than 0.2 Hz for proton Larmor frequencies at ULF. Broadening (which limits resolution) relates to absolute field inhomogeneity. Narrower spectral lines but with lower amplitude can be observed at ULF compared to high field NMR [13]. For example, 300 ppm inhomogeneity at Bm = 50 μT results in line width broadening equivalent to 0.015 ppm inhomogeneity at Bm = 1T.

Our findings highlight the importance of precise rotational adjustments both to achieve high field homogeneity and accurate control of Bm. Bm does not need to be switched off during an experiment since its magnitude is at least three orders of magnitude lower than Bp. This simplifies the design of motion controllers and facilitates adjustments to enhance field homogeneity. Switching off Bp rapidly is crucial and can be achieved by the use of high-quality actuators, such as the SHA25A-81 (Harmonic Drive, Massachusetts, USA).

Our model predicted that fast switching (within 6 ms) from the pre-polarisation to the measurement state would be possible given the total magnetic energy difference between the Halbach pattern and the tangential pattern. The rapid switching time is comparable to current ULF instruments using resistive coils and customized switch boxes [9, 12]. It is plausible that even faster switching can be achieved through the use of additional hydraulic or pneumatic actuator systems. However, mechanical vibration may need to be considered and was beyond the scope of this study.

Unlike resistive coils, energy is not dissipated into heat due to current flow, obviating sample heating. Furthermore, undesired signal generation due to transient currents, induced in conductors by rapid switching, is reduced because the conductivity of magnet alloys is much lower than for conductors like copper used in coils. Compared to superconducting magnets, the energy consumption of the SPMA is significantly lower because cooling is not required, further reducing cost.

A manual SPMA was built and compared with predictions from simulations. It demonstrated the ability to generate varying magnetic fields and the validity of the numerical approach. All the main field parameters were correctly described with our computational model and experimentally verified. This is in agreement with previous studies in which COMSOL was used to simulate magnetic field parameters to optimise the design of instruments based on Halbach arrays [20, 25]. Further optimization to maximize field homogeneity and to match the magnetic fields generated by array B and C was not attempted here since the primary focus was to verify the new features of the SPMA. However, shimming techniques are avenues for future research to optimise the performance of the SPMA [19, 25].

Although this study focuses on the application of the SPMA for of ULF NMR relaxometry, the flexible and modular design of the SPMA allows additional magnet arrays to be added, for instance, to generate the dynamic gradient fields required for imaging with ULF-MRI. This is achievable since the absolute field homogeneity requirements for ULF instrumentation are quite moderate. Notably, other than the switching between Bp and Bm with the SPMA, sequences for signal generation and acquisition are similar to conventional approaches with instrumentation using resistive magnets. Measurement strategies like iterative sample pre-polarisation to enhance signal-to-noise ratio through signal averaging can be applied with the SPMA similar to high field relaxometry applications.

Generating linear gradient fields with permanent magnets is more challenging at ULF due to the presence of concomitant fields [30, 31]. These are predicted by Maxwell’s equations since the field gradients are comparable in magnitude to Bm [13] and result in image distortion that requires correction during reconstruction. New approaches for spatial signal encoding that utilize intrinsic field inhomogeneity in combination with non-linear image reconstruction methods have been described for static Halbach arrays, allowing spatial encoding in 1D [25]. Rotation of the array is required for 2D images and the use of additional excitation hardware has been proposed for spatial encoding in the third dimension, using Bloch-Siebert Spatial Encoding (BS-SET) or TRansient Array Spatial Encoding (TRASE) [25]. We posit that 3D spatial encoding can be achieved with additional dynamically adjustable SPMAs [32]. However, in ULF-MRI slice selection remains an open question and applications needing to acquire a single slice are currently faced with having to acquire the entire 3D image. Furthermore, slice selection may require very long duration excitation pulses at ULF. The idea that the high homogeneity of Bp produced by the SPMA may be harnessed to create fields for slice selection requires further exploration.

Conclusions

The proposed SPMA is substantially different from all the resistive coil-based approaches reported in the literature to date. Flexible manipulation of the magnetic field while achieving very high field homogeneities through rearrangement of a concentric SPMA was empirically demonstrated. Pre-polarisation and measurement fields relevant for ULF NMR could be generated by a combination of rotation of individual permanent magnets and rotation of Halbach arrays. Our findings may benefit future developments in ULF-NMR/MRI by eliminating the need for resistive coils, increasing the potential for compact, low energy, portable instrumentation.

