Loop-Mediated Isothermal Amplification for Laboratory Confirmation of Buruli Ulcer Disease—Towards a Point-of-Care Test

  • Marcus Beissner ,

    beissner@lrz.uni-muenchen.de

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Richard Odame Phillips,

    Affiliation Komfo Anokye Teaching Hospital (KATH), Kwame Nkrumah University of Science and Technology (KNUST), Kumasi, Ghana

  • Florian Battke,

    Affiliation Battke Scientia GmbH, Taufkirchen, Germany

  • Malkin Bauer,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Kossi Badziklou,

    Affiliation Institut National d’Hygiène (INH), Ministère de la Santé, Lomé, Togo

  • Fred Stephen Sarfo,

    Affiliation Komfo Anokye Teaching Hospital (KATH), Kwame Nkrumah University of Science and Technology (KNUST), Kumasi, Ghana

  • Issaka Maman,

    Affiliation Institut National d’Hygiène (INH), Ministère de la Santé, Lomé, Togo

  • Agata Rhomberg,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Ebekalisai Piten,

    Affiliation Centre Hospitalier Régional Maritime (CHR-Maritime), Tsévié, Togo

  • Michael Frimpong,

    Affiliation Kumasi Centre for Collaborative Research in Tropical Medicine (KCCR), Komfo Anokye Teaching Hospital (KATH), Kwame Nkrumah University of Science and Technology (KNUST), Kumasi, Ghana

  • Kristina Lydia Huber,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Dominik Symank,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Moritz Jansson,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Franz Xaver Wiedemann,

    Affiliation German Leprosy and Tuberculosis Relief Association, Togo office (DAHW-T), Lomé, Togo

  • Abiba Banla Kere,

    Affiliation Institut National d’Hygiène (INH), Ministère de la Santé, Lomé, Togo

  • Karl-Heinz Herbinger,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • Thomas Löscher,

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  •  [ ... ],
  • Gisela Bretzel

    Affiliation Department of Infectious Diseases and Tropical Medicine (DITM), University Hospital, Ludwig-Maximilians-University, Munich, Germany

  • [ view all ]
  • [ view less ]

Abstract

Background

As the major burden of Buruli ulcer disease (BUD) occurs in remote rural areas, development of point-of-care (POC) tests is considered a research priority to bring diagnostic services closer to the patients. Loop-mediated isothermal amplification (LAMP), a simple, robust and cost-effective technology, has been selected as a promising POC test candidate. Three BUD-specific LAMP assays are available to date, but various technical challenges still hamper decentralized application. To overcome the requirement of cold-chains for transport and storage of reagents, the aim of this study was to establish a dry-reagent-based LAMP assay (DRB-LAMP) employing lyophilized reagents.

Methodology/Principal Findings

Following the design of an IS2404 based conventional LAMP (cLAMP) assay suitable to apply lyophilized reagents, a lyophylization protocol for the DRB-LAMP format was developed. Clinical performance of cLAMP was validated through testing of 140 clinical samples from 91 suspected BUD cases by routine assays, i.e. IS2404 dry-reagent-based (DRB) PCR, conventional IS2404 PCR (cPCR), IS2404 qPCR, compared to cLAMP. Whereas qPCR rendered an additional 10% of confirmed cases and samples respectively, case confirmation and positivity rates of DRB-PCR or cPCR (64.84% and 56.43%; 100% concordant results in both assays) and cLAMP (62.64% and 52.86%) were comparable and there was no significant difference between the sensitivity of the assays (DRB PCR and cPCR, 86.76%; cLAMP, 83.82%). Likewise, sensitivity of cLAMP (95.83%) and DRB-LAMP (91.67%) were comparable as determined on a set of 24 samples tested positive in all routine assays.

Conclusions/Significance

Both LAMP formats constitute equivalent alternatives to conventional PCR techniques. Provided the envisaged availability of field friendly DNA extraction formats, both assays are suitable for decentralized laboratory confirmation of BUD, whereby DRB-LAMP scores with the additional advantage of not requiring cold-chains. As validation of the assays was conducted in a third-level laboratory environment, field based evaluation trials are necessary to determine the clinical performance at peripheral health care level.

Author Summary

Buruli ulcer disease (BUD) mainly occurs in remote rural areas of Sub-Saharan Africa, affects skin and soft tissue, and may lead to severe disabilities. Therefore, early diagnosis and treatment with antimycobacterial therapy are essential whereby the WHO recommends laboratory confirmation of 70% of the cases. As the current diagnostic gold standard (polymerase chain reaction [PCR]) is restricted to third-level laboratories, development of confirmatory point-of-care (POC) tests for BUD applicable at primary health care level has become a research priority to bring diagnosis closer to where the patients are. Loop-mediated isothermal amplification (LAMP) has been selected by the WHO as one of the promising candidate technologies for POC tests. The aim of this study was to establish and validate a LAMP assay applying lyophilized reagents which are stable at ambient temperature, thus avoiding the need for cold-chains. The results from this study suggest that the assay provides a valuable alternative to other PCR tests as currently used for laboratory confirmation of BUD.

Introduction

Buruli ulcer disease (BUD), caused by Mycobacterium ulcerans, is an infectious disease affecting skin, soft tissues and sometimes the bones. The major endemic foci occur in rural areas of Sub-Saharan Africa where BUD mainly affects children below the age of 15 years.

Antimycobacterial therapy can cure up to 80% of patients diagnosed in early stages of the disease. If treated in advanced stages or left untreated, extensive destruction of tissue followed by fibrous scarring and contractures may lead to severe sequelae such as functional limitation of affected joints, which occur in up to 25% of cases. In the absence of proven preventive strategies, early diagnosis and treatment are therefore crucial to avoid disease related disabilities [12].

