Contrast Sensitivity and Spherical Aberration in Eyes Implanted with AcrySof IQ and AcrySof Natural Intraocular Lens: the Results of a Meta-Analysis

  • Jianping Liu,

    Affiliations Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang, People’s Republic of China, Eye Hospital of China Medical University, Shenyang, People’s Republic of China, Key Lens Research Laboratory of Liaoning Province, Shenyang, People’s Republic of China

  • Jiangyue Zhao,

    Affiliations Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang, People’s Republic of China, Eye Hospital of China Medical University, Shenyang, People’s Republic of China, Key Lens Research Laboratory of Liaoning Province, Shenyang, People’s Republic of China

  • Liwei Ma,

    Affiliations Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang, People’s Republic of China, Eye Hospital of China Medical University, Shenyang, People’s Republic of China

  • Guangcong Liu,

    Affiliation School of Public Health, China Medical University, Shenyang, People’s Republic of China

  • Di Wu,

    Affiliations Eye Hospital of China Medical University, Shenyang, People’s Republic of China, Key Lens Research Laboratory of Liaoning Province, Shenyang, People’s Republic of China

  • Jinsong Zhang

    ZhangJS_CMU@163.com

    Affiliations Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang, People’s Republic of China, Eye Hospital of China Medical University, Shenyang, People’s Republic of China, Key Lens Research Laboratory of Liaoning Province, Shenyang, People’s Republic of China

Contrast Sensitivity and Spherical Aberration in Eyes Implanted with AcrySof IQ and AcrySof Natural Intraocular Lens: the Results of a Meta-Analysis

  • Jianping Liu, 
  • Jiangyue Zhao, 
  • Liwei Ma, 
  • Guangcong Liu, 
  • Di Wu, 
  • Jinsong Zhang
PLOS
x

Abstract

Background

To systematically evaluate the visual performance of aspheric AcrySof IQ and spherical AcrySof Natural intraocular lens (IOL) after cataract surgery.

Methodology/Principal Findings

Potential randomized controlled trials (RCTs) that involved implanting AcrySof IQ and AcrySof Natural were searched from PubMed, Web of science, EMBASE, Chinese Science and Technology Periodicals Databases and Cochrane Central Register of Controlled Trials. The methodological quality of the studies was assessed by the Jadad method. Standardized mean differences (SMDs) with 95% confidence intervals (CIs) of best-corrected visual acuity (BCVA), contrast sensitivity and spherical aberration were pooled using a random-effects model. Seven studies were identified and analyzed to compare AcrySof IQ (236 eyes) with AcrySof Natural (232 eyes) after phacoemulsification. There was no significant difference in postoperative BCVA between AcrySof IQ and AcrySof Natural (p =0.137) after a follow up of 3 months. For contrast sensitivity, these differences reached statistical significance under photopic conditions at two spatial frequencies (3 cycles per degree (cpd), 6 cpd, 12 cpd, and 18 cpd; p =0.022, p =0.017, p = 0.065, and p=0.191, respectively) and under mesopic conditions at three spatial frequencies (3 cpd, 6 cpd, 12 cpd, and 18 cpd; p =0.007, p =0.033, p =0.030, and p =0.080, respectively). Eyes with AcrySof IQ also had statistically significant less spherical aberration than eyes with AcrySof Natural (p<0.001). Sensitivity analysis showed that the results were relatively stable and reliable.

Conclusions/Significance

The overall findings indicate that AcrySof IQ with a modified aspheric surface induced significantly less spherical aberration than AcrySof Natural. Contrast sensitivity in eyes with AcrySof IQ is better than that in eyes with AcrySof Natural, especially under mesopic conditions.

Introduction

Cataract is the leading reason of blindness in the world, and usually has to be treated surgically [1].

At present, the main surgical procedures are phacoemulsification and intraocular lens (IOL) implantation [2].With the improvement of life quality, cataract surgery has developed from a procedure for the safe removal of the cataract to one aimed at refining to achieve the best possible postoperative refractive result [3]. Although surgical technique as well as IOL materials and designs has improved significantly, there is now an increasing emphasis on improving visual quality and functional vision provided by IOL [4]. Previous studies have demonstrated that contrast sensitivity and wavefront aberration are the potent indicators of functional vision [5,6]. Contrast sensitivity function, measured under varying conditions of luminance and glare, establishes the limits of visual perception across the spectrum of spatial frequencies [7,8]. It determines the relationship between the optical efficiency of the eye and minimal retinal threshold for pattern detection, which is not detected by the measurement of Snellen visual acuity [8].

