Signal-to-noise ratio, the ratio between signal and noise, is a quantity that has been well established for MRI data but is still subject of ongoing debate and confusion when it comes to fMRI data. fMRI data are characterised by small activation fluctuations in a background of noise. Depending on how the signal of interest and the noise are identified, signal-to-noise ratio for fMRI data is reported by using many different definitions. Since each definition comes with a different scale, interpreting and comparing signal-to-noise ratio values for fMRI data can be a very challenging job. In this paper, we provide an overview of existing definitions. Further, the relationship with activation detection power is investigated. Reference tables and conversion formulae are provided to facilitate comparability between fMRI studies.
In science and engineering, the signal-to-noise ratio (SNR) is a measure that compares the level of a desired signal to the level of background noise. For data acquired through magnetic resonance imaging (MRI), this quantification is typically used to allow comparison between imaging hardware, imaging protocols and acquisition sequences. In this context, SNR is conceptualised by comparing the signal of the MRI image to the background noise of the image
Translating SNR of MRI images to fMRI images is not as straightforward as it may seem. First of all, the noise in fMRI images does not correspond to the background noise of MRI images (see
How to define SNR for MRI and fMRI data is documented quite well in terms of its relationship with the MRI acquisition parameters. Several studies have demonstrated the dependence on these parameters and illustrated the necessary conditions to obtain higher SNR
These different conceptualisations of SNR make it hard to compare results over studies. Therefore, in this paper, we provide an overview of the most current SNR definitions for fMRI. First, we demonstrate the use of these definitions in the literature and discuss thoroughly how the definitions can be applied. Second, we show analytic results that will enable the comparison between the several definitions. Third, we also analyzed the relationship with activation detection power and present simulation results that clarifies this relation. Finally, an application on real data illustrates how the definitions can be used in practice.
To retrieve the range of possible values of SNR and CNR, we looked at the reported values of SNR and CNR in fMRI studies. NeuroImage published in 2012 about 458 fMRI studies that presented original fMRI data. These papers were manually screened if they reported SNR/CNR values of their data. In total, 50 of these studies mentioned the role of SNR/CNR for their experiment or method, while only 18 papers also reported SNR or CNR values.
Since the determination of the SNR and CNR of real data can be a demanding job and is not standardly reported, we also looked at the SNR/CNR values that were reported in simulation studies. In simulation studies, the range of the reported SNR/CNR values was determined based on a representative sample of 119 articles describing at least one simulation study. This sample containes papers that were published between 1996 and 2012 in about 49 scientific journals. The reported values varied widely across studies and were almost exclusively labelled as SNR.
A simulation study was conducted to investigate the relationship between SNR/CNR levels and the power to detect activation in a basic GLM analysis.
An activation signal of 200s was modelled for each design. The block design consisted of alternating task and rest blocks that lasted 20s each. For the ER design, 25 events were randomly distributed over the whole time series. For the contrast, two alternating block conditions of 20s each were modelled with a rest period of 20s after each sequence
The empirical power was determined by fitting a standard GLM model to each of the simulated time series. In both the block and the ER design, the power was assessed by testing
As an illustration of the discussed definitions, we applied the SNR/CNR definitions in the context of experimental fMRI data and resting-state fMRI data. The experimental fMRI dataset is based on a houses-faces object recognition paradigm
Reported SNR values ranged over studies from 0.35 to 203.6 for real fMRI data
In the simulation studies, the reported values varied widely across studies and were almost exclusively labelled as SNR. For example, the SNR for the simulations varied from 1 to 10 in one study, while the range was 0.01 to 1 in another, and in yet other studies, we found SNR values that could be negative, for instance, ranging from −13 to 30.
Both in the experimental and simulation studies in our literature search, the reported values demonstrated a range that was much wider than can be explained by natural variation only. There is only one reason that could account for the found variation, namely, the use of different definitions to calculate SNR or CNR. Indeed, several definitions can be found in the literature, especially for CNR. All these CNR measurements model some form of relative signal change, related to the contrast of interest, relative to the noise level. However, there is no consensus on how this contrast of interest should be conceptualised. Therefore, the scale of the CNR definitions varies widely and this makes comparing studies very hard.