The WHO recommends laboratory confirmation of at least 70% of clinically suspected BUD cases per country [3]. Application of the 100% M. ulcerans specific diagnostic reference standard for clinical samples, i.e. amplification of the multicopy insertion sequence (IS) 2404 by dry-reagent-based (DRB) PCR, conventional gel-based PCR (cPCR), or quantitative real-time PCR (qPCR) requires fully equipped molecular biology units with highly-skilled personnel and is thus mostly restricted to tertiary (reference) level laboratories or national research centres [49]. However, as the major burden of BUD exists in (remote) rural areas of endemic countries and up to one-third of BUD cases are diagnosed in advanced category III stages [1012], molecular IS2404 detection formats applicable as point-of-care (POC) tests are urgently needed to bring diagnosis closer to where the patients live [13].

Behind this background, an expert group convened by the Foundation for New Innovative Diagnostics (FIND) and the WHO in November 2013 selected loop-mediated isothermal amplification (LAMP) as promising nucleic acid based candidate POC technology applicable for decentralized diagnosis at primary health care level [14].

The salient features of LAMP technology are attributable to the Bacillus stearothermophilus-derived Bst polymerase, which is characterized by strand-displacement activity (without 5’-3’ exonuclease activity), enzyme activity at constant temperature (~ 65 +/- 3°C) without the need of steps for denaturation of double-stranded DNA or primer annealing at different temperatures, high amplification efficiency (up to 1010 copies in 60 minutes) and low susceptibility to classical PCR inhibitors (e.g. melanin, collagen, humic acids). Furthermore, the ability to specifically amplify target sequences by the use of four distinct primers recognizing 6 distinct regions in a single step without the need for sophisticated laboratory equipment made this nucleic acid detection method promising as POC test. LAMP applications were thus established and validated for the diagnosis of various human pathogens such as (protozoan) parasites (e.g. Plasmodium falciparum, Leishmania spp., Trypanosoma brucei, Giardia duodenalis, Schistosoma mansoni/haematobium, Taenia solium), bacteria (e.g. Listeria monocytogenes, Staphylococcus aureus, Mycobacterium tuberculosis) as well as viruses and fungi in settings with limited resources [1525].

To date, three different LAMP assays for laboratory confirmation of BUD were published. The assay described by de Souza et al. targets the enoyl reductase gene of the M. ulcerans virulence plasmid, technical validation of the assay however was conducted only on a limited number of samples [26]. Njiru et al. and Ablordey et al. reported two LAMP assays amplifying different regions of the IS2404 of M. ulcerans. Both assays underwent validation on various clinical and environmental samples of BUD patients and infected animals from Ghana and Australia, were 100% M. ulcerans specific (without any false positive result) and revealed analytical sensitivities of 20 [27] as well as 30–300 [28] copies of the respective IS2404 target sequence, which equals 0.1 to 1.5 genome equivalents of M. ulcerans, respectively. These analytical sensitivities approach that of cPCR [6, 2728], but not that of qPCR [7, 29]. However, both assays were evaluated under optimal laboratory conditions applying high-standard DNA extraction and purification procedures in third level laboratories or national research centers, which may not be practicable at primary health care level. To simulate technical feasibility under field conditions, crude (i.e. boiled) DNA extracts were used without further purification for LAMP testing of clinical samples and led to a significant decrease in sensitivity [28]. Moreover, all LAMP assays described so far require unlimited cold-chains as well as shipment of reagents on dry-ice, which is a major cost factor for endemic settings and not always feasible at decentralised facilities. Therefore, technical advancement of LAMP technology and DNA extraction into utterly field friendly formats is unanimously recommended [2728].

Against this background, the aim of this study was to establish an IS2404 detection based LAMP assay employing lyophilized reagents (dry-reagent-based [DRB] LAMP) which provides significant benefit for application under tropical climate conditions, to validate the assay on clinical samples including fine needle aspirates (FNA) which were largely omitted in previous studies, and to provide a prototype assay for future large-scale field testing.

Materials and Methods

Ethical statement

The study was approved by the Ghanaian KNUST (CHRPE/91/10) and the national Togolese (14/2010/CRBS) ethics committees. All samples analyzed in this study were collected for diagnostic purposes. Written informed consent was obtained from all study participants and/or their legal representative, if aged below 18 years.

Study participants, clinical samples and data collection

Clinically suspected BUD patients were recruited from two study sites in Ghana (Agogo Presbyterian Hospital, Asante Akim North District, n = 12; Tepa Government Hospital, Ahafo Ano North District, n = 20) and one study site in Togo (“Centre Hospitalier Régional de Tsévié”, region “Maritime”, n = 59) and 140 diagnostic samples (FNA, n = 66; swab samples, n = 32; punch biopsy samples, n = 42) were collected according to standardized procedures. Briefly, swabs were taken by circling the undermined edges of ulcerative lesions, and FNA or 3mm punch biopsies were obtained from the center of non-ulcerative lesions. Samples were transported to the Kumasi Center for Collaborative Research in Tropical Medicine (KCCR, Kumasi, Ghana) or the “Institut National d’Hygiène” (INH, Lomé, Togo) in 2 ml screw cap tubes containing 700 μl (swab and punch biopsy samples) or 300 μl (FNA samples) cell lysis solution (CLS; Qiagen, Hilden, Germany) within one day at ambient temperature [10, 3033].

Clinical, epidemiological and routine laboratory data were collected by means of WHO BU 01.N forms [34] and standardized project specific laboratory data entry forms, and were entered in a web-based database as previously described [10].

Laboratory confirmation by PCR

Whole genome DNA was extracted from clinical samples in CLS at KCCR or INH using the Gentra Puregene DNA extraction kit (Qiagen) with minor modifications of the manufacturer’s instructions as described in S1 Protocol [6]. DNA extracts were stored at 4–8°C (up to one week) or -18°C (long-term storage).