Spherical aberration is one of the most important higher-order aberrations in the human eye [9].Some scientific studies have shown the decrease in visual quality that occurs with age is usually attributed to lens change. In youth, the generally negative spherical aberration of the crystalline lens largely neutralizes the average positive spherical aberration of the prolate cornea, resulting in an optimized retinal image. In the aging eye, there is a loss of this cornea–lens balancing, resulting in a deterioration of retinal image quality [10-12]. Based on these findings, the AcrySof IQ (model SN60WF, Alcon Surgical Laboratories), is a single-piece foldable acrylic IOL with a posterior aspheric surface (-0.20um, asphericity value) designed to decrease the total amount of ocular spherical aberration after cataract surgery. However, these previous trials generally had small sample sizes and showed conflicting results, which greatly hindered researchers drawing correct conclusions.

Hence, we conducted a meta-analysis of published randomized controlled trials (RCTs) to assess the quality of vision provided by the aspheric IOL (AcrySof IQ) after cataract surgery.

Materials and Methods

Literature search strategy

An extensive literature review was searched through PubMed, EMBASE, Chinese Science and Technology Periodicals Databases and Cochrane Controlled Trials Register (most recently updated in 2013 Mar). We performed the literature search with no language limitations. The search terms included were “AcrySof IQ”, “AcrySof Natural”, “comparison”, “contrast sensitivity”, and “spherical aberration”. Abstracts were read and full texts were retrieved if they seemed to meet the objective of this review. Related references and articles were checked and analyzed in depth. Considering all the study design, no previous systematic review or meta-analysis on this topic was found. The study and data accumulation were carried out with approval from the Institutional Review Board of The Fourth Affiliated Hospital of China Medical University and the study complied with the tenets of the Declaration of Helsinki.

Inclusion Criteria and Exclusion Criteria

The following inclusion criteria were used to identify published studies for this meta-analysis: (1) study design: prospective, randomized clinical controlled trials; (2) population and intervention: patients who were diagnosed as age-related cataract underwent cataract surgery with AcrySof IQ and AcrySof Natural IOL implantation; (3)outcome measurement: variables of the report containing sufficient information on contrast sensitivity and spherical aberration. Exclusion criteria included double reporting, in vitro studies, unrelated outcome measurement, use of refractive surgery, delivering no baseline data and no aggregated results. Decisions regarding which trials to be included were made independently by two reviewers (Jianping.Liu, Guangcong Liu).

Data extraction

All available data from the selected articles were extracted by two independent reviewers (Jianping Liu, Di Wu) cautiously. The following categories of information were extracted: each study’s author, publication year, study design, study location, number of patients at final follow-up, quality control, duration of follow-up. Whenever any disagreements occurred, they were resolved through discussion till a consensus was made.

Statistical analysis

The statistical analysis was carried out through stata software version 10 (StataCorp, College Station, TX). Forest plots were used to present the results, and the results were expressed as standardized mean difference (SMD) and 95% confidence interval (CI). The center of each square indicated the SMD and the size of the square was proportional to percent weight each study contributed to the pooled estimates .The horizonal line bisecting each square represented the 95% CI for the SMD. Heterogeneity among studies was tested using the Chi-squared statistic. If the significant evidence of statistical heterogeneity or clinical diversity was not found (p >0.10), fixed-effects model was used [13].However, for the result showing significant heterogeneity (p <0.10), we used random-effects model to account for inter-study heterogeneity and tested for statistically significant difference between the estimates with respect to AcrySof IQ and AcrySof Natural IOL. Funnel plot was used to observe the included studies’ publication bias, asymmetry plot implied possible existence of publication bias. The asymmetry degree was measured by Egger’ test, and a p value<0.05 was considered as an evidence of publication bias. To explore the steadiness of our results, sensitivity analysis investigating the influence of each individual study on the overall meta-analysis summary estimates was carried out to identify potential outliners [13,14].All statistical tests were two-sided.