Both SNR and CNR definitions have in common that a signal measure is compared to the noise level. The distinction between SNR versus CNR and the differences between the CNR definitions will be the result of how the signal measure and the noise is defined. While discussing the definitions, we will consider fMRI time series as the result of an addition of an activation signal time course and a noise signal time course.
In the overview of the SNR and CNR definitions below, we will focus on those definitions that are commonly found in the literature. In fMRI simulation studies, values for SNR/CNR are often chosen to give an indication of the strength of the modelled signal relative to the modelled noise. Six different definitions were found in total. We will discuss their definition and whether they should be referred to as SNR or CNR. Note that, although in most papers these formulae were labelled as SNR, the majority of them are in fact CNR measurements.
The first definition models SNR based on the mean signal of the fMRI time series and the standard deviation of the noise in the time series
As such, the global signal level, comprised of the baseline and activation, is related to the noise.
In contrast, an amplitude measurement can be related to the standard deviation of the noise
The amplitude of the signal is generally defined as the absolute difference between the baseline of the signal and the signal peak (
The previous definition of the CNR can also be transformed in decibel (dB) scale, which is a common scale in signal processing
Another possibility is to model the strength of the signal based on the standard deviation of the activation signal
This definition is also implemented in the DCM simulator
Again, the ratio of the standard deviations is also found in dB scale
Definition 1 is actually a measurement of tSNR
In contrast, the remainder of the definitions all include some measurement of the activation signal strength. Therefore, these definitions are referred to as CNR formulae. It should be clear that, in theory, the value for these CNR definitions will always be zero for non-active voxels and larger than zero for active voxels. Consequently, theoretically it would be possible to detect active voxels based on their CNR value. In practice however, the activation signal is stricto sensu unknown and it may be complicated to calculate CNR values for single voxels.
For the CNR definitions, two different sets can be distinguished; the first set (Definition 2–3) focuses on the amplitude of the activation signal,
Note that in our literature search, we also found possible negative SNR values, which might be confusing to many neuroscientists. These negative values stem from definitions 3 and 6 that define CNR on the dB scale, which are often used in signal processing. On this logarithmic scale, a value of 0 means that an equal amount of signal compared to noise is present in the data. Therefore, negative values are an indication of less signal than noise, while positive CNR values according to these definitions represent more activation signal than noise.
In the case of a block activation signal (
So far, the definitions described above were only discussed based on a single condition experiment. As soon as multiple conditions are considered in a experiment, it is not quite clear anymore how to calculate the SNR or CNR of the fMRI data. One option could be to determine the SNR/CNR for each condition separately, which would be valid when distinct regions are activated by the conditions. Another option could be to first create an expected activation signal based on a contrast between the conditions, and then to calculate the SNR/CNR of the contrast signal in the same manner as for single condition time series. In this way, the signal of interest is directly based on the contrast that will be tested.
In essence all of these definitions have the same denominator (i.e.