For routine on-site laboratory confirmation DNA extracts were subjected to IS2404 DRB PCR at KCCR and INH as previously described [6, 10, 35]. For comparative testing in the context of external quality assurance programs with conventional, gel-based IS2404 PCR (cPCR) [45, 33, 35, 36] and a recently described modified IS2404 qPCR based on the assay published by Fyfe et al. [7, 29] aliquots of DNA extracts were shipped to the Department of Infectious Diseases and Tropical Medicine (DITM), Munich, Germany by courier service at ambient temperature [10].

Development of an IS2404 detection-based LAMP assay

The development and validation of the LAMP assay was conducted in the laboratories of DITM.

DNA extracts.

For establishment and technical validation of an IS2404 LAMP assay, sequencing confirmed DNA extracts of five M. ulcerans strains from cultures (i.e. “must detect” samples), twelve mycobacterial species (i.e. “must not detect” samples: M. avium, M. chelonae, M. fortuitum, M. gordonae, M. intracellulare, M. kansasii, M. marinum, M. smegmatis, M. szulgai, M. tuberculosis, M. xenopi and M. lentiflavum) were available at DITM [29, 3738].

IS2404 plasmid standard.

To generate a plasmid standard applicable as positive control and calibration template with known copy numbers for regions amplified by gel-based IS2404 (DRB) PCR, IS2404 qPCR as well as the novel IS2404 LAMP assays, respectively, the complete IS2404 sequence was amplified by conventional PCR from a sequencing confirmed M. ulcerans culture extract. The primers were IS2404-fwd (5`-3`: ATG GCT TTG TTG GCG ATC GC) and IS2404-rev (5`-3`: TTA GCA GGC TTG TGA GCT GG). The reaction mixture contained 13.5 μl molecular grade H2O (Carl Roth, Karlsruhe, Germany), 2.5 μl 10-fold PCR buffer for Thermococcus kodakaraenis (KOD) derived DNA polymerase (Merck, Darmstadt, Germany), 2.0 μl MgSO4, 2.5 μl dNTP mix (2 mM each), 10 pmol of each primer, 2 μl DNA template and 0.5 μl KOD Hot-Start Polymerase (Merck). The amplification was performed at 95°C for 2 min., followed by 35 cycles of 95°C for 20 sec., 57°C for 15 sec., 70°C for 20 sec. and a final incubation at 70°C for 2 min. The PCR product was purified from a 1.2% agarose TAE gel by means of the Double Pure Kit (Bio&SELL, Feucht, Germany) according to the manufacturer’s instructions. Then, a 3’A-overhang was added to the purified PCR product by incubating the following reaction mixture for 20 minutes at 72°C: 38 μl purified PCR product, 5 μl PCR-reaction buffer (10-fold), 5 μl MgCl2 (25 mM), 1 μl dATP (10 mM), 1 μl Taq polymerase (Bio&SELL, each). The ‘A’-tailed PCR product was again purified with the Double Pure Kit and then ligated into a pGEM-T-vector (Promega, Mannheim, Germany) using the following reaction mixture incubated at 4°C overnight: 5 μl 2-fold Rapid Ligation Buffer, 3 μl DNA (with 3’A-overhangs), 1 μl pGEM-T-vector and 1 μl T4 DNA Ligase. For cloning the plasmid to bacteria, 2 μl of the ligation reaction were mixed with 50 μl E. coli JM109 z-competent cells (Zymo, Freiburg, Germany), inoculated onto LB agar plates with 100 mg/L ampicillin (Carl Roth) and incubated at 37°C overnight. The selected E. coli clone was cultivated in 5 ml LB medium with 100 mg/L ampicillin (Carl Roth), incubated at 37°C overnight and then subjected to plasmid preparation using the matrix-based HiYield Plasmid Mini kit (Süd-Laborbedarf, Gauting, Germany) according to the manufacturer’s specification. The cloned plasmid sequence was confirmed by direct DNA sequencing as previously described [37]. Purity of extracted IS2404 plasmids was assessed by photometry on a BioPhotometer plus (Eppendorf, Wesseling-Berzdorf, Germany) and agarose gel-electrophoresis on a 1% TAE gel. Quantification of plasmid DNA extracts was done using the fluorescence quantification kit “Quant-It dsDNA Broad Range” on a Qubit (Life Technologies, Karlsruhe, Germany) according to the manufacturer’s instruction and plasmid standards were adjusted to 106 copies per μl.

Analytical sensitivity was determined as lower limit of detection (LOD), i.e. lowest template concentration rendering positive amplification of 95% of samples. The LOD of each assay (including DRB PCR, cPCR and qPCR) was determined at DITM using the plasmid standard in 10-fold serial dilutions [39].

IS2404 LAMP primers.

A set of four primers was designed for amplification of the M. ulcerans specific IS2404 by manually analyzing the target sequence and designing the primers according to the needs for LAMP amplification. Specificity of the primers for M. ulcerans was confirmed in silico by means of the basic local alignment search tool (BLAST, GenBank, NCBI) [40]. The primer set consisted of two smaller oligonucleotides with forward (MU2-F3) and reverse complementary (MU2-B3) sequences and two larger oligonucleotides (MU2-FIP and MU2-BIP) with a complex sequence construction. Primer sequences are provided in Table 1 and binding sites within the IS2404 are displayed in Fig 1. Primers MU2-B3 and MU2-BIP were first published by Ablordey et al. [28].

thumbnail
Fig 1. Binding sites of LAMP primers within the IS2404.

Fig 1 shows binding sites of LAMP primers within the IS2404 of M. ulcerans strain Agy 99 (Genbank accession number: CP000325.1). Primer MU2-FIP consists of a first reverse complementary region “F1” and a second forward region “F2”. Primer MU2-BIP (first described by Ablordey et al. [28]) consists of a first forward region “B1” and a second reverse complementary region “B2”. Primers and corresponding regions are highlighted in colors; red: primer MU2-F3 (region “F3”); yellow: primer MU2-FIP (region “F2”); light blue: primer MU2-FIP (region “F1”); dark blue: primer MU2-BIP (region “B1”); green: primer MU2-BIP (region “B2”) and pink: primer MU2-B3 (region “B3”; first described by Ablordey et al. [28]).

https://doi.org/10.1371/journal.pntd.0004219.g001

Conventional IS2404 LAMP (cLAMP) protocol.