Results

Search Results

Initial electronic search retrieved 74 articles after removing duplicates, in which 56 articles were excluded after first-pass review of titles and abstracts. 11 studies were eliminated after full text review according to the inclusion and exclusion criteria specified earlier. Hence a total of 7 prospective RCTs [15-21] were identified. We evaluated the methodological quality of eligible studies using the Jadad method [22]. A checklist contained five items(1): with or without randomization(2); with or without a double-blind design(3); the appropriateness of the randomization methods if used(4); the appropriateness of double-blinding design if used(5); the analysis and reasons for withdrawals and dropouts. Each of the item’s scale is assigned 0 or1.Thus, excellent quality of studies can receive a Jadad score of five, good qualities if the score was 3 or 4, and poor quality if the score was≤2.The characteristics of eligible studies are described in Table 1.

AuthorCountryPublished yearDesignNo.of eyesTime of follow upQuality assessment (Jadad score)
Awwad ST etal.USA2007RCT273 months5
253 months
Nanavaty MA etal.England2009RCT473 months4
473 months
Chen YY etal.China2012RCT283 months3
273 months
ROCHA KM etal.Brazil2006RCT403 months4
403 months
Tzelikis PF etal.Brazil2007RCT253 months4
253 months
Pandita D etal.India2007RCT363 months4
373 months
Li HW etal.China 2009RCT333 months3
313 months

Table 1. Characteristics of included studies in the meta-analysis.

RCT,randomized controlled trial.
CSV
Download CSV

Meta-Analysis Results

Differences in mean changes in best-corrected visual acuity (BCVA) after implantation, along with SMD and 95% CI, are presented in Figure 1.The analysis results of BCVA of each study were presented in a forest plot. Values to the left of the vertical line at 0 show greater change in the subjects implanted with AcrySof Natural and values to the right of the vertical line show greater change in AcrySof IQ. The subtotal rows show the meta-analysis summary values for each time point. No significant difference in BCVA was found in AcrySof IQ group compared with AcrySof Natural group (SMD=0.399, 95%CI-0.127 to 0.924, p =0.137).

thumbnail
Figure 1. Standardized mean difference (SMD) in best-corrected visual acuity (BCVA): This forest plot showed the SMD in BCVA (log MAR) along with associated 95% confidence interval (CI), comparing AcrySof IQ and AcrySof Natural intraocular lens.

https://doi.org/10.1371/journal.pone.0077860.g001

Changes in contrast sensitivity were assessed under photopic and mesopic conditions at low spatial frequency (3 cycles per degree [cpd]), intermediate frequency (6 cpd), high frequencies (12 cpd and 18 cpd). Differences in mean changes in contrast sensitivity after implantation, along with SMD and 95% CI are presented. The analysis results of contrast sensitivity of each study were presented in a forest plot. Values to the left of the vertical line at 0 show greater change in contrast sensitivity in the subjects implanted with AcrySof Natural and values to the right of the vertical line show greater change in AcrySof IQ. Significant improvements in contrast sensitivity under photopic conditions(Figure 2) at 3cpd,6cpd spatial frequencies (SMD=0.486,95%CI 0.069 to 0.903, p = 0.022;SMD=0.291,95%CI 0.052 to 0.531, p = 0.017) and under mesopic conditions (Figure 3) at 3cpd,6cpd ,12cpd spatial frequencies (SMD=0.564, 95%CI 0.155 to 0.973, p =0.007;SMD=0.585,95%CI 0.048 to1.122, p =0.033; SMD=0.311, 95%CI 0.030 to 0.592, p =0.030) were found in AcrySof IQ group compared with AcrySof Natural group. There were no statistically significant differences in contrast sensitivity between AcrySof IQ and AcrySof Natural in photopic conditions at 12cpd, 18cpd (SMD =0.433, 95% CI -0.026 to 0.892, p = 0.065; SMD=0.362, 95% CI -0.181 to 0.904, p = 0.191) and mesopic conditions at 18 cpd (SMD=0.423, 95%CI -0.056 to 0.902, p =0.423).

thumbnail
Figure 2. Standardized mean difference (SMD) in contrast sensitivity: This forest plot showed the SMD in postoperative contrast sensitivity under photopic conditions at 3 cycles per degree (cpd), 6 cpd, 12 cpd, and 18 cpd along with associated 95% confidence interval (CI), comparing AcrySof IQ and AcrySof Natural intraocular lens.