Supporting Information

S1 File. SPMA Model Report.

Model documentation generated by COMSOL with implemented parameters for the SPMA.

https://doi.org/10.1371/journal.pone.0157040.s001

(ZIP)

S2 File. Prototype Model Report.

Model documentation generated by COMSOL with implemented parameters for the manual prototype.

https://doi.org/10.1371/journal.pone.0157040.s002

(ZIP)

Acknowledgments

This journal paper is dedicated to the memory of Prof Graeme Hanson (1955–2015), a great scientist and a dear colleague and friend.

Author Contributions

Conceived and designed the experiments: MWV AG VV RPG DCR. Performed the experiments: MWV AG VV RPG. Analyzed the data: MWV AG VV RPG. Contributed reagents/materials/analysis tools: MWV AG VV RPG DCR. Wrote the paper: MWV AG VV RPG DCR. Developed the computational SPMA design: AG. Conducted the numerical study: AG. Developed the original SPMA design: MWV. Conducted the numerical study for the SPMA prototype: MWV. Performed the analysis on magnetic forces and torque and energy consumption: MWV V.V. developed the original idea of SPMA, performed the analysis on magnetic forces and contributed to manuscript preparation. R.P. built the SPMA prototype and performed the magnetic field measurements. D.R. developed the original idea of SPMA, co-designed the numerical studies.