Each cLAMP reaction mix consisted of 1 μl Bst DNA polymerase (large fragment, 8 U/μl; New England Biolabs [NEB], Frankfurt am Main, Germany), 1.0 μl dNTP mix (2 mM each, Merck), 1.0 μl of primers MU2-F3 and MU2-B3 (5 pmol/μl) and 2.0 μl of primers MU2-FIP and MU2-BIP (10 pmol/μl), respectively (TibMolBiol, Berlin, Germany), 2.0 μl betaine (5 M; Sigma-Aldrich, Taufkirchen, Germany), 2.5 μl 10-fold Thermopol buffer for Bst DNA polymerase (NEB) and 11.5 μl molecular grade H2O (Carl Roth). Following the addition of 1 μl DNA extract (template), cLAMP reactions (final volume: 25 μl) were carried out in 1.5 ml SafeSeal reaction tubes (Sarstedt, Nümbrecht, Germany) at 65°C for 60 minutes in a conventional thermoblock (HLC Thermomixer MKR 13, HLC BioTech, Bovenden, Germany) and a final step at 80°C for 10 minutes terminated the amplification.

Each run included negative extraction and no template (H2O) as well as positive controls.

IS2404 dry-reagent-based LAMP (DRB LAMP) protocol.

The DRB LAMP reaction mix contained the same concentrations of reagents and primers as described for the conventional LAMP assay. However, as the wildtype Bst polymerase large fragment contained glycerol it was not possible to lyophilize. Therefore, a customized Bst polymerase dissolved in H2O and reaction buffer (NEB) were applied for the DRB LAMP assay. The DRB LAMP reaction mix was prepared for each reaction in 1.5 ml reaction tubes and subjected to lyophilization by means of a RVC 2–25 CD plus Vacuum Concentrator (Christ, Osterode, Germany) at 1.0 mbar and a safety pressure of 1.000 mbar according to the manufacturer’s specifications. During the process of validation, DRB LAMP reaction tubes were stored at ambient temperature in the dark and reactions were carried out within one week after lyophilization as described for cLAMP following the addition of 1 μl template DNA and adjustment with 24 μl molecular grade H2O (Carl Roth) to a final reaction volume of 25 μl.

LAMP product visualization.

LAMP products were detected by two different methods: i) gel-electrophoresis on a 0.5x TBE gel containing 0.01% GelRed (Biotium, Cologne, Germany) and ii) SYBR Green I (Life Technologies) staining (0.5 μl of 1:5 diluted SYBR Green I staining solution and 12.5 μl of LAMP product) followed by UV-transillumination.

Comparative testing of clinical samples

Confirmation rates of routine assays (PCR [DRB PCR, cPCR] and qPCR) compared with cLAMP.

The confirmation rate was defined as the number of patients with a positive PCR, qPCR or cLAMP test result divided by the number of all suspected BUD cases.

Sensitivity rates of PCR (DRB PCR, cPCR) and cLAMP among confirmed BUD patients.

The sensitivity rate was defined as the number of patients with a positive PCR or cLAMP test result divided by the number of all qPCR confirmed patients.

Positivity rates of routine assays (PCR [DRB PCR, cPCR] and qPCR) and cLAMP among clinical samples from suspected BUD cases.

The positivity rate was defined as the number of clinical samples from suspected BUD cases with a positive PCR, qPCR or cLAMP test result divided by the number of all samples tested.

Sensitivity rates of routine assays (PCR [DRB PCR, cPCR]) compared with cLAMP among clinical samples from confirmed BUD cases.

The sensitivity rate was defined as the number of clinical samples from confirmed BUD patients with a positive PCR or cLAMP test result divided by the number of all samples with a positive qPCR result.

Positivity rates of DRB LAMP compared with routine assays (PCR [DRB PCR, cPCR] and qPCR) and cLAMP among clinical samples from suspected BUD cases.

The positivity rate was defined as the number of clinical samples with a positive PCR, qPCR or cLAMP test result divided by the number of all samples tested.

Sensitivity rates of cLAMP and DRB LAMP among clinical samples from confirmed BUD cases.

The sensitivity rate was defined as the number of clinical samples from confirmed BUD patients with a positive cLAMP or DRB LAMP test result divided by the number of all samples with a positive result in DRB PCR, cPCR and qPCR each.

Statistical analysis

The study was observational and transversal (cross-sectional study design). An approximative test (McNemar chi-square test for matched pairs of samples with categorical test results) and estimation of standard error of proportion (to calculate 95 percent confidence intervals [95%-CI] of categorical test results) were conducted. Significant differences were defined as P-values below 0.05 or as not overlapping of 95%-CI of proportions.

Results

Performance characteristics of routine assays (DRB PCR, cPCR and qPCR) compared with cLAMP

In silico analysis of the novel IS2404 LAMP primers and testing of 17 DNA extracts from mycobacterial cultures (M. ulcerans, n = 5; other mycobacteria, n = 12) by cLAMP and DRB LAMP revealed 100% specificity of both assays for M. ulcerans. The LODs were 50 (DRB PCR and cPCR), 3 (qPCR) and 100 (cLAMP) copies of the target sequence IS2404, corresponding to 0.2, 0.01, and 0.5 M. ulcerans genome equivalents, respectively.

Laboratory confirmation of suspected BUD cases

Out of the 91 patients with suspected BUD, 68 were laboratory confirmed as BUD patients (74.73%) by routine methods. Among 68 confirmed BUD patients, 40 patients (58.82%) were in age group 5–14 years (age range 5–56 years, mean 14 years, median 11 years), 33 patients (48.53%) were male, and 36 patients (52.94%) presented with non-ulcerative lesions.