https://doi.org/10.1371/journal.pone.0077860.g002

thumbnail
Figure 3. Standardized mean difference (SMD) in contrast sensitivity: This forest plot showed the SMD in postoperative contrast sensitivity under mesopic conditions at 3 cycles per degree ( cpd ), 6 cpd, 12 cpd, and 18 cpd along with associated 95% confidence interval (CI), comparing AcrySof IQ and AcrySof Natural intraocular lens.

https://doi.org/10.1371/journal.pone.0077860.g003

Differences in mean changes in spherical aberration after implantation, along with SMD and 95% CI are presented in Figure 4. The analysis results of spherical aberration of each study were presented in a forest plot. Values to the left of the vertical line at 0 show greater change in spherical aberration in the subjects implanted with AcrySof IQ and values to the right of the vertical line show greater change in AcrySof Natural. Significant improvement in spherical aberration was found in the AcrySof IQ group compared with AcrySof Natural group (SMD=-3.435, 95% CI-4.487 to -2.384, p <0.001).

thumbnail
Figure 4. Standardized mean difference (SMD) in spherical aberration: This forest plot showed the SMD in spherical aberration along with associated 95% confidence interval (CI), comparing AcrySof IQ and AcrySof Natural intraocular lens.

https://doi.org/10.1371/journal.pone.0077860.g004

Sensitivity analysis was carried out by removal of one RCT at a time and computing the pooled SMDs for the remaining studies. We found no individual study could alter the pooled results of BCVA (SMD = 0.399, 95% CI -0.09 to 0.73, p =0.137) ,contrast sensitivity under photopic conditions (SMD= 0.389, 95%CI 0.19 to 0.53, p <0.001) and mesopic conditions (SMD= 0.467,95CI%0.26 to 0.59, p <0.001) as well as spherical aberration (SMD=-3.435, 95% CI -3.62 to -2.41, p <0.001),which indicated that our results were stable and reliable.

Publication bias

Funnel plot was performed to assess the publication bias of literatures (Figure 5). There was low possibility of asymmetry in the funnel plot because a limited number of trials were involved in the final analysis, but the result of Egger’s test did not show any evidence of publication bias (t=0.79, p = 0.476).

thumbnail
Figure 5. Funnel plot for the results between AcrySof IQ and AcrySof Natural intraocular lens.

Egger’ publication bias test result (p =0.476).

https://doi.org/10.1371/journal.pone.0077860.g005

Discussion

Focus on the quality of life today, with advances in cataract surgery and IOL implantation, the visual outcome is measured not only by means of visual acuity, but also by quality of vision [23]. Contrast sensitivity and wavefront aberration are the main characteristics of the quality of vision [24,25]. Our study evaluated the visual outcomes between aspheric IOL (AcrySof IQ) and spherical IOL (AcrySof Natural) implantation. In this meta-analysis, we identified a number of relevant studies (n=7), and it should be noted that all of the studies did randomize treatments, which could be considered as sufficient evidence reporting.

Currently, evidence of vision quality based on implanting AcrySof IQ and AcrySof Natural IOLremains lacking. Despite a prospective RCT of AcrySof IQ and AcrySof Natural implants might go some way to provide the answer to the question. However, such a trial would require a great deal of patients to provide conclusive result. So we carried out the meta-analysis to evaluate the visual outcomes of AcrySof IQ IOL. Meta-analysis is a method of using statistical analysis that combines or integrates the results of several independent studies considered by the analyst to be combinable [13,26]. By combining the results of these RCTs through the meta-analysis, our study has a greater statistical power than the powers of the studies as individuals. Based on our result, there was no clinically relevant difference in BCVA (logMAR) between AcrySof IQ and AcrySof Natural implantation, which indicates that measuring visual acuity at different contrast levels is a more sensitive measure as it offers the various visual cues of daily life and may display more preferable information about functional vision than the information provided by the high contrast letters of Snellen acuity charts [19].