Due to the fact that there is no consensus on how to define the SNR or CNR for fMRI data, interpreting a value can be an almost impossible job. Dependent on how the SNR/CNR is calculated, the values will be on a different scale. This impedes comparability between fMRI studies and consequently delays convergence of conclusions. In order to facilitate the comprehension of SNR and CNR values, three reference tables were assembled (
% Sig. ch. | Power | |||||||
1 | 0.1 | 1003 | 10 | 20 | 4.46 | 19.85 | 12.98 | 1.00 |
0.2 | 502 | 5 | 14 | 2.23 | 4.96 | 6.96 | 1.00 | |
0.5 | 201 | 2 | 6 | 0.89 | 0.79 | −1.00 | 1.00 | |
1 | 100 | 1 | 0 | 0.45 | 0.20 | −7.02 | 0.99 | |
2 | 50 | 0.5 | −6 | 0.22 | 0.050 | −13.04 | 0.58 | |
5 | 20 | 0.2 | −14 | 0.089 | 0.0079 | −21.00 | 0.14 | |
10 | 10 | 0.1 | −20 | 0.045 | 0.0020 | −27.02 | 0.07 | |
2 | 0.1 | 1007 | 20 | 26 | 8.91 | 79.42 | 19.00 | 1.00 |
0.2 | 503 | 10 | 20 | 4.46 | 19.85 | 12.97 | 1.00 | |
0.5 | 201 | 4 | 12 | 1.78 | 3.18 | 5.02 | 1.00 | |
1 | 101 | 2 | 6 | 0.89 | 0.79 | −1.00 | 1.00 | |
2 | 50 | 1 | 0 | 0.45 | 0.20 | −7.02 | 0.99 | |
5 | 20 | 0.4 | −8 | 0.18 | 0.032 | −14.98 | 0.42 | |
10 | 10 | 0.2 | −14 | 0.089 | 0.0079 | −21.00 | 0.15 | |
5 | 0.1 | 1017 | 50 | 34 | 22.28 | 496.35 | 26.96 | 1.00 |
0.2 | 508 | 25 | 28 | 11.14 | 124.09 | 20.94 | 1.00 | |
0.5 | 203 | 10 | 20 | 4.46 | 19.85 | 12.98 | 1.00 | |
1 | 102 | 5 | 14 | 2.23 | 4.96 | 6.96 | 1.00 | |
2 | 51 | 2.5 | 8 | 1.11 | 1.24 | 0.94 | 1.00 | |
5 | 20 | 1 | 0 | 0.45 | 0.1985 | −7.02 | 0.99 | |
10 | 10 | 0.5 | −6 | 0.22 | 0.0496 | −13.04 | 0.59 |
% Sig. ch. | Power | |||||||
1 | 0.1 | 1002 | 10 | 20 | 3.07 | 9.41 | 9.74 | 1.00 |
0.2 | 501 | 5 | 14 | 1.53 | 2.35 | 3.72 | 1.00 | |
0.5 | 200 | 2 | 6 | 0.61 | 0.38 | −4.24 | 0.99 | |
1 | 100 | 1 | 0 | 0.31 | 0.094 | −10.26 | 0.67 | |
2 | 50 | 0.5 | −6 | 0.15 | 0.024 | −16.28 | 0.22 | |
5 | 20 | 0.2 | −14 | 0.06 | 0.0038 | −24.24 | 0.08 | |
10 | 10 | 0.1 | −20 | 0.03 | 0.00094 | −30.26 | 0.06 | |
2 | 0.1 | 1004 | 20 | 26 | 6.14 | 37.64 | 15.76 | 1.00 |
0.2 | 502 | 10 | 20 | 3.07 | 9.41 | 9.74 | 1.00 | |
0.5 | 201 | 4 | 12 | 1.23 | 1.51 | 1.78 | 1.00 | |
1 | 100 | 2 | 6 | 0.61 | 0.38 | −4.24 | 0.99 | |
2 | 50 | 1 | 0 | 0.31 | 0.094 | −10.26 | 0.75 | |
5 | 20 | 0.4 | −8 | 0.13 | 0.015 | −18.23 | 0.17 | |
10 | 10 | 0.2 | −14 | 0.06 | 0.0038 | −24.24 | 0.08 | |
5 | 0.1 | 1010 | 50 | 34 | 15.34 | 235.26 | 23.72 | 1.00 |
0.2 | 505 | 25 | 28 | 7.67 | 58.81 | 17.69 | 1.00 | |
0.5 | 202 | 10 | 20 | 3.07 | 9.41 | 9.74 | 1.00 | |
1 | 101 | 5 | 14 | 1.54 | 2.35 | 3.72 | 1.00 | |
2 | 51 | 2.5 | 8 | 0.77 | 0.59 | −2.31 | 0.99 | |
5 | 20 | 1 | 0 | 0.31 | 0.15 | −10.26 | 0.64 | |
10 | 10 | 0.5 | −6 | 0.15 | 0.024 | −16.28 | 0.21 |
% Sig. ch. | Power | |||||||
1 | 0.1 | 1001 | 10.56 | 20.47 | 3.02 | 9.14 | 9.61 | 1.00 |
0.2 | 501 | 5.28 | 14.45 | 1.51 | 2.28 | 3.59 | 1.00 | |
0.5 | 200 | 2.11 | 6.49 | 0.60 | 0.37 | −4.37 | 0.96 | |
1 | 100 | 1.06 | 0.47 | 0.30 | 0.09 | −10.39 | 0.46 | |
2 | 50 | 0.53 | −5.55 | 0.15 | 0.02 | −16.41 | 0.15 | |
5 | 20 | 0.21 | −13.51 | 0.06 | 0.0037 | −24.37 | 0.07 | |
10 | 10 | 0.11 | −19.53 | 0.03 | 0.0009 | −30.39 | 0.05 | |