Confirmation rates of routine assays (DRB PCR, cPCR and qPCR) compared with cLAMP among suspected BUD cases

DRB PCR (on-site at KCCR or INH) and cPCR (DITM, 100% concordance between DRB and cPCR results) confirmed 59/91 (64.84%; 95%-CI: 55.02%-74.65%) of the suspected BUD cases, the qPCR confirmed 68/91 (74.73%; 95%-CI: 65.80%-83.65%), thus added an additional diagnostic yield of 9.89%. The confirmation rate for cLAMP was 62.64% (95%-CI: 52.70%-72.58%; n = 57). Neither DRB PCR nor cPCR or cLAMP had false positive results compared with qPCR, and confirmation rates were not significantly different. According to McNemar test, there was no significant difference between DRB PCR or cPCR (100% concordant results) compared with qPCR (ORcrude = 1.60; 95%-CI: 0.81–3.20; P-value = 0.15), between cLAMP compared with qPCR (ORcrude = 1.76; 95%-CI: 0.89–3.50; P-value = 0.08) and between DRB PCR or cPCR compared with cLAMP (ORcrude = 1.10; 95%-CI: 0.57–2.11; P-value = 0.76).

Sensitivity rates of DRB PCR, cPCR and cLAMP among confirmed BUD cases

Among the 68 BUD cases confirmed by qPCR, the sensitivity was 86.76% (95%-CI: 78.71%-94.82%; n = 59) for DRB PCR and cPCR and 83.82% (95%-CI: 75.07%-92.58%; n = 57) for cLAMP. According to McNemar test, there was no significant difference between DRB PCR and cPCR compared with cLAMP (ORcrude = 1.27; 95%-CI: 0.44–3.63; P-value = 0.63).

Positivity rates of routine assays (DRB PCR, cPCR and qPCR) and cLAMP among clinical samples from suspected BUD cases

Among the 140 samples from 91 clinically suspected BUD cases, the positivity rate was 56.43% (95%-CI: 48.21%-64.64%; n = 79) for DRB PCR and cPCR, 67.14% (95%-CI: 59.36%-74.92%; n = 94) for qPCR, and 52.86% (95%-CI: 44.59%-61.13%; n = 74) for cLAMP. Neither DRB PCR nor cPCR or cLAMP revealed false positive results compared with qPCR. According to McNemar test, there was no significant difference between DRB PCR or cPCR compared with qPCR (ORcrude = 1.58; 95%-CI: 0.94–2.64; P-value = 0.07) and between DRB PCR or cPCR compared with cLAMP (ORcrude = 1.16; 95%-CI: 0.70–1.90; P-value = 0.55), whereas the difference between cLAMP compared with qPCR was significant (ORcrude = 1.82; 95%-CI: 1.09–3.05; P-value = 0.02).

Stratification into sample types did not reveal significant differences in positivity rates of DRB PCR, cPCR, qPCR and cLAMP among FNA, swab or punch biopsy samples.

Sensitivity rates of routine assays (DRB PCR and cPCR) compared with cLAMP among clinical samples from confirmed BUD cases

Among the 94 samples from 68 BUD cases confirmed by qPCR, the sensitivity was 84.04% (95%-CI: 76.64%-91.45%; n = 79) for DRB PCR and cPCR, and 78.72% (95%-CI: 44.59%-61.13%; n = 74) for cLAMP. According to McNemar test, there was no significant difference between DRB PCR or cPCR compared with cLAMP (ORcrude = 1.42; 95%-CI: 0.64–3.18); P-value = 0.35).

Performance characteristics of DRB LAMP

DRB LAMP revealed the same performance characteristics as determined for cLAMP (i.e. 100% M. ulcerans specificity and a LOD of 0.5 M. ulcerans genome equivalents).

Positivity rates of DRB LAMP compared with routine assays (DRB PCR, cPCR and qPCR) and cLAMP among clinical samples from suspected BUD cases

To compare DRB LAMP with DRB PCR, cPCR, qPCR and cLAMP, 32 samples (FNA and swab samples, n = 16, respectively) from 32 suspected BUD cases were subjected to the assays. The positivity rate was 75.0% (95%-CI: 62.41%-87.59%; n = 24) for DRB PCR, cPCR and qPCR, 71.88% (95%-CI: 87.84%-100%; n = 23) for cLAMP, and 68.75% (95%-CI: 80.61%-100%; n = 22) for DRB LAMP. Neither cLAMP nor DRB LAMP revealed false positive results compared with DRB PCR, cPCR and qPCR.

According to McNemar test, there was no significant difference neither between DRB PCR, cPCR or qPCR compared with cLAMP (ORcrude = 1.17; 95%-CI: 0.34–4.10; P-value = 0.78), nor between DRB PCR, cPCR or qPCR compared with DRB LAMP (ORcrude = 1.36; 95%-CI: 0.40–4.49; P-value = 0.58), nor between cLAMP compared with DRB LAMP (ORcrude = 0.86; 95%-CI: 0.26–2.87; P-value = 0.79).

Sensitivity rates of cLAMP and DRB LAMP among clinical samples from confirmed BUD cases

Among the 24 samples from 24 BUD patients confirmed by DRB PCR, cPCR and qPCR, the sensitivity was 95.83% (95%-CI: 87.84%-100%; n = 23) for cLAMP and 91.67% (95%-CI: 80.61%-100%; n = 22) for DRB LAMP. According to McNemar test, there was no significant difference between cLAMP compared with DRB LAMP (ORcrude = 0.48; 95%-CI: 0.02–7.54); P-value = 0.56).

IS2404 LAMP detection methods

Out of 74 amplicons derived from cLAMP reactions, 74/74 (100%) were judged positive by gel-electrophoresis and 73/74 (98.65%) by SYBR Green I staining. All products derived from DRB LAMP were likewise analyzed and the concordance rate was 22/22 (100%) between both detection methods.