Cataract surgery is becoming more of a refractive procedure, patients who acquire good vision after surgery, at the same time, parts of the sufferings have appeared the phenomenon of glare, halos, and poor night vision [27,28]. Improvement in contrast sensitivity could influence the quality of vision. Measuring contrast sensitivity under different luminance and glare conditions is a more sensitive measure as it gives much better perception of visual status and contrast testing determines contrast thresholds at each spatial frequency in normal eyes as well as in eyes with ocular disease [29,30]. Traditional AcrySof Natural IOL that is designed without prolate surface modification decreases the optical transfer function after cataract surgery, which is leading to a reduction in the quality of the retinal image. In a prospective study by Kennis et al [31], contrast sensitivity was significantly better at almost all spatial frequencies in the Tecnis Z9000 compared with Sensar AR40e, or AcrySof Natural SN60AT. Contrary to the findings of Kennis et al, a study by Mun˜oz [32], spherical aberration and contrast sensitivity with the Tecnis Z9000 IOL, found no statistically significant differences between the AR40e IOL and Z9000 IOL in photopic and mesopic contrast sensitivity.

With the development of wavefront aberration techniques, in the design of IOLs, an attempt has been made to mirror natural crystal as closely as possible [33]. Spherical aberration is a property of all spherical IOLs, which is one of the most important aberrations contributing to visual deterioration of the pseudophakic eye [34,35]. It occurs when the lens bends peripheral rays more strongly (positive spherical aberrations). The flatter curve of the peripheral cornea refracts light less strongly than the steeper central area [36]. Conventional spherical IOLs can result in positive spherical aberration and are unable to offset corneal positive spherical aberration, ultimately leading to a decline in image quality.

Wang et al [37] conducted a meta-analysis to compare the visual performance of aspheric IOL with spherical IOL, but IOLs made from different materials and optical designs may have different optical properties, which is difficult to get a conclusive recommendation in the selection of a particular IOL for an individual [38]. In our meta-analysis, the intraindividual comparative study was implanted with aspheric IOL (AcrySof IQ) and the design-equivalent spherical IOL (AcrySof Natural).The results showed that AcrySof IQ provided significant improvement in contrast sensitivity under mesopic conditions at almost all spatial frequencies. We found a statistically significant decrease in spherical aberration between AcrySof IQ group and AcrySof Natural group.

There were some limitations to this study: (1) Only a small number of trials were enrolled in this meta-analysis, (2) There was a follow up of 3 months. It may be that different measurements are obtained in a longer term study at least one year, and (3) We only evaluated spherical aberration, which might not represent the visual performance of total higher-order aberrations.

In summary, findings of the present study indicate that visual quality can be improved by implantation of an IOL with a modified prolate surface. Patients will be more generally satisfied after implantation of an AcrySof IQ IOL because they will have less spherical aberration and better contrast sensitivity, especially under dim light. Future multi-center RCTs of AcrySof IQ versus AcrySof Natural IOL are warranted for further research and development.

Supporting Information

Author Contributions

Conceived and designed the experiments: JPL JSZ GCL. Performed the experiments: JPL JYZ DW. Analyzed the data: JPL LWM JSZ DW. Contributed reagents/materials/analysis tools: JPL GCL LWM JSZ. Wrote the manuscript: JPL LWM JYZ.