References

  1. 1. Abragam A. The Principles of Nuclear Magnetism: Clarendon Press; 1961.
  2. 2. Haake E, Brown R, Thompson M, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design: Wiley, New York; 1999.
  3. 3. Keeler J. Understanding NMR spectroscopy: John Wiley & Sons; 2013.
  4. 4. McDermott R, Trabesinger AH, Mück M, Hahn EL, Pines A, Clarke J. Liquid-state NMR and scalar couplings in microtesla magnetic fields. Science. 2002;295(5563):2247–9. pmid:11910105
  5. 5. Matlachov AN, Volegov PL, Espy MA, George JS, Kraus RH Jr. SQUID detected NMR in microtesla magnetic fields. J Magn Reson 2004;170(1):1–7. pmid:15324752
  6. 6. McDermott R, Kelso N, Lee S, Mößle M, Mueck M, Myers W, et al. SQUID-detected magnetic resonance imaging in microtesla magnetic fields. J Low Temp Phys. 2004;135(5–6):793–821.
  7. 7. Burghoff M, Hartwig S, Trahms L, Bernarding J. Nuclear magnetic resonance in the nanoTesla range. Appl Phys Lett. 2005;87(5):054103.
  8. 8. Lee SK, Möβle M, Myers W, Kelso N, Trabesinger AH, Pines A, et al. SQUID‐detected MRI at 132 μT with T1‐weighted contrast established at 10 μT–300 mT. Magn Reson Med 2005;53(1):9–14. pmid:15690496
  9. 9. Vadim SZ, Andrei NM, Petr LV, Algis VU, Michelle AE, Robert HK Jr. SQUID-based instrumentation for ultralow-field MRI. Supercond Sci Technol. 2007;20(11):S367.
  10. 10. Espy M, Flynn M, Gomez J, Hanson C, Kraus R, Magnelind P, et al. Applications of Ultra-Low Field Magnetic Resonance for Imaging and Materials Studies. IEEE Trans Appl Supercond. 2009;19(3):835–8.
  11. 11. Trahms L, Burghoff M. NMR at very low fields. Magn Reson Imag 2010;28(8):1244–50.
  12. 12. Espy M, Matlashov A, Volegov P. SQUID-detected ultra-low field MRI. J Magn Reson 2013;228(0):1–15.
  13. 13. Kraus RH Jr, Espy M, Magnelind P, Volegov P. Ultra-Low Field Nuclear Magnetic Resonance: A New MRI Regime: Oxford University Press; 2014.
  14. 14. Sarracanie M, LaPierre CD, Salameh N, Waddington DEJ, Witzel T, Rosen MS. Low-Cost High- Performance MRI. Scientific Reports. 2015;5:15177. pmid:26469756
  15. 15. Macovski A, Conolly S. Novel approaches to low‐cost MRI. Magn Reson Med 1993;30(2):221–30. pmid:8366803
  16. 16. Ledbetter M, Theis T, Blanchard J, Ring H, Ganssle P, Appelt S, et al. Near-zero-field nuclear magnetic resonance. Phys Rev lett. 2011;107(10):107601. pmid:21981529
  17. 17. Vogel MW, Vegh V, Reutens DC. Numerical study of ultra-low field nuclear magnetic resonance relaxometry utilizing a single axis magnetometer for signal detection. Med Phys. 2013;40(5):052301. pmid:23635288
  18. 18. Vesanen PT, Nieminen JO, Zevenhoven KC, Dabek J, Parkkonen LT, Zhdanov AV, et al. Hybrid ultra‐low‐field MRI and magnetoencephalography system based on a commercial whole‐head neuromagnetometer. Magnetic Resonance in Medicine. 2013;69(6):1795–804. pmid:22807201
  19. 19. Zalesskiy SS, Danieli E, mich B, Ananikov VP. Miniaturization of NMR systems: Desktop spectrometers, microcoil spectroscopy, and “NMR on a chip” for chemistry, biochemistry, and industry. Chem Rev. 2014;114(11):5641–94. pmid:24779750
  20. 20. Bjørk R, Bahl CRH, Smith A, Pryds N. Comparison of adjustable permanent magnetic field sources. J Magn Magn Mater. 2010;322(22):3664–71.
  21. 21. Zhu Z, Howe D. Halbach permanent magnet machines and applications: a review. IEE Proc-Electr Power Appl. 2001;148(4):299–308.
  22. 22. Moresi G, Magin R. Miniature permanent magnet for table‐top NMR. Concepts Magn Reson Part B. 2003;19(1):35–43.
  23. 23. Raich H, Blümler P. Design and construction of a dipolar Halbach array with a homogeneous field from identical bar magnets: NMR Mandhalas. Concepts Magn Reson Part B. 2004;23(1):16–25.
  24. 24. Nan S, Tae-Jong Y, Hakho L, Andress W, Weissleder R, Ham D. Palm NMR and 1-Chip NMR. IEEE J Solid-State Circuit. 2011;46(1):342–52.
  25. 25. Cooley CZ, Stockmann JP, Armstrong BD, Sarracanie M, Lev MH, Rosen MS, et al. Two‐dimensional imaging in a lightweight portable MRI scanner without gradient coils. Magn Reson Med 2015;73(2):872–83. pmid:24668520
  26. 26. http://www.harmonicdrive.net/products/rotary-actuators/hollow-shaft-actuators/sha-sg.
  27. 27. Espy MA, Magnelind PE, Matlashov AN, Newman SG, Sandin HJ, Schultz LJ, et al. Progress Toward a Deployable SQUID-Based Ultra-Low Field MRI System for Anatomical Imaging. IEEE Trans Appl Supercond. 2015;25(3):1–5.
  28. 28. Volegov P, Flynn M, Kraus R, Magnelind P, Matlashov A, Nath P, et al. Magnetic Resonance Relaxometry at Low and Ultra Low Fields. In: Supek S, Sušac A, editors. 17th International Conference on Biomagnetism Advances in Biomagnetism–Biomag2010. IFMBE Proceedings. 28: Springer Berlin Heidelberg; 2010. p. 82–7.
  29. 29. Soltner H, Blümler P. Dipolar Halbach magnet stacks made from identically shaped permanent magnets for magnetic resonance. Concepts Magn Reson Part A. 2010;36(4):211–22.
  30. 30. Nieminen JO, Ilmoniemi RJ. Solving the problem of concomitant gradients in ultra-low-field MRI. J Magn Reson 2010;207(2):213–9. pmid:20884262
  31. 31. Volegov P, Mosher J, Espy M, Kraus R. On concomitant gradients in low-field MRI. J Magn Reson 2005;175(1):103–13. pmid:15869890
  32. 32. Hennig J, Welz AM, Schultz G, Korvink J, Liu Z, Speck O, et al. Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept study. Magn Reson Mat Phys Biol Med. 2008;21(1–2):5–14.