Table 2 shows confirmation rates, sensitivity, specificity and significance of the applied molecular tests.

thumbnail
Table 2. Confirmation rates, sensitivity, specificity and significance of the applied molecular tests.

https://doi.org/10.1371/journal.pntd.0004219.t002

Discussion

BUD belongs to the currently five neglected diseases in line for the IDM (innovative and intensified disease management) approach demanding a major scaling up of active detection, treatment, monitoring and surveillance. Development of diagnostic tests that bring health services closer to where NTDs are is considered a research priority. LAMP, a technology that features cost effectiveness, robustness and modest needs in terms of equipment, has recently been selected by the WHO as one of the promising tools for decentralized diagnostics [1314, 41].

Several investigators recently succeeded in developing M. ulcerans specific LAMP assays which showed performance characteristics comparable to conventional PCR formats [2628]. Based on longstanding experience with a DRB PCR format for laboratory confirmation of BUD in Ghana and Togo [6, 10, 30, 33, 36, 42], the development of a DRB LAMP assay applicable under tropical climate conditions at primary health care level was envisaged in this study. Thermodynamic reasons (i.e. leaving out an initial denaturation step for annealing of primers) required the design of modified primers, therefore as a first step a new cLAMP assay was established that constituted the basis for the DRB format. During development of the DRB assay lyophilization of the reaction mix initially constituted a major challenge. Due to the glycerol content of Bst polymerase and reaction buffer as employed in previous cLAMP formats (including our own), customized glycerol-free reagents had to be obtained and adequate lyophilization protocols had to be established.

The comparable performance of cLAMP, DRB LAMP and DRB PCR as well as cPCR suggests that both LAMP formats constitute a reliable alternative to conventional routine assays. Our data also show that the LAMP assays and the DRB PCR as well as cPCR have equal sensitivity for FNA samples. Both LAMP formats are applicable at primary health care level, the DRB format however provides significant advantages such as a simplified test layout and the possibility of storage of reagents at ambient temperature. Decentralized utilization of LAMP technology furthermore would lead to cost saving due to reduced expenditures for transportation of samples to a reference center as well as reduced test costs, i.e. US$ 1–2 per LAMP reaction as compared to US$ 8–10 per DRB PCR or cPCR reaction.

In this study it was not possible to assess long-term storage of DRB-LAMP reaction tubes under tropical conditions. Long-term storability of DRB PCR reaction tubes was however previously proven [6, 36] which allows the conclusion that maximum storage periods of up to 12 months also apply for LAMP reagents.

Although in our study routine PCR and LAMP assays for the most part did not perform significantly different from qPCR, it must be assumed that qPCR renders an additional diagnostic yield of approximately 10% [10]. Therefore, regardless of the method used, confirmation of negative samples by qPCR e.g. through the global network of laboratories for confirming M. ulcerans infection [43] should be attempted. Likewise, participation of laboratories in external quality assurance programs as implemented by Eddyani et al. in collaboration with the WHO is strongly recommended [44].

While the amplification procedure of LAMP technology especially in the DRB format can be considered field friendly without restriction, current DNA extraction procedures are not yet entirely appropriate for POC testing and need optimization. As shown by Ablordey et al. the use of boiled crude DNA extracts led to a significant decrease in sensitivity [28]. Other options such as one-tube silica-membrane based extraction protocols [45] or one-tube enzyme-based lyophilized reactions are yet to be evaluated. A field friendly approach to storage of DNA extracts for purposes of quality assurance could be the filter paper technology as successfully applied for TBC [46].

In conclusion, the cLAMP and DRB LAMP formats evaluated in this study are equivalent alternatives to conventional PCR techniques and, provided the availability of field friendly DNA extraction formats, constitute valuable tools for decentralized laboratory confirmation of BUD. As in the case of other investigators who previously developed BUD specific LAMP assays, the validation of the LAMP assays presented in this study was conducted in a third-level laboratory environment, therefore field based evaluation trials are necessary to determine the clinical performance at peripheral health care level.

Supporting Information

S1 Protocol. Extraction of mycobacterial DNA from clinical specimens.

https://doi.org/10.1371/journal.pntd.0004219.s001

(PDF)

Acknowledgments

The authors thank Kerstin Helfrich and Carolin Mengele (DITM) for laboratory assistance, as well as Dr. Basile Kobara (“Programme National de Lutte contre l’Ulcère de Buruli, Lèpre et Pian”, Lomé, Togo) for excellent collaboration. This manuscript contains parts of the doctoral thesis of Dominik Symank and Moritz Jansson.

Author Contributions

Conceived and designed the experiments: MBe ROP FB GB. Performed the experiments: MBe MBa MF IM AR KLH DS MJ. Analyzed the data: MBe ROP FB MBa KB FSS FXW ABK KHH TL GB. Contributed reagents/materials/analysis tools: ROP FB KB FXW ABK KHH TL GB. Wrote the paper: MBe ROP FB MBa TL GB. Patient management: ROP FSS EP. Study logistics: FXW.