References

  1. 1. Resnikoff S, Pascolini D, Etya’ale D, Kocur I, Pararajasegaram R, et al. (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82: 844-851.
  2. 2. Pesudovs K, Dietze H, Stewart OG, Noble BA, Cox MJ et al. (2005) Effect of cataract surgery incision location and intraocular lens type on ocular aberrations. J Cataract Refract Surg 31: 725–734. doi:10.1016/j.jcrs.2004.09.028. PubMed: 15899449.
  3. 3. Elliott DB (1993) Evaluating visual function in cataract. Optom Vis Sci70: 896–902. doi:10.1097/00006324-199311000-00006. PubMed: 8302525.
  4. 4. Olawoye O, Ashaye A, Bekibele C, Ajuwon AJ (2012) Quality-of-Life and Visual Function after Manual Small Incision Cataract Surgery in South Western Nigeria .West Afr J Med 31: 114-119. PubMed: 23208481.
  5. 5. Ginsburg AP, Evans DW, Sekule R, Harp SA (1982) Contrast sensitivity predicts pilots’ performance in aircrafts simulators. Am J Optom Physiol Opt 59: 105-109. doi:10.1097/00006324-198201000-00020. PubMed: 7055195.
  6. 6. Rubin GS, Bandeen-Roche K,Huang GH, Schein OD , Fried LP et al. (2001) The association or multiple visual impairments with self-reported visual disability: SEE project. Invest Ophthalmol Vis Sci 42:64-72.
  7. 7. Arden GB (1978) The importance of measuring contrast sensitivity in cases of visual disturbance. Br J Ophthalmol 62: 198-209. doi:10.1136/bjo.62.4.198. PubMed: 348230.
  8. 8. Santhiago MR, Netto MV, Barreto J Jr, Gomes BA, Mukai A et al. (2010) Wavefront analysis, contrast sensitivity, and depth of focus after cataract surgery with aspherical intraocular lens implantation. Am J Ophthalmol 149: 383-389. doi:10.1016/j.ajo.2009.09.019. PubMed: 20035922.
  9. 9. Wang JM, Liu CL, Luo YN, Liu YG, Hu BJ (2012) Statistical virtual eye model based on wavefront aberration. Int J Ophthalmol 5: 620-624. PubMed: 23173112.
  10. 10. Guirao A, Gonzalez C, Redondo M, Geraghty E , Norrby S, et al. (1999) Average optical performance of the human eye as function of age in a normal population. Invest Ophthalmol Vis Sci 40: 203–213. PubMed: 9888445.
  11. 11. Glasser A, Campbell MC (1998) Presbyopia and the optical changes in the human crystalline lens with age. Vision Res 38: 209–229. doi:10.1016/S0042-6989(97)00102-8. PubMed: 9536350.
  12. 12. Guirao A, Redondo M, Artal P (2000) Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis 17: 1697–1702. doi:10.1364/JOSAA.17.001697. PubMed: 11028517.
  13. 13. Thomas J,Graziosi S, Hiqqins , Coe R, Torgerson C et al. (2012) Teaching meta-analysis using MetaLight. BMC Res Notes 18: 571. PubMed: 23078762.
  14. 14. Egger M, Smith GD, Phillips AN (1997) Meta-analysis: principles and procedures. BMJ 315: 1533–1537. doi:10.1136/bmj.315.7121.1533. PubMed: 9432252.
  15. 15. Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottós K et al. (2006) Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 142: 750-756. doi:10.1016/j.ajo.2006.06.031. PubMed: 17056359.
  16. 16. Chen YY, Tan SJ, Liang H, Liu YB (2012) Clinical study on the Visual Quality of eyes Implanted with Acrysof IQ Aspheric Intraocular Lenses. Guangxi Yi Ke Xue Xue Bao 34: 284-287.
  17. 17. Li HW, Guo HK, Zhang HY, Jin HY, Chen YL ,et al. ( 2009) Comparative analysis of visual quality after aspheric intraocular lens implantation. Int J Ophthalmol Sentyabr:2092-2096.
  18. 18. Tzelikis PF, Akaishi L, Trindade FC, Boteon JE (2007) Ocular aberrations and contrast sensitivity after cataract surgery with AcrySof IQ intraocular lens implantation Clinical comparative study. J Cataract Refract Surg 33: 1918–1924. doi:10.1016/j.jcrs.2007.06.053. PubMed: 17964398.
  19. 19. Pandita D, Raj SM, Vasavada VA, Kazi NS, Vasavada AR (2007) Contrast sensitivity and glare disability after implantation of AcrySof IQ Natural aspheric intraocular lens: prospective randomized masked clinical trial. J Cataract Refract Surg 33: 603-610. doi:10.1016/j.jcrs.2007.01.009. PubMed: 17397731.