2 | 0.1 | 1003 | 21.12 | 26.49 | 6.05 | 36.56 | 15.63 | 1.00 |
0.2 | 501 | 10.56 | 20.47 | 3.02 | 9.14 | 9.61 | 1.00 | |
0.5 | 201 | 4.22 | 12.51 | 1.21 | 1.46 | 1.65 | 1.00 | |
1 | 100 | 2.11 | 6.49 | 0.60 | 0.37 | −4.37 | 1.00 | |
2 | 50 | 1.06 | 0.47 | 0.30 | 0.091 | −10.39 | 0.92 | |
5 | 20 | 0.42 | −7.49 | 0.12 | 0.015 | −18.35 | 0.27 | |
10 | 10 | 0.21 | −13.51 | 0.06 | 0.004 | −24.37 | 0.10 | |
5 | 0.1 | 1007 | 52.79 | 34.45 | 15.12 | 228.50 | 23.59 | 1.00 |
0.2 | 504 | 26.40 | 28.43 | 7.56 | 57.12 | 17.57 | 1.00 | |
0.5 | 201 | 10.56 | 20.47 | 3.02 | 9.14 | 9.61 | 1.00 | |
1 | 101 | 5.28 | 14.45 | 1.51 | 2.28 | 3.59 | 1.00 | |
2 | 50 | 2.64 | 8.43 | 0.76 | 0.57 | −2.43 | 0.99 | |
5 | 20 | 1.06 | 0.47 | 0.30 | 0.09 | −10.39 | 0.47 | |
10 | 10 | 0.53 | −5.55 | 0.15 | 0.02 | −16.41 | 0.16 |
The results in
The CNR definitions based on the amplitude of the signal (Definition 2 and Definition 3) are also partially determined by the baseline since the (maximal) amplitude of the signal will always correspond to the % signal change relative to the baseline. However, given the relative % signal change of the activation or contrast signal, the amplitude is constant over experimental designs. This is not true for the CNR definitions based on the standard deviation of the activation signal (Definition 4, Definition 5 and Definition 6). Although these CNR definitions are completely independent from the baseline, the activation standard deviation will be influenced by the number of events in an ER design or by the length of the epochs in a block design. The reference tables (
Of course, the conversion of one definition to another can also be solved analytically in some cases. For completeness, we explicitly demonstrate here the analytic similarities between the SNR/CNR definitions. Given the percent signal change
Vice versa, a dB CNR value
Since the standard deviation of the activation signal (as in Definition 4–6) will be partially determined by the experimental design, there is no direct way to go from the percent signal change to the standard deviation. To compare these CNR values, either the reference tables, listed here, can be used to provide a rough estimate, or the values have to be calculated specifically for each design.
There is no discussion on the fact that SNR or CNR is somehow related to activation detection power. Indeed, the higher the signal or the lower the noise (i.e. higher values for the SNR/CNR), the higher the power will be. Naively, one could expect that, when, for example, SNR
Looking at the results, we can immediately conclude that the simple rule
For both datasets we calculated the tSNR (Definition 1), the amplitude-based CNR (Definition 2) and the standard deviation CNR (Definition 4) on the raw data. The tSNR was calculated similarly for both datasets. For each voxel the mean and the standard deviation of the corresponding time series were calculated and then divided to determine the tSNR. Note that the interpretation of tSNR is only useful in gray matter. The results are presented in
Upper panel: tSNR results for a block design. Lower panel: tSNR results for the resting-state data.