References

  1. 1. WHO. Treatment of Mycobacterium ulcerans disease (Buruli ulcer): guidance for health workers. Geneva, Switzerland: World Health Organization. 2012.
  2. 2. WHO. Buruli Ulcer. Fact sheet 199. Geneva, Switzerland: World Health Organization. 2007 updated 2015.
  3. 3. WHO. Recommendations for control of buruli ulcer. Geneva, Switzerland: World health organisation. 2013.
  4. 4. Ross BC, Johnson PDR, Oppedisano F, Marino L, Sievers A, Stinear T, et al. Detection of Mycobacterium ulcerans in Environmental Samples during an Outbreak of Ulcerative Disease. Applied and Environmental Microbiology. 1997;63(10):4135–38. pmid:9327583
  5. 5. Stinear T, Ross BC, Davies JK, Marino L, Robins-Browne RM, Oppedisano F, et al. Identification and Characterization of IS2404 and IS2606: Two Distinct Repeated Sequences for Detection of Mycobacterium ulcerans by PCR. Journal of Clinical Microbiology. 1999;37(4):1018–23. pmid:10074520
  6. 6. Siegmund V, Adjei O, Racz P, Berberich C, Klutse E, van Vloten F, et al. Dry-reagent-based PCR as a novel tool for laboratory confirmation of clinically diagnosed Mycobacterium ulcerans-associated disease in areas in the tropics where M. ulcerans is endemic. J Clin Microbiol. 2005;43(1):271–6. pmid:15634982
  7. 7. Fyfe JAM, Lavender CJ, Johnson PDR, Globan M, Sievers A, Azuoloas J, et al. Development and Application of Two Multiplex Real-Time PCR Assays for the Detection of Mycobacterium ulcerans in Clinical and Environmental Samples. Applied and Environmental Microbiology. 2007;73(15):4733–40. pmid:17526786
  8. 8. Beissner M, Herbinger KH, Bretzel G. Laboratory diagnosis of buruli ulcer disease. Future Microbiol. 2010;5(3):363–70. pmid:20210548
  9. 9. WHO. Ed Portaels F. Laboratory diagnosis of buruli ulcer: a manual for health care providers. Geneva, Switzerland: World Health Organization; 2014.
  10. 10. Beissner M, Huber KL, Badziklou K, Halatoko WA, Maman I, Vogel F, et al. Implementation of a National Reference Laboratory for Buruli Ulcer Disease in Togo. PLoS Negl Trop Dis. 2013;7(1):e2011. pmid:23359828
  11. 11. Vincent QB, Ardant MF, Adeye A, Goundote A, Saint-André JP, Cottin J, et al. Clinical epidemiology of laboratory-confirmed Buruli ulcer in Benin: a cohort study. Lancet Glob Health. 2014;2(7):e422–30. pmid:25103396
  12. 12. Philips RO, Sarfo FS, Abass MK, Abotsi J, Wilson T, Forson M, et al. Clinical and Bacteriological Efficacy of Rifampin-Streptomycin Combination for Two Weeks followed by Rifampin and Clarithromycin for Six Weeks for Treatment of Mycobacterium ulcerans Disease. Antimicrob. Agents Chemother. 2014;58(2):1161–66. pmid:24323473
  13. 13. WHO. Investing to overcome the global impact of neglected tropical diseases: Third WHO report on neglected tropical diseases. Geneva, Switzerland: World Health Organization; 2015.
  14. 14. WHO. Report of a WHO-FIND consulatative meeting on diagnostics for Buruli ulcer. Geneva, Switzerland: World Health Organization; 2013.
  15. 15. Poon LL, Wong BW, Ma EH, Chan KH, Chow LM, Abeyewickreme W, et al. Sensitive and inexpensive molecular test for falciparum malaria: detecting Plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clin Chem. 2006;52(2):303–6. pmid:16339303
  16. 16. Paris DH, Imwong M, Faiz AM, Hasan M, Yunus EB, Silamut K, et al. Loop-Mediated Isothermal PCR (LAMP) for the Diagnosis of Falciparum Malaria. Am J Trop Med Hyg. 2007;77(5):972–76. pmid:17984362
  17. 17. Pandey BD, Poudel A, Yoda T, Tamaru A, Oda N, Fukushima Y, et al. Development of an in-house loop-mediated isothermal amplification (LAMP) assay for detection of Mycobacterium tuberculosis and evaluation in sputum samples of Nepalese patients. J Med Microbiol. 2008;57(4):439–43.
  18. 18. Abbasi I, King CH, Muchiri EM, Hamburger J. Detection of Schistosoma mansoni and Schistosoma haematobium DNA by Loop-Mediated Isothermal Amplification: Identification of Infected Snails from Early Prepatency. Am J Trop Med Hyg. 2010;83(2):427–32. pmid:20682894
  19. 19. Nkouawa A, Sako Y, Li T, Chen X, Wandra T, Swastika IK, et al. Evaluation of a loop-mediated isothermal amplification method using fecal specimens for differential detection of Taenia species from humans. J Clin Microbiol. 2010;48(9):3350–52. pmid:20631114
  20. 20. Tang MJ, Zhou S, Zhang XY, Ou XY, Ge QL, Tang XJ, et al. Rapid and sensitive detection of Listeria monocytogenes by loop-mediated isothermal amplification. Curr Microbiol. 2011;63(6):511–16. pmid:21935669
  21. 21. Neonakis IK, Spandidos DA, Petinaki E. Use of loop-mediated isothermal amplification of DNA for the rapid detection of Mycobacterium tuberculosis in clinical specimens. Eur J Clin Microbiol Infect Dis. 2011;30:937–42. pmid:21331481
  22. 22. Khan MG, Bhaskar KR, Salam MA, Akther T, Pluschke G, Mondal D. Diagnostic accuracy of loop-mediated isothermal amplification (LAMP) for detection of Leishmania DNA in buffy coat from visceral leishmaniasis patients. Parasit Vectors. 2012;5:280. pmid:23206441
  23. 23. Poole CB, Tanner NA, Zhang Y, Evans TC Jr, Carlow CK. Diagnosis of brugian filariasis by loop-mediated isothermal amplification. PLoS Negl Trop Dis. 2012;6(12):e1948. pmid:23272258
  24. 24. Lim KT, The CS, Thong KL. Loop-mediated isothermal amplification assay for the rapid detection of Staphylococcus aureus. Biomed Res Int. 2013;2013:895816. pmid:23509796