  20. 20. Awwad ST, Warmerdam D, Bowman RW, Dwarakanathan S, Cavanagh HD et al. (2008) Contrast sensitivity and higher order aberrations in eyes implanted with AcrySof IQ SN60WF and AcrySof SN60AT intraocular lenses. J Refract Surg 24: 619-625. PubMed: 18581789.
  21. 21. Nanavaty MA, Spalton DJ, Boyce J, Saha S, Marshall J (2009) Wavefront aberrations, depth of focus, and contrast sensitivity with aspheric and spherical intraocular lenses: fellow-eye study. J Cataract Refract Surg 35: 663-671. doi:10.1016/j.jcrs.2008.12.011. PubMed: 19304086.
  22. 22. Jadad AR , Moore RA,Carroll D, Jenkinson C, Reynolds DJ (1996) et al.. Accessing the Quality of Rep of Randomized Clinical Trials: Is blinding Necessary? Control Clin Trials17:1-12.
  23. 23. Holladay JT, Piers PA, Koranyi G, van der Mooren M, Norrby NE (2002) A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg18: 683-691. PubMed: 12458861.
  24. 24. Artal P, Berrio E, Guirao A, Piers P (2002) Contribution of the cornea and internal surface to the change of ocular aberrations with age. J Opt Soc Am A Opt Image Sci Vis19: 137-143. doi:10.1364/JOSAA.19.000137. PubMed: 11778716.
  25. 25. Packer M, Fine IH, Hoffman RS, Piers PA (2004) Improved functional vision with a modified prolate intraocular lens. J Cataract Refract Surg30: 986-992. doi:10.1016/j.jcrs.2003.10.022. PubMed: 15130633.
  26. 26. DerSimonian R (1996) Meta-analysis in the design and monitoring of clinical trials. Stat Med 15: 1237-1248. doi:10.1002/(SICI)1097-0258(19960630)15:12. PubMed: 8817798.
  27. 27. Artal P, Guirao A, Berrio E, Williams DR (2001) Compensation of corneal aberrations by the internal optics in the human eye. J Vis 1: 1-8. doi:10.1167/1.2.1. PubMed: 12678609.
  28. 28. Sekuler R, Owsley C, Berenberg R (1986) Contrast sensitivity during provoked visual impairment in multiple sclerosis. Ophthal Physiol Opt 6: 229–232. doi:10.1111/j.1475-1313.1986.tb01147.x. PubMed: 3748570.
  29. 29. Oen FT, Lim TH, Chung MP (1994) Contrast sensitivity in a large adult population. Ann Acad Med Singapore 23: 322–326. PubMed: 7944242.
  30. 30. Nochez Y, Favard A, Majzoub S, Pisella PJ (2010) Measurement of corneal aberrations for customisation of intraocular lens asphericity: impact on quality of vision after micro-incision cataract surgery. Br J Ophthalmol 94: 440-444. doi:10.1136/bjo.2009.167775. PubMed: 19833613.
  31. 31. Kennis H, Huygens M, Callebaut F (2004) Comparing the contrast sensitivity of a modified prolate anterior surface IOL and of two spherical IOLs. Bull Soc Belge Ophtalmol 294: 49-58. PubMed: 15682919.
  32. 32. Muñoz G, Albarrán-Diego C, Monte´s-Mico´ R, Rodrı´guez-Galietero A,Alio´JL (2006) Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg 32: 1320–1327. doi:10.1016/j.jcrs.2006.02.055. PubMed: 16863968.
  33. 33. Portney V (2012) New bi-sign aspheric IOL and its application. Optom Vis Sci 89: 80-89. doi:10.1097/OPX.0b013e3182376ba8. PubMed: 22041588.
  34. 34. Guirao A, Tejedor J, Artal P (2004) Corneal aberrations before and after small-incision cataract surgery. Invest Ophthalmol Vis Sci 45: 4312-4319. doi:10.1167/iovs.04-0693. PubMed: 15557437.
  35. 35. Packer M, Fine IH, Hoffman RS, Piers PA (2002) Prospective randomized trial of an anterior surface modified prolate intraocular lens. J Refract Surg 18: 692–696. PubMed: 12458862.
  36. 36. Nakagawa T, Maeda N, Kosaki R, Hori Y, Inoue T et al. (2009) High-order aberrations due to the posterior corneal surface in patients with keratoconus. Invest Ophthalmol Vis Sci 50: 2660-2665. doi:10.1167/iovs.08-2754. PubMed: 19029032.
  37. 37. Wang L, Tang X (2010) Meta-analysis of clinical randomized controlled trials on aspherical intraocular lens. Zhonghua Yan Ke Zhi 46: 254-257.
  38. 38. Erie JC, Bandhauer MH, McLaren JW (2001) Analysis of postoperative glare and intraocular lens design. J Cataract Refract Surg 27: 614-621. doi:10.1016/S0886-3350(00)00781-1. PubMed: 11311633.