CNR values are only informative for the task fMRI data, so they were not determined for the resting state data. To calculate them, we first had to create an activation signal. For the task-based data, the activation signal was calculated by averaging the time series of all voxels within the activation contrast mask (i.e. the mask indicating the mean response over all categories). A second activation signal was calculated for all voxels outside the mask. These activation signals were then subtracted from the individual voxels time series to eliminate the contribution of activation to the time series and isolate the noise. The resulting signals, both activation signals and the noise signal, were used to calculate the parameters for the CNR definitions.
The amplitude-based CNR measure (left) is on average 37.06 (with a range from 8.17 to 95.73) for in-mask voxels and 0.48 (with a range from 0.01 to 95.73) for out-mask voxels. Similarly, for the CNR measure using the standard deviation of the signal, in-mask voxels had an average CNR of 0.029 (ranging from 0.06 to 0.76), while out-mask voxels had an average CNR of 0.04 (ranging from 0.01–0.76). Of course, it should be noted that these values are highly dependent on how the mask or ROIs are determined.
fMRI data are often characterised by their SNR or CNR. SNR measurements are, for example, used to compare scanner hardware or the quality of scanning sequences, while CNR can be indicative of the quality (i.e. detectability) of the contrast of interest. In this paper, an overview was provided of common SNR and CNR definitions in an fMRI time series context. It was established that the literature lacks consensus on how to define SNR/CNR for fMRI data. Consequently, reported SNR and CNR values are hard to compare, possibly hindering the convergence of conclusions based on fMRI studies.
Based on how the signal of interest is defined, an explicit distinction was made between SNR and CNR. SNR compares the global signal level to the amount of noise and can be applied to either MRI images or task-related and resting-state fMRI (e.g. tSNR). The main purpose of determining the SNR of the data will be to assess the quality of the data (e.g. influence of noise). However, when applied to task-related fMRI data, the SNR of the data will most likely miss out on the small fluctuations present in the activation signal that are caused by the task. Therefore, in the case of these particular data, in which the signal of interest is a specific contrast that models the influence of certain conditions, it would be better to consistently use the concept of CNR. The CNR value will also give an indication of the quality of the data in terms of noise, but additionally it contains information on the strength of the activation signal for a specific task. This information can be related to activation detection sensitivity.
A sceptical reader would argue that it might be meaningless to capture the information present in 4D fMRI data, which are characterised by very high inter- and intra-subject and -scanner variability, in one single number (either SNR or CNR). Indeed, for real data, SNR or CNR values are seldom reported. Moreover, screening of published simulation studies teaches us that no less than 62.2% of these studies avoid reporting an SNR/CNR value. Instead, they reported separate parameters for the activation strength and the noise level. A second problem might be that the same value of SNR/CNR can indicate different levels of activation strength and noise, which can have a different impact on the detection accuracy. Despite the justly scepticism, determining the SNR or CNR of fMRI data can still hold useful information, because it provides an assesment of the quality of the data at a glance. However, we recommend to calculate the values only for small regions that are likely to have the same value of SNR/CNR based on anatomy or function. For simulation studies in particular, it would be interesting to report the SNR/CNR of the simulated value along with the specific values of activation strength and noise level. As such, generalising the conclusions from these studies to real data will be facilitated.
Consensus on a common SNR/CNR definition for fMRI data might be difficult to achieve, because the measurement depends very much on how the signal of interest and the noise is defined. Therefore, we strongly recommend that authors reporting SNR/CNR values, at least mention the type of definition they use and provide an interpretation on the meaning of the reported values. The tables presented in this chapter can then be a reference allowing easy comparison from one definition to another. Furthermore, these tables are an excellent tool to provide an estimate of the maximal power that can be expected to detect activation in data with a given SNR/CNR value. Finally, a better understanding of the SNR/CNR values might encourage fMRI researchers to report these measurements in a more systematic way. Consequently, the ability to compare these reported values will facilitate the convergence of fMRI based knowledge.