  25. 25. Dhama K, Karthik K, Chakraborty S, Tiwari R, Kapoor S, Kumar A, et al. Loop-mediated isothermal amplification of DNA (LAMP): a new diagnostic tool lights the world of diagnosis of animal and human pathogens: a review. Pak J Biol Sci. 2014;17(2):151–66. pmid:24783797
  26. 26. de Souza DK, Quaye C, Mosi L, Addo P, Boakye DA. A quick and cost effective method for the diagnosis of Mycobacterium ulcerans infection. BMC Infect Dis. 2012;12:8. pmid:22257432
  27. 27. Njiru ZK, Yeboah-Manu D, Stinear TP, Fyfe JAM. Rapid and sensitive detection of Mycobacterium ulcerans using a Loop-Mediated Isothermal Amplification test. J Clin Microbiol. 2012; 50(5):1737–41. pmid:22357495
  28. 28. Ablordey A, Amissah DA, Aboagye IF, Hatano B, Yamazaki T, Sata T, et al. Detection of Mycobacterium ulcerans by the Loop Mediated Isothermal Amplification Method. PLoS Neglected Tropical Diseases. 2012;6(4):e1590. pmid:22509415
  29. 29. Beissner M, Symank D, Phillips RO, Amoako YA, Awua-Boateng NY, Sarfo FS, et al. Detection of Viable Mycobacterium ulcerans in Clinical Samples by a Novel Combined 16S rRNA Reverse Transcriptase/IS2404 Real- Time qPCR Assay. PLoS Negl Trop Dis. 2012;6(8):e1756. pmid:22953006
  30. 30. Herbinger KH, Adjei O, Awu-Boateng NY, Nienhuis WA, Kunaa L, Siegmund V, et al. Comparative study of the sensitivity of different diagnostic methods for the laboratory diagnosis of Buruli ulcer disease. Clin Infect Dis. 2009;48(8):1055–64. pmid:19275499
  31. 31. Phillips RO, Sarfo FS, Osei-Sarpong F, Boateng A, Tetteh I, Lartey A, et al. Sensitivity of PCR Targeting Mycobacterium ulcerans by Use of Fine-Needle Aspirates for Diagnosis of Buruli Ulcer. Journal of Clinical Microbiology. 2009;47(4):924–26. pmid:19204098
  32. 32. Herbinger KH, Beissner M, Huber K, Awua-Boateng NY, Nitschke J, Thompson W, et al. Efficiency of Fine-Needle Aspiration Compared with Other Sampling Techniques for Laboratory Diagnosis of Buruli Ulcer Disease. Journal Of Clinical Microbiology. 2010;48(10):3732–34. pmid:20739480
  33. 33. Bretzel G, Huber KL, Kobara B, Beissner M, Piten E, Herbinger KH, Wiedemann FX, et al. Laboratory Confirmation of Buruli Ulcer Disease in Togo, 2007–2010. PLoS Negl Trop Dis. 2011;5(7):e1228. pmid:21811641
  34. 34. WHO. Weekly epidemiological record No.17: Buruli ulcer: progress report, 2004–2008. Geneva, Switzerland: World Health Organization. 2008;83:145–56.
  35. 35. Bretzel G, Siegmund V, Nitschke J, Herbinger KH, Thompsopn W, Klutse E, et al. A stepwise approach to the laboratory diagnosis of Buruli ulcer disease. Trop Med Int Health. 2007;12(1):89–96. pmid:17207152
  36. 36. Siegmund V, Adjei O, Nitschke J, Thompson W, Klutse E, Herbinger KH, et al. Dry reagent-based polymerase chain reaction compared with other laboratory methods available for the diagnosis of Buruli ulcer disease. Clin Infect Dis. 2007;45(1):68–75. pmid:17554703
  37. 37. Beissner M, Awua-Boateng NY, Thompson W, Nienhuis WA, Klutse E, Agbenorku P, et al. A Genotypic Approach for Detection, Identification, and Characterization of Drug Resistance in Mycobacterium ulcerans in Clinical Samples and Isolates from Ghana. Am J Trop Med Hyg. 2010;83(5):1059–65. pmid:21036838
  38. 38. Jansson M, Beissner M, Phillips RO, Badziklou K, Piten E, Maman I, et al. Comparison of Two Assays for Molecular Determination of Rifampin Resistence in Clinical Samples from Patients with Buruli Ulcer Disease. J. Clin. Microbiol. 2014;52(4):1246. pmid:24478404
  39. 39. Taylor S, Wakem M, Dijkman G, Alsarraj M, Nguyen M. A practical approach to RT-qPCR—Publishing data that conform to the MIQE guidelines. Methods. 2010;50(4):1–5.
  40. 40. Benson DA, Karsch-Mizrachi I, Lipman DJ, Ostell J, Wheeler DL. GenBank. Nucleic Acids Res. 2008;36(D25-30).
  41. 41. Parida M, Sannarangaiah S, Dash PK, Rao PVL, Morita K. Loop mediated isothermal amplification (LAMP): a new generation of innovative gene amplification technique; perspectives in clinical diagnosis of infectious diseases. Rev Med Virol. 2008;18:407–21. pmid:18716992
  42. 42. Phillips RO, Phanzu DM, Beissner M, Badziklou K, Luzolo EK, Sarfo FS, et al. Effectiveness of routine BCG vaccination on buruli ulcer disease: a case-control study in the Democratic Republic of Congo, Ghana and Togo. PLoS Negl Trop Dis. 2015; 9(1):e3457. pmid:25569674
  43. 43. World Health Organization, Laboratory Support Network. Global network of laboratories for confirming Mycobacterium ulcerans infection (Buruli ulcer). Geneva, Switzerland: World Health Organization. Available: http://www.who.int/buruli/ Global_network_laboratories_PCR.pdf. Accessed 28 Mai 2015.
  44. 44. Eddyani M, Lavender C, de Rijk WB, Bomans P, Fyfe J, de Jong B, et al. Multicenter External Quality Assessment Program for PCR Detection of Mycobacterium ulcerans in Clinical and Environmental Specimens. PLoS One. 2014;9(2):e89407. pmid:24586755
  45. 45. McKenna . Refining Sample Prep for Molecular Dx. Genetic Engineering & Biotechnology News. 2013;33(17):26–7.
  46. 46. Tansuphasiri U. Detection of Mycobacterium tuberculosis from sputum collected on filter paper and stored at room temperature for 5 days by PCR assay and culture. J Med Assoc Thai. 2001;84(8):1183–9. pmid:11758856